Fri Nov 1 17:54:47 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection   
|
Tue Nov 5 18:37:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS Testpoint Spectrum Data
|
Sat Nov 9 19:28:36 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Loop nonlinearity Test  
|
Mon Nov 11 20:42:49 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Test  
|
Tue Nov 12 18:36:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test
|
Thu Nov 14 12:19:32 2019, rana, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test
|
Tue Nov 19 18:51:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Second-order Distortion Calculation 
|
Mon Dec 9 19:54:44 2019, anchal, DailyProgress, FSS, FSS Diagnostics - How much distortion affects the functioning of PDH?
|
Wed Nov 27 13:59:24 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS TP1 TimeSeries
|
Wed Nov 27 15:17:51 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS
|
Wed Nov 27 16:31:20 2019, rana, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS
|
Wed Dec 4 18:11:09 2019, anchal, DailyProgress, FSS, Free running laser frequency noise spectrum
|
Thu Dec 5 18:25:11 2019, rana, DailyProgress, FSS, Free running laser frequency noise spectrum
|
Tue Dec 17 18:29:58 2019, anchal, DailyProgress, FSS, Need for ISS?
|
Wed Dec 18 18:40:32 2019, anchal, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all
|
Wed Dec 18 22:03:17 2019, rana, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all
|
Fri Dec 20 15:00:30 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement
|
Fri Dec 20 18:08:44 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement Through FSS
|
Tue Jan 7 17:59:20 2020, anchal, DailyProgress, FSS, NPMC TF
|
Thu Dec 12 19:15:55 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Beatnote Spectrum wrt Laser Power and Modulation Index
|
Tue Dec 17 16:55:07 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS North U7 Boost Stage Input-Output TimeSeries
|
Tue Nov 12 11:39:48 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection
|
Fri Dec 13 16:24:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD Circuit Latest Available Photos   
|
Fri Dec 13 17:44:05 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Coupled Output Offset
|
Mon Jan 13 19:04:38 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improvement Attempt  
|
Tue Jan 14 18:19:33 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improved!  
|
Thu Jan 16 11:17:15 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improvement Attempt  
|
Thu Jan 16 20:18:44 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improved  
|
Mon Jan 27 18:29:19 2020, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Out Spectrum
|
|
Message ID: 2497
Entry time: Fri Dec 13 17:44:05 2019
In reply to: 2496
Reply to this: 2509
|
Author: |
anchal |
Type: |
DailyProgress |
Category: |
FSS |
Subject: |
FSS Diagnostics - RFPD RF Coupled Output Offset |
|
|
I rechecked the 20dB coupled RF output of SN009 (RFPD on north Cavity reflection) and SN010 (RFPD on south Cavity refleciton). The following are the mean values over 10s taken at DC coupling with 50 Ohm input impedance, +/- 40 mV input range with two different oscilloscopes.
SN009 RF Out 20dB Coupled
|
Shorted |
SN009 RF Out 20dB Coupled |
SN010 RF Out 20dB Coupled |
TDS 3052B |
-0.2 +/- 0.01 mV |
7.0 +/- 0.1 mV |
7.3 +/- 0.1 mV |
TDS 3034C |
0.5 +/- 0.01 mV |
7.4 +/- 0.1 mV |
7.6 +/- 0.1 mV |
Note that these are 20dB coupled values, so the actual offset reaching the FSS board is about 70 mV. The PD input is AC coupled through a transformer so it shouldn't be reaching further than there but ideally MAX4107 at the RFPD is supposed to have a maximum input offset voltage of 3 mV which at gain 10 should look like 30 mV. So what we are seeing (7.0 mV) is more than twice the rated maximum input offset. I'm not sure if this means our MAX4017 is busted or something is wrong in the loop. Help needed in understanding this result. |