Entry  Thu Jul 31 09:40:24 2014, Evan, DailyProgress, optic, BRDF of AlGaAs mirror 114 114brdf.pdf
    Reply  Mon Aug 4 15:23:06 2014, Evan, DailyProgress, optic, BRDF of AlGaAs mirror 114 after cleaning 114_cleaned_brdf.pdf
Message ID: 1465     Entry time: Mon Aug 4 15:23:06 2014     In reply to: 1463
Author: Evan 
Type: DailyProgress 
Category: optic 
Subject: BRDF of AlGaAs mirror 114 after cleaning 

[Tara, Evan]

Tara also took a BRDF measurement of #114 after cleaning it.

After cleaning, TIS from 14° to 71° is 2.7(5) ppm. Much improved.

Data and code are on the SVN at CTNlab/measurements/2014_07_31.