ID |
Date |
Author |
Type |
Category |
Subject |
7396
|
Mon Sep 17 07:57:58 2012 |
rana | Update | PEM | BS and AS tables for REFL change |
This was in response to my suggestion to move the REFL beam path to the table containing the BS/PRM Oplevs. From this seismic data it is clear that the BS table is no worse than the AS table, so we should plan to make the layout change during the next vent. |
7429
|
Sat Sep 22 01:03:30 2012 |
Den | Update | PEM | ETMY table |
I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94). +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.
Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions. |
Attachment 1: X.pdf
|
|
Attachment 2: Y.pdf
|
|
Attachment 3: Z.pdf
|
|
7433
|
Mon Sep 24 17:03:39 2012 |
Jenne | Update | PEM | 1X1 rack power |
Quote: |
I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94). +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.
|
I'm not sure what the problem is here. Den and I looked at it for a few minutes, before I went back to helping with putting doors on. The Sorensons are not supplying the rack power for 1X1. There are some flat cables which go from the fuses on the side of the rack up to the cable tray, and go elsewhere. Den is going to continue looking into this, but I think it's a moderately high priority, since lots of things should be getting served by that same power. |
7459
|
Mon Oct 1 19:21:03 2012 |
rana | Configuration | PEM | changed PEM DQ channels |
Changed the list of channels to be written to frames from having the IN1 suffix to OUT. Now we can load the calibration of the channel into the filter module and the DQ channel will be calibrated.
We should do this wherever possible so that our channels will have real calibrations associated with them.
SEIS_GUR1_X_OUT 256
SEIS_GUR1_Y_OUT 256
SEIS_GUR1_Z_OUT 256
SEIS_GUR2_X_OUT 256
SEIS_GUR2_Y_OUT 256
SEIS_GUR2_Z_OUT 256
SEIS_STS_1_X_OUT 256
SEIS_STS_1_Y_OUT 256
SEIS_STS_1_Z_OUT 256
SEIS_STS_2_X_OUT 256
SEIS_STS_2_Y_OUT 256
SEIS_STS_2_Z_OUT 256
SEIS_STS_3_X_OUT 256
SEIS_STS_3_Y_OUT 256
SEIS_STS_3_Z_OUT 256
MIC_1_OUT 2048
MIC_2_OUT 2048
MIC_3_OUT 2048
MIC_4_OUT 2048
MIC_5_OUT 2048
MIC_6_OUT 2048
ACC_MC1_X_OUT 2048
ACC_MC1_Y_OUT 2048
ACC_MC1_Z_OUT 2048
ACC_MC2_X_OUT 2048
ACC_MC2_Y_OUT 2048
ACC_MC2_Z_OUT 2048
XARM_DIFFERENTIAL_MOTION_IN1 256
XARM_DIFFERENTIAL_MOTION_OUT 256
YARM_DIFFERENTIAL_MOTION_IN1 256
YARM_DIFFERENTIAL_MOTION_OUT 256
Next we should up the rate at which the model runs up to 16 kHz so that we can record the microphones at 16 kHz. FM radio has information up to 20 kHz. AM radio goes up to ~8 kHz. We should be at least as modern as AM radio. How do we make the change? How do we make sure the FOTON file stays OK?
I have made some changes to the daily summary file to compensate. New files is /users/public_html/40m-summary/share/c1_summary_page.ini. |
7468
|
Wed Oct 3 15:37:04 2012 |
Steve | Update | PEM | ants on the PSL table |
We observed one or two ants climbing over PMC optics without booties and safety glasses.
The floor was mopped with strong Bayer Home Pest Control solution in the Vertex area.
Do not work inside the 40m lab if you are sensitive to chemicals!
|
Attachment 1: IMG_1690.JPG
|
|
7470
|
Wed Oct 3 16:26:58 2012 |
Manasa | Update | PEM | ants on the PSL table |
I spotted around 4 within 30 minutes working at the PSL table even after the deathly spray . They seem to be running down from the cables on the oscilloscope rack to the table and the optics. |
7487
|
Fri Oct 5 00:29:34 2012 |
Den | Update | PEM | readout box power |
Guralp readout box received +13.7 /0/ -15V instead +15V because of the broken fuse. Power provided by the source is normal.
