Message ID: 244     Entry time: Sat Dec 17 06:29:15 2016     In reply to: 237
Author: rana 
Type: Electronics 
Category: Configuration 
Subject: Timing issue: is it a DAC issue? 

I find sometimes that the probe configuration can give these distorted signals. For the Tektronix probes, its best to use a 500 MHz probe instead of the BNC clip leads. The probe also should be compensated by attaching to the gold fingers square wave generator on the scope front and adjusting the capacitor in the probe with a little screwdriver until the square wave becomes perfect.