40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  PSL  Not logged in ELOG logo
Entry  Tue Jul 2 11:21:24 2013, Erica, DailyProgress, fiber optic, mode matching snip_of_beam_profiling_6_27.JPGP1020518.JPG
    Reply  Wed Jul 3 09:46:48 2013, Erica, DailyProgress, fiber optic, mode matching 130703-095332.jpg
       Reply  Thu Jul 11 01:40:19 2013, Erica, DailyProgress, fiber optic, mode matching P1020690.JPGP1020692.JPG
          Reply  Fri Jul 12 00:51:39 2013, Erica, Notes, fiber optic, mode matching P1020705.JPGP1020703.JPG7_11_13_beam_profile.PNG7_11_13_beam_from_fixed_lens.PNG
             Reply  Tue Jul 16 10:27:47 2013, Erica, Notes, fiber optic, mode matching, beam recombination 
                Reply  Tue Jul 16 11:02:47 2013, Erica, Notes, fiber optic, mode matching, beam recombination 130805-120628.jpg
                   Reply  Wed Jul 17 11:58:47 2013, Erica, Notes, fiber optic, mode matching, beam recombination 7_16_13_determining_polarization_axis.PNGTEK0000.PNG
                      Reply  Wed Jul 17 18:29:37 2013, Erica, DailyProgress, fiber optic, reducing drift in the recombined beam 
                         Reply  Sat Jul 20 23:39:09 2013, Erica, DailyProgress, fiber optic, reducing drift in the recombined beam 
                            Reply  Tue Jul 23 10:29:12 2013, Erica, DailyProgress, fiber optic, reducing drift in the recombined beam shielded_10s_TEK0005.PNGunshielded_10s_TEK0027.PNG
                               Reply  Tue Jul 23 11:43:00 2013, Evan, DailyProgress, Electronics Equipment, Zach's op-amp recommendations 
                               Reply  Fri Jul 26 11:07:44 2013, Erica, DailyProgress, BEAT, circuit for measuring temperature fluctations on CTN table P1030056.JPGP1030057.JPGP1030059.JPGP1030056.JPG
                                  Reply  Fri Jul 26 11:22:45 2013, Erica, DailyProgress, BEAT, circuit for measuring temperature fluctations on CTN table 
                                     Reply  Fri Jul 26 11:45:28 2013, Erica, DailyProgress, BEAT, circuit for measuring temperature fluctations on CTN table 
                                        Reply  Mon Jul 29 13:30:26 2013, Erica, DailyProgress, BEAT, adding resistor, capacitor, and sockets w_o_capacitor_7_26_13.JPGcapacitor_7_26_13.JPGP1030066.JPGP1030070.JPG
                                           Reply  Tue Jul 30 12:23:41 2013, Erica, DailyProgress, BEAT, addressed noise problem, DC signal 
                                              Reply  Thu Aug 1 00:25:02 2013, Erica, DailyProgress, fiber optic, insulating foam 731_7graph.PNG731_11graph.PNGP1030318.JPG
                                                 Reply  Fri Aug 2 12:38:41 2013, tara, DailyProgress, fiber optic, insulating foam chmbrtmp.JPGfringe.JPG
                                                    Reply  Tue Aug 6 12:31:24 2013, Erica, DailyProgress, fiber optic, insulating foam P1030403.JPG
                                                       Reply  Wed Aug 7 13:27:00 2013, Erica, DailyProgress, fiber optic, more foam, calculations for convertion from voltage to frequency P1030409.JPGP1030417.JPGP1030420.JPG
                                                          Reply  Wed Aug 7 18:53:20 2013, Erica, DailyProgress, fiber optic, noise budget, Matlab notes. noise_budget.PNGP1030427.JPG
                                                             Reply  Tue Aug 13 12:39:36 2013, Erica, DailyProgress, fiber optic, rewrote calculations for converting from voltage noise to frequency noise P1030458.JPGP1030459.JPGP1030460.JPG
                                                       Reply  Tue Aug 13 12:28:20 2013, Erica, Notes, Electronics Equipment, new AD 590 circuit and its noise calculation, beam alignment P1030461.JPGP1030454.JPGP1030456.JPGP1030457.JPGAD590circuit2.PNG
                                                          Reply  Fri Aug 16 10:35:44 2013, Erica, Notes, Electronics Equipment, AD590 msrmt, conversion to freq noise 
                   Reply  Tue Jul 30 16:19:32 2013, Erica, Notes, fiber optic, setup for beam recombination new_setup.PNGbeam_recombination2.PNG
                   Reply  Mon Aug 5 12:21:00 2013, Erica, Notes, fiber optic, calculation: phase change for fiber 130805-121449.jpg
    Reply  Tue Jul 30 15:48:47 2013, Erica, DailyProgress, fiber optic, noise for GYRO 
Message ID: 1284     Entry time: Wed Aug 7 13:27:00 2013     In reply to: 1282     Reply to this: 1286
Author: Erica 
Type: DailyProgress 
Category: fiber optic 
Subject: more foam, calculations for convertion from voltage to frequency 
august 6, 2013


We better insulated the exposed fiber by sandwiching it on both sides with foam. Instead of the aluminum foil, we used the bubble wrap that we used previously to cover the holes. We did this because the aluminum could be vibrating, which causes more noise.


I was able to take spectrum data since the drift was slow enough, at least 30s on the linear region.
The laser was not locked to the cavity when I took the data. I'll be putting it into matlab
The data was taken with the span at 400 Hz, 1.6 kHz, and 12.8 kHz. I took 50 averages
I tried to take the data so the pk-pk voltage would be the same but it still drifted a little bit.
For the data taken wtih the 1.6 kHz and 12.8 kHz, deltaV = 2.24V, Vmax = 2.7V, Vmin= 460 mV
For the 400 Hz, deltaV = 2.10V, Vmax = 2.56 V, V min = 460 mV.

The y-axis from the graph is in V/rt(Hz).
Here is the calculation done to calibrate between V and length change. see the correct calculation in 1286



*NOTE: I made an error in the calculations here. \phi = kx. But when I plugged in pi/4 to solve for dV/dx at pi/4, I did not multiply it by two. See

Here is the conversion from length to frequency.
ELOG V3.1.3-