Message ID: 399     Entry time: Wed Aug 9 12:10:47 2017
Author: Zach 
Type: Electronics 
Category: Modeling 
Subject: Preliminary improvement from ESD optimization 

2017-08-09

  • I created a plot of the ratio of the force in the optimized design to the force in the original design. The improvement factor is huge, some modes are excited by more than a factor of 100. I took the same ratio keeping the gap between the ESD and the sample constant and it decreased the excitation by almost a factor of 10. Keeping that gap constant, the geometric modifications to the ESD give an improvement factor ranging from almost 2 to almost 4 for most of the modes. Modes 10 and 25 are outliers but in the original geometry they are barely excited at all, so this could easily be a numerical artifact where those modes were excited at a minimum in the original geometry.

Attachment 1: Ratio.jpg  13 kB  Uploaded Wed Aug 9 16:20:02 2017  | Hide | Hide all
Ratio.jpg