40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 98 of 339  Not logged in ELOG logo
ID Date Author Type Category Subject Text Attachments
  12166   Fri Jun 10 12:09:01 2016 jamieConfigurationCDSIRIG-B debuggingLooks like we might have a problem with
the IRIG-B output of the GPS receiver.

Rolf came over this morning to
  
  12165   Fri Jun 10 09:52:51 2016 SteveUpdateSUS EQ 5.2 mag Borego SpringEQ  5.2 mag at Jun 10, 8:04 am UTC,
Borego Spring, CA  ~150 mi away.........no
obvoius damage, damping restored, MC is locking,
 5.2mBoregoSpring.pngeqBSs.png 
  12164   Thu Jun 9 19:08:58 2016 VarunUpdateElectronicsAnti-Aliasing Filter updateEric gave me a psd plot of a signal which
would be the input of a channel of the AA
filter. the Nyquist freq. is about 32.8kHz.
 32to65.jpg65to98.png 
  12163   Thu Jun 9 18:54:40 2016 AakashUpdateGeneralAbout Acromag | SURF 2016Today I tried to setup Acromag Busworks
card. I was able to calibrate and test it
over USB but I couldn't test it over
  
  12162   Thu Jun 9 15:14:46 2016 jamieUpdateCDSold fb restarted, test of new daqdon fb1 aborted for time beingI've restarted the old daqd on fb until
I can figure out what's going on with
the symmetricom driver on fb1.
 Jamie.jpg 
  12161   Thu Jun 9 13:28:07 2016 jamieConfigurationCDSSpectracom IRIG-B card installed on fb1Something is wrong with the timing we're
getting out of the symmetricom driver, associated
with the new spectracom card.
  
  12160   Thu Jun 9 09:57:06 2016 AakashUpdateGeneralSeismometer Enclosure Development
Me and Gautam yesterday
opened the tilt-free seismometer enclosure
to see if we could use the thermocouples
  
  12159   Wed Jun 8 16:12:38 2016 VarunUpdateGeneralDAFI updateSummary: I am implementing
digital audio filtering on various interferometer
 dafi.pngagcin.pdfagcout.pdf 
  12158   Wed Jun 8 13:50:39 2016 jamieConfigurationCDSSpectracom IRIG-B card installed on fb1[EDIT: corrected name of installed card]

We just installed a Spectracom
TSyc-PCIe timing card on fb1.  The
  
  12157   Wed Jun 8 10:20:14 2016 SteveUpdatesafetySURF 2016 safetyAakash Patil received 40m specific basic
safety training.

  
  12156   Wed Jun 8 08:34:55 2016 jamieUpdateCDSDAQD work ongoing38 restarts overnight.  Problem definitely
not fixed.  I'll be reverting back
to old daqd and fb this morning.  Then
  
  12155   Tue Jun 7 20:49:50 2016 jamieUpdateCDSDAQD work ongoingSummary: new daqd code
running overnight test on fb1.  Stability
issues persist.
  
  12154   Tue Jun 7 18:20:18 2016 VarunUpdateGeneralDAFI updateTried to implement AGC on FE. Had some
trouble bringing the code into the correct
form. It looks okay now. However, this agc
  
  12153   Tue Jun 7 17:21:13 2016 AakashUpdateGeneralSURF 2016Hi!

I am Aakash Patil. I will be working
at the 40m lab as a SURF student with Gautam
  
  12152   Tue Jun 7 11:12:47 2016 jamieUpdateCDSDAQD UPGRADE WORK UNDERWAYI
am re-starting work on the daqd upgrade again
now.  Expect the daqd
  
  12151   Mon Jun 6 16:41:36 2016 ericqUpdateCDSFB upgrade workBarring objections, starting tomorrow morning, Jamie
will be testing the new FB code. The IFO
will not be available for other use while
  
  12150   Fri Jun 3 17:56:14 2016 VarunUpdateGeneralDAFI updateWrote and tested a phase vocoder, with
two of its applications:

1) Time scaling: This enables change
  
  12149   Fri Jun 3 14:24:13 2016 SteveUpdateSUSlocal EQ 3.1 mLocal EQ 3.1 mag at Jun 2, 2016 11:06:16
PM UTC, Muscoy CA........no damage

Our STS should seen this shake.
 eq3.1mMuscoyCa.png 
  12148   Fri Jun 3 13:05:18 2016 ericqUpdateCDSCDS NotesSome CDS related things:


Keith Thorne has told us
about a potential fix for our framebuilder woes.
  
  12147   Fri Jun 3 12:53:44 2016 ericqUpdateElectronicsCommon board Op amp input offsetsI replaced some of the AD829s with other
AD829s, but the offset situation didn't
improve.
 CMboard_offsets.png 
  12146   Thu Jun 2 16:35:44 2016 KojiUpdateSUS wire standoffs updateGap of the prism from the mirror

Sag: s = R(1-Cos[ArcSin[d/2/R]])

- Mirror curvature sag for 2mm
 sapphire_stand_off_v3.pdfarrangement-v3.pdf 
  12145   Wed Jun 1 16:28:28 2016 ericqUpdateElectronicsCommon board Op amp input offsetsI used a Eurocard extension board to peek
at the inputs and outputs of each of the
gain-ladder AD829s on input B of the CM board
  
  12144   Wed Jun 1 15:01:56 2016 SteveUpdateSUS wire standoffs updateThere are some issues with 5 mm sapphire
prism Atm5. It will cause  interference
between one of the prisms and the Side OSEM.
 06011603.PDF06011602.PDFsapphire_stand_off_v2.pdfside_OSEM_ETMY.jpg5p2wsf.png 
  12143   Wed Jun 1 11:19:14 2016 VarunUpdateGeneralUpdate of work till nowCompleted:

Wrote and tested a code for AGC
using cavity transmission signal and length
  
  12142   Wed Jun 1 09:06:38 2016 SteveUpdateLSCNew stands for TransMon/Oplev QPDsMachined from I-beam 6061 T6 Aluminum 5"
x 0.5 x 3.25

 pdIb.PDFpdstand.pdf 
  12141   Tue May 31 16:52:58 2016 SteveUpdatesafetyNONOPlease do not place anything on the top
of the cabinets that is not tied down. It
will end up on our head in an earth quake.
 nono.jpg 
  12140   Mon May 30 18:19:50 2016 JohannesUpdateCDSASS medm screen updateI noticed that the TRY button in the ASS
main screen was linking to LSC_TRX instead
of LSC_TRY. Gautam fixed it.
  
