  13692   Tue Mar 20 19:48:10 2018 gautamUpdatePEMtest setupaccording to the temp sensor readout, which
was ~-3.35V which corresponds to ~335K, the
temperature of the can is now 60
  
  13691   Tue Mar 20 16:56:01 2018 KiraUpdatePEMtest setupThe MOSFET was getting pretty hot, so I
switched it out to a larger heat sink and
it's fine now. I then used a function
 IMG_20180320_154516.jpgIMG_20180320_145957.jpg 
  13690   Tue Mar 20 16:53:03 2018 gautamUpdateIOOIMC loop checkupRe-measured the demod board noises after
replacing the suspect ERA-5SMs, with LO driven
by a marconi at the "nominal" level
 IMC_RF_noise.pdf 
  13689   Mon Mar 19 23:44:00 2018 gautamUpdateIOOIMC loop checkup[koji,
gautam]


I began my investigations
 IMC_RF_noise.pdfIMC_OLTF_20180320.pdf 
  13688   Mon Mar 19 15:02:29 2018 gautamUpdateALSNoisy MC sensingThe working hypothesis, since the excess
noise in single arm locks is coherent between
both arms, the excess sensing noise is frequency
  
  13687   Mon Mar 19 14:39:09 2018 johannesConfigurationComputersc1auxex replacement[gautam, johannes]

The temperature control output
channel for the XEND seismometer wasn't
  
  13686   Mon Mar 19 07:37:00 2018 Angelina PanSummary Proposed QPD Optical ArrangementI am currently working on an optical arrangement
consisting of a QPD that measures the fluctuations
of an incoming HeNe laser beam that
 angelinaCode.py.tar.bz2 
  13685   Fri Mar 16 09:36:56 2018 SteveUpdateVACRGA scan at day 511,218dPumpdown 80 at 511 days and pd80b at 218
days

Valve configuration:  special
 RgaScan511d.png08.png 
  13684   Thu Mar 15 17:33:56 2018 KiraUpdatePEMtest setupI have attached the setup I completed today.
The metal box contains the heater circuit
and the board for the temperature sensor
 IMG_20180315_172512.jpg 
  13683   Thu Mar 15 16:00:25 2018 Larry WallaceSummaryComputersCert renewal for NODUSThe cert for nodus has been renewed for
another 2 years.

The following is the basic
  
  13682   Wed Mar 14 23:58:30 2018 johannesConfigurationComputersc1auxex replacementI replaced the borrowed server with the
permanent one today. Before Removing the
current server, Before, I performed several
  
  13681   Tue Mar 13 20:03:16 2018 johannesConfigurationComputersc1auxex replacementI assembled the rack-mount server that
will long-term replace c1auxex, so we can
return the borrowed unit to Larry.
 IMG_20180313_105154890.jpgIMG_20180313_133031002.jpg 
  13680   Mon Mar 12 23:57:31 2018 gautamUpdateALSNoisy POX[kevin,
gautam]
 NoisyPOXandPOY_20180312.pdfnoisyPOX_Jan2018.pdf 
  13679   Mon Mar 12 22:08:31 2018 gautamUpdateALSNoisy POX[kevin, gautam]

we tested my noisy POX hypothesis
tonight. By locking the single arm with POX,
 NoisyPOX_20180312.pdf 
  13678   Mon Mar 12 13:58:37 2018 gautamUpdateGeneralprojector light bulb blownBulb went out ~10am today. Looks like the
lifetime of this bulb was <100 days.

