Message ID: 187     Entry time: Mon Nov 14 14:41:08 2016
Author: Gabriele 
Type: General 
Category: Measurements 
Subject: S1600474 

2016-11-14

  • 2:40pm, in chamber, balanced
  • 2:41pm, roughing pump on
  • 2:50pm, turbo pump on
  • 4:12pm, excitation (2kV, 60s), quiet time before 1163203908, quiet time after 1163204078
  • More excitations:

    • Quiet time before excitation: 1163213574
      Excitation (1100 Hz) at 1163213607 (60 s)
      Excitation (broad band) at 1163213668 (60 s)
      Quiet time after excitation: 1163213730

    • Quiet time before excitation: 1163224560
      Excitation (1100 Hz) at 1163224592 (60 s)
      Excitation (broad band) at 1163224654 (60 s)
      Quiet time after excitation: 1163224716

    • Quiet time before excitation: 1163235546
      Excitation (1100 Hz) at 1163235578 (60 s)
      Excitation (broad band) at 1163235640 (60 s)
      Quiet time after excitation: 1163235702

2016-11-15

  • 8:49am pumps off

Results

2016-11-14 A

% Freq        Q                Q (C.I. 95%)        Q (C.I. 95%)
1107.7        4.6075e+06        4.6037e+06        4.6113e+06
2540.0        5.4810e+06        5.4785e+06        5.4836e+06
4490.2        2.5331e+05        2.5304e+05        2.5359e+05
6743.0        3.3684e+06        3.3662e+06        3.3707e+06
6746.9        3.3685e+06        3.3665e+06        3.3706e+06
6757.6        7.1350e+06        7.1255e+06        7.1445e+06
6772.0        5.8021e+06        5.4002e+06        6.2684e+06
6824.1        2.4899e+06        2.4871e+06        2.4928e+06
9501.9        5.4497e+06        5.4322e+06        5.4674e+06
10187.8        6.4223e+06        6.4184e+06        6.4262e+06
10343.0        2.3547e+06        2.3256e+06        2.3845e+06
12679.1        3.9818e+06        3.9685e+06        3.9952e+06
12806.0        3.2209e+05        3.1252e+05        3.3227e+05
14145.0        4.1585e+06        4.1033e+06        4.2152e+06
16050.7        2.2872e+06        2.2699e+06        2.3047e+06
16058.8        6.6646e+06        6.6544e+06        6.6748e+06
16281.1        2.3068e+06        2.2741e+06        2.3405e+06
18601.1        5.3760e+06        5.3475e+06        5.4049e+06
18776.0        5.7876e+05        5.7874e+05        5.7877e+05
21315.5        8.6998e+06        8.5404e+06        8.8652e+06
23520.3        3.4527e+06        3.4258e+06        3.4799e+06
24658.2        1.3013e+06        1.3012e+06        1.3013e+06
28810.3        6.4869e+06        6.4174e+06        6.5579e+06
28909.1        4.2011e+06        4.1313e+06        4.2734e+06
29001.9        8.7561e+06        8.7012e+06        8.8118e+06
29526.3        2.1542e+06        2.1311e+06        2.1777e+06
29563.5        5.9460e+06        5.8940e+06        5.9989e+06
30463.0        2.0306e+06        2.0306e+06        2.0307e+06
31307.8        2.0177e+06        2.0176e+06        2.0177e+06
31915.5        2.4119e+06        2.4118e+06        2.4119e+06
32178.4        6.3761e+06        6.2763e+06        6.4791e+06

 

2015-11-14 B

% Freq        Q                 Q (C.I. 95%)        Q (C.I. 95%)
1107.5        3.2516e+06        3.2494e+06        3.2538e+06
2539.9        4.5383e+06        4.5368e+06        4.5397e+06
4423.0        2.2248e+06        2.2238e+06        2.2257e+06
4490.2        2.6724e+05        2.6695e+05        2.6754e+05
6688.5        3.5492e+06        3.4597e+06        3.6434e+06
6746.8        3.5335e+06        3.5324e+06        3.5345e+06
6757.6        7.5849e+06        7.5835e+06        7.5864e+06
6824.1        2.4839e+06        2.4824e+06        2.4853e+06
9501.8        4.6405e+06        4.6330e+06        4.6480e+06
10187.7        6.1494e+06        6.1467e+06        6.1520e+06
10342.9        2.5426e+06        2.5181e+06        2.5676e+06
12679.0        3.3433e+06        3.3374e+06        3.3492e+06
14144.6        3.9980e+06        3.8729e+06        4.1316e+06
16050.5        1.9010e+06        1.8941e+06        1.9081e+06
16058.6        6.4936e+06        6.4881e+06        6.4991e+06
16281.1        2.7856e+06        2.7785e+06        2.7928e+06
18215.4        2.2829e+06        2.2828e+06        2.2829e+06
18600.9        7.3657e+06        7.3568e+06        7.3746e+06
18775.6        5.8072e+05        5.8070e+05        5.8073e+05
23520.1        1.7679e+06        1.7538e+06        1.7822e+06
24658.4        8.3031e+05        8.3030e+05        8.3033e+05
27082.0        6.7421e+06        6.5672e+06        6.9266e+06
28810.1        6.1339e+06        6.1000e+06        6.1682e+06
28909.0        4.3651e+06        4.3651e+06        4.3652e+06
29001.7        9.2108e+06        9.1518e+06        9.2705e+06
29526.1        2.0948e+06        2.0765e+06        2.1133e+06
30463.0        1.5875e+06        1.5875e+06        1.5876e+06
31308.0        1.9777e+06        1.9777e+06        1.9778e+06
31915.8        2.5156e+06        2.5156e+06        2.5157e+06

