40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  Coating Ring-down Measurement Lab elog, Page 4 of 18  Not logged in ELOG logo
ID Date Authorup Type Category Subject
  50   Mon Jul 18 17:48:58 2016 GabrieleElectronicsConfigurationStatus of real time system and workstation

[EricQ, Gabriele]

The real time system seems to be working properly, except for the excitations: we can't activate any excitation using awggui or diaggui

Eric rebuilt the workstation from scratch installing Debian 8.5. All CDS software seem to be working. We setup a ssh-key for ssh'ing into cymac3 and configured the automatic mount of the remote /opt/rtcds.

  51   Mon Jul 18 18:05:46 2016 GabrieleOpticsDaily ProgressOptical setup

Small modifications to the optical setup:

  • added a 1m focal lenght lens (AR coated for 1064nm...) right before the beam enters the viewport, to reduce the beam spot size on the QPD
  • added a 0.3 ND right after the laser, to avoid saturation of the QPD (about 75 microwatts total on the QPD)

  52   Tue Jul 19 17:15:37 2016 GabrieleFacilityDaily ProgressVacuum pressure acquired

[Alena, Gabriele]

We connected the analog output of the vacuum gauge controller to one of the ADC channels. The signal is calibrated so that the pressure is 10^(X3:CR1-PRESSURE_LOGTORR_OUT). Unfortunately the RTG does not know how to compute 10^x...

  53   Tue Jul 19 17:21:48 2016 GabrieleGeneralDaily ProgressFirst modes excited!

We can't generate any arbitary signal with the real time model, since awg is not working properly yet. For the moment being I added a uniform random number generator in the model (only option I found for noise) and send it into the ESD filter bank. In this way I can generate band-passed noise.

I plugged in the DAC output to the HV amplifier input, and I could send white noise to the electrostatic drive. Behold: I was able to excite quite a few modes. In the following trace blue is a reference and red is right after I sent white noise (3 V peak to peak) to the disk for a while (less than 1 minute). Excitation stopped at 4:56pm LT.

Using COMSOL and tuning the disk thickness at 1.018 mm I could hit the frequency of the first butterfly mode (1109 Hz) and get a reasobly good estimate of the other modes.

After about half a hour of ring down, most of the modes are gone but the two lowest are still going strong.

Note that the flattish background noise seems to be generated by some sort of glitches. I tried to swap the laser and the power supply, without change. More investigations are needed.

Note that the roughing pump was still on during the test.

  54   Tue Jul 19 20:19:38 2016 GabrieleElectronicsConfigurationGlitches are a digital artifact

The glitches I saw in the data happens roughly every second, even though not exactly on the second. They are suddend jumps on the signal values over one sample, so of clear digital origin

 

Attachment 1: glitches.png
glitches.png
  55   Wed Jul 20 10:26:40 2016 GabrieleElectronicsConfigurationAre glitches a digital artifact?

Not so sure anymore...

  • those glitches do not happen at regular times
  • I tried to send a sinusoid into another ADC channel, and I couldn't see any jump
  • using DTT and zooming into the jumps, they don't seem a clean one-sample jump anymore... There might be a bit of ringing before and/or after the jump
  • The jumps have different sizes and directions, and they seems to happen at the same time in all four quadrants, but not in the same direction or with the same amplitude

I was suspecting dust crossing the beam, so I build a very rough enclosure, that should help with dust. I don't think I saw any change in the glitches. 

Also:

  • glitches are there even with the HeNe off, with ambient light only. So it's not the laser.
  • I tried sending a sinusoid into one of the same ADC channels used to acquire the QPD signals. I couldn't see any glitch

So one might conclude that the glictches are produced by the analog QPD electronics. However, I plugged in a scope and I couldn't see any in the analog signal. But I checked only before the DRV135 stages. I'll need some sort of breakout board to test the output of the DRV135.

Quote:

The glitches I saw in the data happens roughly every second, even though not exactly on the second. They are suddend jumps on the signal values over one sample, so of clear digital origin

 

 

  56   Thu Jul 21 12:10:14 2016 GabrieleGeneralDaily ProgressGlitch hunting

Yesterday I could clearly see the glitches as jumps in the time domain plot of X and Y signals, and trace them to somewhat harder to see jumps in the quadrant signals. 

  • I tried to catch them with an oscilloscope looking at the analog signals, but couldn't see any. However, for a brief period I had a persistent square wave oscillation in all transimpedance stages
  • I hooked the single ended output of the whitening filter to an ADC, and I could see the jumps there too. So I can exclude it's a problem of the differential drivers

One suspect was an intermittend oscillation of the transimpendance amplifier, so I looked into the schematics (D1600196) to see what could be the optimal value of the compensating capacitor C7. Following some useful notes online I computed the optimal value of C7 to be close to 2pF (instead of 10pF). I used 30-35 pF as the QPD capacitance, and 10 MHz has the gain-bandwidth product of the opamp. I swapped all 10pF capacitors with 2pF. After this I can still see the glitches in the spectra, but I can't find them anymore in the time domain. So things seem to have improved, although I still have annoying glitches.

Rich suggested to test the stability of the transimpendance stage by driving the output with a square wave and looking for the signal ringing. Here's his note:

I tried this for both the TI stage and the whitening stage, using 1k and 1uF and a square wave at 10 kHz. Here are the results, which look reasonable to me (firts is the TI, ringing at about 0.5 MHz, second is the whitening, almost no ringing):

So now I'm quite confident that the electronics is working. In the first trace you can see some intermittend background noise, due to the ambient light leaking into the QPD.

More investigations will follow.

