Entry  Tue Aug 5 08:14:14 2014, Evan, DailyProgress, optic, BRDF of AlGaAs mirror 141 after cleaning 141brdf.pdf
    Reply  Tue Aug 5 14:54:36 2014, Evan, DailyProgress, optic, BRDF of AlGaAs mirror 132 after cleaning 132brdf.pdf
Message ID: 1466     Entry time: Tue Aug 5 08:14:14 2014     Reply to this: 1467
Author: Evan 
Type: DailyProgress 
Category: optic 
Subject: BRDF of AlGaAs mirror 141 after cleaning 

Incident power: 20.0(1) mW

Exposure times used: 25 ms, 50 ms, 200 ms, 500 ms, 1000 ms

Transmitted power: 3.34(2) µW. This gives a transmission of 167(1) ppm for this mirror.

TIS from 16° to 73° is 18(1) ppm.

Data and code are on the SVN at CTNLab/measurements/2014_08_05.