Message ID: 137     Entry time: Wed Oct 19 16:40:24 2016
Author: sults: 
Type: General 
Category: Measurements 
Subject: S1600349 

S1600439, not annealed, as received from Mark Optics

Installation

  • installed and balanced, pump down started at 4:38pm (roughing pump), turbo started at 4:51pm
  • removed from the chamber on 10/20 at about 11:00am

Measurements

2016_10_19

  • QPD centered at 9:25pm, excited (2000 V, 10 s) at 9:27pm, measurement ongoing.

The spectrum was noise than usual due to the roughing pump. I already found out in the past that I can reduce the noise by tweaking the position of the pump. This time however I wasn't successful.

Here are the results:

% Freq        Q                 Qlow (C.I. 95%)   Qhi (C.I. 95%)
1111.5        6.0958e+06        6.0641e+06        6.1279e+06
2549.6        3.6416e+06        3.6305e+06        3.6528e+06
4441.7        2.6916e+06        2.6879e+06        2.6953e+06
4512.9        2.1090e+05        2.0993e+05        2.1188e+05
6777.5        1.1797e+06        1.1790e+06        1.1803e+06
6790.9        3.8621e+06        3.8540e+06        3.8701e+06
6858.4        1.2434e+06        1.2295e+06        1.2577e+06
9548.0        1.2177e+05        1.1878e+05        1.2491e+05
10234.6       3.9934e+05        3.9351e+05        4.0535e+05
10399.0       2.8135e+05        2.7455e+05        2.8849e+05
12744.2       1.3145e+06        1.3115e+06        1.3174e+06
14211.3       3.8414e+06        3.8136e+06        3.8696e+06
16123.3       2.0276e+06        2.0181e+06        2.0372e+06
16135.8       5.2811e+06        5.2770e+06        5.2852e+06
16370.0       1.4976e+06        1.4922e+06        1.5031e+06
18689.3       4.2954e+06        4.2738e+06        4.3172e+06
23632.0       2.2796e+06        2.2528e+06        2.3070e+06
24797.4       8.7545e+05        8.5819e+05        8.9341e+05
27214.5       1.2566e+06        1.2374e+06        1.2764e+06
28947.5       1.7367e+06        1.7130e+06        1.7611e+06
29144.0       3.2148e+06        3.1213e+06        3.3142e+06

 

2016_10_20

  • Roughing pump was creating too much noise, switched it off at 8:30am
  • Excitation at 8:35am (2000 V, 30 s)

Here are the results:

And the measured Q values:

% Freq        Q                 Qlow (C.I. 95%)   Qhi (C.I. 95%)
1111.4        6.2206e+06        6.2061e+06        6.2351e+06
2549.6        4.1538e+06        4.1498e+06        4.1579e+06
2592.9        1.0952e+06        1.0887e+06        1.1017e+06
4441.7        2.8521e+06        2.8513e+06        2.8529e+06
4513.0        2.0592e+05        2.0455e+05        2.0730e+05
6777.5        1.1013e+06        1.1006e+06        1.1020e+06
6790.9        3.9913e+06        3.9879e+06        3.9947e+06
6858.2        1.3100e+06        1.3078e+06        1.3122e+06
9547.8        1.2492e+06        1.2469e+06        1.2514e+06
10234.7       3.4336e+06        3.4209e+06        3.4463e+06
10398.7       4.3217e+05        3.2860e+05        6.3104e+05
12744.2       4.5274e+05        4.5177e+05        4.5371e+05
14211.3       2.0254e+06        2.0081e+06        2.0430e+06
16123.0       2.1225e+06        2.1173e+06        2.1276e+06
16135.8       4.6547e+06        4.6529e+06        4.6564e+06
16370.3       1.4781e+06        1.4762e+06        1.4800e+06
18689.2       4.5978e+06        4.5891e+06        4.6064e+06
20299.5       4.1859e+05        4.1375e+05        4.2353e+05
20364.6       1.1099e+06        1.0630e+06        1.1612e+06
21418.2       4.5814e+06        4.5718e+06        4.5911e+06
23631.9       2.8907e+06        2.8700e+06        2.9116e+06
24797.4       1.1241e+06        1.1121e+06        1.1363e+06
27214.5       2.1015e+06        2.0752e+06        2.1285e+06
28947.3       2.1285e+06        2.1162e+06        2.1410e+06
29054.7       1.4873e+06        1.4631e+06        1.5124e+06
29143.8       4.1250e+06        4.1050e+06        4.1453e+06
29647.5       4.5456e+05        4.4215e+05        4.6768e+05
31135.4       7.5136e+05        7.3331e+05        7.7033e+05
32013.4       1.1846e+06        1.1565e+06        1.2140e+06
Attachment 9: ringdown_rawdata_2016_10_19.m  8 kB  Uploaded Fri Oct 21 09:33:37 2016  | Hide | Hide all
%% Parameters
prefix = '2016_10_19';  % name of the folder where result will be saved
gps0 = 1160972805;      % GPS time of clean data before excitation
gps1 = 1160972852;      % GPS time right after excitation
dt = 30;                % how much data to be used to search peaks

minsnr = 6;             % minimum peak SNR
minfr = 1000;           % minimum peak frequency
Dt = 3600;              % total amount of time for the ringdown measurement

... 274 more lines ...
Attachment 10: ringdown_rawdata_2016_10_20.m  8 kB  Uploaded Fri Oct 21 09:33:37 2016  | Hide | Hide all
%% Parameters
prefix = '2016_10_20';  % name of the folder where result will be saved
gps0 = 1161012900;      % GPS time of clean data before excitation
gps1 = 1161012972;      % GPS time right after excitation
dt = 30;                % how much data to be used to search peaks

minsnr = 6;             % minimum peak SNR
minfr = 1000;           % minimum peak frequency
Dt = 3600;              % total amount of time for the ringdown measurement

... 274 more lines ...