Entry  Fri Nov 1 17:54:47 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection NFSS_RFPD_Output_Oscilloscope.pdfSFSS_RFPD_Output_Oscilloscope.pdfNFSS_RFPD_RF_OUT_COUPLED.pdfSFSS_RFPD_RF_OUT_COUPLED.pdf
    Reply  Tue Nov 5 18:37:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS Testpoint Spectrum Data FSS_TP_Spectrum.pdf
       Reply  Sat Nov 9 19:28:36 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Loop nonlinearity Test NFSS_NonLinearity_Test.pngSFSS_NonLinearity_Test.pngsignal-attachment-2019-11-09-195635.jpeg
          Reply  Mon Nov 11 20:42:49 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Test NFSS_RFPD_SN009_TF_11-11-2019_183131.pdfSFSS_RFPD_SN010_TF_11-11-2019_203310.pdfFSStwoToneTest.pdf
             Reply  Tue Nov 12 18:36:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test FSStwoToneThirdOrderIM.pdf
                Reply  Thu Nov 14 12:19:32 2019, rana, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test 
                   Reply  Tue Nov 19 18:51:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Second-order Distortion Calculation FSSDistortionCalc.pdfFSSDistortion.zip
                      Reply  Mon Dec 9 19:54:44 2019, anchal, DailyProgress, FSS, FSS Diagnostics - How much distortion affects the functioning of PDH? 
       Reply  Wed Nov 27 13:59:24 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS TP1 TimeSeries NFSS_TP1_Oscilloscope.pdf
          Reply  Wed Nov 27 15:17:51 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS NFSS_OLTF_HF.pdf
             Reply  Wed Nov 27 16:31:20 2019, rana, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS 
                Reply  Wed Dec 4 18:11:09 2019, anchal, DailyProgress, FSS, Free running laser frequency noise spectrum PMCErrorSignals.pdf
                   Reply  Thu Dec 5 18:25:11 2019, rana, DailyProgress, FSS, Free running laser frequency noise spectrum 
                   Reply  Tue Dec 17 18:29:58 2019, anchal, DailyProgress, FSS, Need for ISS? 
                      Reply  Wed Dec 18 18:40:32 2019, anchal, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all NorthLaserRINAnalysis.pdf
                         Reply  Wed Dec 18 22:03:17 2019, rana, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all 
                            Reply  Fri Dec 20 15:00:30 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement NorthLaserPZTEOMtoRINTF.pdf
                               Reply  Fri Dec 20 18:08:44 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement Through FSS NorthLaserPZTEOMtoRINTFThruFSS.pdf
                                  Reply  Tue Jan 7 17:59:20 2020, anchal, DailyProgress, FSS, NPMC TF NPMC_TF.pdf
       Reply  Thu Dec 12 19:15:55 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Beatnote Spectrum wrt Laser Power and Modulation Index BeatnoteSpectrumwrtPowerandModindex.pdf
       Reply  Tue Dec 17 16:55:07 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS North U7 Boost Stage Input-Output TimeSeries NFSS_UT_TS_Analysis.pdf
    Reply  Tue Nov 12 11:39:48 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection 
    Reply  Fri Dec 13 16:24:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD Circuit Latest Available Photos SN009.jpgSN010.jpgSN009.jpgSN010.jpg
    Reply  Fri Dec 13 17:44:05 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Coupled Output Offset 
       Reply  Mon Jan 13 19:04:38 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improvement Attempt SN009.jpgSN009_D980454-00-C.pdfSN009_Notch_Improvement_Attempt.pdf
          Reply  Tue Jan 14 18:19:33 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improved! SN009.jpgSN009_D980454-00-C.pdfSN009_Notch_Improvement_Attempt.pdf
             Reply  Thu Jan 16 11:17:15 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improvement Attempt SN010.jpgSN010_D980454-00-C.pdfSN010_Notch_Improvement_Attempt.pdf
                Reply  Thu Jan 16 20:18:44 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improved SN010.jpgSN010_D980454-00-C.pdfSN010_Notch_Improved.pdf
                   Reply  Mon Jan 27 18:29:19 2020, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Out Spectrum FSS_RFPD_RFOut_Spectrum.pdf
Message ID: 2512     Entry time: Tue Jan 14 18:19:33 2020     In reply to: 2509     Reply to this: 2513
Author: anchal 
Type: DailyProgress 
Category: FSS 
Subject: FSS Diagnostics - SN009 (North RFPD) Notch Improved! 

Circuit changes (Updated Schematic)

I made the following changes to SN009:

  • R1   ->  100 Ohms
  • R2   ->  10 Ohms
  • Added Ln, 220 nH fixed inductor. (Coilcraft 1206CS-221X_L_)
  • Added Cn 4.5-50 pF Trim Tunable Capactior (GKY50086).
  • Ln and Cn are in series such that Ln + Cn is parallel to R1.

Updated Photo

 


Measurements

  • I used the setup in 40m to take light transfer function of the modified SN009.
  • Method described in CTN:2247 was used.

Inference

  • The ratio of resonance peak at 1-Omega (36 MHz) to 2-Omega (72 MHz) has been imporved from about 30 to about 200!
  • Overall gain has been reduced but we can just increased the laser power to compensate for that.
  • The problem from last time got resolved by chosing a combination of L and C such that C is significantly more than any stray capacitances.

Data

Attachment 2: SN009_D980454-00-C.pdf  961 kB  | Hide | Hide all
SN009_D980454-00-C.pdf SN009_D980454-00-C.pdf
Attachment 3: SN009_Notch_Improvement_Attempt.pdf  79 kB  Uploaded Wed Jan 15 13:51:07 2020  | Hide | Hide all
SN009_Notch_Improvement_Attempt.pdf SN009_Notch_Improvement_Attempt.pdf
ELOG V3.1.3-