Wed Apr 24 16:01:55 2019, anchal, DailyProgress, FSS, Time series and spectrum analysis of RFout of SCavReflRFPD near resonance 
|
Fri Apr 26 14:34:03 2019, anchal, DailyProgress, FSS, Time series and spectrum analysis of RFout of SCavReflRFPD near resonance
|
Sun Apr 28 22:31:39 2019, rana, DailyProgress, FSS, RF beats
|
Tue Apr 30 18:07:49 2019, anchal, DailyProgress, FSS, Spectrum of RFout from South Cavity Reflection RFPD
|
Wed May 8 20:03:43 2019, anchal, DailyProgress, FSS, FSS PDH Error Signal on South Path is in good health 
|
|
Message ID: 2330
Entry time: Sun Apr 28 22:31:39 2019
In reply to: 2328
Reply to this: 2331
|
Author: |
rana |
Type: |
DailyProgress |
Category: |
FSS |
Subject: |
RF beats |
|
|
but if all systems are linear, where does this beat physically get generated? What are the actual modulation frequencies for the cavities?
Quote: |
I think I figured the source of this 281 kHz peak. 281 kHz ≈ 37 MHz - 36.72 MHz, so it is the beat signal between 37 MHz and the 36.72 MHz signals. I think I should tune the RFPD more next time I open the cage to bring its resonance closer to 37 MHz.
|
|
|