  57   Thu Jul 21 14:58:26 2016 GabrieleElectronicsGeneralGlitches are of digital origin

So here's the final proof that the glitches I see are digital:

  • There is a positive jump (of random size) every second, every time 0.45s after the beginning of the second. There is a negative jump at a less constant time, but between 0.65s and 0.8s after the beginning of each second
  • I moved my whole setup to the crackle lab. This included: HeNe laser with power supply, QPD with cable, interface board and power supply. I don't see any glitch there

I swapped the ADC board with a second one in the new cymac, but no change: glitches are still there.

 

  58   Fri Jul 22 14:22:10 2016 GabrieleElectronicsConfigurationInvestigations on the signal jumps

[Rolf, Ben, Rich, Gabriele]

Rolf couldn't find any good explanation on the software side for the signals jumps. He investigated a bit the reason why IOP takes a long execution time, without success. It's still mysterious why it ran with low time for a while.

One effect of Rolf activity is that now the signal jumps happen at ~0.8 seconds (after the start of each second) instead that ~0.45. This is suspiciously pointing to a software issue...

Ben and I spent a few hours trying to better understand the origin of the glitches. Finally, we plugged in a function generator to the four ADC channels, and we could find the glitches again, at the expected time. So we could rule out completely that it's a problem of the QPD analog electronics.

Some more ivnestigations:

  • swapped the SR DS345 with an Agilent 33210A for the timing signal. No change: jumps are still there
  • rebooted and powered off the cymac many time, this never changed the glitches or the position in time
  • swapped the ADC interface board (D1600196) with a new one similar to that of the crackle lab (D1500402): no change, jumps are still there
  • switched off the roughin pump to reduce ambient acoustic noise. Not surprisingly, no change, still jumps
  • according to the motherboad manual, PCIe slots 4 and 6 are PCI-EX16, while the other are PCI-EX8. I moved the ADC and DAC to those those slots. Guess what? No change, still jumps

On a bright side, Rolf recompiled the awgman software, and now excitation channels are working.

  59   Mon Jul 25 12:24:06 2016 GabrieleElectronicsConfigurationMore on the signal jumps

Here are two more tests I did this morning

  • removed the DAC card from the cymac and the models, and restarted everything: signal jumps are still there
  • move the cymac2 from the crackling noise lab to the CRIME lab, ans set it up with all the same hardware I was using with the cymac3: there are no jumps, signals look perfect (see plots at the end)

So in conclusion: the problem is in the cymac3 computer, either software or hardware. I tend to excluse an I/O hardware problem, since I used two different ADCs and removed the DAC, without improving the situation.

  60   Mon Jul 25 15:35:31 2016 GabrieleOpticsCharacterizationCalibration of the QPD in physical units

The equations below are wrong. Please refer to https://nodus.ligo.caltech.edu:8081/CRIME_Lab/116 for the correct results

I measured the properties of the beam on the QPD. The total power is 31 uW. The beam shape is not gaussian, since we are seeing the interference of the reflection from the two surfaces:

The X and Y diameters are 1400 and 1300 microns, so I take the average of the two as an estimate of the beam size: 1300 +- 100 um. I also estimated the lever arm length to be 1.03 +- 0.02 m.

This allows me to esitmate the response of the normalized QPD signal to a tilt of the disk surface:

\frac{s}{P_0}=\frac{4}{\pi w^2} \cdot 2L \cdot \theta_{disk}

Plugging in the numbers gives a gain of (1900 +- 300) /rad for the normalized signals. I implemented those numbers in the filter banks: now X_NORM and Y_NORM have units of radians, and measure the disk surface angular motion. I also calibrated the SUM channel in microwatts, using the nominal responsivity of 0.45 A/W and the transimpedance of 200k (gain 11.1 uW/V)

Here's teh calibrated spectrum: note that the background noise is much larger than the real one because of the signal jumps.

Attachment 1: crime_beam.PNG
crime_beam.PNG
Attachment 2: crime_beam.PNG
crime_beam.PNG
  61   Mon Jul 25 17:44:13 2016 GabrieleElectronicsCharacterizationHigh frequency noise budget

I measured the noise sources limiting the QPD sensitivity. Unfortunately, I had to do some MATLAB tricks to get rid of the glitches: basically I load the data directly from the raw frames (NDS access to data is not working yet) and remove all jumps in the signals that happen in one single sample and are larger than a manually tuned threshold. This is not perfect, but it's enough to give us a rough idea of the spectrum of the QPD signal. The following plot shows the QPD_X signal (in units of disk motion, radians) in a few situations:

  • Blue: normal (laser on, room lights on)
  • Orange: laser on, room lights off (including the vacuum gauge)
  • Yellow: laser and room light off
  • Purple: same as above, but I switched off the PC monitors too
  • Green: QPD electronics off (this is the ADC noise)

The total power on the QPD is 30 uW, which correspond to a shot noise limited sensitivity of 4.3e-12 W/rHz. Considering that the signal is the quadrant asymmetry normalized by the total power, the shot noise limited sensitivity is sqrt(2) * SN / Power which once calibrated corresponds to 1.1e-10 rad/rHz.

The following plot shows that shot noise is the dominant source, followed closely by the electronics dark noise. The total agrees perfectly with the measured background noise above 2 kHz. Below that we have some leakage due to the large turbopump peak: this is due to FFT limitations but mostly to unsuppressed glitches.

From the QPD datasheet (Hamamatsu S5981) I learn that the noise equivalent power should be of the order of 2e-14 W/rHz at the sensitivity peak, so probably a factor of two or so worse at the HeNe frequency. It's still much lower than the measured dark noise. 

