Fri Sep 7 23:23:00 2018, awade, anchal, DailyProgress, BEAT, Back to beat
|
Wed Sep 12 16:27:19 2018, anchal, DailyProgress, BEAT, Back to beat
|
Fri Sep 14 00:57:14 2018, rana, DailyProgress, BEAT, Back to beat
|
Fri Sep 21 18:10:53 2018, rana, DailyProgress, BEAT, Back to beat
|
|
Message ID: 2237
Entry time: Fri Sep 21 18:10:53 2018
In reply to: 2232
|
Author: |
rana |
Type: |
DailyProgress |
Category: |
BEAT |
Subject: |
Back to beat |
|
|
I don't understand what that means. Please provide 10x more details on how the measurement was made.
Also, clearly one of these traces is not like the other. What does that mean ???
Quote: |
Attached are transfer function measurements of the North and South Cavity Reflection RFPD (14.75MHz resonant RFPD) along with dark noise around 14.75 MHz.
Edit:[09/14/2018, 16:12] Changed plots to physical units. Used 2k Transimpedance for Bode Plot and 2.5kHz bandwidth (801 points in 2MHz) for noise plots.
|
|