2016-11-08

  • 11:23, in chamber, balanced
  • 11:25 roughing pump on
  • 11:38 turbo pump on
  • 1:38pm excitation (3kV, ~20s). Quiet time before: 1162677138, quiet time after 1162677462
  • 3:27pm excitation. (3kV, 30s). Quiet time after 1162682879
  • 4:37pm, pumps stopped

Results

2016-11-08 A

% Freq        Q                 Q (C.I. 95%)      Q (C.I. 95%)
1108.3        1.5638e+07        1.5570e+07        1.5708e+07
2503.6        7.2192e+06        6.8796e+06        7.5940e+06
2542.2        7.3800e+06        7.3760e+06        7.3839e+06
4389.0        5.3514e+06        5.3121e+06        5.3913e+06
4427.6        5.3885e+06        5.3869e+06        5.3901e+06
4466.2        5.4399e+06        5.4000e+06        5.4803e+06
4495.0        2.1932e+05        2.1899e+05        2.1966e+05
6754.2        3.8355e+06        3.8326e+06        3.8383e+06
6767.2        1.3227e+07        1.3213e+07        1.3240e+07
6793.5        2.1749e+06        2.1747e+06        2.1751e+06
6832.1        2.1373e+06        2.1366e+06        2.1380e+06
6870.6        2.2276e+06        2.2274e+06        2.2278e+06
9512.8        2.9775e+06        2.9723e+06        2.9828e+06
10200.3        2.5862e+06        2.5812e+06        2.5913e+06
10357.0        1.7162e+06        1.7161e+06        1.7163e+06
12694.6        5.9843e+05        5.9841e+05        5.9845e+05
14162.4        4.9621e+06        4.9386e+06        4.9860e+06
16069.7        1.1980e+06        1.1979e+06        1.1980e+06
16080.1        2.6316e+06        2.6237e+06        2.6395e+06
16301.9        1.4163e+06        1.4162e+06        1.4163e+06
18623.9        3.6294e+06        3.6258e+06        3.6330e+06
20280.4        8.2385e+05        8.2383e+05        8.2387e+05
23548.3        4.6763e+06        4.6762e+06        4.6764e+06
24690.0        1.1723e+06        1.1723e+06        1.1724e+06
25694.0        5.4900e+05        5.4899e+05        5.4901e+05
27116.0        2.5536e+06        2.5536e+06        2.5537e+06
28845.1        1.6430e+06        1.6430e+06        1.6430e+06
29041.5        2.5349e+06        2.5348e+06        2.5349e+06
29519.0        2.5811e+06        2.5811e+06        2.5812e+06
30419.5        1.8440e+06        1.8439e+06        1.8440e+06
31263.4        1.5940e+06        1.5940e+06        1.5941e+06
32136.9        4.5330e+06        4.5330e+06        4.5331e+06

2016-11-08 B

% Freq        Q                 Q (C.I. 95%)      Q (C.I. 95%)
1108.4        1.3404e+07        1.3360e+07        1.3447e+07
2542.2        6.9483e+06        6.9392e+06        6.9574e+06
4389.0        4.5933e+06        4.5528e+06        4.6345e+06
4427.6        5.0578e+06        5.0565e+06        5.0591e+06
4466.1        4.7063e+06        4.6660e+06        4.7473e+06
4495.1        2.2376e+05        2.2324e+05        2.2427e+05
6754.3        3.5768e+06        3.5759e+06        3.5777e+06
6767.1        1.2441e+07        1.2411e+07        1.2470e+07
6793.7        1.9583e+06        1.9581e+06        1.9584e+06
6832.2        2.0323e+06        2.0318e+06        2.0329e+06
6870.7        2.0027e+06        2.0026e+06        2.0029e+06
9512.8        2.7522e+06        2.7497e+06        2.7546e+06
10200.4        2.4866e+06        2.4793e+06        2.4938e+06
10357.0        1.6884e+06        1.6837e+06        1.6931e+06
12694.7        6.2234e+05        6.2231e+05        6.2236e+05
14162.3        5.4381e+06        5.4299e+06        5.4463e+06
16069.8        1.1596e+06        1.1595e+06        1.1596e+06
16080.2        2.5075e+06        2.5002e+06        2.5148e+06
16302.0        1.3303e+06        1.3302e+06        1.3303e+06
18624.0        3.6265e+06        3.6207e+06        3.6322e+06
20280.6        7.8640e+05        7.8638e+05        7.8642e+05
21342.5        6.1946e+06        6.1945e+06        6.1948e+06
23548.5        3.9191e+06        3.9190e+06        3.9191e+06
24690.1        1.0713e+06        1.0713e+06        1.0713e+06
28845.3        1.6909e+06        1.6909e+06        1.6910e+06
29041.7        2.6668e+06        2.6668e+06        2.6669e+06
29518.6        2.4022e+06        2.4022e+06        2.4022e+06
30419.3        1.5929e+06        1.5929e+06        1.5930e+06
31263.0        1.4750e+06        1.4750e+06        1.4751e+06
32136.8        4.1712e+06        4.1711e+06        4.1713e+06

 