40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 96 of 354  Not logged in ELOG logo
ID Date Author Type Categorydown Subject
  7396   Mon Sep 17 07:57:58 2012 ranaUpdatePEMBS and AS tables for REFL change

 

 This was in response to my suggestion to move the REFL beam path to the table containing the BS/PRM Oplevs. From this seismic data it is clear that the BS table is no worse than the AS table, so we should plan to make the layout change during the next vent.

  7429   Sat Sep 22 01:03:30 2012 DenUpdatePEMETMY table

I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94).  +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.

Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions.

Attachment 1: X.pdf
X.pdf
Attachment 2: Y.pdf
Y.pdf
Attachment 3: Z.pdf
Z.pdf
  7433   Mon Sep 24 17:03:39 2012 JenneUpdatePEM1X1 rack power

Quote:

I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94).  +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.

 I'm not sure what the problem is here.  Den and I looked at it for a few minutes, before I went back to helping with putting doors on.  The Sorensons are not supplying the rack power for 1X1.  There are some flat cables which go from the fuses on the side of the rack up to the cable tray, and go elsewhere.  Den is going to continue looking into this, but I think it's a moderately high priority, since lots of things should be getting served by that same power.

  7459   Mon Oct 1 19:21:03 2012 ranaConfigurationPEMchanged PEM DQ channels

Changed the list of channels to be written to frames from having the IN1 suffix to OUT. Now we can load the calibration of the channel into the filter module and the DQ channel will be calibrated.

We should do this wherever possible so that our channels will have real calibrations associated with them.
SEIS_GUR1_X_OUT 256
SEIS_GUR1_Y_OUT 256
SEIS_GUR1_Z_OUT 256
SEIS_GUR2_X_OUT 256
SEIS_GUR2_Y_OUT 256
SEIS_GUR2_Z_OUT 256
SEIS_STS_1_X_OUT 256
SEIS_STS_1_Y_OUT 256
SEIS_STS_1_Z_OUT 256
SEIS_STS_2_X_OUT 256
SEIS_STS_2_Y_OUT 256
SEIS_STS_2_Z_OUT 256
SEIS_STS_3_X_OUT 256
SEIS_STS_3_Y_OUT 256
SEIS_STS_3_Z_OUT 256
MIC_1_OUT 2048
MIC_2_OUT 2048
MIC_3_OUT 2048
MIC_4_OUT 2048
MIC_5_OUT 2048
MIC_6_OUT 2048
ACC_MC1_X_OUT 2048
ACC_MC1_Y_OUT 2048
ACC_MC1_Z_OUT 2048
ACC_MC2_X_OUT 2048
ACC_MC2_Y_OUT 2048
ACC_MC2_Z_OUT 2048
XARM_DIFFERENTIAL_MOTION_IN1 256
XARM_DIFFERENTIAL_MOTION_OUT 256
YARM_DIFFERENTIAL_MOTION_IN1 256
YARM_DIFFERENTIAL_MOTION_OUT 256

Next we should up the rate at which the model runs up to 16 kHz so that we can record the microphones at 16 kHz. FM radio has information up to 20 kHz. AM radio goes up to ~8 kHz. We should be at least as modern as AM radio. How do we make the change? How do we make sure the FOTON file stays OK?

I have made some changes to the daily summary file to compensate. New files is /users/public_html/40m-summary/share/c1_summary_page.ini.

  7468   Wed Oct 3 15:37:04 2012 SteveUpdatePEMants on the PSL table

We observed one or two ants climbing over PMC optics without booties and safety glasses.

The floor was mopped with strong Bayer Home Pest Control solution in the Vertex area.

Do not work inside the 40m lab if you are sensitive to chemicals!

 

Attachment 1: IMG_1690.JPG
IMG_1690.JPG
  7470   Wed Oct 3 16:26:58 2012 ManasaUpdatePEMants on the PSL table

I spotted around 4 within 30 minutes working at the PSL table even after the deathly spray. They seem to be running down from the cables on the oscilloscope rack to the table and the optics.

