Message ID: 1504
Entry time: Sun Sep 7 11:54:21 2014
In reply to: 1502
Reply to this: 1507
2443
|
Author: |
Evan |
Type: |
DailyProgress |
Category: |
FSS |
Subject: |
TTFSS OLTFs, residual frequency noise |
|
|
Acutally it does look like it's a 50 Ω loading issue. I find that when 50 Ω inline terminators are added to OUT1 and OUT2, the measured OLTF is reduced by a factor of 1.6. This explains the discrepancy between the SR785 and HP4395 measurements. I've attached the corrected OLTF plots, along with plots of a vector fit, and the expected residual frequency noise [assuming a free-running NPRO noise of 104 Hz/Hz1/2 × (1 Hz / f)].
South UGF is at 200 kHz with almost no phase margin. We need to fix this. |
|
|
|
|
|
|
|
|
|
|