This sensitivity is already pretty good, but we can improve it by increasing the power on the diode. Indeed, 30 uW corresponds to about 2.7 V after the transimpedance, so we could increase the power by a factor 4 and win a factor 2 in the shot noise to dark noise ratio. Probably not worth it, since it will give us only a 30% gain in high frequency noise.

  62   Tue Jul 26 11:45:03 2016 GabrieleElectronicsCharacterizationHigh frequency noise budget

Just to confirm that my noise estimates make sense, here's a plot of the not-normalized QPD signal that gives the X motion (sum and difference of all four quadrants):

This is the signal after compensating for the whitening filter. If I remove this compensation, the following plot gives the noises in terms of the voltage directly in input to the ADC (or in output of the analog board):

So the total "dark" electronic noise is about 13 uV/rHz.

I did a roughly estimate of the sources of electronic noise:

  • QPD dark current noise, from datasheet, at the peak sensitivity is equivalent to 2e-14 W/rHz, or 2 nV/rHz at the output of the TI stage
  • First stage: Johnson-Nyquist noise of the TI resistor: 58 nV/rHz
  • First stage: output voltage noise of the LT1124: 3 nV/rHz
  • First stage: input curent noise of the LT1124, converted to the output: 60 nV/rHz

So the total noise at the outoput of the first stage is about 84 nV/rHz. The second stage adds a gain of 30 at high frequency, and negligible noise. So at the output of the whitening we have 2.5 uV/rHz. The DRV135 adds another gain of 2 and a neglegible output noise. 

So the total electronic noise at the output of each quadrant is 5 uV/rHz. Since we are combining four of them, the total expected electronic noise is 10 uV/rHz, which is not too far from the measured value. 

We are basically dominated equally by the Johnson-Nyquist noise of the TI resistor and by the input current noise of the LT1124. No gain to be obtained by changing the whitening.

Quote:

I measured the noise sources limiting the QPD sensitivity. Unfortunately, I had to do some MATLAB tricks to get rid of the glitches: basically I load the data directly from the raw frames (NDS access to data is not working yet) and remove all jumps in the signals that happen in one single sample and are larger than a manually tuned threshold. This is not perfect, but it's enough to give us a rough idea of the spectrum of the QPD signal. The following plot shows the QPD_X signal (in units of disk motion, radians) in a few situations:

  • Blue: normal (laser on, room lights on)
  • Orange: laser on, room lights off (including the vacuum gauge)
  • Yellow: laser and room light off
  • Purple: same as above, but I switched off the PC monitors too
  • Green: QPD electronics off (this is the ADC noise)

The total power on the QPD is 30 uW, which correspond to a shot noise limited sensitivity of 4.3e-12 W/rHz. Considering that the signal is the quadrant asymmetry normalized by the total power, the shot noise limited sensitivity is sqrt(2) * SN / Power which once calibrated corresponds to 1.1e-10 rad/rHz.

The following plot shows that shot noise is the dominant source, followed closely by the electronics dark noise. The total agrees perfectly with the measured background noise above 2 kHz. Below that we have some leakage due to the large turbopump peak: this is due to FFT limitations but mostly to unsuppressed glitches.

From the QPD datasheet (Hamamatsu S5981) I learn that the noise equivalent power should be of the order of 2e-14 W/rHz at the sensitivity peak, so probably a factor of two or so worse at the HeNe frequency. It's still much lower than the measured dark noise. 

This sensitivity is already pretty good, but we can improve it by increasing the power on the diode. Indeed, 30 uW corresponds to about 2.7 V after the transimpedance, so we could increase the power by a factor 4 and win a factor 2 in the shot noise to dark noise ratio. Probably not worth it, since it will give us only a 30% gain in high frequency noise.

 

Attachment 1: noise_budget.png
noise_budget.png
  63   Wed Jul 27 08:59:12 2016 GabrieleElectronicsGeneralPressure now in torr

The old channel X3:CR1-PRESSURE_LOGTORR does not exist anymore. The new channel is now directly in torr and it is called X3:CR1-PRESSURE_TORR.

I had to write a C function to compute the 10^x operation, since it is not included in the RCG routines. Also it's not possible to include library functions, so I had to write an ad-hoc function, which first compute the integer part of the exponent, and then approximate the fractional part with linear interpolation and a look-up table. Code is in /opt/rtcds/userapps/release/models/pow10.c

  64   Wed Jul 27 10:19:36 2016 GabrieleElectronicsConfigurationLast finding in the signal jump saga

This plot shows a signal which is generated purely digitally inside a code that I was testing.

Jumps are present even in this case.

Another addition: I was testing some simple code I was writing. Forget about OUT1 and OUT2, but OUT3 and OUT4 should be simple counters. Here's what I get by running the code on cymac3:

Since it didn't make any sense, I ran the same code on cymac2. Here's the (correct) result:

So there's something wrong even when all-digital signals are generated.

  65   Wed Jul 27 11:27:11 2016 GabrieleFacilityGeneralPumps stopped and venting

11:25am LT: closed valve between roughing and turbo pumps, switched off both pumps. Turbo pump is slowing down

After lunch I opened the chamber and removed everything from the inside.

The chamber around the vacuum gauge is really dirty now, see picture:

In addition, the electrostatic driver shows some signs of "burn" even though it was still working quite well. Unfortunately, whatever happened contaminated our sample:

  66   Wed Jul 27 15:43:02 2016 GabrieleFacilityDaily ProgressVacuum chamber dismantled and ready for cleaning

[Alena, Gabriele]

As decided at out Red Door meeting, we're going to clean the vacuum chamber and move it to the large table, which will be enclosed in a clean room.