Edit by Den: I've found a similar fuse on one of the tables and borrowed it. Guralp is not working again. |
7488
|
Fri Oct 5 01:36:49 2012 |
Den | Configuration | PEM | changed PEM DQ channels |
Quote: |
We should do this wherever possible so that our channels will have real calibrations associated with them.
Next we should up the rate at which the model runs up to 16 kHz so that we can record the microphones at 16 kHz. FM radio has information up to 20 kHz. AM radio goes up to ~8 kHz. We should be at least as modern as AM radio. How do we make the change? How do we make sure the FOTON file stays OK?
|
I've added calibration gains to Guralp (to um/sec) and EM172 (to Pa) channels.
We can run PEM at 16 kHz. I think Foton file stores both sos-representation and filter commands which are independent of the sampling frequency, so it should be possible to change model sampling frequency quickly.
In fact, we can save data at 64 kHz from iop models. I've done this once with MC_F channel. However, I did not test EM172 noise at frequencies > 1 kHz. |
7490
|
Fri Oct 5 11:11:00 2012 |
Den | Update | PEM | readout box power |
Quote: |
Guralp readout box received +13.7 /0/ -15V instead +15V because of the broken fuse. Power provided by the source is normal.
Edit by Den: I've found a similar fuse on one of the tables and borrowed it. Guralp is not working again.
|
I've meant Guralp is NOW working again  |
7539
|
Fri Oct 12 22:44:49 2012 |
Den | Update | PEM | ETMY table |
Quote: |
Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions.
|
Accelerometers were installed on the ETMY table and nearby ground to measure amplification of the seismic noise due to the table. During this experiment ground and table motions were measured simultaneously.

|
Attachment 1: etmy_x_psd.pdf
|
|
Attachment 2: etmy_y_psd.pdf
|
|
Attachment 3: etmy_z_psd.pdf
|
|
Attachment 4: etmy_coh.pdf
|
|
7540
|
Sun Oct 14 11:41:42 2012 |
Den | Update | PEM | ETMY table |
Quote: |
Accelerometers were installed on the ETMY table and nearby ground to measure amplification of the seismic noise due to the table. During this experiment ground and table motions were measured simultaneously
|
I've added xml file with measurement settings and data to 40m svn at directory 40m_seismic/etmy.
|
Attachment 2: 14OCT2012.pdf
|
|
Attachment 3: 14OCT2012.xml
|
<?xml version="1.0"?>
<!DOCTYPE LIGO_LW [
<!ELEMENT LIGO_LW ((LIGO_LW|Comment|Param|Time|Table|Array|Stream)*)>
<!ATTLIST LIGO_LW Name CDATA #IMPLIED Type CDATA #IMPLIED>
<!ELEMENT Comment (#PCDATA)>
<!ELEMENT Param (#PCDATA)>
<!ATTLIST Param Name CDATA #IMPLIED Type CDATA #IMPLIED Dim CDATA #IMPLIED
Unit CDATA #IMPLIED>
<!ELEMENT Table (Comment?,Column*,Stream?)>
<!ATTLIST Table Name CDATA #IMPLIED Type CDATA #IMPLIED>
... 231759 more lines ...
|
7541
|
Sun Oct 14 16:44:20 2012 |
Den | Update | PEM | ETMX table |
I've measured ETMX table motion compared to ground motion using accelerometers. Data and settings in the xml file are at the svn directory 40m_seismic/etmx.

