40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Tue Apr 24 14:37:39 2018, Anchal and Andrew, DailyProgress, BEAT, Beam Profiling Beatnote detector BeamradiusProfile.pngBeam_Profile_Beatnote_-_Sheet1.csvfitatz500.png
    Reply  Wed Apr 25 12:25:09 2018, awade, DailyProgress, BEAT, Beam Profiling Beatnote detector 
       Reply  Thu May 3 22:05:23 2018, awade and anchal, DailyProgress, BEAT, Beam Profiling Beatnote detector 20180503_193329.jpg
          Reply  Wed May 9 00:51:45 2018, anchal, DailyProgress, BEAT, Beam Profiling Beatnote detector FitGraphs.pdf
             Reply  Wed May 9 08:56:54 2018, Neo, DailyProgress, BEAT, Beam Profiling Beatnote detector 
                Reply  Thu May 10 18:18:06 2018, Anchal, DailyProgress, BEAT, Beam Profiling Beatnote detector ModifiedFitWithNewData.pdfDataAndCode.zip
Message ID: 2182     Entry time: Thu May 10 18:18:06 2018     In reply to: 2181
Author: Anchal 
Type: DailyProgress 
Category: BEAT 
Subject: Beam Profiling Beatnote detector 

Today I took more measurements after reflecting off the beam by 90 degrees to another direction and using the Beam Profiler Dataray Beamr2-DD. I used the InGaAs detector with motor spee dof 11 rps and averaging over 100 values.

Following is the fit with and without the new data taken. Data1 in the graph is the earlier data taken using razor blade and Data2 is the data taken today using beam profiler.

The two fits estimate same waist positions and waist sizes within error bars of each other. However, the reduced chi-square is still pretty high.

I've also added the data file and code in the zip.

Attachment 1: ModifiedFitWithNewData.pdf  22 kB  Uploaded Fri May 11 11:38:37 2018  | Hide | Hide all
ModifiedFitWithNewData.pdf
Attachment 2: DataAndCode.zip  90 kB  Uploaded Fri May 11 11:38:45 2018