So today we disassembled and packaged the vacuum chamber, which is now ready to be trasnspoted to be cleaned and baked.

  69   Thu Jul 28 17:15:22 2016 GabrieleElectronicsDaily ProgressADC/DAC interfaces upgrade

Installed the ADC and DAC boards into a proper box. Also, swapped the temporary DAC board (with cale hack) with the final one. Schematics and PCB are in the DCC: D1600196 and D1600301

The box is sitting on top of the cymac computer, on the back, since I don't have any long cable to connect the ADC. 

  70   Fri Jul 29 14:08:31 2016 GabrieleGeneralDaily ProgressPeak tracking code for real time model

Summary

I wrote a C function to reconstruct the amlplitude and frequency of a line. It can be added as a block into a real time monitor. The idea is to use it to track in real time the frequency and amplitude of the disk modes, during the ring down. I did some tests and finally managed to get the function to compile and run on the cymac2 (the crackling lab cymac).

The following plot shows a simulation, since I can't run the code on the new cymac and I don't have the disk installed anymore. The top panel gives the amplitude of a decaying line, and the bottom panel the frequency offset from a reference local oscillator (more below). The nominal values are an initial amplitude of 1, frequency of 1109.375 Hz to be compared to a 1109.0 local oscillator. The fitted decay time is 10.005 seconds, to be comapred with 10 seconds nominal. There is some additive gaussian noise, that causes the ring down to be unmesurable after about 70 seconds of data.

This code will be used for real time estimation of the disk modes, once theot frequency has been roughly estimated with FFTs. The estimation of the frequency work remarkably well. In the first 20 seconds the mean value is 0.3747 Hz, with a standard deviation of 1.5 mHz. When the SNR gets worse (between t=30s and t=50s) the mean value is 0.3745 with a standard deviation of 20 mHz. 

Because of the way it's built (see below) the code is sensitive to DC offsets, so the input signal must be high-passed.  

Details on the implementation

The code is based on demodulation of the input signal with a reference local oscillator thta must have a frequency as close as possible to the line we want to track.  The inputs to the block are: the signal to be monitored, sine of the local oscillator, cosine of the local oscillator. The outputs are: amplitude squared of the peak, frequency offset in Hz from the local oscillator.

Here's the math. Let's assume that the signal is

x(t)=A \sin (2\pi f t + \phi)

and the local oscillator has a frequency f0:

s(t)= \sin (2\pi f_0 t) \qquad c(t)= \cos (2\pi f_0 t)

The code multiply the signal by the two local oscillators and average the result over 65kHz / 8 Hz samples. Therefore we get two output streams at 8 Hz which are

\left< x(t) s(t) \right> = -\frac{A}{2} \sin\left[2\pi(f-f_0)t + \phi \right]

\left< x(t) c(t) \right> = \frac{A}{2} \sin\left[2\pi(f-f_0)t + \phi \right]

Then the sum of the two squared 8 Hz streams give an estimate of the amplitude squared. The code computes this every second

\hat A = 4 \left[ \left<x(t) s(t)\right>^2+\left<x(t) c(t)\right>^2\right]

while the arctangent of the ratio gives a phase that varies linearly with time.

\arctan \frac{\left< x(t) c(t)\right>}{\left< x(t) s(t)\right>} = \phi + 2\pi(f-f_0)t

For each of the 8Hz samples the code computes the arctangent (using a home-brewed function based on a lookup table, since we can't import math.h in the RCG). It unwraps it, and then every second fit a line to the unwrapped arctangent, to estimate the frequency offset with respect to the local oscillator.

The C function has some parameters hard coded: the main sampling frequency (65536 Hz), the number of points per second to use for the frequency estimation (8 Hz), the fact that the output is computed every second. The first two parameters can be changed, the third one cannot for the moment being.

Attachment 2: Screen_Shot_2016-07-29_at_2.02.24_PM.png
Screen_Shot_2016-07-29_at_2.02.24_PM.png
  71   Mon Aug 1 16:56:57 2016 GabrieleElectronicsDaily ProgressPopulated four new QPD boards

Today I populated four more QPD boards. All components are installed, except for the QPD themselves.

  72   Tue Aug 2 15:58:12 2016 GabrieleElectronicsDaily ProgressSix QPD functional and tested

I finished populating the new four QPD boards, and fixed the first one I populated weeks ago. I tested all five new boards: the output of the transimpendance respond correctly to the ambient light; the output of the whitening also respond correctly and has increased high frequency noise; the differential driver stages are all functional and balanced.

In summary, we have six QPD circuits ready: serial 02 is installed into the box and it has been used for the previous tests. Serial number 01, 03, 04, 05, 06 are not yet into a box, but fully functional. Boxes are ready.

For testing purposed, I also built another ADC interface board: it's complete with the exception of the connector that goes to the ADC.

  73   Tue Aug 2 16:00:32 2016 GabrieleElectronicsCharacterizationMore on the signal jumps

So I’m sending a 1.123 Hz sinusoid into the ADC, and generating 3.123 Hz sines and cosines in the model. Frequencies are reasonably incommensurable with one second.

The plots below shows all three signals. In the top panels, I just separate the two segments in each second: the “normal” segment in blue, and the “jumped” segment in orange.
In the bottom panels, I took the jumped segments (orange) and shifted them “in the future” by exactly one second. The signals are perfect!
 
