Subject: PMC servo problem fixed 

found the problem why the PMC servo card wasn't working: the medm screens on the sun workstation didn't have a toggle button for the disable signal for the variable gain amplifier (AD602). this signal was set to "disabled" by default and because there was no button to enable it Tara couldn't get the servo to work. added a toggle button on the screen, PMC is now locked stable

the PMC seems to be very dirty. If we lock it in s-pol the transmitted power is only 7%, even if 60% are going into it (didn't try to optimize it). the rest is dumped in the pmc. switched back to p-pol but should think about cleaning the mirrors. we also have first contact. i think it's worth a try and we can't loose much. peter has also spare mirrors (and a spare spacer)