40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Mon Oct 14 17:01:43 2019, anchal, Notes, Laser, South Laser Switched Off! SouthLaserStopped.pdf
    Reply  Mon Oct 14 18:30:14 2019, anchal, Notes, PMC, South PMC Error Signal Oscillations SPMCMixerOutOscillations.pdf
       Reply  Tue Oct 15 18:54:16 2019, anchal, Notes, PMC, South PMC further problems 
          Reply  Wed Oct 16 16:56:43 2019, anchal, Notes, PMC, South PMC won't work with it's own Servo Card only 
             Reply  Wed Oct 16 17:33:08 2019, Koji, Notes, PMC, South PMC won't work with it's own Servo Card only 
                Reply  Wed Oct 16 18:46:58 2019, anchal, Notes, PMC, South PMC started working somehow :( 
                   Reply  Wed Oct 23 17:39:44 2019, anchal, Notes, PMC, South PMC started working somehow :( 
                      Reply  Fri Oct 25 16:50:12 2019, anchal, Notes, PMC, South PMC problem debugging efforts 
                         Reply  Tue Oct 29 19:10:59 2019, anchal, Notes, PMC, South PMC problem debugging efforts SPMC_Diagnostics.pdf
                            Reply  Wed Oct 30 15:15:28 2019, anchal, Notes, PMC, Replacing U5A AD602 didn't solve the issue. 
                               Reply  Wed Oct 30 18:35:31 2019, anchal, Notes, PMC, Issue closed 
                                  Reply  Mon Nov 4 12:15:01 2019, anchal, Notes, PMC, Changing the autolocking method 
Message ID: 2451     Entry time: Mon Oct 14 17:01:43 2019     Reply to this: 2452
Author: anchal 
Type: Notes 
Category: Laser 
Subject: South Laser Switched Off! 

On Oct 11th at 15:04:04, the south laser switched off on its own. I would like to know if anyone entered the lab around this time. Koji did mention that our Laser Safety sign outside was blinking, but I have no more information than that. Attached is the data of south PMC reflection DC, which is the first photodiode that measures the laser. It suddenly went to zero, indicating the laser was switched off and the locks did not drive it to this point. I'm also finding that the laser intensity is reducedasit used to saturate the South PMC reflection photodiode when unlocked but presently shows around 5V. I'm trying to put the experiment back to same parameters as before.


Code and Data

Attachment 1: SouthLaserStopped.pdf  28 kB  Uploaded Mon Oct 14 18:02:12 2019  | Hide | Hide all
SouthLaserStopped.pdf SouthLaserStopped.pdf SouthLaserStopped.pdf
ELOG V3.1.3-