I took a swept-sine measurement of the photothermal TF just as Tara and I did for the north cavity. To get a better measurement, I made some configuration changes:
- I turned the power incident on the south cavity up to 8.5 mW by adjusting the post-laser HWP from 318° to 286°.
- I placed an OD2.0 in front of the beat PD to prevent RF saturation.
Settings/values:
- The beat was at 13.8 MHz.
- The PLL Marconi was on 50 kHz FM deviation, and the SR560 gain was 100 V/V.
- South transmission PD was 460(5) mV dc.
- South transmission power (directly out of vacuum chamber) was 2.20(5) mW dc.
The results are attached. I'm not sure why there's a discrepancy around 200 Hz between the two traces. Below 100 Hz the measurement looks relatively clean.
The light rejected out of the post-EOAM PBS is only 2 mW (compared with 9 mW transmitted), which makes me suspicious that the post-EOAM QWP is not rotated properly, or else the input polarization into the EOAM is wrong. We should check this before redoing this measurement.
As with the north cavity, I find that an absorption of 6 ppm is needed make the measured curve lie on top of the theory curve.
For the time being, I have left the input power at 8 mW in case we want to take this again tomorrow. There's currently a dump upstream of the PMC to block the beam.
|