Not logged in |
 |
|
Thu Mar 29 02:51:35 2018, Koji, Summary, BEAT, The design for a new beat PD 6x
|
Mon Apr 2 00:15:34 2018, rana, Summary, BEAT, The design for a new beat PD
|
Wed Apr 4 02:06:45 2018, Koji, Summary, BEAT, First article of the CTN beat PD 7x
|
Thu Apr 5 02:04:49 2018, Koji, Summary, BEAT, Compensation for MAX4107 at G=4.5    
|
Fri Apr 6 04:48:02 2018, Koji, Summary, BEAT, Compensation for MAX4107 at G=4.5  
|
Sun Apr 22 17:46:55 2018, awade, Summary, BEAT, Compensation for MAX4107 at G=4.5
|
|
Message ID: 2162
Entry time: Wed Apr 4 02:06:45 2018
In reply to: 2161
Reply to this: 2164
|
Author: |
Koji |
Type: |
Summary |
Category: |
BEAT |
Subject: |
First article of the CTN beat PD |
|
|
Made the first article of the 25MHz resonant PD for CTN experiment.
Performance:
Resonant frequency f_res = 26.0MHz
Transimpedance R_res = 2.3kOhm
Input referred curent noise = 10pA/rtHz
Shot noise intercept current = 0.29mA
Remarks:
MAX4107 at G=10 gives a gain peaking at 260MHz. It is not tamed as the dark spectrum shows the peak does not have significant RMS.
But this might cause a trouble later. We can consider to replace the opamp with ones suggested by Rana. Rana bought "THS4271" to give it a try. |
|
|
|
|
|
|
|
|
|
|
|
|
|