|
Attachment 2: 14OCT2012.pdf
|
|
Attachment 3: 14OCT2012.xml
|
<?xml version="1.0"?>
<!DOCTYPE LIGO_LW [
<!ELEMENT LIGO_LW ((LIGO_LW|Comment|Param|Time|Table|Array|Stream)*)>
<!ATTLIST LIGO_LW Name CDATA #IMPLIED Type CDATA #IMPLIED>
<!ELEMENT Comment (#PCDATA)>
<!ELEMENT Param (#PCDATA)>
<!ATTLIST Param Name CDATA #IMPLIED Type CDATA #IMPLIED Dim CDATA #IMPLIED
Unit CDATA #IMPLIED>
<!ELEMENT Table (Comment?,Column*,Stream?)>
<!ATTLIST Table Name CDATA #IMPLIED Type CDATA #IMPLIED>
... 231779 more lines ...
|
7542
|
Sun Oct 14 17:26:03 2012 |
Den | Update | PEM | ETMX table |
Quote: |
I've measured ETMX table motion compared to ground motion using accelerometers. Data and settings in the xml file are at the svn directory 40m_seismic/etmx
|
High frequency (>60 Hz) resonances that are present at the ETMX motion spectrum seem to be understandable. Amplification ETMX/GROUND of a factor of 2 at 1 Hz is interesting. I've monitored ACC DQ channels for a few hours and noticed that usually spectrum looks like in the previous elog. But every ~40 min ETMX motion is much higher then ground motion at low frequencies (<5 Hz). I wonder if this a reaction of a table to outside disturbances or accelerometer issue.
|
Attachment 1: etmx.pdf
|
|
7543
|
Sun Oct 14 20:51:20 2012 |
Den | Update | PEM | AA board |
Quote: |
But every ~40 min ETMX motion is much higher then ground motion at low frequencies (<5 Hz). I wonder if this a reaction of a table to outside disturbances or accelerometer issue.
|
This could come from AA board, its range is +/- 2.5 V, RMS of the ETMX table motion is a few times higher then ground motion, so ETMX accelerometer signal was corrupted.

As this small AA range has already caused problems before, I decided to increase it. I've looked through the board scheme and found that all its differential line receives and output amplifiers have absolute maximum range of 40V. We used KEPKO power supply for this board with a voltage range up to 6 V. So I've replaced it with a BK PRECISION power supply and set it to +/- 15 V. Now AA board range is 7.5 V.
I'll leave accelerometers near ETMX table. It's interesting to measure table motion in the morning when trucks drive by. |
7546
|
Mon Oct 15 10:14:22 2012 |
Den | Update | PEM | AA board |
Quote: |
I'll leave accelerometers near ETMX table. It's interesting to measure table motion in the morning when trucks drive by.
|
That low frequency effect was due to AA board, now it is gone. |
7562
|
Tue Oct 16 21:57:47 2012 |
Den | Update | PEM | accelerometers |
All accelerometers are now at the table behind 1X4, cables are near readout box. |
7567
|
Wed Oct 17 09:01:37 2012 |
Steve | Update | PEM | ants on the PSL table |
Quote: |
We observed one or two ants climbing over PMC optics without booties and safety glasses.
The floor was mopped with strong Bayer Home Pest Control solution in the Vertex area.
Do not work inside the 40m lab if you are sensitive to chemicals!
|
The floor is mopped again with strong PEST CONTROL SOLUTION in water in the Vertex area.
Do not plane to work in the IFO-room till noon if you are sensitive to chemicals! |
7626
|
Thu Oct 25 21:02:34 2012 |
Den | Update | PEM | 1x7 dc power |
We now stop using bench DC power supplies for microphone preamp and PEM AA board. DC power is wired from 1x5 rack suppliers. I've installed a beam to mount fuse houses in the 1x7 as we did not have one.

|
7645
|
Wed Oct 31 14:31:34 2012 |
Steve | Update | PEM | high particle count |
High particle count confirmed with #2 counter |
Attachment 1: highparticlec.png
|
|
7650
|
Wed Oct 31 22:56:41 2012 |
Den | Update | PEM | acoustic noise |
Microphone preamp box had a low-pass filter at 2kHz, Ayaka changed it to 20 kHz by replacing 100pF capacitor with a 10pF.
We've measured frequency response of the box. Signal from the microphone was split into two. One path went to the box, while another was amplified by the gain 20 (and bandpass filter 1Hz - 300kHz) and sent to spectrum analyzer. Coherence and frequency response were measured using box output and amplified input. Low-pass filter in the box does not limit our sensitivity.
