40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  Coating Ring-down Measurement Lab elog  Not logged in ELOG logo
Message ID: 183     Entry time: Sun Nov 13 19:12:56 2016
Author: Gabriele 
Type: General 
Category: Measurements 
Subject: S1600471 

2016-11-13

Note: some scratches or contamination visible on the surface near the egde

  • 7:03pm, in chamber, balanced
  • 7:04pm, roughing pump on
  • 7:12pm turbo pump on
  • Excitations:
    • Quiet time before excitation: 1163133663
      Excitation (1100 Hz) at 1163133695 (60 s)
      Excitation (broad band) at 1163133757 (60 s)
      Quiet time after excitation: 1163133819
    • Quiet time before excitation: 1163144649
      Excitation (1100 Hz) at 1163144682 (60 s)
      Excitation (broad band) at 1163144744 (60 s)
      Quiet time after excitation: 1163144806
    • Quiet time before excitation: 1163155636
      Excitation (1100 Hz) at 1163155669 (60 s)
      Excitation (broad band) at 1163155731 (60 s)
      X3:CR1-ESD_GAIN => 0
      Quiet time after excitation: 1163155793

2016-11-14

  • 8:50am pumps off

Results

2016-11-13 A

% Freq        Q               Q (C.I. 95%)        Q (C.I. 95%)
1098.2        7.4598e+06        7.4472e+06        7.4724e+06
2196.1        8.9864e+06        8.5757e+06        9.4383e+06
2520.2        4.3443e+06        4.3400e+06        4.3486e+06
2558.1        6.0816e+05        6.0143e+05        6.1503e+05
4390.6        2.7832e+06        2.7819e+06        2.7844e+06
6698.1        2.4391e+06        2.4377e+06        2.4404e+06
6720.1        1.2186e+07        1.2174e+07        1.2198e+07
6771.3        1.2193e+06        1.2187e+06        1.2199e+06
6838.3        3.8835e+05        3.8533e+05        3.9143e+05
9434.8        2.0534e+06        2.0501e+06        2.0568e+06
10129.0        1.1046e+05        1.1046e+05        1.1047e+05
10275.4        7.3179e+05        7.2786e+05        7.3576e+05
12589.7        9.8392e+05        9.6718e+05        1.0012e+06
12709.8        3.4484e+05        3.4164e+05        3.4809e+05
14061.9        6.0359e+06        6.0310e+06        6.0408e+06
15955.7        1.8406e+06        1.8350e+06        1.8462e+06
15969.9        7.9895e+06        7.9796e+06        7.9995e+06
16167.0        1.1887e+06        1.1858e+06        1.1916e+06
17933.7        6.7847e+05        6.7845e+05        6.7849e+05
18348.0        1.1646e+06        1.1645e+06        1.1646e+06
18491.0        7.9603e+05        7.9601e+05        7.9605e+05
18658.5        9.0778e+04        9.0775e+04        9.0780e+04
20459.0        1.9219e+06        1.9218e+06        1.9219e+06
21191.2        3.1915e+06        3.1677e+06        3.2157e+06
23383.6        3.5399e+06        3.5198e+06        3.5602e+06
24489.7        9.6402e+05        9.5706e+05        9.7108e+05
24509.4        1.0474e+06        1.0474e+06        1.0474e+06
24553.5        1.4376e+06        1.4375e+06        1.4376e+06
26923.0        7.2079e+06        7.1783e+06        7.2378e+06
28653.0        2.5777e+06        2.5717e+06        2.5838e+06
28747.7        2.2403e+06        2.1968e+06        2.2855e+06
28842.1        1.1938e+07        1.1927e+07        1.1948e+07
29765.5        8.8519e+06        8.7781e+06        8.9269e+06
30712.0        1.6995e+06        1.6848e+06        1.7145e+06
31467.0        4.8235e+05        4.8235e+05        4.8236e+05
31489.0        1.1729e+06        1.1649e+06        1.1809e+06
32120.5        9.6230e+05        9.6229e+05        9.6232e+05

