The cause of the peaks around 1kHz in RIN is solved, PMC is the reason. After damping it, the peaks disappears.
Short notes from tonight measurement:
- RIN in ACAV is better after PMC is damped, no peaks around 1kHz anymore.
- The peaks in the beat measurement also disappear, so we really see photo thermal noise.
- PD for RIN measurement behind RCAV is added, RIN is measured.
- I Will add the effect from RIN and compare with beat measurement soon.
need to buy:
- new PBS for PDH locking. the one for RCAV is not good because there is an unwanted reflected beam going to the RFPD.