  7487   Fri Oct 5 00:29:34 2012 DenUpdatePEMreadout box power

Guralp readout box received +13.7 /0/ -15V instead +15V because of the broken fuse. Power provided by the source is normal.

Edit by Den: I've found a similar fuse on one of the tables and borrowed it. Guralp is not working again.

  7488   Fri Oct 5 01:36:49 2012 DenConfigurationPEMchanged PEM DQ channels

Quote:

We should do this wherever possible so that our channels will have real calibrations associated with them.

Next we should up the rate at which the model runs up to 16 kHz so that we can record the microphones at 16 kHz. FM radio has information up to 20 kHz. AM radio goes up to ~8 kHz. We should be at least as modern as AM radio. How do we make the change? How do we make sure the FOTON file stays OK?

 I've added calibration gains to Guralp (to um/sec) and EM172 (to Pa) channels.

We can run PEM at 16 kHz. I think Foton file stores both sos-representation and filter commands which are independent of the sampling frequency, so it should be possible to change model sampling frequency quickly.

In fact, we can save data at 64 kHz from iop models. I've done this once with MC_F channel. However, I did not test EM172 noise at frequencies > 1 kHz.

  7490   Fri Oct 5 11:11:00 2012 DenUpdatePEMreadout box power

Quote:

Guralp readout box received +13.7 /0/ -15V instead +15V because of the broken fuse. Power provided by the source is normal.

Edit by Den: I've found a similar fuse on one of the tables and borrowed it. Guralp is not working again.

 I've meant Guralp is NOW working again 

  7539   Fri Oct 12 22:44:49 2012 DenUpdatePEMETMY table

Quote:

Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions.

Accelerometers were installed on the ETMY table and nearby ground to measure amplification of the seismic noise due to the table. During this experiment ground and table motions were measured simultaneously.

DSC_4734.JPG     DSC_4736.JPG

Attachment 1: etmy_x_psd.pdf
etmy_x_psd.pdf
Attachment 2: etmy_y_psd.pdf
etmy_y_psd.pdf
Attachment 3: etmy_z_psd.pdf
etmy_z_psd.pdf
Attachment 4: etmy_coh.pdf
etmy_coh.pdf etmy_coh.pdf etmy_coh.pdf
  7540   Sun Oct 14 11:41:42 2012 DenUpdatePEMETMY table

Quote:

 

Accelerometers were installed on the ETMY table and nearby ground to measure amplification of the seismic noise due to the table. During this experiment ground and table motions were measured simultaneously

 I've added xml file with measurement settings and data to 40m svn at directory 40m_seismic/etmy.

DSC_4739.JPG 

Attachment 2: 14OCT2012.pdf
14OCT2012.pdf 14OCT2012.pdf 14OCT2012.pdf 14OCT2012.pdf
Attachment 3: 14OCT2012.xml
<?xml version="1.0"?>
<!DOCTYPE LIGO_LW [
<!ELEMENT LIGO_LW ((LIGO_LW|Comment|Param|Time|Table|Array|Stream)*)>
<!ATTLIST LIGO_LW Name CDATA #IMPLIED Type CDATA #IMPLIED>
<!ELEMENT Comment (#PCDATA)>
<!ELEMENT Param (#PCDATA)>
<!ATTLIST Param Name CDATA #IMPLIED Type CDATA #IMPLIED Dim CDATA #IMPLIED
                Unit CDATA #IMPLIED>
<!ELEMENT Table (Comment?,Column*,Stream?)>
<!ATTLIST Table Name CDATA #IMPLIED Type CDATA #IMPLIED>
... 231759 more lines ...
  7541   Sun Oct 14 16:44:20 2012 DenUpdatePEMETMX table

 I've measured ETMX table motion compared to ground motion using accelerometers. Data and settings in the xml file are at the svn directory 40m_seismic/etmx.

