Subject: PDH shot noise estimate 

Ignoring for the time being the issue of offsets in the PDH error signal, here's my prediction for the new PDH shot noise level, assuming a visibility of 0.92 × 0.7 = 0.64 and an incident power of 2 mW.

So our beat will be slightly worse around 1 kHz, but we aren't completely hosed by the shot noise. I'd think the true solution here is to find (or buy) two large-aperture Faraday isolators to replace the PBS+QWP setup (according to alm, the spot size in this region is about 1.1 mm). E.g., we might consider a large-aperture ThorLabs isolator.