I think we could explain this if the order of the samples in each second is scrambled, like we have a circular buffer but we point to the wrong starting point…
 
 
  74   Mon Aug 8 10:42:46 2016 GabrieleGeneralGeneralDependency of eigenmode frequencies on thickness and diameter

I ran a set of COMSOL simulations to determine the dependency of the frequency of each eigenmode on the disk thickness and diameter, within the tolerances. I chose wide ranges: diameter 75.0 +- 0.1 mm and thickness 1.0 +- 0.1 mm, much more than the expected tolerances. It turns out that the frequencies depends almost exactly linearly on both variables: mostly on the thickness and negligibly on the diameter. The following plots shows: the mode shape and frequency (left), the frequency dependency on the two variables (center), the residual of a linear fit and the functional form of the fit (right).

I'm including only the modes that will be measurable by our system (no motion in the center, frequency below 32kHz. Since the disks in my simulation is completely round, I'm showing only one mode for each doublet.

Attachment 19: mode_24159.png
mode_24159.png
Attachment 24: mode_28832.png
mode_28832.png
  75   Wed Aug 10 09:39:38 2016 GabrieleElectronicsDaily ProgressAnnuntio vobis gaudium magnum: habemus Cymac

Yesterday I cloned the cymac2 disk and installed it into the cymac3.

Jamie tweaked a few things (I can't really give more details) and now cymac3 is up an running with the same software as cymac2.

I compiled and installed the CR1 model, to readout the QPD. No more jumps in the signals!

To be able to access testpoints and have AWG working I had to follow the hack explained here: Cryo_Lab/781

I tested the following features:

  • the model is running and signals look good
  • I can access test points and DQ channels in real time with both dataviewer and DTT
  • I can access old data with dataviewer, using NDS
  • awggui is working and I can inject noise, that is properly going to the DAC channel (I check with a scope in the analog world)
  • I can upload filters using foton

The plot below shows the spectrum of the QPD. For the moment being I'm just sending a straight HeNe beam into the QPD, since the test setup with the disk is no more available. Units are arbitrary

  76   Wed Aug 10 10:04:35 2016 GabrieleMechanicsDaily ProgressThe prototype of the disk retain system is here

Yesterday we received the prototype of the disk suspension and retain system. Everything looks good. I checked that the disk fits in the holder, and all dimensions are good. The coil holders are out for winding, so I couldn't test the movimentation yet.

Attachment 3: 2016-08-09_14.18.55.jpg
2016-08-09_14.18.55.jpg
Attachment 4: 2016-08-09_14.19.02.jpg
2016-08-09_14.19.02.jpg
  77   Thu Aug 11 08:45:19 2016 GabrieleGeneralGeneralQPD signals with dielectric filter

I installed a dielectric laser line filter (Thorlabs FLH05633-5, center wavelength 633 nm, FWHM 5 nm) in front of the QPD. In this way we are no more affected by the room light. In the plot below blue is without filter, red with filter. A lot of peaks at high frequency are eliminated by the filter.

The plot below shows the QPD signal quadrant signals in a few different configurations: blue with the room light on and laser off, red with the room light off and laser off, green with the laser on. With the filter installed, when the laser is on we are dominated by its intensity noise, which shows a lot of peaks at high frequency. Those peaks are not completely eliminated by the difference of the quadrants.

  78   Thu Aug 11 09:10:01 2016 GabrieleElectronicsConfigurationReal time model

Compiled and installed an updated real time model. It acquires the four QPD segments, compute sum and differences, and normalize. I also added three experimental peak tracking components, that will be used to track in real time the amplitude and frequency of the disk modes.

I also created a summary medm screen that can be used to control all the QPD readout. It can be opened from any terminal with the command 'cr1'. I haven't added the peak tracking parts yet.

The X_NORM_OUT and Y_NORM_OUT signals are high passed at 500 Hz, to get rid of low frequency stuff that can affect the peak tracking.

 

  79   Thu Aug 11 10:59:39 2016 GabrieleElectronicsConfigurationQuadrant calibration

I cross checked the calibration of the sum channel with a  power meter. Now all quadnant signals X3:Q?_OUT and the sum signal X3:CR1-SUM_OUT are correctly calibrated in microwatts.

  80   Fri Aug 12 10:33:14 2016 GabrieleElectronicsConfigurationStorage added to cymac3

Installed two new 2TB disks into the cymac3. Also, the main disk has a 1TB partition with the operating system, so I created a new 1TB partition. I created a logic volume that spans the three partitions, for a total of about 5TB. This partition is mounted in /mnt/data and linked to the /frames folder. Frames are written to this new logic volume.

  81   Sat Aug 13 11:17:17 2016 GabrieleFacilityDaily ProgressSpot the differences

  82   Sat Aug 13 14:05:02 2016 GabrieleGeneralDaily ProgressFirst work on the measurement user interface

Today I started programming part of the user interface that will be used to perform the measurements. Not much implemented so far, but you can get an idea of the look:

Buttons on the left sidebar will allow the user to perform some basic tasks. The main panel has a plot (which will show spectra or ring down measurements) and a log section.

