40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
PSL |
Not logged in |
 |
|
Mon Dec 17 22:13:03 2018, anchal, DailyProgress, TempCtrl, Proposed circuit for cavity temperature sensor Part 2 
|
Wed Dec 19 11:01:00 2018, awade, DailyProgress, TempCtrl, Proposed circuit for cavity temperature sensor Part 2
|
Wed Dec 19 14:00:55 2018, anchal, DailyProgress, TempCtrl, Proposed circuit for cavity temperature sensor Part 2
|
|
Message ID: 2265
Entry time: Mon Dec 17 22:13:03 2018
Reply to this: 2266
|
Author: |
anchal |
Type: |
DailyProgress |
Category: |
TempCtrl |
Subject: |
Proposed circuit for cavity temperature sensor Part 2 |
|
|
The second stage of the amplifier for temperature sensor would be subtractor circuits using OP27G.
- PFA the circuit schematic consisting of the instrumentation amplifier and different circuits for acromag channels.
- There would be 1 low-resolution channel with a range from about 15
to 95 with a response: 0.238053435115*T - 13.167938931297705
- Rest would high-resolution channels with a smaller range. These channels would give us about 1mV/mK slope so that we can read with mK precision with 16 bit
V range.
- Channel 1: From 15
to 35 with a response: 0.952875*T - 22.5
- Channel 2: From 35
to 55 with a response: 0.943758551683*T - 43.329447115384596
- Channel 3: From 55
to 75 with a response: 0.955040625*T - 61.9375
- Channel 4: From 75
to 95 with a response: 0.9429496875*T - 79.74466991341995
I'm attaching calculations done in the jupyter notebook as zip file.
Wed Dec 19 17:13:14 2018
There is an error in this schematic. This is not the final schematic.
Refer this only for the calculations. Use the updated schematic at DCC. |
|
|
|