  12139   Fri May 27 11:54:22 2016 VarunUpdateGeneralPackage deliveryA package labelled 'UPS Ground'
has arrived.

-Varun
  
  12138   Fri May 27 02:52:53 2016 ericqUpdateLSCRestoring high BW single arm controlI've been futzing with the common mode
servo, trying to engage the AO path with
POY for high bandwidth control of a single
  
  12137   Thu May 26 18:10:48 2016 VarunUpdateGeneralSURF 2016Wrote and tested a function for downconversion.
It contains a mixer with a sinusoidal input
for modulation with the desired frequency
 input.pngoutput.png 
  12136   Wed May 25 14:29:31 2016 VarunUpdateGeneralSURF 2016Edited the AGC to include overlapping frames
yesterday. forgot to put an elog on it!

  
  12135   Wed May 25 14:21:29 2016 Max IsiUpdateGeneralSummary page configurationI have modified the c1summary.ini and c1lsc.ini
configuration files slightly to avoid overloading
the system and remove the errors that were
  
  12134   Wed May 25 11:51:40 2016 SteveUpdatesafetySURF 2016 safety


  
  12133   Wed May 25 08:32:55 2016 SteveUpdateSUSlocal EQ 3.5mLocal EQ 3.5 mag  at 2:28 UTC May
24, 2016 Rancho Cucamonga, Ca.....no damage

 
 3.5Cucam.pnglocal3.5cucam.png 
  12132   Wed May 25 02:54:09 2016 ericqUpdateGeneralOdds and endsWFS locking point seemed degraded; I hand
aligned and reset the WFS offsets as usual.

ITMX oplev recentered. While doing
  
  12131   Tue May 24 23:17:37 2016 ericqUpdateCOCFinesse modelling - mode overlap scansI think you should use the current actual
PRC & SRC cavity lengths as measured,
as it would be simplest to simply replace
  
  12130   Tue May 24 22:49:02 2016 gautamUpdateCOCFinesse modelling - mode overlap scansSummary:

Having played around with a toy finesse
model, I went about setting up a model in
 9x 
  12129   Tue May 24 17:55:17 2016 VarunUpdateElectronicsUsing AltiumContacted Charles regarding use of Altium.
Got to know that Altium is installed
on cit40m iMac in Win7 on VirtualBox.
  
  12128   Tue May 24 10:21:36 2016 ericqSummarySUSITMX Oplev loopsI did a quick measurement of the ITMX oplev
loops, both pitch and yaw have about the
same upper UGF as previous measurements with
  
  12127   Mon May 23 17:47:51 2016 VarunUpdateGeneralSURF 2016Tested the AGC today with LSC cavity transmission
signal and error signal. Not in real time
still.
 cav_tr-eps-converted-to.pdfcav_tr_out-eps-converted-to.pdferr-eps-converted-to.pdferr_out-eps-converted-to.pdf 
  12126   Mon May 23 15:51:32 2016 steveSummarySUSoplev laser summary updated 




 oplSum.png 
  12125   Mon May 23 10:55:49 2016 steveSummarySUSITMX oplev laser replaced 

     
May 23, 2016            
 ITMXoplev.png 
  12124   Fri May 20 17:36:06 2016 gautamUpdateLSCNew stands for TransMon/Oplev QPDsAs we realized during the EX table switch,
the transmitted beam height from the arm
is not exactly 4" relative to the endtable,
  
  12123   Fri May 20 00:06:19 2016 VarunUpdateGeneralSURF 2016I have written a basic version of AGC,
and have done some tests with a data file.
will do tests on whitening and agc today.
  
  12122   Thu May 19 16:29:20 2016 SteveUpdateendtable upgradeOptical layout almost complete 

 
 ETMX_4x3_closed.jpgsealedETMXenclosure.jpg 
  12121   Wed May 18 17:42:52 2016 VarunUpdateGeneralSURF 2016Finished writing the code on whitening.
I have to still test it. uploaded on github
noise cancellation repo. @eric could you
  
  12120   Wed May 18 01:10:22 2016 gautamUpdateCOCFinesse modellingI've been working on putting together
a Finesse model for the current 40m configuration.
The idea was to see if I could reproduce
 arms.pdfPRC.pdfSRC.pdfFinesse_model.zip 
  12119   Tue May 17 14:46:51 2016 SteveUpdateVACRGA scan at day 595We have good RGA scan now. There was no
scan for 3 months.
 RGAscan595d.pngpd78-560Hz-d600.png 
  12118   Tue May 17 05:50:43 2016 Varun KelkarUpdateGeneralSURF 2016Hello, I am Varun Kelkar. I will be working
at the 40m lab as a SURF student this summer
with Eric Quintero on Audio processing for
  
  12117   Sun May 15 19:48:08 2016 SteveUpdateVACrun out of N23-4 hrs ago we run out of nitrogen. We
are back to Vacuum Normal

 
 noN2.png 
ELOG V3.1.3-