Entry  Sat Sep 6 12:54:05 2014, Tara, Evan, DailyProgress, optic, North photothermal TF nPT.pdfnpttf.zip
    Reply  Sun Sep 7 19:33:40 2014, Tara, Evan, DailyProgress, optic, North photothermal TF photothermalTF.pdfnorthPT.zip
       Reply  Sun Nov 16 23:32:49 2014, Evan, DailyProgress, optic, North photothermal TF 
Message ID: 1532     Entry time: Sun Nov 16 23:32:49 2014     In reply to: 1505
Author: Evan 
Type: DailyProgress 
Category: optic 
Subject: North photothermal TF 

I believe the factor of π / F here is an error. It should instead be the transmission T. That lowers the absorption estimates to more like 5 ppm.

 