  13721   Fri Mar 30 06:14:31 2018 ranaUpdateIOOMCREFL_PD Optical response measurementthe noise eater on/off measurements should
be done for 0-100 kHz and from the demod
board output monitor
  
  13720   Fri Mar 30 03:23:50 2018 KojiUpdateGeneralaLIGO EOM workI have been working on the aux beat setup
on the PSL table between 9PM-3AM.

This work involved:
  
  13719   Thu Mar 29 17:57:36 2018 arijitUpdateIOOMCREFL_PD Optical response measurementKevin, Gautam and Arijit

Today we performed the in-loop
noise measurements of the MCREFL-PD using
 NOISE_EATER_On_OFF.pdf 
  13718   Thu Mar 29 17:14:42 2018 KiraUpdatePEMPID test[Kira, Gautam]

We closed the loop today and implemented
the PID script. I have attached the StripTool
 PID_test.png 
  13717   Thu Mar 29 12:03:37 2018 Jon RichardsonSummaryGeneralProof-of-Concept SRC Gouy Phase MeasurementI've been developing an idea for making
a direct measurement of the SRC Gouy phase at
RF. It's a very different approach from
 src_gouy_phase_v3.pdfmeasurement_block_diagram.png 
  13716   Wed Mar 28 21:47:37 2018 arijitUpdateIOOMCREFL_PD Optical response measurementKevin, Gautam and Arijit

We did a optical measurement of
the MCREFL_PD transimpedance using the Jenny
 LISO_fit_with_info.pdfLISO_fit.zip 
  13715   Wed Mar 28 21:31:39 2018 gautamUpdateIOOMC REFL PD removedI re-installed the MC REFL photodiode.
Centered beam on the PD by adjusting steering
mirror to maximize the DC signal level (on
  
  13714   Wed Mar 28 17:28:58 2018 SteveUpdatePSLnoise eater on or offTill RIN measurement noise eater is off
on 2W laser. The inno 1W  has no noise
eater.
 inno2W.pnginno1W.png 
  13713   Wed Mar 28 16:44:27 2018 SteveUpdateGeneralAP table todayMCRefl is absent, it is under investigation.
I removed a bunch of hardware and note all
spare optics along the edges.
 AP_Table_20180328.png 
  13712   Tue Mar 27 23:37:35 2018 gautamUpdateIOOMC REFL PD removedI've
removed the MC REFL PD unit from the AS table
 IMG_6955.JPG 
  13711   Tue Mar 27 19:32:03 2018 arijitUpdateIOOPSL noise eater was offKevin, Gautam and Arijit

We made a measurement of the MC_REFL
photodiode transfer function using the network
 MCREFL_TF.pdfMCREFL_SPECTRUM.pdf 
  13710   Tue Mar 27 11:11:16 2018 KiraUpdatePEMChannel setup[Kira, Gautam]

We setup the channels for PID control
of the seismometer can. First, we ssh into
  
  13709   Tue Mar 27 08:58:21 2018 SteveUpdateVACVM1 opened.......scan fine 




 rga2018march27.png 
  13708   Tue Mar 27 01:39:44 2018 gautamUpdateIOOPSL noise eater was offWhile Kevin and Arijit were doing their
MC_REFL PD noise measurements (which they
will elog about separately shortly), I noticed
 IR_ALS_20180326.pdf 
  13707   Mon Mar 26 23:49:27 2018 gautamUpdateGeneralNew ADC Adaptor Board installed in C1LSC expansion chassisTodd informed me that the ADC Timing adaptor
boards we had ordered arrived today. I had
to solder on the components and connectors
 CDS_20180326.png 
  13706   Mon Mar 26 21:40:26 2018 gautamSummaryIOOMC2 classical radiation pressure noise[rana, gautam]

we measured the RIN of the MC2
transmission using the PDA255 I had put on
 IMC_RIN.pdfIMC_RadPress.pdfMC2_radPress.tgz 
  13705   Mon Mar 26 21:25:55 2018 ranaSummaryIOOMC2 Trans table has issuesGautam, Rana