Attachment 1: Screenshot-C.Ri.Me..png
Screenshot-C.Ri.Me..png
  83   Sun Aug 14 11:40:29 2016 GabrieleElectronicsDaily ProgressSome python code

Today I wrote some auxiliary functions that will be useful for the measurement system:

  • noise.py: set of functions to generate band-limited noise (using inverse FFT) and multiple band noise. Using the awg python interface I can also start and stop the noise injection. Some examples of the result:

     
  • readdata.py: read data online, compute a spectrum, and additionally a function that find peaks in a whitened spectrum. All peaks above a minimum SNR are returned: the central frequency is computed by an average of the bin SNR:

     
  • diskmodel.py: reads a list of mode frequencies from txt files. In each file, the first row is the disk diameter in mm, the second is the disk thickness in mm, all other rows are the modes are computed by COMSOL. I also produced a whole bunch of such files, with diameters within 75 +- 0.1 mm and thickness within 1 +- 0.1 mm
  84   Sun Aug 14 18:13:52 2016 GabrieleElectronicsDaily ProgressProgress on the user interface

Some progress on writing the user interface:

Now the user can open a file that defines the nominal mode frequencies (from COMSOL simulations) and select which modes to search for:

The plan is that the automation will then inject band limited noise around each nominal frequency, to excite the mode, and then find the exact frequency. The user can set some additional parameters like

  • the duration of the excitation (I plan to use this as a maximum: if the mode rings up with large enough SNR after a shorter time, the injection will stop earlier)
  • the amplitude of the excitation (units to be determined)
  • the bandwitdh, otr in other words how many Hz the noise will be spread around the nominla frequency to search for the mode.
  • the minimum SNR to detect a peak (here the automation will get a spectrum before any excitation, to be used as the baseline to whiten all further spectra)
  85   Sun Aug 14 18:26:19 2016 GabrieleGeneralGeneralPlans for the automatic mode search

Here's how I imagine the mode search to proceed:

  1. the user select the list of nominal frequencies and the frequency range for the search
  2. the automation creates an excitation which is a sum of band-limited white noise, centered around each of the modes, with the desired bandwidth. The excitation is normalized to match a reasonable fraction of the DAC range
  3. the excitation is injected for a given amount of time and each band is searched for the mode. In each band, if a mode is found, the frequency is estimated together with the maximum SNR obtained and the results are saved
  4. if some modes are not found, the automation will try to improve the noise, by scaling all band amplitudes in such a way to have the same resulting SNR for the modes. In this way we should get more dynamical range and could increase the excitation for the modes that were not found
  5. if some modes are still unidentified, then the user will be asked to choose between a few options:
    1. skip the mode
    2. try to excite it with an excitation focused on that band only
    3. increase the bandwidth for the search
    4. maintain the same bandwidth but move the band to adjacent frequencies
  6. At the end all identified modes will be stored in memory and saved to a file defined by the user
  87   Tue Aug 16 11:57:03 2016 GabrieleElectronicsConfigurationFixed cymac3 network and time issue

The cymac3 internal clock was off by about 10 seconds. When I tried to start the NTP service, I found out that the cymac3 couldn't reach any external server. It turned out that the gateway in /etc/network/interfaces was set to the wrong address. I fixed it and rebooted. Now NTP is working and the time is correct.

This fixed a small issue with diaggui, which always complained about a data receiving error when starting a measurement (although after the complian the measurement could continue)

  91   Fri Aug 19 19:05:01 2016 GabrieleOpticsDaily ProgressDisk installed into the chamber

I assembled the disk suspension sytem and installed into the chamber. Although I don't have the magnets and coils, I installed the movable retaining disk, and used it to center the disk.

I first aligned the input laser using the reflection off the black glass, which turns out to be quite bright and very well visible. Tomorrow I'm going to measure how much power we have in the black glass.

The reflection from the disk is slighlty separated from the reflection from the black glass, so I can block it using an iris.

At 6:50pm I closed the chamber and started the roughing pump. At 7:05pm pressure was below 1 Tor so I started the turbo pump. When leaving pressure is about 1.6e-5 Tor.

  92   Sat Aug 20 17:01:10 2016 GabrieleGeneralDaily ProgressFirst ring down measurements

Everything is working pretty well. This morning the pressure was about 1.2e-6 Torr. I connected the high voltage amplifier and I could drive the disk without problems.

I measured the beam shape and size at the QPD. We have about 50 uW, we see a TEM01-like mode due to the interference of the two disk surfaces (this is normal). The beam is about 3 mm in diameter. using this information and the estimated optical lever length of 1.2 m, I calbrated the QPD NORM signals in units of angular motion of the disk surface. The computation posted in CRIME_Lab/60 is actually wrong. I'll post the correct one later.

Injecting broadband white noise I could excite all the modes that are visible up to about 30kHz. I tuned the COMSOL model, by changing the thickess of the disk to 1.017 mm, to fit the frequency of the first few modes. Here are the modes I could measure:

Nominal frequency [Hz] Actual frequency [Hz]
1108 1109
2543 2543
4427 4424
6748 6739
6792 6781
9494 9475
10239 10216
12656 12623
14209 14168
16150 16107
16226 16173
18673 18606
20195 20116
21414 21338
23611 23510
24559 24708
27203 27084
29006 28859
29180 29053
29309 29151

Take a look at the attached PDF file for the shape of all the modes, including all that are not visible. We see all the modes we expect to be able to excite with the central suspension of the disk.

The roughing pump is making a lot of non stationary low frequency noise. I turned it off, and the pressure stayed constant at 1.2e-6 Torr over about 1.5 hours. Here's the difference in the QPD spectrum:

It turned out that I have enough excitation authority to knock the disk out of the right place. So I had to vent to recover the situation. I'll open the chamber tomorrow and see what happened.

Here's a first bird eye look at the ring downs. We see beating of the two almost denegerate modes in some cases. Fits will follow, using the procedure I used for the LMA measurements.

Attachment 1: modes.pdf
modes.pdf modes.pdf modes.pdf modes.pdf modes.pdf modes.pdf modes.pdf modes.pdf
  93   Sat Aug 20 22:18:26 2016 GabrieleGeneralDaily ProgressFirst measured Qs

Using the first ring down of the day (GPS 1155754513 + 3600 seconds), I computed the amplitude of each of the modes already identified, using a short FFT spectrogram (each FFT is 1 second long, overlap of 0.5 s).

