40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  Coating Ring-down Measurement Lab elog  Not logged in ELOG logo
Message ID: 182     Entry time: Sun Nov 13 10:40:16 2016
Author: Gabriele 
Type: General 
Category: Measurements 
Subject: S1600469 

2016-11-13

  • 10:39am in chamber, balanced
  • 10:40am, roughing pump on
  • 10:49am, turbo pump on
  • 1:21pm, excitation (3 kV, 30s), quiet time before 1163107279, quiet time after 1163107415
  • 4:20pm, excitation (3kV, 30s) quiet time before 1163118070, quiet time after 1163118145
  • 6:34pm, pumps off

Results

2016-11-13 A

% Freq        Q                 Q (C.I. 95%)      Q (C.I. 95%)
1109.1        1.2009e+07        1.1962e+07        1.2057e+07
2543.6        5.5300e+06        5.5229e+06        5.5371e+06
4429.4        3.5636e+06        3.5626e+06        3.5646e+06
4495.8        1.9892e+05        1.9849e+05        1.9934e+05
6755.9        2.7949e+06        2.7935e+06        2.7963e+06
6769.6        8.0386e+06        8.0366e+06        8.0406e+06
6833.0        2.0677e+06        2.0671e+06        2.0684e+06
9514.3        2.1571e+06        2.1525e+06        2.1617e+06
10205.2        2.2730e+06        2.2706e+06        2.2755e+06
10360.6        2.0024e+06        2.0012e+06        2.0036e+06
12695.7        1.2769e+06        1.2595e+06        1.2947e+06
14168.6        3.4389e+06        3.4347e+06        3.4430e+06
16085.8        7.6129e+06        7.5677e+06        7.6586e+06
16302.1        1.1512e+06        1.1403e+06        1.1623e+06
18632.1        3.4509e+05        3.4508e+05        3.4510e+05
18806.5        8.7268e+05        8.6962e+05        8.7575e+05
20223.4        6.1496e+05        6.1494e+05        6.1498e+05
21351.3        2.9288e+06        2.9173e+06        2.9403e+06
23559.0        3.1627e+06        3.1237e+06        3.2027e+06
23712.0        4.2045e+05        4.2045e+05        4.2046e+05
24212.1        1.2451e+06        1.2451e+06        1.2451e+06
24687.7        1.1293e+06        1.1293e+06        1.1293e+06
25352.0        3.2923e+06        3.2922e+06        3.2923e+06
27127.4        9.1114e+06        8.9219e+06        9.3091e+06
28854.0        3.0897e+06        3.0269e+06        3.1552e+06
28952.0        2.6625e+06        2.6624e+06        2.6625e+06
29052.0        7.2887e+06        7.0946e+06        7.4937e+06
29523.0        5.4690e+05        5.4689e+05        5.4691e+05
32122.0        1.8145e+06        1.8145e+06        1.8146e+06

2016-11-13 B

% Freq        Q                 Q (C.I. 95%)      Q (C.I. 95%)
1130.0        1.7017e+06        1.6756e+06        1.7287e+06
2543.6        5.1969e+06        5.1941e+06        5.1998e+06
4429.4        3.7969e+06        3.7957e+06        3.7980e+06
4495.8        2.0313e+05        2.0241e+05        2.0386e+05
6755.9        2.8919e+06        2.8893e+06        2.8945e+06
6769.6        8.0715e+06        8.0699e+06        8.0730e+06
6833.0        2.1233e+06        2.1227e+06        2.1239e+06
9514.4        2.2500e+06        2.2489e+06        2.2512e+06
10205.3        2.4315e+06        2.4298e+06        2.4332e+06
10360.6        2.2703e+06        2.2641e+06        2.2766e+06
12695.8        1.3177e+06        1.3105e+06        1.3250e+06
14168.6        1.5517e+06        1.5485e+06        1.5549e+06
16074.0        1.1441e+06        1.1441e+06        1.1442e+06
16085.8        7.5344e+06        7.5003e+06        7.5688e+06
16302.1        1.1328e+06        1.1262e+06        1.1394e+06
17549.4        7.2792e+05        7.2790e+05        7.2794e+05
18632.4        4.3522e+05        4.3520e+05        4.3523e+05
18806.6        9.4602e+05        9.4289e+05        9.4916e+05
20144.0        1.3355e+06        1.3355e+06        1.3356e+06
20280.5        2.6059e+05        2.6058e+05        2.6060e+05
21351.0        2.7260e+06        2.7259e+06        2.7261e+06
23559.0        4.1056e+06        3.9740e+06        4.2461e+06
23711.6        4.0080e+05        4.0079e+05        4.0081e+05
24212.0        1.1849e+06        1.1849e+06        1.1849e+06
24687.7        9.9194e+05        9.9192e+05        9.9196e+05
28854.3        3.1246e+06        3.1246e+06        3.1247e+06
28952.0        2.4117e+06        2.4116e+06        2.4117e+06
29052.3        6.6583e+06        6.6582e+06        6.6585e+06

 

ELOG V3.1.3-