While at the MC2 table, we noticed
that it has some optical problems:
 20180326_201929.jpg 
  13704   Mon Mar 26 16:10:33 2018 KiraUpdatePEMfinal setup sketchI made sketches of the final setup. There
will be a box in the rack that contains both
the heater circuit and the temperature sensor
 heater_1_new.pngheater_2_new.pngHeaterCircuit.pdftemp_sensor.png 
  13703   Mon Mar 26 10:15:20 2018 ArijitUpdateIOOPMC and IMC re-locked[Gautam, Arijit]

PMC and IMC re-aligned and re-locked.
Both cavities are staying locked. Arm cavities
  
  13702   Mon Mar 26 09:25:18 2018 SteveUpdateVACVM1 openedCC1 old MKS cold cathode gauge randomly
turns on- off. This makes software interlock
close VM1 to protect RGA  So the closed
 CC4VM1.png 
  13701   Fri Mar 23 12:45:08 2018 KiraUpdatePEMtest setupI fit the data that we got from the test.
The time constant for the cooling came out
to be about 4.5 hours. The error is quite
 seis_fit.png 
  13700   Fri Mar 23 12:00:20 2018 ranaUpdatePEMtest setupwe don't ever want to use our 16 kHz
real time system for such low frequency action;
its main purpose is for real-time controls,
  
  13699   Thu Mar 22 17:47:16 2018 Angelina PanSummary Proposed QPD Optical Arrangement   IMG_0869.jpg 
  13698   Wed Mar 21 21:13:44 2018 gautamUpdateIOOIMC noise budgetI've added two curves to the NB. Both
are measured (with FET preamp) at the output
of the demod board, with the LO driven at
 IMC_RF_noise_calib.pdf 
  13697   Wed Mar 21 17:31:21 2018 gautamUpdateIOOMC error point calibrationI did a preliminary noise budget of the
transmitted frequency noise of the IMC. Attachment
#1 shows the NB. I'm going to use this
 IMC_RF_noise_calib.pdf 
  13696   Wed Mar 21 15:52:45 2018 gautamUpdateIOOMC error point calibrationAs discussed at the meeting, I decided
to calibrate the MC error point into physical
units of Hz/rtHz (a.k.a. the 
 IMC_PDHdisc_20180321.pdfIMC_PDHdisc785_20180321.pdfIMC_oscope.pdf 
  13695   Wed Mar 21 10:00:35 2018 steveUpdateGeneralprojector light bulb replacedLight bulb replaced.




  
  13694   Tue Mar 20 22:44:45 2018 gautamUpdateIOOIMC loop checkupAfter some persistence, I managed to get
the mixers off.


Having gotten the mxiers
  
  13693   Tue Mar 20 21:08:03 2018 gautamUpdateIOOIMC loop checkupThis elog
by koji inspired me to consider power supply
as a possible issue.
  
  13692   Tue Mar 20 19:48:10 2018 gautamUpdatePEMtest setupaccording to the temp sensor readout, which
was ~-3.35V which corresponds to ~335K, the
temperature of the can is now 60
  
  13691   Tue Mar 20 16:56:01 2018 KiraUpdatePEMtest setupThe MOSFET was getting pretty hot, so I
switched it out to a larger heat sink and
it's fine now. I then used a function
 IMG_20180320_154516.jpgIMG_20180320_145957.jpg 
  13690   Tue Mar 20 16:53:03 2018 gautamUpdateIOOIMC loop checkupRe-measured the demod board noises after
replacing the suspect ERA-5SMs, with LO driven
by a marconi at the "nominal" level
 IMC_RF_noise.pdf 
  13689   Mon Mar 19 23:44:00 2018 gautamUpdateIOOIMC loop checkup[koji,
gautam]


