The second EOM (an amplitude modulator: Newfocus-4104) was placed at the next waist and optics installed were lambda/2->AEOM(4104)->PBS. Location of the AEOM (referenced to laser head) was 1012 mm.
The beam was reameasured after this point (as a check) with a AR coated window pickoff. Fit was:
Horz. beam waist = 183.5091 um
Horz. beam waist position = 957.1859 mm
Vert. beam waist = 226.7887 um
Vert. beam waist position = 856.7048 mm.
Accounting for the AEOM and PBS length and RI this is pretty much right for placement of the AEOM.
--> Edit, fixed authour name. Fri Jul 15 17:06:45 2016 |