40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
PSL |
Not logged in |
 |
|
Thu Apr 18 19:54:01 2013, tara, Notes, DAQ, EPICS channel for slow feedback to laser
|
Mon Apr 22 17:13:24 2013, tara, Notes, DAQ, slow feedback to laser via EPICS is on  
|
Wed May 8 01:11:09 2013, tara, Notes, DAQ, slow feedback to laser via EPICS is on
|
|
Message ID: 1172
Entry time: Wed May 8 01:11:09 2013
In reply to: 1155
|
Author: |
tara |
Type: |
Notes |
Category: |
DAQ |
Subject: |
slow feedback to laser via EPICS is on |
|
|
Slow feedback for 2nd laser is ready.
EPICs channel:
- C3:PSL-RCAV-FMON was created for fast mon to laser.
- C3:PSL-RCAV_SLOWOUT was created for SLOW feedback. The channel was originally named C3:PSL-FSS_VCOMODLEVEL J9 input 11 and 12, VMIVME-4116, C2 S4.
The output of EPICS channels have capacitors installed in parallel for low pass filter.
|