Fri Sep 8 00:32:20 2017, awade, Summary, FSS, Schematic block diagram FSS loops
|
Thu Sep 14 21:54:26 2017, awade, Summary, FSS, FSS modifications: series resistance (R1 and R2) to monitor points on servo board D040105
|
Mon Sep 18 00:46:08 2017, rana, Summary, FSS, FSS modifications: series resistance (R1 and R2) to monitor points on servo board D040105
|
Mon Sep 18 13:34:13 2017, awade, Summary, FSS, FSS modifications: series resistance (R1 and R2) to monitor points on servo board D040105
|
Tue Oct 10 00:49:34 2017, awade, Craig, Summary, FSS, Replacing U3 on north path (2010:005) FSS servo board.  
|
Mon Oct 2 12:30:15 2017, rana, Summary, FSS, LISO & mfil
|
Wed Oct 18 11:19:15 2017, awade, Summary, FSS, Comparing south TF to LISO model  
|
Wed Oct 18 15:42:59 2017, awade, Summary, FSS, Comparing south TF to LISO model
|
Mon Oct 23 15:05:44 2017, awade, Summary, FSS, Comparing south TF to LISO model  
|
Wed Oct 25 23:15:51 2017, awade, Craig, Summary, FSS, Comparing south and north TF to LISO model  
|
|
Message ID: 1953
Entry time: Wed Oct 18 15:42:59 2017
In reply to: 1952
Reply to this: 1954
|
Author: |
awade |
Type: |
Summary |
Category: |
FSS |
Subject: |
Comparing south TF to LISO model |
|
|
probably saturating in the PZT path. Repeat with -10 dBm source and look at outputs with a scope to check if its internally saturating. |