Message ID: 168     Entry time: Tue Nov 8 17:09:20 2016
Author: Gabriele 
Type: General 
Category: Measurements 
Subject: S1600460 

2016-11-08

  • 5:08pm, in chamber, balanced
  • 5:09pm, roughing pump on
  • 5:18pm, turbo pump on
  • 7:56: excitation. Quiet time before excitation 1162698977, excitation ampl. 3kV duration ~30s, quiet time after 1162700079
  • Three more excitation planned for the night
    • Quiet time before excitation: 1162710917
      Excitation at 1162710961 (30 s)
      Quiet time after excitation: 1162710991

    • Quiet time before excitation: 1162721821
      Excitation at 1162721865 (30 s)
      Quiet time after excitation: 1162721895

    • Quiet time before excitation: 1162732725
      Excitation at 1162732769 (30 s)
      Quiet time after excitation: 1162732799

  • The last two automatic excitation didn't go through. It seems to be the same problem as yesterday (which I "solved" by restarting the models)

2016-11-09

  • 8:37am, pumps off

Results

2016-11-08 A

% Freq        Q                 Q (C.I. 95%)      Q (C.I. 95%)
1119.5        6.6404e+05        6.4444e+05        6.8484e+05
2504.0        5.3952e+06        5.3663e+06        5.4245e+06
2505.5        5.4008e+06        5.3791e+06        5.4226e+06
2510.6        1.0237e+06        9.9922e+05        1.0495e+06
2537.5        5.5690e+05        5.5679e+05        5.5701e+05
2545.0        9.5810e+05        9.5676e+05        9.5944e+05
2552.0        7.5631e+05        7.5616e+05        7.5646e+05
2582.3        1.0468e+06        1.0373e+06        1.0565e+06
3377.5        1.2974e+06        1.2972e+06        1.2975e+06
4361.5        2.9905e+06        2.9870e+06        2.9940e+06
4390.0        1.5871e+05        1.5869e+05        1.5873e+05
4428.9        1.6234e+05        1.6226e+05        1.6242e+05
4467.5        1.6373e+05        1.6371e+05        1.6375e+05
6653.7        3.0794e+06        3.0779e+06        3.0809e+06
6670.1        1.0338e+07        1.0326e+07        1.0350e+07
6730.4        1.4921e+06        1.4893e+06        1.4949e+06
6798.0        3.5459e+05        3.5456e+05        3.5461e+05
9372.3        2.1885e+06        2.1875e+06        2.1895e+06
10055.0        1.6518e+06        1.6517e+06        1.6519e+06
10210.3        2.7007e+05        2.7005e+05        2.7008e+05
12507.3        1.6391e+06        1.6391e+06        1.6392e+06
13961.7        7.0939e+06        7.0387e+06        7.1500e+06
15837.2        2.0382e+06        2.0365e+06        2.0399e+06
15850.6        5.9047e+06        5.8559e+06        5.9543e+06
16064.1        1.8247e+06        1.8246e+06        1.8247e+06
18532.6        3.1042e+05        3.1041e+05        3.1043e+05
18797.0        1.8528e+06        1.8527e+06        1.8528e+06
19983.2        1.3315e+06        1.3196e+06        1.3436e+06
21043.9        6.7756e+06        6.3946e+06        7.2047e+06
22640.6        2.6064e+06        2.6063e+06        2.6064e+06
23218.7        4.4360e+06        4.4359e+06        4.4361e+06
24331.4        1.2956e+06        1.2955e+06        1.2956e+06
26268.9        5.9110e+05        5.9109e+05        5.9111e+05
26740.4        3.9234e+06        3.9234e+06        3.9235e+06
28436.5        2.8936e+06        2.8858e+06        2.9014e+06
28636.0        2.4942e+06        2.4941e+06        2.4942e+06
29140.0        5.9494e+05        5.9493e+05        5.9495e+05
30906.6        2.0991e+06        2.0991e+06        2.0991e+06
32340.0        1.9250e+06        1.9250e+06        1.9250e+06

2016-11-08 B

% Freq        Q                 Q (C.I. 95%)      Q (C.I. 95%)
1112.1        1.0416e+06        1.0354e+06        1.0479e+06
2224.6        1.0526e+06        1.0498e+06        1.0555e+06
2506.2        6.2467e+06        6.2222e+06        6.2715e+06
2511.0        9.1074e+05        8.9608e+05        9.2589e+05
2544.9        9.2936e+05        9.2745e+05        9.3127e+05
2579.0        9.3199e+05        9.2088e+05        9.4336e+05
2583.3        9.1765e+05        9.1228e+05        9.2308e+05
2778.1        2.6436e+06        2.6213e+06        2.6664e+06
3657.0        9.4905e+05        9.4892e+05        9.4918e+05
4323.0        2.7441e+06        2.6975e+06        2.7924e+06
4361.4        2.9870e+06        2.9852e+06        2.9887e+06
5618.3        9.7528e+05        9.7519e+05        9.7536e+05
6653.7        2.9851e+06        2.9799e+06        2.9903e+06
6670.0        1.0065e+07        1.0049e+07        1.0081e+07
6691.7        1.4940e+06        1.4774e+06        1.5110e+06
6723.0        9.4356e+04        9.4349e+04        9.4363e+04
6730.5        1.5010e+06        1.5000e+06        1.5021e+06
6737.7        9.3824e+04        9.3817e+04        9.3831e+04
6769.0        1.4817e+06        1.4683e+06        1.4954e+06
6797.8        3.4621e+05        3.4618e+05        3.4623e+05
7842.6        1.3638e+06        1.3637e+06        1.3639e+06
9372.2        2.1002e+06        2.0979e+06        2.1025e+06
10055.0        1.6575e+06        1.6417e+06        1.6735e+06
10210.0        2.7696e+05        2.7694e+05        2.7697e+05
12507.3        1.5841e+06        1.5824e+06        1.5858e+06
13961.5        6.9382e+06        6.8656e+06        7.0122e+06
15837.2        1.9985e+06        1.9964e+06        2.0006e+06
15850.6        5.9234e+06        5.9232e+06        5.9236e+06
16064.0        1.7705e+06        1.7704e+06        1.7705e+06
18361.4        2.5816e+06        2.5815e+06        2.5817e+06
19983.2        1.3972e+06        1.3907e+06        1.4038e+06
21043.9        9.7066e+06        8.8724e+06        1.0714e+07
22640.9        2.8173e+06        2.8173e+06        2.8174e+06
23218.7        4.2100e+06        4.2099e+06        4.2101e+06
24331.1        1.2336e+06        1.2336e+06        1.2336e+06
28436.4        2.7860e+06        2.7860e+06        2.7861e+06
28635.6        2.7569e+06        2.7568e+06        2.7569e+06

 

ELOG V3.1.3-