Steve:
  
  13677   Fri Mar 9 20:35:41 2018 Udit KhandelwalSummaryGeneralSummary 2018/03/091. Optical Table Layout 

I had discussed with Koji a way
to record coordinates of optical table equipments
  
  13676   Fri Mar 9 12:59:53 2018 KiraUpdatePEMADC noise measurement[Kira, Gautam]

I ceated a simple circuit that
takes in 15V and outputs precisely 5V by
 IMG_20180309_114345.jpgIMG_20180309_125153.jpg 
  13675   Fri Mar 9 01:07:01 2018 gautamUpdateALSFirst look at new ALS electronics[koji,
gautam]
 ALS_vs_POXnoise.pdf 
  13674   Thu Mar 8 23:50:27 2018 gautamUpdateALSFirst look at new ALS electronics
Locked single arms, dither aligned,
and saved offsets to EPICS (slow) sliders
in anticipation of having to reboot all vertex
 BeatMouth_OOL.pdf 
  13673   Thu Mar 8 19:38:37 2018 gautamUpdateALSdigital unwhitening of daughter boardI made a LISO fit of the measured TF of
the daughter board, so that I can digitally
invert the daughter board whitening. Results
 TFfit.pdf 
  13672   Thu Mar 8 18:15:42 2018 gautamUpdateGeneralCDS recovery after work at LSC rackI was forced into a simultaneous power-cycle
rebooting of the three vertex FEs just now.
I took the opportunity to completely disconnect
  
  13671   Thu Mar 8 15:23:16 2018 gautamUpdateElectronicsNew DC power ports at c1lsc[Koji,
Gautam]
  
  13670   Thu Mar 8 14:41:25 2018 gautamUpdateGeneralCDS recovery after work at LSC rackAs I had found before,
restarting the c1oaf
 CDS-recovery.pngc1sus_timing.pngc1sus_crashed.png 
  13669   Thu Mar 8 01:10:22 2018 gautamUpdateGeneralCDS recovery after work at LSC rackThis required multiple hard reboots, but
seems like all the RT models are back for
now. The only indicator I can't explain
 CDS-recovery.png 
  13668   Thu Mar 8 00:40:25 2018 gautamUpdateALSnew look ALS electronics - characterizationI am almost ready for a digital test of
the new ALS electronics. Today, Koji and
I spent some time tapping new +/-24VDC DIN
 BeatMouthX_20180305_diffOut.pdf 
  13667   Wed Mar 7 12:04:14 2018 gautamUpdateElectronicsThree opamps walked onto an AA boardHere are the plots. Comments:


Measurement
and model agree quite well 
 AA_TF.pdfAA_noise.pdfD070081_LTspiceFiles.zip 
  13666   Mon Mar 5 17:27:34 2018 gautamUpdateALSnew look ALS electronics - characterizationI did a quick test of the noise of the
new ALS electronics with the X arm ALS. Attachment
#1 shows the results - but something looks
 BeatMouthX_20180305_diffOut.pdf 
  13665   Mon Mar 5 11:58:24 2018 gautamUpdateElectronicsThree opamps walked onto an AA boardFor testing the new IR ALS noise, we had
decided that we would like to use the differential
output of the demodulated ALS beat signal,
 AA_TF.pdfAA_noise.pdf 
  13664   Mon Mar 5 10:13:21 2018 gautamUpdateVACvacuum healthIn Steve's absence, I've tried
to keep an eye on the health of the vacuum
system. From Attachment #1, the pressure
 VacuumStatus.png 
  13663   Fri Mar 2 01:45:06 2018 gautamUpdateALSnew look ALS electronicsI spent today making another daughter board
(so that we can use the new scheme for I
and Q for one arm), testing it (i.e. measuring
 IMG_6916.JPG 
  13662   Wed Feb 28 21:14:34 2018 gautamSummaryPEMChannel adminSince we decided to use the Acromag for
readback of the temperature sensor for Kira's
seismometer temperature control, I enabled
 tempSensData.png 
  13661   Wed Feb 28 19:13:25 2018 gautamUpdateALSADC test for differential receiving in c1lsc[koji,
gautam]

we did a bunch of tests to figure
 c1lsc_ADCtest.png 
  13660   Wed Feb 28 12:31:28 2018 KiraUpdatePEMtemp sensor inputI switched out the DIN fuses for the long
cables and it fixed the issue of them
not showing any votage on the other end.
  
  13658   Tue Feb 27 21:10:45 2018 gautamUpdateALSDaughter board testingI thought a little bit about the next steps
in testing the daughter board. The idea is
to install this into the existing 1U chassis
 IMG_6912.JPGIMG_6913.JPGIMG_6914.JPG 
  13657   Mon Feb 26 20:55:56 2018 ranaUpdateALSDaughter board prototypingLooks good.