Acoustic noise significantly decreases at frequencies higher then 2kHz. So we need to modify the circuit by adding whitening filter.

I've plugged in PMC length channel into PEM board CH15 through and amplifier (gain=200) that is AC coupled to avoid ~2.5 DC V coming from PMC servo. I measured coherence with microphone that was located ~30 cm higher. Measurements show contribution of acoustic noise to PMC length in the frequency range 20-50 Hz. In this range PMC length / MC length coherence is ~0.5.
Acoustic noise couples to PMC length in a non-stationary way. 5 minutes after the first measurement I already see much higher contribution. This was already discussed here. I've made C1:X02-MADC3_TP_CH15 a DQ channel at 64kHz. This a fast PMC length channel.
Next step will be to use several microphones located around PMC for acoustic noise cancellation.

|
Attachment 3: pmc.pdf
|
|
Attachment 4: pmc_high.pdf
|
|
7680
|
Wed Nov 7 11:43:29 2012 |
Steve | Update | PEM | high particle count ALART |
The BS camber is open only. We should close ASAP
Outside air quality is 1.7- 2.2 million particles / cf min of 0.5 micron |
Attachment 1: highpartcount.png
|
|
7681
|
Wed Nov 7 14:59:35 2012 |
Steve | Update | PEM | high particle count ALART & oplevs |
Quote: |
The BS camber is open only. We should close ASAP
Outside air quality is 1.7- 2.2 million particles / cf min of 0.5 micron
|
Air is still bad and the chambers are closed. Before lunch Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.
ETMX and ETMY are fine.
SRM and ITMY oplevs needs more work. |
Attachment 1: oplevs.png
|
|
7733
|
Tue Nov 20 15:24:52 2012 |
Ayaka | Update | PEM | Seismometers and a microphone |
I got two seismometers and one microphone back from Tara.
They are now near the Gurlap under the MC. |
7734
|
Tue Nov 20 16:03:37 2012 |
janosch | Update | PEM | Seismometers and a microphone |
Quote: |
I got two seismometers and one microphone back from Tara.
They are now near the Gurlap under the MC.
|
If anybody wants a fancy single-axis seismometer for a while (GS-13), then please let me know. |
7744
|
Mon Nov 26 10:58:20 2012 |
Steve | Update | PEM | air cond maitenance tomorrow |
Air conditioning maintenance is scheduled for tomorrow morning till noon. |
7745
|
Mon Nov 26 18:36:17 2012 |
Jenne | Update | PEM | BLRMS back |
Quote: |
I got two seismometers and one microphone back from Tara.
They are now near the Gurlap under the MC.
|
I have finally plugged GUR1 back in....it is down at ETMY for now, since that's where the cable was. BLRMS are back up on the projector. |
7754
|
Tue Nov 27 11:27:22 2012 |
Steve | Update | PEM | particle count is going up |
Jeff is still working on the filter banks. |
Attachment 1: 5yparticles.png
|
|
Attachment 2: particle_6m_6m.png
|
|
Attachment 3: 1y_particlec.png
|
|
Attachment 4: 10y_particle_c.png
|
|
7759
|
Wed Nov 28 23:18:35 2012 |
Charles | Update | PEM | Decreased RMS in Seismometers |
The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal. |
Attachment 1: 8-31_11-29_PEM-RMS.jpg
|
|
7762
|
Thu Nov 29 02:43:48 2012 |
Ayaka | Update | PEM | Decreased RMS in Seismometers |
Quote: |
The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal.
|
The Image is replaced
[Den, Ayaka]
We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.
We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers.
 
RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room... |
Attachment 2: seismometers1129.jpg
|
|
7763
|
Thu Nov 29 09:58:06 2012 |
Den | Update | PEM | Decreased RMS in Seismometers |
Quote: |
[Den, Ayaka]
We found that seismometer was working and the calibration in the filter banks should have been wrong.
We turned off the all FM2 filter in RMS filter banks.