DSC_4742.jpg

Attachment 2: 14OCT2012.pdf
14OCT2012.pdf 14OCT2012.pdf 14OCT2012.pdf 14OCT2012.pdf
Attachment 3: 14OCT2012.xml
<?xml version="1.0"?>
<!DOCTYPE LIGO_LW [
<!ELEMENT LIGO_LW ((LIGO_LW|Comment|Param|Time|Table|Array|Stream)*)>
<!ATTLIST LIGO_LW Name CDATA #IMPLIED Type CDATA #IMPLIED>
<!ELEMENT Comment (#PCDATA)>
<!ELEMENT Param (#PCDATA)>
<!ATTLIST Param Name CDATA #IMPLIED Type CDATA #IMPLIED Dim CDATA #IMPLIED
                Unit CDATA #IMPLIED>
<!ELEMENT Table (Comment?,Column*,Stream?)>
<!ATTLIST Table Name CDATA #IMPLIED Type CDATA #IMPLIED>
... 231779 more lines ...
  7542   Sun Oct 14 17:26:03 2012 DenUpdatePEMETMX table

Quote:

 I've measured ETMX table motion compared to ground motion using accelerometers. Data and settings in the xml file are at the svn directory 40m_seismic/etmx

 High frequency (>60 Hz) resonances that are present at the ETMX motion spectrum seem to be understandable. Amplification ETMX/GROUND of a factor of 2 at 1 Hz is interesting. I've monitored ACC DQ channels for a few hours and noticed that usually spectrum looks like in the previous elog. But every ~40 min ETMX motion is much higher then ground motion at low frequencies (<5 Hz). I wonder if this a reaction of a table to outside disturbances or accelerometer issue.

 

Attachment 1: etmx.pdf
etmx.pdf etmx.pdf etmx.pdf etmx.pdf
  7543   Sun Oct 14 20:51:20 2012 DenUpdatePEMAA board

Quote:

 

 But every ~40 min ETMX motion is much higher then ground motion at low frequencies (<5 Hz). I wonder if this a reaction of a table to outside disturbances or accelerometer issue. 

 This could come from AA board, its range is +/- 2.5 V, RMS of the ETMX table motion is a few times higher then ground motion, so ETMX accelerometer signal was corrupted.

acc.png

As this small AA range has already caused problems before, I decided to increase it. I've looked through the board scheme and found that all its differential line receives and output amplifiers have absolute maximum range of 40V. We used KEPKO power supply for this board with a voltage range up to 6 V. So I've replaced it with a BK PRECISION power supply and set it to +/- 15 V. Now AA board range is 7.5 V.

DSC_4747.JPG 

I'll leave accelerometers near ETMX table. It's interesting to measure table motion in the morning when trucks drive by.

  7546   Mon Oct 15 10:14:22 2012 DenUpdatePEMAA board

Quote:

 

 I'll leave accelerometers near ETMX table. It's interesting to measure table motion in the morning when trucks drive by.

 That low frequency effect was due to AA board, now it is gone.

  7562   Tue Oct 16 21:57:47 2012 DenUpdatePEMaccelerometers

All accelerometers are now at the table behind 1X4, cables are near readout box.

  7567   Wed Oct 17 09:01:37 2012 SteveUpdatePEMants on the PSL table

Quote:

We observed one or two ants climbing over PMC optics without booties and safety glasses.

The floor was mopped with strong Bayer Home Pest Control solution in the Vertex area.

Do not work inside the 40m lab if you are sensitive to chemicals!

 

 The floor is mopped again with strong PEST CONTROL SOLUTION in water in the Vertex area.

Do not plane to work in the IFO-room till noon if you are sensitive to chemicals!

  7626   Thu Oct 25 21:02:34 2012 DenUpdatePEM1x7 dc power

 We now stop using bench DC power supplies for microphone preamp and PEM AA board. DC power is wired from 1x5 rack suppliers. I've installed a beam to mount fuse houses in the 1x7 as we did not have one.

DSC_4779.JPG

  7645   Wed Oct 31 14:31:34 2012 SteveUpdatePEMhigh particle count

High particle count confirmed with #2 counter

Attachment 1: highparticlec.png
highparticlec.png
  7650   Wed Oct 31 22:56:41 2012 DenUpdatePEMacoustic noise

Microphone preamp box had a low-pass filter at 2kHz, Ayaka changed it to 20 kHz by replacing 100pF capacitor with a 10pF.

