Message ID: 185     Entry time: Mon Nov 14 11:13:33 2016
Author: Gabriele 
Type: General 
Category: Measurements 
Subject: S1600467 

2016-11-14

Note: marked as "bad edge" by Alena

  • 11:00 in chamber, balanced
  • 11:01 roughing pump on
  • 11:13 turbo pump on
  • 12:28 excitation (2kV, 60s), quiet time before 1163190441, quiet time after 1163190685
  • 1:54pm, pumps off

Results

2016-11-14

% Freq        Q               Q (C.I. 95%)        Q (C.I. 95%)
1102.7        6.3619e+06        6.3431e+06        6.3807e+06
1122.2        1.8673e+06        1.8623e+06        1.8722e+06
2530.3        3.7678e+06        3.7639e+06        3.7717e+06
2569.2        1.4261e+06        1.4195e+06        1.4327e+06
4407.7        2.7665e+06        2.7429e+06        2.7904e+06
4470.7        2.6197e+05        2.6166e+05        2.6228e+05
6723.0        1.7620e+06        1.7567e+06        1.7674e+06
6734.0        6.9851e+06        6.9809e+06        6.9893e+06
6794.4        1.8494e+06        1.8476e+06        1.8513e+06
9470.5        6.0578e+04        6.0574e+04        6.0581e+04
10152.1        3.8492e+06        3.8370e+06        3.8615e+06
10301.4        4.0275e+05        4.0075e+05        4.0478e+05
12638.1        7.3205e+05        7.2391e+05        7.4036e+05
12756.2        7.5936e+05        7.5933e+05        7.5939e+05
14096.2        5.9662e+06        5.9355e+06        5.9971e+06
16001.5        1.7340e+06        1.7104e+06        1.7582e+06
16227.1        1.2612e+06        1.2585e+06        1.2639e+06
20142.2        5.4880e+05        5.4878e+05        5.4881e+05
20191.9        5.5732e+05        5.5731e+05        5.5734e+05
21236.3        2.6162e+06        2.6086e+06        2.6238e+06
22255.0        1.5152e+06        1.5152e+06        1.5152e+06
23445.0        1.0949e+06        1.0824e+06        1.1077e+06
24391.4        7.4968e+05        7.4967e+05        7.4970e+05
24545.8        1.8061e+05        1.8060e+05        1.8061e+05
24580.0        7.5466e+05        7.5465e+05        7.5468e+05
25561.4        1.2230e+06        1.2230e+06        1.2230e+06
26982.5        3.8512e+06        3.8511e+06        3.8512e+06
28730.1        1.6735e+06        1.6617e+06        1.6855e+06
28824.0        9.7775e+05        9.7774e+05        9.7777e+05
28896.2        4.3667e+06        4.3190e+06        4.4154e+06
31387.2        5.9592e+05        5.9592e+05        5.9593e+05
32301.5        6.3985e+05        6.3984e+05        6.3986e+05

 