Then I used the same code I developed at LMA to fit the ring down, including the beat between the unresolved mode pairs. The fit is versy sensitive to the initial conditions, so I had to fine tune them for each of the 20 modes. Still, all fits were successful with 30 minutes of work.

Here are all the fits:

And in summary all the measured Qs, which turned out to be larger than what I was expecting, considering that the disk is not annealed.

The analysis code in MATLAB is attached.

Attachment 22: first_ringdown.m
%% Compare data right before and right after the excitation
gps0 = 1155754000;
gps1 = 1155754513;
dt = 180;

c = nds2.connection('cymac3.ligo.caltech.edu', 8088);
data0 = c.fetch(gps0, gps0+dt, {'X3:CR1-X_NORM_OUT_DQ', 'X3:CR1-Y_NORM_OUT_DQ', 'X3:CR1-ESD_OUT_DQ'});
data1 = c.fetch(gps1, gps1+dt, {'X3:CR1-X_NORM_OUT_DQ', 'X3:CR1-Y_NORM_OUT_DQ', 'X3:CR1-ESD_OUT_DQ'});

x0 = data0(1).getData();
... 606 more lines ...
  94   Sun Aug 21 08:51:22 2016 GabrieleGeneralDaily ProgressNew disk installed

Vented the chamber. Installed a new disk (MO 03). The one I measured yesterday is now named MO 02 and it is the one with the "burnt mark" from the previous experiment (due to the electrostatic drive).

Startep roughing pump at 8:50am. Started turbo pump at 9.00am.

Excited the disk at 9:50:30am with white noise, amplitude 10 V. Pumps are still running, pressure is about 2e-6 Torr

At 11:10am I stopped the roughing pump, pressure is 1.4e-6 Torr. Exciting again the disk at 11:10:45am. At 12:25am I checked again the situation, since both measurements look quite weird, especially for the first couple of modes.

I think the reflection from the balck glass is interfering with the reflection from the disk. Probably I wasn't careful enough when I aligned the disk. At about 12:35 I stopped the turbo pump. I'm going to open the chamber and realign everything again.

  95   Sun Aug 21 13:45:56 2016 GabrieleOpticsDaily ProgressWierd behavior of ring downs and Improved setup

The last two ring downs I measured today showed a weird behavior of the lowest modes:

Although I'm not 100% sure, I suspect this is related to the fact that the beam reflected from the black glass was so close to the beam reflected by the disk that I could see interference.

So I broke vacuum and improved the setup, adding a peek washer below one edge of the black glass, to wedge it. In this way the reflection from the black glass is largely separated: it misses the upper periscope mirror and it is dumped on a black panel (together with the viewport reflection).

I realigned everything, installed back the disk and started pumping down at 1:30pm.

  96   Sun Aug 21 15:36:16 2016 GabrieleGeneralDaily ProgressNew test

Roughing pump stopped at about 3:30:30pm. HV amplifier on at 3:33:30pm, excitation at 3:35:30pm. Recentered QPD at 3:36pm

  97   Sun Aug 21 17:16:03 2016 GabrieleGeneralDaily ProgressRing down measurement with new disk

After fixing the setup, the measurement with the new disk looks great. After more than one hour the first two modes are still ringing down, meaning that the Q's are larger than 10 millions.

Here's the comparison of the spectrum before and after the excitation, with the identified modes:

I used about 4200 s of data to fit the ring downs. Most fits are good. In a couple of cases the peak splitting is large and the algorithm fails to fit the beats:

In summary, here are the Q values for all modes. Despite not being annealed, this disk shows very large Q's

  98   Sun Aug 21 17:41:34 2016 GabrieleGeneralDaily ProgressNew test

I checked the status at about 5:20pm, the turbo pump was in error and spinning down, since the roughing pump has been off for about 1.5 hours.

I let the pump switch off.

Quote:

Roughing pump stopped at about 3:30:30pm. HV amplifier on at 3:33:30pm, excitation at 3:35:30pm. Recentered QPD at 3:36pm

 

  99   Mon Aug 22 08:29:24 2016 GabrieleGeneralDaily ProgressNew test

Restarted roughing and turbo pump at about 8:10am.

Quote:

I checked the status at about 5:20pm, the turbo pump was in error and spinning down, since the roughing pump has been off for about 1.5 hours.

I let the pump switch off.

Quote:

Roughing pump stopped at about 3:30:30pm. HV amplifier on at 3:33:30pm, excitation at 3:35:30pm. Recentered QPD at 3:36pm

 

 

  100   Mon Aug 22 12:01:52 2016 GabrieleGeneralGeneralTests of the day

Disk excited at 12:01pm. Exited the room at 12:03pm.

Opened the chamber at about 2:30pm, got the disk out for edge polishing, installed it back at 3:30pm, pumping down at 3:40pm.

Stopped the roughing pump at 4:44:00pm (+60 seconds clean data, GPS 1155944657). Switched on the HV amplifier, excitation at 4:47:30pm. Recentered QPD, clean data from 4:48:30pm (GPS 1155944927)

After a first look at the data, it seems that something went wrong. I restearted the roughing pump and will pump overnight.  I found the QPD miscentered, so I centered it again.