2015-11-14 C

% Freq        Q               Q (C.I. 95%)        Q (C.I. 95%)
1107.9        2.6382e+06        2.6360e+06        2.6404e+06
2539.9        4.3446e+06        4.3432e+06        4.3461e+06
4423.0        2.1544e+06        2.1534e+06        2.1553e+06
4490.2        2.7485e+05        2.7471e+05        2.7499e+05
6746.8        3.2717e+06        3.2705e+06        3.2728e+06
6757.6        6.3809e+06        6.3797e+06        6.3821e+06
6824.1        2.3227e+06        2.3215e+06        2.3238e+06
9501.8        4.6211e+06        4.6073e+06        4.6351e+06
10187.7        6.5099e+06        6.5090e+06        6.5109e+06
10342.9        2.6098e+06        2.5853e+06        2.6347e+06
12679.1        3.2362e+06        3.2276e+06        3.2449e+06
14144.6        4.1799e+06        4.1049e+06        4.2577e+06
16050.5        2.8054e+06        2.8002e+06        2.8106e+06
16058.6        8.1729e+06        8.1564e+06        8.1895e+06
16281.0        3.2400e+06        3.2399e+06        3.2401e+06
18600.9        7.8239e+06        7.7914e+06        7.8565e+06
18775.7        5.9520e+05        5.9518e+05        5.9522e+05
20255.0        1.0929e+06        1.0929e+06        1.0929e+06
21315.4        8.0568e+06        7.9081e+06        8.2112e+06
23520.0        7.7176e+05        7.7174e+05        7.7178e+05
24658.0        6.0199e+05        6.0197e+05        6.0200e+05
28810.0        6.1531e+06        6.0894e+06        6.2180e+06
29001.6        9.2904e+06        9.2285e+06        9.3531e+06
29526.1        2.0902e+06        2.0700e+06        2.1108e+06
31308.0        1.8177e+06        1.8177e+06        1.8177e+06
31916.0        2.4457e+06        2.4457e+06        2.4458e+06
32179.0        6.2208e+06        6.2207e+06        6.2209e+06

2016-11-14 D

% Freq        Q               Q (C.I. 95%)        Q (C.I. 95%)
1107.8        2.5483e+06        2.5470e+06        2.5497e+06
2539.9        4.1479e+06        4.1464e+06        4.1493e+06
4423.0        2.1043e+06        2.1036e+06        2.1050e+06
4490.2        2.8179e+05        2.8161e+05        2.8196e+05
6746.7        3.1144e+06        3.1133e+06        3.1155e+06
6757.5        5.8249e+06        5.8241e+06        5.8257e+06
6761.4        5.8283e+06        5.8275e+06        5.8292e+06
6824.0        2.2641e+06        2.2633e+06        2.2649e+06
9501.8        4.6535e+06        4.6342e+06        4.6730e+06
10187.7        7.0149e+06        7.0125e+06        7.0173e+06
10342.8        2.6508e+06        2.6469e+06        2.6548e+06
12679.0        3.2619e+06        3.2587e+06        3.2650e+06
14144.6        3.7736e+06        3.7205e+06        3.8281e+06
16050.4        2.7995e+06        2.7862e+06        2.8129e+06
16058.5        8.0705e+06        8.0662e+06        8.0747e+06
16280.9        3.3065e+06        3.3007e+06        3.3123e+06
18600.8        7.9582e+06        7.9432e+06        7.9733e+06
18775.6        5.8072e+05        5.7705e+05        5.8443e+05
20254.7        1.2456e+06        1.2295e+06        1.2622e+06
21315.0        8.2763e+06        8.0802e+06        8.4823e+06
23519.9        1.4508e+06        1.4356e+06        1.4663e+06
28810.0        6.1795e+06        6.0919e+06        6.2696e+06
29001.4        9.6787e+06        9.6439e+06        9.7138e+06
29525.7        2.0513e+06        2.0512e+06        2.0513e+06
29564.2        4.8725e+06        4.8325e+06        4.9133e+06
31308.0        1.8091e+06        1.8090e+06        1.8091e+06

 

 

 

Attachment 3: ringdowns.png  185 kB  Uploaded Tue Nov 15 09:11:45 2016  | Hide | Hide all
ringdowns.png
ELOG V3.1.3-