Thu Jan 10 10:39:27 2019, anchal, DailyProgress, NoiseBudget, New noise budget code is ready, but
|
Mon Aug 12 11:56:21 2019, anchal, DailyProgress, NoiseBudget, Adding specifics of discepancy
|
Tue Aug 13 11:59:56 2019, anchal, DailyProgress, NoiseBudget, Checking the differences between old and new noisebudget 
|
Tue Aug 13 17:05:53 2019, aaron, DailyProgress, NoiseBudget, Checking the differences between old and new noisebudget
|
Tue Sep 24 15:40:33 2019, anchal, DailyProgress, NoiseBudget, Adding specifics of discepancy 2
|
Tue Sep 24 17:11:13 2019, anchal, DailyProgress, NoiseBudget, Adding specifics of discepancy 2 (More)
|
Tue Sep 24 19:39:06 2019, anchal, DailyProgress, NoiseBudget, Checking variation in estimated Coating Brownian Noise with Loss Angles 
|
Fri Sep 27 16:40:49 2019, anchal, DailyProgress, NoiseBudget, Checking variation in estimated Coating Brownian Noise with Loss Angles Better Plots 
|
Mon Sep 30 11:14:41 2019, anchal, DailyProgress, NoiseBudget, Checking variation in estimated Coating Brownian Noise with Loss Angles 2D Plots 
|
Mon Sep 30 12:25:24 2019, rana, DailyProgress, NoiseBudget, Checking variation in estimated Coating Brownian Noise with Loss Angles 2D Plots
|
Mon Sep 30 20:08:42 2019, anchal, DailyProgress, NoiseBudget, Checking variation in estimated Coating Brownian Noise with Beam Radius 
|
Tue Oct 1 13:46:49 2019, anchal, DailyProgress, NoiseBudget, Using bayesian inference to estimate loss Angles 
|
Wed Oct 2 15:13:27 2019, anchal, DailyProgress, NoiseBudget, Checking variation in estimated Coating Brownian Noise with Beam Radius 
|
Wed Oct 23 13:21:12 2019, anchal, DailyProgress, NoiseBudget, Fixed all discrepancies.  
|
Fri Jun 12 12:34:08 2020, anchal, DailyProgress, NoiseBudget, Resolving discrepancy #1 concretly
|
|
Message ID: 2433
Entry time: Tue Sep 24 19:39:06 2019
In reply to: 2432
Reply to this: 2434
|
Author: |
anchal |
Type: |
DailyProgress |
Category: |
NoiseBudget |
Subject: |
Checking variation in estimated Coating Brownian Noise with Loss Angles |
|
|
Gabriele suggested that I plot the estimated coating Brownian noise for different values of bulk and shear loss angles and compare with the old code. The expectation is that the estimated noise should vary smoothly and should agree with old result where bulk and shear loss angles are taken the same. In the first plot, I have plotted estimate coating Brownian noise for CTN experiment over the nominal range of bulk and shear loss angles. Since this range doesn't really have an overlap with equal bulk and shear loss angle case, I made another plot in which such a square region is chosen. We can see that the estimated coating Brownian noise is slightly higher (roughly 2.5% from CTN:2390 figure 2).
Let me know if someone can think of a better viewing angle or more insights to take out from this. Code is attached. |
|
|
|