40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
PSL |
Not logged in |
 |
|
Fri Oct 4 13:41:44 2019, anchal, Notes, FSS, FSS Plant Model  
|
Fri Oct 4 14:56:42 2019, Ian MacMillan, Notes, FSS, FSS Plant Model
|
Tue Oct 8 10:41:43 2019, anchal, Notes, FSS, FSS Plant Model
|
Mon Oct 14 12:54:41 2019, anchal, Notes, FSS, FSS Plant Model v2  
|
Tue Oct 22 13:07:45 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2  
|
Thu Oct 24 11:42:06 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2
|
Tue Dec 3 17:27:54 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2 
|
Mon Dec 9 18:08:13 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2 
|
Tue Dec 10 11:57:58 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2
|
Thu Dec 12 11:54:11 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2   
|
Thu Dec 12 19:05:33 2019, rana, Notes, FSS, FSS Plant Model v2
|
Mon Dec 16 09:43:56 2019, Ian MacMillan, Notes, FSS, FSS Plant Model v2  
|
Tue Jan 14 13:30:39 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2  
|
Tue Jan 14 13:32:16 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2  
|
Tue Jan 28 13:27:25 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2  
|
Mon Feb 10 12:46:17 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2
|
Fri Feb 14 16:46:47 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2 
|
Tue Feb 18 15:42:46 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2  
|
Tue Mar 3 17:13:04 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2    
|
Wed Mar 11 15:40:50 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2 
|
Thu Mar 12 16:21:49 2020, Ian MacMillan, Notes, FSS, FSS Plant Model v2 
|
Thu Mar 12 17:18:51 2020, rana, Notes, FSS, FSS Plant Model v2
|
|
Message ID: 2498
Entry time: Mon Dec 16 09:43:56 2019
In reply to: 2494
Reply to this: 2510
|
Author: |
Ian MacMillan |
Type: |
Notes |
Category: |
FSS |
Subject: |
FSS Plant Model v2 |
|
|
In talks with Anchal, we have adapted our model and the transfer function fits rather well. These measurements were taken with the AGmeasure and SRmeasure script on the larger instruments in the CTN lab (I can't remember the exact make of them). Two were used for different frequency bands: one up to 100 kHz and the other one above. In order to have continuous measurements, the high-frequency measurement was shifted to be continuous with the lower frequency measurement. This was not a significant shift and does not impact the overall interpretation. It is merely a correction for different calibrations in the Instruments used to measure the transfer functions.
The Magnitude of the transfer functions matches well and Anchal agrees that they also look good. So the circuit can be used in its current form. It isn't currently in a box but that is an easy fix. I am still making full documentation on it and that should be done over break. I am also looking to make it as a single PCB to eliminate any points of failure that come from attaching multiple PCBs together. It is not entirely necessary because as is seen in the graphs, the circuit works well but designing a PCB is a good task for over break because it does not really require any in lab work and will be highly beneficial down the road. I am still talking to Anchal about the circuit and its implementation. |
|
|
|
|
|