Fri Nov 1 17:54:47 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection   
|
Tue Nov 5 18:37:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS Testpoint Spectrum Data
|
Sat Nov 9 19:28:36 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Loop nonlinearity Test  
|
Mon Nov 11 20:42:49 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Test  
|
Tue Nov 12 18:36:59 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test
|
Thu Nov 14 12:19:32 2019, rana, DailyProgress, FSS, FSS Diagnostics - Two Tone Third Order Intermodulation Test
|
Tue Nov 19 18:51:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Second-order Distortion Calculation 
|
Mon Dec 9 19:54:44 2019, anchal, DailyProgress, FSS, FSS Diagnostics - How much distortion affects the functioning of PDH?
|
Wed Nov 27 13:59:24 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS TP1 TimeSeries
|
Wed Nov 27 15:17:51 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS
|
Wed Nov 27 16:31:20 2019, rana, DailyProgress, FSS, FSS Diagnostics - Quick HF OLTF of NFSS
|
Wed Dec 4 18:11:09 2019, anchal, DailyProgress, FSS, Free running laser frequency noise spectrum
|
Thu Dec 5 18:25:11 2019, rana, DailyProgress, FSS, Free running laser frequency noise spectrum
|
Tue Dec 17 18:29:58 2019, anchal, DailyProgress, FSS, Need for ISS?
|
Wed Dec 18 18:40:32 2019, anchal, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all
|
Wed Dec 18 22:03:17 2019, rana, DailyProgress, FSS, Maybe the NPRO Laser isn't bad after all
|
Fri Dec 20 15:00:30 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement
|
Fri Dec 20 18:08:44 2019, anchal, DailyProgress, FSS, PZT, EOM to RIN TF Measurement Through FSS
|
Tue Jan 7 17:59:20 2020, anchal, DailyProgress, FSS, NPMC TF
|
Thu Dec 12 19:15:55 2019, anchal, DailyProgress, FSS, FSS Diagnostics - Beatnote Spectrum wrt Laser Power and Modulation Index
|
Tue Dec 17 16:55:07 2019, anchal, DailyProgress, FSS, FSS Diagnostics - TTFSS North U7 Boost Stage Input-Output TimeSeries
|
Tue Nov 12 11:39:48 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Ouput under inspection
|
Fri Dec 13 16:24:02 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD Circuit Latest Available Photos   
|
Fri Dec 13 17:44:05 2019, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Coupled Output Offset
|
Mon Jan 13 19:04:38 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improvement Attempt  
|
Tue Jan 14 18:19:33 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN009 (North RFPD) Notch Improved!  
|
Thu Jan 16 11:17:15 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improvement Attempt  
|
Thu Jan 16 20:18:44 2020, anchal, DailyProgress, FSS, FSS Diagnostics - SN010 (South RFPD) Notch Improved  
|
Mon Jan 27 18:29:19 2020, anchal, DailyProgress, FSS, FSS Diagnostics - RFPD RF Out Spectrum
|
|
Message ID: 2478
Entry time: Tue Nov 12 11:39:48 2019
In reply to: 2470
|
Author: |
anchal |
Type: |
DailyProgress |
Category: |
FSS |
Subject: |
FSS Diagnostics - RFPD RF Ouput under inspection |
|
|
I calculated these values by integrating in the 8 MHz neighborhood around the marked harmonic peak, the power spectral density using the frequency at the point as the lower edge of the bin. Slew rate is calculated by multiplying the rms voltage level with the frequency and the fraction is calculated against the datasheet value for Max 4107.
NFSS RFPD Output Slew Rate Usage (MAX 4107, SR: 500 V/us)
Freq (MHz) |
Vrms (mV) |
Required Slew Rate (V/us) |
Fraction of Slew Rate Used (%) |
36
|
59.221
|
2.1319
|
0.4264
|
72 |
9.248
|
0.6659
|
0.1332
|
108 |
1.135
|
0.1226
|
0.0245
|
144 |
0.39
|
0.0562
|
0.0112
|
180 |
0.621
|
0.1117
|
0.0223
|
SFSS RFPD Output Slew Rate Usage (MAX 4107, SR: 500 V/us)
Freq (MHz) |
Vrms (mV) |
Required Slew Rate (V/us) |
Fraction of Slew Rate Used (%) |
37 |
23.205
|
0.8586
|
0.1717
|
74 |
3.248
|
0.2403
|
0.0481
|
111 |
0.495
|
0.055
|
0.011
|
148 |
0.301
|
0.0445
|
0.0089
|
185 |
0.512
|
0.0947
|
0.0189
|
These calculations at least show that MAX 4107 should be much far away from reaching its slew rate limit in both RFPDs. |