We also installed STS seismometer. It is under the BS. Now we have spectrum of three seismometers;
RA: the above plot is kind of unreadable and useless. Please replace with something legible and put in some words about why there is a wrong filter, what exactly it is, etc., etc. etc. And why would you leave in a filter which is not supposed to be on? We might as well leave a few secretly broken chairs in the control room...
|
First of all, STS-2 is in the end of X arm, GUR2 is under BS, GUR1 is in the end of Y arm.
BLRMS were small because we applied calibration from counts to um/s two times. In the past we had calibration in the RMS BP filter bank (vel2vel = FM2). Now we have calibration in the seismometer input filter bank so we can save calibrated _OUT channels. |
7777
|
Mon Dec 3 16:37:09 2012 |
Steve | Update | PEM | EM 172 microphones ordered |
Quote: |
I've put EM 172 microphones inside Steve's isolation box to measure their noise. I've attached mics to each other and aligned them using the tape.
At low frequencies (below 1 Hz) the noise is limited by ADC as there is a 10 Hz high-pass filter inside mic readout box.
ADC noise is measured by splitting the signal from 1 mic into 2 ADC channels.

|
BT EM172 microphones are ordered. |
7780
|
Tue Dec 4 10:44:36 2012 |
Steve | Update | PEM | short power outage |
CALIFORNIA INSTITUTE OF TECHNOLOGY
FACILITIES MANAGEMENT
**PLEASE POST**
Building: Campus
Date: Thursday October 04,2012
This morning at 2:17 a.m. much of the City of Pasadena including our Campus experienced a electric power sag of short duration, approximately 1/10 of a second. The cause was a fault on one of Pasadena’s 17KV circuits. Some sensitive equipment have been impacted.
Contact: Mike Anchondo x-4999 |
7807
|
Tue Dec 11 08:53:52 2012 |
Steve | HowTo | PEM | cables needs care |
How NOT to:
The janitor can not clean in areas like this. He may only steps on these cables accidentally as he dust wiping our chambers. |
Attachment 1: IMG_1839.JPG
|
|
7809
|
Tue Dec 11 10:09:04 2012 |
Ayaka | HowTo | PEM | cables needs care |
Quote: |
How NOT to:
The janitor can not clean in areas like this. He may only steps on these cables accidentally as he dust wiping our chambers.
|
Sorry for the mess. I fixed it. |
7819
|
Thu Dec 13 01:49:53 2012 |
rana | Update | PEM | seismometers |
Quote: |
I've installed Guralp readout box back and it turned out that it does not work with voltage provided from the rack (+13.76 0 -14.94). +/-12 voltage regulators inside the box convert it to -0.9 0 -12. I've connected the box to +/-15 DC voltage supply to measure seismic motion at the ETMY table. Readout box works fine with +/- 15.
Seismic noise on the ETMY table measured to be a few times higher then on the floor in horizontal direction in the frequency range 50 - 200 Hz. Attached are compared spectrums of X, Y and Z motions.
|
Looking at the PEM BLRMS, I noticed that the GUR1Z channel had a much reduced microseism compared to the GUR1X. Looking at the BLRMS screens everything seems ON, although its a mess (too many filters in the banks, etc. - clean this up, PEM people).
Looking at the Z channel in DTT, I see that the Z spectra looks double high pass filtered below ~1 Hz. Needs some attention in the daytime.
From Den and Ayaka's elog entry from Nov 29, its clear that this problem is there at that time. It seems that the seismometer was not even hooked up before then. Perhaps Tara returned the seismometer around Thanksgiving and then someone here hooked it up but neglected to log this work? If so, please make an elog now describing the installation of this sensor at the 40m and log any future work which takes place at the 40m lab even if you think it is inconsequential. |
Attachment 1: a.pdf
|
|
7825
|
Thu Dec 13 21:21:34 2012 |
Den | Update | PEM | seismometers |
Quote: |
Looking at the PEM BLRMS, I noticed that the GUR1Z channel had a much reduced microseism compared to the GUR1X. Looking at the BLRMS screens everything seems ON, although its a mess (too many filters in the banks, etc. - clean this up, PEM people).
