Subject:
Corrected sample gap sweep
2017-07-27
I resolved a couple more data processing bugs and calculated a sweep of the ESD-Sample gap from a distance of .5 mm to 1.5 mm. The resulting data behaves far more like I would expect from a force generated by an electric field, it seems to drop off like distance squared. This is a very strong correlation with a good intuitive explanation, and would suggest that it is prudent to place the ESD as close to the sample as possible.
I also computed a higher resolution sweep of the gap between the arms of the ESD. It did not resolve the strange behavior at all, so I will investigate coupling into the mode pairs as a possible source.