* for bypass type applications,
you don't have to use Wima caps (which
  
  13656   Mon Feb 26 16:22:10 2018 KiraUpdatePEMtemp sensor input[Kira, Gautam]

We began the setup for the lab
temperature sensor today. First, we needed
 1.jpgIMG_20180226_154649.jpgIMG_20180226_154720.jpg 
  13655   Sun Feb 25 00:03:12 2018 gautamUpdateALSDaughter board prototypingUsing one of the prototype  schematic.pdfdaughterBoard_TF.pdfdaughterBoard_noise.pdfTEK00000.PNGdaughterBoard_noise.pdf 
  13654   Fri Feb 23 20:46:04 2018 Udit KhandelwalSummaryGeneralCAD Summary 2018/02/23I have more or less finished cadding the
test mass chamber by referring to the drawings
Steve gave me. Finer details like lugs and
 tmc_20180223.JPG 
  13653   Fri Feb 23 07:47:54 2018 SteveUpdateVACCC1 HornetWe have the IFO pressure logged again!
Thanks Johannes and Gautam

This
 InstruTech_Hornet_CC1.pngCC1_readback_updated.png 
  13652   Thu Feb 22 17:19:47 2018 KojiUpdateGeneralModulation depth measurement for an aLIGO EOMaLIGO
EOM test: Setup


The modulation signal was
 IMG_3436.JPGmodulation_depth.pdfmodulation_linearity.pdfmodulation_24MHz.pdf 
  13651   Thu Feb 22 16:16:43 2018 KiraUpdatePEMtemp sensor inputRewired the temperature sensor inputs to
Molex connectors so that we can now attach
them to the +/- 15V Sorensens for input instead
 IMG_20180222_160602.jpg 
  13650   Thu Feb 22 16:11:14 2018 KojiUpdateGeneralaLIGO EOM crystal replacementaLIGO
EOM crystal replacement


The entire operation has
 IMG_3421.JPGIMG_3426.JPGIMG_3427.JPG 
  13649   Thu Feb 22 10:49:11 2018 SteveUpdateElectronicsrack power supplies checkedAll rack power supplies labeled if their
load changed.

 
 8x 
  13648   Thu Feb 22 00:09:11 2018 gautamUpdateALSD0902745 in-situ testingI thought a little bit about the design
of the preamp we want for the demodulated
ALS signals today. The requirements are:
 daughterBoard_TF.pdfdaughterBoard_noise.pdfIMG_6906.JPG 
  13647   Wed Feb 21 17:20:32 2018 johannesUpdateVACHornet gauge connected to DAQ.I wired the six available BNC connectors
on the front panel of the new XEND slow DAQ
to physical Acromag channels. There were
 VacLog.png 
  13646   Wed Feb 21 12:17:04 2018 gautamUpdateCDSLO Power mon channels added to c1lscTo make this setup more permanent, I modified
the c1lsc
  
  13645   Wed Feb 21 00:04:27 2018 gautamUpdateElectronicsTemporary RF power monitor setupI made a voltage divider using a 20.47kohm
and 1.07kohm (both values measured with a
DMM). The whole thing is packaged inside
 RFbeatAmp.pngBeatMouthX_RFAM_20180221.pdf 
  13644   Tue Feb 20 23:08:27 2018 gautamUpdateALSD0902745 in-situ testingAttachment #1 shows
the ALS noise measurement today. Main differences
from the spectrum posted last week is that
 BeatMouthX_20180220_diffOut.pdf 
  13643   Tue Feb 20 21:14:59 2018 gautamUpdateCDSRFM network errorsI wanted to lock the single arm POX/POY
config to do some tests on the BeatMouth.
But I was unable to.
  
  13642   Tue Feb 20 13:59:30 2018 KojiUpdateGeneralModulation depth measurement for an aLIGO EOMLast night I worked at the PSL table for
the modulation depth measurement for an aLIGO
EOM. Let me know if the IFO behavior is unusual.
  