We've measured frequency response of the box. Signal from the microphone was split into two. One path went to the box, while another was amplified by the gain 20 (and bandpass filter 1Hz - 300kHz) and sent to spectrum analyzer. Coherence and frequency response were measured using box output and amplified input. Low-pass filter in the box does not limit our sensitivity.

Acoustic noise significantly decreases at frequencies higher then 2kHz. So we need to modify the circuit by adding whitening filter.

pre_after_mic.png  freq_resp.png

I've plugged in PMC length channel into PEM board CH15 through and amplifier (gain=200) that is AC coupled to avoid ~2.5 DC V coming from PMC servo.  I measured coherence with microphone that was located ~30 cm higher. Measurements show contribution of acoustic noise to PMC length in the frequency range 20-50 Hz. In this range PMC length / MC length coherence is ~0.5.

Acoustic noise couples to PMC length in a non-stationary way. 5 minutes after the first measurement I already see much higher contribution. This was already discussed here. I've made C1:X02-MADC3_TP_CH15 a DQ channel at 64kHz. This a fast PMC length channel.

Next step will be to use several microphones located around PMC for acoustic noise cancellation.

DSC_4792.JPG    DSC_4793.JPG

Attachment 3: pmc.pdf
pmc.pdf pmc.pdf
Attachment 4: pmc_high.pdf
pmc_high.pdf
  7680   Wed Nov 7 11:43:29 2012 SteveUpdatePEMhigh particle count ALART

The BS camber is open only. We should close ASAP

Outside air quality is 1.7- 2.2  million particles  / cf min of 0.5 micron

Attachment 1: highpartcount.png
highpartcount.png
  7681   Wed Nov 7 14:59:35 2012 SteveUpdatePEMhigh particle count ALART & oplevs

Quote:

The BS camber is open only. We should close ASAP

Outside air quality is 1.7- 2.2  million particles  / cf min of 0.5 micron

 Air is still bad and the chambers are closed. Before lunch  Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.

ETMX and ETMY are fine.

SRM and ITMY oplevs needs more work.

Attachment 1: oplevs.png
oplevs.png
  7733   Tue Nov 20 15:24:52 2012 AyakaUpdatePEMSeismometers and a microphone

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

  7734   Tue Nov 20 16:03:37 2012 janoschUpdatePEMSeismometers and a microphone

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 If anybody wants a fancy single-axis seismometer for a while (GS-13), then please let me know.

  7744   Mon Nov 26 10:58:20 2012 SteveUpdatePEMair cond maitenance tomorrow

Air conditioning maintenance is scheduled for tomorrow morning till noon.

  7745   Mon Nov 26 18:36:17 2012 JenneUpdatePEMBLRMS back

Quote:

 I got two seismometers and one microphone back from Tara.

They are now near the Gurlap under the MC.

 I have finally plugged GUR1 back in....it is down at ETMY for now, since that's where the cable was.  BLRMS are back up on the projector.

  7754   Tue Nov 27 11:27:22 2012 SteveUpdatePEMparticle count is going up

Jeff is still working on the filter banks.
Attachment 1: 5yparticles.png
5yparticles.png
Attachment 2: particle_6m_6m.png
particle_6m_6m.png
Attachment 3: 1y_particlec.png
1y_particlec.png
Attachment 4: 10y_particle_c.png
10y_particle_c.png
  7759   Wed Nov 28 23:18:35 2012 CharlesUpdatePEMDecreased RMS in Seismometers

The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.

Attachment 1: 8-31_11-29_PEM-RMS.jpg
8-31_11-29_PEM-RMS.jpg
  7762   Thu Nov 29 02:43:48 2012 AyakaUpdatePEMDecreased RMS in Seismometers

Quote:

The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.

The Image is replaced

[Den, Ayaka]

We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.