Excited again at about 5:46:35pm. Clean data from 1155948460

  102   Tue Aug 23 08:46:04 2016 GabrieleGeneralDaily ProgressDaily tests

MO03 - edge polished:

Turbo off, QPD centered, before excitation (60 seconds)

PDT: 2016-08-23 08:42:54.514987
PDT UTC: 2016-08-23 15:42:54.514987
UTC GPS: 1156002191.514987

Excitation (white uniform noise, amplitude 5 V)

PDT: 2016-08-23 08:45:01.007626 PDT
UTC: 2016-08-23 15:45:01.007626 UTC
GPS: 1156002318.007626

Clean data for ring-down

PDT: 2016-08-23 08:45:46.448949 PDT
UTC: 2016-08-23 15:45:46.448949 UTC
GPS: 1156002363.448949

Restarted roughing pump, QPD got misaligned

PDT: 2016-08-23 10:00:29.259345 PDT
UTC: 2016-08-23 17:00:29.259345 UTC
GPS: 1156006846.259345

MO03 - edge polished:

Band-limited noise, +-10Hz around eahc nominal frequency, amplitude scaled based on the inverse of the peak height obtained with white noise. See attached code and plot

from numpy import *
from noise import *


x = loadtxt('/home/controls/Measurements/2016_08_23/mo_02_laserpolished_frequencies.txt')

freqs = x[:,0]
ampl = x[:,1]

bw = 10


bands = map(lambda x: [x - bw, x + bw], freqs)
a = 1 / (ampl/max(ampl))
a[a>50] = 50.


x = multi_band_noise(bands, a, 10, fs=65536)
x = x / 30

Ring down after:

PDT: 2016-08-23 11:07:02.661145 PDT
UTC: 2016-08-23 18:07:02.661145 UTC
GPS: 1156010839.661145

 

  103   Tue Aug 23 09:21:25 2016 GabrieleGeneralGeneralStatus of Mark Optics disks

We initially received 20 disks (75 mm diameter, 1 mm thickness) from Mark Optics. Here's their status as of today

  • MO-01 was annelaed and measured at LMA. It was then taken back to Mark Optics by Julie Houser at the time of her visit
  • MO-02 was initially used to test the "dirty" measurement apparatus, then it was installed into the new chamber and measured on Saturday August 20th
    Then this disk was used as a guinea pig for the laser edge polishing, and it's now in a pretty bad shape
  • MO-03 was installed into the new chamber and measured a couple of times on Sunday and Monday. Then we laser polished the edges, with good results, although there are small damages on the surface. It has been installed back into the chamber and a couple of rind-down measurement carried out on Monday and Tuesday.
  • MO-04 is still untouched

All the other disks have been sent back to Mark Optics to grind out flats. 

  104   Tue Aug 23 09:44:11 2016 GabrieleGeneralGeneralMeasurements on MO-02

Two good ring-downs measurements were performed on MO-02. The first one was already reported in a previous elog entry. I performed another measurement, and refined the mode identification. I think I had misidentified some modes in my previous analysis. The following plot shows the difference between the modes as predicted by COMSOL and as measured. A clean quadratic trend is visible and fitted:

Here's the spectrum with all the modes:

And the updated Q measurement plot:

A second ring down was measured on Monday morning . Here are the relevant plots:

This is the same disk as before, but almost all Q values are systematically higher. Here's a direct comparison:

I'm not sure what changed between the two measurements, except for a re-alignment of the QPD. The disk might have moved a bit...

Attachment 4: spectrum_with_modes_12_01pm.png
spectrum_with_modes_12_01pm.png
  105   Thu Sep 1 02:13:33 2016 GabrieleOpticsCharacterizationExpected frequencies for the 75 mm disks with flats

Here are the nominal parameters of the disk with flats

Parameter Value
Diameter (nominal) 75.0 mm
Thickness (nominal) 1.00 mm
Distance of flats from center of disk
(as measured by MO)
36.05 mm
Young's modulus (from G1601850) 72.3 GPa
Poisson's ratio (from G1601850) 0.164
Density 2202 kg/m^3

A COMSOL simulation gives the frequencies and mode shapes shown in the attached PDF file. Following the list of frequencies and a classification of the mode family (numer of radial nodes, number of azimuthal nodes in a half turn):

Frequency [Hz] Radial Azimuthal
1089.7 0 2
1109.1 0 2
1661.1 1 0
2501.3 0 3
2542.3 0 3
3837.7 1 1
3909.8 1 1
4359.9 0 4
4421.2 0 4
6656 0 5
6665.3 1 2
6732.7 0 5
6780 1 2
7283.9 2 0
9381.5 0 6
9470.6 0 6
10054.1 1 3
10210.2 1 3
11244.2 2 1
11414.2 2 1
12528.4 0 7
12655.4 0 7
13970.1 1 4
14148.9 1 4
15854.8 2 2
15856.8 2 2
16101.1 0 8
16227.8 3 0
16853.9 3 0
18387.2 1 5
18562.6 1 5
19984.9 0 9
20045.8 0 9
21067.2 2 3
21358.1 2 3
22300.9 3 1
22848.8 3 1
23270.5 1 6
23419.4 1 6
24374.1 0 10
24425.2 0 10
26793.8 2 4
27083.9 2 4
28533.5 1 7
28637.8 1 7
28692 3 2
28998.5 4 0
29228.3 0 11
29271.9 0 11
29870.1 4 0
33040 2 5
33301.4 2 5
33949.1 0 12
33986.5 0 12
Attachment 1: allmodes_aggregated.pdf
allmodes_aggregated.pdf allmodes_aggregated.pdf allmodes_aggregated.pdf allmodes_aggregated.pdf allmodes_aggregated.pdf allmodes_aggregated.pdf
  106   Sun Sep 11 10:05:33 2016 GabrieleGeneralDaily ProgressNew substrate in the chamber

I installed one of the new substrates (with flats) into the chamber, and started the pumpdown at about 9:45am LT.

Before that, I removed the retaining ring: tomorrow I'm going to glue the magnet to it.

ELOG V3.1.3-