Using AD590s for vacuum can sensing |
|
|
that's not how ADC noise works. #RMFT
Quote: |
The acromag cards are configured to have ±10 V over 2^15 bits. The least significant bit (LSB) is therefore 20V/2^15 bit = 0.61 mV/bit. For an AD590 hooked up to a tranimpedance amplifier one would select a 30.96 kΩ transresistance gain to achieve 10 V @ 50 C. This make the total gain from temperature to volgate G_{temp to V} = 1µA/K * 30.96 kV/A = 30.96 mV/K. The minimum possible resolution is therefore LSB/G_{temp to V} = 19.7 mK/bit. Not great if we are going for 1mK temperature stabilization.
|
|