We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers.
GUR1Xfilterbank.pngseismometers1129.pdf


 

RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...

Attachment 2: seismometers1129.jpg
seismometers1129.jpg
  7763   Thu Nov 29 09:58:06 2012 DenUpdatePEMDecreased RMS in Seismometers

Quote:

[Den, Ayaka]

We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.
 

We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers;
 

RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...

 First of all, STS-2 is in the end of X arm, GUR2 is under BS, GUR1 is in the end of Y arm.

BLRMS were small because we applied calibration from counts to um/s two times. In the past we had calibration in the RMS BP filter bank (vel2vel = FM2). Now we have calibration in the seismometer input filter bank so we can save calibrated _OUT channels.

  7777   Mon Dec 3 16:37:09 2012 SteveUpdatePEMEM 172 microphones ordered

Quote:

I've put EM 172 microphones inside Steve's isolation box to measure their noise. I've attached mics to each other and aligned them using the tape.

At low frequencies (below 1 Hz) the noise is limited by ADC as there is a 10 Hz high-pass filter inside mic readout box.

ADC noise is measured by splitting the signal from 1 mic into 2 ADC channels.

em172.png

 BT EM172 microphones  are ordered.

  7780   Tue Dec 4 10:44:36 2012 SteveUpdatePEMshort power outage

 

CALIFORNIA INSTITUTE OF TECHNOLOGY

                 FACILITIES MANAGEMENT

 

**PLEASE POST**

 

 

Building:         Campus

 

Date:             Thursday October 04,2012

 

This morning at 2:17 a.m. much of the City of Pasadena including our Campus experienced a electric power sag of short duration, approximately 1/10 of a second. The cause was a fault on one of Pasadena’s 17KV circuits. Some sensitive equipment have been impacted.

                 

Contact:          Mike Anchondo x-4999

  7807   Tue Dec 11 08:53:52 2012 SteveHowToPEMcables needs care

How NOT to:

The janitor can not clean in areas like this. He may only steps on these cables accidentally as he dust wiping our chambers.

Attachment 1: IMG_1839.JPG
IMG_1839.JPG
  7809   Tue Dec 11 10:09:04 2012 AyakaHowToPEMcables needs care

Quote:

How NOT to:

The janitor can not clean in areas like this. He may only steps on these cables accidentally as he dust wiping our chambers.

 Sorry for the mess. I fixed it.

  7819   Thu Dec 13 01:49:53 2012 ranaUpdatePEMseismometers

Quote:

I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94).  +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.

Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions.

 Looking at the PEM BLRMS, I noticed that the GUR1Z channel had a much reduced microseism compared to the GUR1X. Looking at the BLRMS screens everything seems ON, although its a mess (too many filters in the banks, etc. - clean this up, PEM people).

 Looking at the Z channel in DTT, I see that the Z spectra looks double high pass filtered below ~1 Hz.     Needs some attention in the daytime.

From Den and Ayaka's elog entry from Nov 29, its clear that this problem is there at that time. It seems that the seismometer was not even hooked up before then. Perhaps Tara returned the seismometer around Thanksgiving and then someone here hooked it up but neglected to log this work? If so, please make an elog now describing the installation of this sensor at the 40m and log any future work which takes place at the 40m lab even if you think it is inconsequential.

Attachment 1: a.pdf
a.pdf
  7825   Thu Dec 13 21:21:34 2012 DenUpdatePEMseismometers

Quote:

 

 Looking at the PEM BLRMS, I noticed that the GUR1Z channel had a much reduced microseism compared to the GUR1X. Looking at the BLRMS screens everything seems ON, although its a mess (too many filters in the banks, etc. - clean this up, PEM people).

 Looking at the Z channel in DTT, I see that the Z spectra looks double high pass filtered below ~1 Hz.     Needs some attention in the daytime.

From Den and Ayaka's elog entry from Nov 29, its clear that this problem is there at that time. It seems that the seismometer was not even hooked up before then. Perhaps Tara returned the seismometer around Thanksgiving and then someone here hooked it up but neglected to log this work? If so, please make an elog now describing the installation of this sensor at the 40m and log any future work which takes place at the 40m lab even if you think it is inconsequential.