Looking at the Z channel in DTT, I see that the Z spectra looks double high pass filtered below ~1 Hz. Needs some attention in the daytime.
From Den and Ayaka's elog entry from Nov 29, its clear that this problem is there at that time. It seems that the seismometer was not even hooked up before then. Perhaps Tara returned the seismometer around Thanksgiving and then someone here hooked it up but neglected to log this work? If so, please make an elog now describing the installation of this sensor at the 40m and log any future work which takes place at the 40m lab even if you think it is inconsequential.
|
Yesterday I wanted to recenter Guralps. I turned them off, understood that would be able to center them because we do not have power cable to Guralp box from Tara yet and turned them back on.
I've switched Guralp cables and spectrums are fine now. |
Attachment 1: gur.pdf
|
|
Attachment 2: gur_fix.pdf
|
|
7831
|
Fri Dec 14 09:05:04 2012 |
steve | Update | PEM | TT after eq 6.1 |
SRM, ITMY and MC3 moved some what, but how are the TT? It may worth looking at them before we vent. |
Attachment 1: eq6.1.png
|
|
7841
|
Mon Dec 17 19:47:15 2012 |
rana | Update | PEM | Seismic StripTool config updated |
I have updated the Seismic Striptool display which is plotted on the wall in the control room. Please take a look and make comments. We should finalize it and not change it anymore.
By having an unchanging display, we can get used to small changes in the seismic environment which disrupt our locking.
- y-scale is now linear; the log-scale was suppressing the factor of 2-3 variations which are important to us.
- Just as the rainbow does, the colors now go from red to purple to represent the noise from 0.1 - 30 Hz: the red traces are 0.1-0.3 Hz, the green/yellow traces are 0.3-3 Hz, and the blue/purple traces are 3-30 Hz.
- This is just showing GUR1. Let's try to keep this seismometer working so that we can have some long term record of the seismicity here. This means don't click off the buttons, disconnect the sensor, reboot the machine, etc. When you do do these things, elog them.
|
Attachment 1: SeismicRainbow.png
|
|
7923
|
Tue Jan 22 09:10:19 2013 |
Steve | Update | PEM | big foot is dirty |
Please wipe, clean car wheels and wear booties entering the 40m lab.
Obviously this person has no idea about our clean room rules. |
Attachment 1: IMG_1873.JPG
|
|
Attachment 2: IMG_1874.JPG
|
|
7942
|
Thu Jan 24 16:31:46 2013 |
Steve | Update | PEM | building exterier wall painted |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69 |
Attachment 1: IMG_1880.JPG
|
|
7954
|
Tue Jan 29 14:34:42 2013 |
Jenne | Update | PEM | Secret Seis filters |
The BLRMS have been bad again, since the computer crash of last week. Finally getting around to looking into it, I discovered that there are filter banks that have the microns/sec calibration filters, which are not accessible from the sitemap. I have added links to them for GUR1 and GUR2. We need to make the PEM/BLRMS screens macro-expansion-y, so that I don't have to change each screen individually.
Anyhow, the BLRMS are back.
|
7999
|
Tue Feb 5 09:08:11 2013 |
Steve | Update | PEM | exterier doors to be painted |
Quote: |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69
|
The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily. They will be painted on the out side only. This job will be done by tomorrow noon
Do not open chamber if you smell the paint ! |
Attachment 1: IMG_0055.JPG
|
|
8001
|
Tue Feb 5 10:18:54 2013 |
Steve | Update | PEM | high particle count ALART |
Quote: |
Quote: |
The BS camber is open only. We should close ASAP
Outside air quality is 1.7- 2.2 million particles / cf min of 0.5 micron
|
Air is still bad and the chambers are closed. Before lunch Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.
ETMX and ETMY are fine.
SRM and ITMY oplevs needs more work.
|
The bad outside air quality is pushing up the inside counts.
