Entry  Tue Jul 2 11:21:24 2013, Erica, DailyProgress, fiber optic, mode matching snip_of_beam_profiling_6_27.JPGP1020518.JPG
    Reply  Wed Jul 3 09:46:48 2013, Erica, DailyProgress, fiber optic, mode matching 130703-095332.jpg
       Reply  Thu Jul 11 01:40:19 2013, Erica, DailyProgress, fiber optic, mode matching P1020690.JPGP1020692.JPG
          Reply  Fri Jul 12 00:51:39 2013, Erica, Notes, fiber optic, mode matching P1020705.JPGP1020703.JPG7_11_13_beam_profile.PNG7_11_13_beam_from_fixed_lens.PNG
             Reply  Tue Jul 16 10:27:47 2013, Erica, Notes, fiber optic, mode matching, beam recombination 
                Reply  Tue Jul 16 11:02:47 2013, Erica, Notes, fiber optic, mode matching, beam recombination 130805-120628.jpg
                   Reply  Wed Jul 17 11:58:47 2013, Erica, Notes, fiber optic, mode matching, beam recombination 7_16_13_determining_polarization_axis.PNGTEK0000.PNG
                      Reply  Wed Jul 17 18:29:37 2013, Erica, DailyProgress, fiber optic, reducing drift in the recombined beam 
                         Reply  Sat Jul 20 23:39:09 2013, Erica, DailyProgress, fiber optic, reducing drift in the recombined beam 
                            Reply  Tue Jul 23 10:29:12 2013, Erica, DailyProgress, fiber optic, reducing drift in the recombined beam shielded_10s_TEK0005.PNGunshielded_10s_TEK0027.PNG
                               Reply  Tue Jul 23 11:43:00 2013, Evan, DailyProgress, Electronics Equipment, Zach's op-amp recommendations 
                               Reply  Fri Jul 26 11:07:44 2013, Erica, DailyProgress, BEAT, circuit for measuring temperature fluctations on CTN table P1030056.JPGP1030057.JPGP1030059.JPGP1030056.JPG
                                  Reply  Fri Jul 26 11:22:45 2013, Erica, DailyProgress, BEAT, circuit for measuring temperature fluctations on CTN table 
                                     Reply  Fri Jul 26 11:45:28 2013, Erica, DailyProgress, BEAT, circuit for measuring temperature fluctations on CTN table 
                                        Reply  Mon Jul 29 13:30:26 2013, Erica, DailyProgress, BEAT, adding resistor, capacitor, and sockets w_o_capacitor_7_26_13.JPGcapacitor_7_26_13.JPGP1030066.JPGP1030070.JPG
                                           Reply  Tue Jul 30 12:23:41 2013, Erica, DailyProgress, BEAT, addressed noise problem, DC signal 
                                              Reply  Thu Aug 1 00:25:02 2013, Erica, DailyProgress, fiber optic, insulating foam 731_7graph.PNG731_11graph.PNGP1030318.JPG
                                                 Reply  Fri Aug 2 12:38:41 2013, tara, DailyProgress, fiber optic, insulating foam chmbrtmp.JPGfringe.JPG
                                                    Reply  Tue Aug 6 12:31:24 2013, Erica, DailyProgress, fiber optic, insulating foam P1030403.JPG
                                                       Reply  Wed Aug 7 13:27:00 2013, Erica, DailyProgress, fiber optic, more foam, calculations for convertion from voltage to frequency P1030409.JPGP1030417.JPGP1030420.JPG
                                                          Reply  Wed Aug 7 18:53:20 2013, Erica, DailyProgress, fiber optic, noise budget, Matlab notes. noise_budget.PNGP1030427.JPG
                                                             Reply  Tue Aug 13 12:39:36 2013, Erica, DailyProgress, fiber optic, rewrote calculations for converting from voltage noise to frequency noise P1030458.JPGP1030459.JPGP1030460.JPG
                                                       Reply  Tue Aug 13 12:28:20 2013, Erica, Notes, Electronics Equipment, new AD 590 circuit and its noise calculation, beam alignment P1030461.JPGP1030454.JPGP1030456.JPGP1030457.JPGAD590circuit2.PNG
                                                          Reply  Fri Aug 16 10:35:44 2013, Erica, Notes, Electronics Equipment, AD590 msrmt, conversion to freq noise 
                   Reply  Tue Jul 30 16:19:32 2013, Erica, Notes, fiber optic, setup for beam recombination new_setup.PNGbeam_recombination2.PNG
                   Reply  Mon Aug 5 12:21:00 2013, Erica, Notes, fiber optic, calculation: phase change for fiber 130805-121449.jpg
    Reply  Tue Jul 30 15:48:47 2013, Erica, DailyProgress, fiber optic, noise for GYRO 
Message ID: 1258     Entry time: Fri Jul 26 11:45:28 2013     In reply to: 1257     Reply to this: 1261
Author: Erica 
Type: DailyProgress 
Category: BEAT 
Subject: circuit for measuring temperature fluctations on CTN table 
July 25, 2013

finished building the circuit today. Had the positive, negative, and ground wires running above the board, while the one jumper wire from the output to the negative input under.
Twisted the positive, negative, and ground wires together using a drill, as well as the positive and negative wires that will connect to the AD 590. We made these longer so we can connect to the power supply and place the AD590 at opposite ends of the table.

Tested the circuit, used an amplitude of 0.01Vpp and 0.1 Hz for frequency to drive circuit, which was what we did on Monday. The output signal is a square wave which was strange but found out the problem: the bnc cable driving the circuit was put in the sync output of the function generator, instead of the function output.
Fixed this, and the circuit behaves as we expect.

Discovered that I used a 36 kohm resistor instead of a 33 k ohm resistor, so now we have a gain of about 110, which is close to that of 100.

We used aluminum tape to connect the AD590 and insulating tape to prevent shorting. The output signal was at some DC voltage, which we expect at first due to the power supply turning on, and it should die away, but it didn't, or was very very slowly. So Evan placed a 15k Ohm resistor in parallel w/ the 2M ohm resistor in the high pass filter to lower the time constant, which brought the signal close to zero. Once he took the resistor away, then the signal would drift up to the previous DC level. The circuit was responding as expected when he placed a cooler object by it, so the signal went down, and the signal went up when we held the AD 590.

We tried this in the CTN lab but it didn't seem to work; there was a lot of noise. I'll test it again tomorrow.
A possibility is the power supply could be noisy.