 Yesterday I wanted to recenter Guralps. I turned them off, understood that would be able to center them because we do not have power cable to Guralp box from Tara yet and turned them back on.

I've switched Guralp cables and spectrums are fine now.

Attachment 1: gur.pdf
gur.pdf
Attachment 2: gur_fix.pdf
gur_fix.pdf
  7831   Fri Dec 14 09:05:04 2012 steveUpdatePEMTT after eq 6.1

 

 SRM, ITMY and MC3 moved some what, but how are the TT? It may worth looking at them before we vent.

Attachment 1: eq6.1.png
eq6.1.png
  7841   Mon Dec 17 19:47:15 2012 ranaUpdatePEMSeismic StripTool config updated

I have updated the Seismic Striptool display which is plotted on the wall in the control room. Please take a look and make comments. We should finalize it and not change it anymore.

By having an unchanging display, we can get used to small changes in the seismic environment which disrupt our locking.

  1. y-scale is now linear; the log-scale was suppressing the factor of 2-3 variations which are important to us.
  2. Just as the rainbow does, the colors now go from red to purple to represent the noise from 0.1 - 30 Hz: the red traces are 0.1-0.3 Hz, the green/yellow traces are 0.3-3 Hz, and the blue/purple traces are 3-30 Hz.
  3. This is just showing GUR1. Let's try to keep this seismometer working so that we can have some long term record of the seismicity here. This means don't click off the buttons, disconnect the sensor, reboot the machine, etc. When you do do these things, elog them.
Attachment 1: SeismicRainbow.png
SeismicRainbow.png
  7923   Tue Jan 22 09:10:19 2013 SteveUpdatePEMbig foot is dirty

Please wipe, clean car wheels and wear booties entering the 40m lab.

Obviously this person has no idea about our clean room rules.

Attachment 1: IMG_1873.JPG
IMG_1873.JPG
Attachment 2: IMG_1874.JPG
IMG_1874.JPG
  7942   Thu Jan 24 16:31:46 2013 SteveUpdatePEMbuilding exterier wall painted

The wood exteier walls, gutters and doors were painted at CES-Annex building #69

Attachment 1: IMG_1880.JPG
IMG_1880.JPG
  7954   Tue Jan 29 14:34:42 2013 JenneUpdatePEMSecret Seis filters

 

The BLRMS have been bad again, since the computer crash of last week.  Finally getting around to looking into it, I discovered that there are filter banks that have the microns/sec calibration filters, which are not accessible from the sitemap.  I have added links to them for GUR1 and GUR2.  We need to make the PEM/BLRMS screens macro-expansion-y, so that I don't have to change each screen individually.

Anyhow, the BLRMS are back.

 

  7999   Tue Feb 5 09:08:11 2013 SteveUpdatePEM exterier doors to be painted

Quote:

The wood exteier walls, gutters and doors were painted at CES-Annex building #69

 The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily.  They will be painted on the out side only.  This job will be done by tomorrow noon

 Do not open chamber if you smell the paint !

Attachment 1: IMG_0055.JPG
IMG_0055.JPG
  8001   Tue Feb 5 10:18:54 2013 SteveUpdatePEMhigh particle count ALART

Quote:

Quote:

The BS camber is open only. We should close ASAP

Outside air quality is 1.7- 2.2  million particles  / cf min of 0.5 micron

 Air is still bad and the chambers are closed. Before lunch  Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.

ETMX and ETMY are fine.

SRM and ITMY oplevs needs more work.

 The bad outside air quality is pushing up the inside counts.

The outside air is 5 million counts / cf min for 0.3 micron and 2 million counts / cf min for 0.5 micron particles

Do not open chamber over 10,000 counts / cf min of 0.5 micron

Attachment 1: badair.png
badair.png
Attachment 2: 100d_airq.png
100d_airq.png
  8010   Wed Feb 6 15:10:22 2013 SteveUpdatePEMnew safety signs on exterier doors

Quote:

Quote:

The wood exteier walls, gutters and doors were painted at CES-Annex building #69

 The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily.  They will be painted on the out side only.  This job will be done by tomorrow noon

 Do not open chamber if you smell the paint !

