40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
PSL |
Not logged in |
 |
|
Thu Aug 1 03:34:35 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
Thu Aug 1 18:33:12 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
Fri Aug 2 18:15:06 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
Mon Aug 5 18:47:04 2013, Chloe, DailyProgress, ECDL, Assembling ECDL  
|
Tue Aug 6 19:48:19 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
Wed Aug 7 18:13:21 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
Thu Aug 8 18:43:44 2013, Chloe, DailyProgress, ECDL, Assembling ECDL  
|
Fri Aug 9 18:01:47 2013, Chloe, DailyProgress, ECDL, Assembling ECDL 
|
Tue Aug 13 18:13:35 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
Thu Aug 15 18:53:45 2013, Chloe, DailyProgress, ECDL, Assembling ECDL
|
|
Message ID: 1299
Entry time: Thu Aug 15 18:53:45 2013
In reply to: 1297
|
Author: |
Chloe |
Type: |
DailyProgress |
Category: |
ECDL |
Subject: |
Assembling ECDL |
|
|
Quote: |
Made some modifications to the Solidworks design. All of these have been changed on the SVN.
- Made heat sink for TEC adjustable so it can clamp the TEC element between the heat sink and the laser diode mount, with no risk of it touching the laser diode mount. This was achieved by making a longer hole on the base plate so the heat sink can be screwed on anywhere. The heat sink will be copper because this is very thermally conductive compared to aluminum or other metals.
- The laser diode will be thermally isolated from the rest of the setup using a piece of plastic (most likely delrin).
- Tara and I found a lens mount we can use which allows for small height adjustments (Newport LA1V-XY, http://search.newport.com/?x2=sku&q2=LA1V-XY). This is at an optic height of 1 inch which works well with our setup. We are going to purchase this (http://search.newport.com/?q=*&x2=sku&q2=LPLH-25T) which is an adapter so that it can hold the 0.25" optic. The collimating lens' distance from the laser diode will be adjustable by the same method as the heat sink - there is a longer hole on the base plate so that the lens mount can be screwed down anywhere.
- Added 2 D-sub holes that are properly sized (last time was a bit too small) so that we can plug in the TEC and current driver onto the ECDL box.
Tomorrow morning I will go to the machine shop to get the base plate and left plate modified, and get them to machine a heat sink and plastic insulator.
|
Today I got the newly machined parts. I put together the TEC element and stuff again and will calibrate the next time I get a chance.
Erica and I practiced our presentations in front of Tara. I got a lot of feedback and I'm going to edit my presentation in my free time outside of lab. It was also useful to see someone else's work to get an idea of how to present.
I'm working on putting together a Michelson interferometer to measure the laser diode free running noise. I don't have the actual collimating lens, so I'm using a f=5mm lens from a fiber optic. I have mirrors and I borrowed a beam splitter from the GYRO experiment. Picture below. I'm working on getting the beams to combine by adjusting the mirrors. Will continue doing this tomorrow. |
|