The outside air is 5 million counts / cf min for 0.3 micron and 2 million counts / cf min for 0.5 micron particles
Do not open chamber over 10,000 counts / cf min of 0.5 micron |
Attachment 1: badair.png
|
|
Attachment 2: 100d_airq.png
|
|
8010
|
Wed Feb 6 15:10:22 2013 |
Steve | Update | PEM | new safety signs on exterier doors |
Quote: |
Quote: |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69
|
The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily. They will be painted on the out side only. This job will be done by tomorrow noon
Do not open chamber if you smell the paint !
|
The east and south end of the 40m emergency exit doors received new safety signs. |
Attachment 1: IMG_0068.JPG
|
|
8202
|
Thu Feb 28 14:51:43 2013 |
Jenne | Update | PEM | "Rock Monster" causing low frequency noise |
Manasa and I have been wondering why the low frequency seismic noise has been different in the last few days/weeks. I went over to visit the Rock Monster, a.k.a. the Earth Surface Dynamics Laboratory next door, and the grad student turned it up for me (increased water flow significantly) for a few minutes. Then I came back to look at the BLRMS, and you can see the low frequency increase a few minutes ago, when he turned it up on high. He said that they run it on low almost all the time, 9am-5pm ish, and on high for an hour or so at a time periodically during the day.

|
8219
|
Mon Mar 4 11:30:47 2013 |
Steve | Update | PEM | air cond problem |
The air cond is out of order at the area covered by racks 1 X 1 through 1 X 7
The arm X and Y AC units are working. |
Attachment 1: 40dAC.png
|
|
8270
|
Mon Mar 11 16:29:29 2013 |
Steve | Update | PEM | proposed seismometer locations |
Granite base 20" x 20" x 5" locations are on the CES side of our IFO arms: as shown ETMY_ south-west, ETMX_north-east, ITMX_south-east . No height limitation. This side of the tube has no traffic.
SS cover McMaster# 41815T4 (H) SS container cov |
Attachment 1: ETMY_sw.jpg
|
|
Attachment 2: ITMX_se.jpg
|
|
Attachment 3: ETMX_ne.jpg
|
|
8272
|
Mon Mar 11 19:01:21 2013 |
Jenne | Update | PEM | proposed seismometer locations |
This is my interpretation of where Steve is proposing to place the seismometers (he wrote ITMX southwest, but I'm pretty sure from the photo he means southeast).
I think his point is that these locations are on the less-used side of the beam tube, so they will not be in the way. Also, they are not underneath the tube, so we will not have any problems putting the covers on/taking them off.

|
8314
|
Wed Mar 20 08:59:24 2013 |
Steve | Update | PEM | air cond maitenance is today |
Outside air quality at 8:30am 1.1 million of 0.5 micron particles cf / min with MET counter #2
AC unit output measured directly at the outlet of east end 6,000 and south end 3,000 of 0.5 micron particles cf / min with counter #2
Lab measurement at the top of IOOC 20K of 0.5 micron particles cf / min with MET #1
HEPA filters: 2 mobile units at the middle of the east arm at speed 100%, PSL enclosure 20% and south end flow bench are running.
The PSL, south flow bench and mobile unit (Envirco sn69406003 ) measured zero counts of 0.5 & 0.3 micron particles
Envirco HEPA filter unit sn 69406001 measured 50 particles of 0.5micron and 320 of 0.3 micron particles cf / min with counter #2
This filter will be replaced.
NOTE: inside air quality peaks we can not control. It depends on the weather and our house keeping.
When your work is particle sensitive keep your eyes on the particle counter.
|
Attachment 1: 100dPEM.png
|
|
8456
|
Mon Apr 15 16:10:52 2013 |
Den | Update | PEM | seismometer isolation kit |
We got granite bases today from the manufacturer. We plan to set them up on Wednesday, 8 am. Please note, there will be an installation mess at Xend, Yend and corner during ~4 hours. Let us know if you have any objections to do this at this particular time.
Installation locations are specified in elog 8270, scheme attached is valid except for Xend. Instrument will be installed on the place of nitrogen containers.
( next to the wall at corner sout-east of the south end ) |