I began my investigations
 IMC_RF_noise.pdfIMC_OLTF_20180320.pdf 
  13688   Mon Mar 19 15:02:29 2018 gautamUpdateALSNoisy MC sensingThe working hypothesis, since the excess
noise in single arm locks is coherent between
both arms, the excess sensing noise is frequency
  
  13687   Mon Mar 19 14:39:09 2018 johannesConfigurationComputersc1auxex replacement[gautam, johannes]

The temperature control output
channel for the XEND seismometer wasn't
  
  13686   Mon Mar 19 07:37:00 2018 Angelina PanSummary Proposed QPD Optical ArrangementI am currently working on an optical arrangement
consisting of a QPD that measures the fluctuations
of an incoming HeNe laser beam that
 angelinaCode.py.tar.bz2 
  13685   Fri Mar 16 09:36:56 2018 SteveUpdateVACRGA scan at day 511,218dPumpdown 80 at 511 days and pd80b at 218
days

Valve configuration:  special
 RgaScan511d.png08.png 
  13684   Thu Mar 15 17:33:56 2018 KiraUpdatePEMtest setupI have attached the setup I completed today.
The metal box contains the heater circuit
and the board for the temperature sensor
 IMG_20180315_172512.jpg 
  13683   Thu Mar 15 16:00:25 2018 Larry WallaceSummaryComputersCert renewal for NODUSThe cert for nodus has been renewed for
another 2 years.

The following is the basic
  
  13682   Wed Mar 14 23:58:30 2018 johannesConfigurationComputersc1auxex replacementI replaced the borrowed server with the
permanent one today. Before Removing the
current server, Before, I performed several
  
  13681   Tue Mar 13 20:03:16 2018 johannesConfigurationComputersc1auxex replacementI assembled the rack-mount server that
will long-term replace c1auxex, so we can
return the borrowed unit to Larry.
 IMG_20180313_105154890.jpgIMG_20180313_133031002.jpg 
  13680   Mon Mar 12 23:57:31 2018 gautamUpdateALSNoisy POX[kevin,
gautam]
 NoisyPOXandPOY_20180312.pdfnoisyPOX_Jan2018.pdf 
  13679   Mon Mar 12 22:08:31 2018 gautamUpdateALSNoisy POX[kevin, gautam]

we tested my noisy POX hypothesis
tonight. By locking the single arm with POX,
 NoisyPOX_20180312.pdf 
  13678   Mon Mar 12 13:58:37 2018 gautamUpdateGeneralprojector light bulb blownBulb went out ~10am today. Looks like the
lifetime of this bulb was <100 days.

Steve:
  
  13677   Fri Mar 9 20:35:41 2018 Udit KhandelwalSummaryGeneralSummary 2018/03/091. Optical Table Layout 

I had discussed with Koji a way
to record coordinates of optical table equipments
  
  13676   Fri Mar 9 12:59:53 2018 KiraUpdatePEMADC noise measurement[Kira, Gautam]

I ceated a simple circuit that
takes in 15V and outputs precisely 5V by
 IMG_20180309_114345.jpgIMG_20180309_125153.jpg 
  13675   Fri Mar 9 01:07:01 2018 gautamUpdateALSFirst look at new ALS electronics[koji,
gautam]
 ALS_vs_POXnoise.pdf 
  13674   Thu Mar 8 23:50:27 2018 gautamUpdateALSFirst look at new ALS electronics
Locked single arms, dither aligned,
and saved offsets to EPICS (slow) sliders
in anticipation of having to reboot all vertex
 BeatMouth_OOL.pdf 
  13673   Thu Mar 8 19:38:37 2018 gautamUpdateALSdigital unwhitening of daughter boardI made a LISO fit of the measured TF of
the daughter board, so that I can digitally
invert the daughter board whitening. Results
 TFfit.pdf 
  13672   Thu Mar 8 18:15:42 2018 gautamUpdateGeneralCDS recovery after work at LSC rackI was forced into a simultaneous power-cycle
rebooting of the three vertex FEs just now.
I took the opportunity to completely disconnect
  