 The east and south end of the 40m emergency exit doors received new safety signs.

Attachment 1: IMG_0068.JPG
IMG_0068.JPG
  8202   Thu Feb 28 14:51:43 2013 JenneUpdatePEM"Rock Monster" causing low frequency noise

Manasa and I have been wondering why the low frequency seismic noise has been different in the last few days/weeks.  I went over to visit the Rock Monster, a.k.a. the Earth Surface Dynamics Laboratory next door, and the grad student turned it up for me (increased water flow significantly) for a few minutes.  Then I came back to look at the BLRMS, and you can see the low frequency increase a few minutes ago, when he turned it up on high.  He said that they run it on low almost all the time, 9am-5pm ish, and on high for an hour or so at a time periodically during the day.

RockMonsterOn_ForFewMinutes.png

  8219   Mon Mar 4 11:30:47 2013 SteveUpdatePEMair cond problem

 

 The air cond is out of order at the area covered by racks 1 X 1  through 1 X 7

The arm X and Y AC units are working.

Attachment 1: 40dAC.png
40dAC.png
  8270   Mon Mar 11 16:29:29 2013 SteveUpdatePEMproposed seismometer locations

Granite base 20" x 20" x 5"  locations are on the CES side of our IFO arms:  as shown ETMY_ south-west, ETMX_north-east, ITMX_south-east . No height limitation. This side of the tube has no traffic.

SS cover McMaster#  41815T4  (H) SS container cov

Attachment 1: ETMY_sw.jpg
ETMY_sw.jpg
Attachment 2: ITMX_se.jpg
ITMX_se.jpg
Attachment 3: ETMX_ne.jpg
ETMX_ne.jpg
  8272   Mon Mar 11 19:01:21 2013 JenneUpdatePEMproposed seismometer locations

This is my interpretation of where Steve is proposing to place the seismometers (he wrote ITMX southwest, but I'm pretty sure from the photo he means southeast).

I think his point is that these locations are on the less-used side of the beam tube, so they will not be in the way.  Also, they are not underneath the tube, so we will not have any problems putting the covers on/taking them off.

 

 

SeismometerProposedPositions_11March2013.pdf

  8314   Wed Mar 20 08:59:24 2013 SteveUpdatePEMair cond maitenance is today

 Outside air quality at 8:30am  1.1 million of 0.5 micron particles cf / min with MET counter #2

AC unit output measured  directly at the outlet of east end  6,000  and south end 3,000 of 0.5 micron particles cf / min with counter #2

Lab measurement at the top of IOOC 20K of 0.5 micron particles cf / min with MET #1

HEPA filters: 2 mobile units  at the middle of the east arm at speed 100%, PSL enclosure 20% and south end flow bench are running.

The PSL, south flow bench and mobile unit (Envirco  sn69406003 ) measured zero counts of 0.5 & 0.3 micron particles

Envirco HEPA filter unit sn 69406001 measured 50 particles of 0.5micron and 320 of 0.3 micron particles cf / min with counter #2

This filter will be replaced.

                   

NOTE: inside air quality peaks we can not control. It depends on the weather and our house keeping.

When your work is particle sensitive keep your eyes on the particle counter.

 

 

 

 

Attachment 1: 100dPEM.png
100dPEM.png
  8456   Mon Apr 15 16:10:52 2013 DenUpdatePEMseismometer isolation kit

 We got granite bases today from the manufacturer. We plan to set them up on Wednesday, 8 am. Please note, there will be an installation mess at Xend, Yend and corner during ~4 hours. Let us know if you have any objections to do this at this particular time.

Installation locations are specified in elog 8270, scheme attached is valid except for Xend. Instrument will be installed on the place of nitrogen containers.

(  next to the wall at corner sout-east of the south end )

ELOG V3.1.3-