40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
PSL |
Not logged in |
 |
|
Mon Aug 19 15:54:51 2019, anchal, Summary, TempCtrl, Vac Can Heater Driver not working!
|
Tue Aug 20 19:38:36 2019, anchal, Summary, Computers, Framebuilder re-configured
|
Wed Aug 21 10:28:58 2019, anchal, Summary, TempCtrl, Vacuum Can Step Response
|
|
Message ID: 2397
Entry time: Wed Aug 21 10:28:58 2019
In reply to: 2394
|
Author: |
anchal |
Type: |
Summary |
Category: |
TempCtrl |
Subject: |
Vacuum Can Step Response |
|
|
Utilizing the fact that the heater stopped last Friday, I fitted the day to extract out a time constant for cooling down of the Vacuum Can through the present insulation. Attached is the fit. The time constant has come out to 2.298 +/- 0.001 hr.
Quote: |
Infact, I should use this data in future to calculate current time constant of temperature decay of the can through the insulation.
|
Code and Data |
|
|