2016-11-13 B

% Freq        Q                Q (C.I. 95%)        Q (C.I. 95%)
1098.1        5.4817e+06        5.4778e+06        5.4856e+06
2520.1        2.7906e+06        2.7896e+06        2.7916e+06
2558.1        6.0809e+05        6.0570e+05        6.1050e+05
4390.6        1.7952e+06        1.7949e+06        1.7955e+06
6698.1        1.4580e+06        1.4576e+06        1.4584e+06
6720.0        8.2579e+06        8.2563e+06        8.2595e+06
6771.4        9.1260e+05        9.1091e+05        9.1429e+05
6838.4        3.8042e+05        3.7936e+05        3.8148e+05
9434.7        1.2099e+06        1.2081e+06        1.2116e+06
10275.5        6.7566e+05        6.7360e+05        6.7773e+05
12589.7        6.8681e+05        6.8318e+05        6.9047e+05
12709.8        2.8213e+05        2.7783e+05        2.8657e+05
14061.9        3.7724e+06        3.7706e+06        3.7741e+06
15955.8        1.2708e+06        1.2636e+06        1.2780e+06
15969.9        5.9870e+06        5.9818e+06        5.9922e+06
16167.0        8.0447e+05        7.9517e+05        8.1399e+05
21191.3        2.7333e+06        2.7051e+06        2.7620e+06
22345.0        2.3568e+06        2.3567e+06        2.3568e+06
23383.5        2.1083e+06        2.1082e+06        2.1083e+06
24489.5        6.6570e+05        6.6569e+05        6.6571e+05
24509.5        7.9636e+05        7.9634e+05        7.9638e+05
24553.5        8.3880e+05        8.3878e+05        8.3881e+05
26922.9        4.6187e+06        4.6004e+06        4.6372e+06
28652.9        1.4483e+06        1.4347e+06        1.4622e+06
28747.6        1.2629e+06        1.2629e+06        1.2629e+06
28842.1        9.5317e+06        9.5238e+06        9.5396e+06
29766.0        6.6922e+06        6.5397e+06        6.8520e+06
30712.1        1.0888e+06        1.0888e+06        1.0888e+06
31489.0        6.8125e+05        6.8124e+05        6.8126e+05
32120.5        7.8667e+05        7.8666e+05        7.8668e+05

2016-11-13 C

% Freq        Q        Q (C.I. 95%)        Q (C.I. 95%)
1098.1        5.4335e+06        5.4304e+06        5.4365e+06
2196.2        6.0782e+06        5.6658e+06        6.5552e+06
2486.0        3.0319e+06        2.6660e+06        3.5141e+06
2515.7        2.7819e+06        2.7810e+06        2.7828e+06
2520.2        2.7871e+06        2.7863e+06        2.7879e+06
2558.0        5.8364e+05        5.7358e+05        5.9405e+05
2578.0        2.7145e+06        2.4935e+06        2.9785e+06
3584.0        2.9087e+06        2.6452e+06        3.2303e+06
4390.5        1.7822e+06        1.7818e+06        1.7825e+06
6698.1        1.4479e+06        1.4472e+06        1.4487e+06
6720.1        8.2604e+06        8.2559e+06        8.2648e+06
6771.3        9.1201e+05        9.1142e+05        9.1260e+05
6838.4        3.8280e+05        3.8172e+05        3.8389e+05
9434.8        1.1975e+06        1.1957e+06        1.1994e+06
10275.4        6.7234e+05        6.7108e+05        6.7360e+05
11358.4        5.5649e+04        5.5646e+04        5.5651e+04
12589.7        6.7443e+05        6.7079e+05        6.7812e+05
12710.0        2.9710e+05        2.8917e+05        3.0549e+05
14061.8        3.7379e+06        3.7343e+06        3.7415e+06
15955.7        1.2690e+06        1.2679e+06        1.2700e+06
15969.9        5.9884e+06        5.9823e+06        5.9945e+06
16167.1        7.7811e+05        7.7118e+05        7.8516e+05
17934.1        4.9574e+05        4.9572e+05        4.9575e+05
18348.0        9.0824e+05        9.0822e+05        9.0827e+05
18491.0        7.8942e+05        7.8940e+05        7.8944e+05
18658.1        8.8711e+04        8.8709e+04        8.8714e+04
21191.1        2.3393e+06        2.3179e+06        2.3612e+06
22345.0        2.2103e+06        2.2103e+06        2.2104e+06
23383.5        2.1953e+06        2.1740e+06        2.2171e+06
24489.6        6.4955e+05        6.4954e+05        6.4956e+05
24553.5        8.6897e+05        8.6896e+05        8.6899e+05
26922.9        4.7003e+06        4.6832e+06        4.7176e+06
28652.9        1.4372e+06        1.4312e+06        1.4433e+06
28747.6        1.2544e+06        1.2440e+06        1.2650e+06
28842.0        9.5549e+06        9.5483e+06        9.5616e+06
29766.0        6.9324e+06        6.7433e+06        7.1325e+06
30712.1        1.1017e+06        1.0911e+06        1.1126e+06
32121.0        8.1697e+05        8.1696e+05        8.1698e+05

 

 

 

Attachment 3: ringdowns.png  187 kB  Uploaded Mon Nov 14 10:05:44 2016  | Hide | Hide all
ringdowns.png
ELOG V3.1.3-