  16207   Wed Jun 16 20:32:39 2021 YehonathanUpdateCDSOpto-isolator for c1auxeyI installed 2 additional isolators in the
Acromag chassis. I set all the input channels
to PNP. I ran the digital inputs (EnableMon
  
  16208   Thu Jun 17 11:19:37 2021 Ian MacMillanUpdateCDSCDS UpgradeJon and I tested the ADC and DAC cards
in both of the systems on the test stand.
We had to swap out an 18-bit DAC for a 16-bit
  
  16217   Mon Jun 21 17:15:49 2021 Ian MacMillanUpdateCDSCDS UpgradeAnchal and I wrote a script (Attachment
1) that will test the ADC and DAC connections
with inputs on the INMON
 DAC2ADC_Test.py 
  16220   Tue Jun 22 16:53:01 2021 Ian MacMillanUpdateCDSFront-End Assembly and TestingThe channels on both the C1BHD and C1SUS2
seem to be frozen: they arent updating and
are holding one value. To fix this Anchal
  
  16224   Thu Jun 24 17:32:52 2021 Ian MacMillanUpdateCDSFront-End Assembly and TestingAnchal and I ran tests on the two systems
(C1-SUS2 and C1-BHD). Attached are the results
and the code and data to recreate them.
 C1-SU2_Channel_Responses.pdfC1-BHD_Channel_Responses.pdfCDS_Channel_Test.zip 
  16225   Fri Jun 25 14:06:10 2021 JonUpdateCDSFront-End Assembly and TestingSummary

Here is the final
summary (from me) of where things stand with
the new front-end systems. With Anchal and
 test_stand.JPG 
  16230   Wed Jun 30 14:09:26 2021 Ian MacMillanUpdateCDSSUS simPlant modelI have looked at my code from the previous
plot of the transfer function and realized
that there is a slight error that must be
 SingleSusPlantTF.pdfPlant_TF_Test.zip 
  16243   Fri Jul 9 18:35:32 2021 YehonathanUpdateCDSOpto-isolator for c1auxeyFollowing Koji's channel list review,
we made changes to the
wiring spreadsheet.
  
  16244   Mon Jul 12 18:06:25 2021 YehonathanUpdateCDSOpto-isolator for c1auxeyI edited /cvs/cds/caltech/target/c1auxey1/ETMYaux.db
(after creating a backup) and added the spare
coil driver channels.
  
  16263   Wed Jul 28 12:47:52 2021 YehonathanUpdateCDSOpto-isolator for c1auxeyTo simulate a differential output I used
two power supplies connected in series. The
outer connectors were used as the outputs
 SUS-ETMY_SparePDMon0_NoRef.pngSUS-ETMY_SparePDMon0_Ref_WithGND.pngDifferentialOutputTest.png 
  16276   Wed Aug 11 12:06:40 2021 YehonathanUpdateCDSOpto-isolator for c1auxeyI redid the differential input experiment
using the DS360 function generator we recently
got. I generated a low frequency (0.1Hz)
 SUS-ETMY_SparePDMon0_2.png 
  16280   Mon Aug 16 23:30:34 2021 PacoUpdateCDSAS WFS commissioning; restarting models[koji, ian, tega, paco]

With the remote/local assistance
of Tega/Ian last friday I made changes on
  
  16283   Thu Aug 19 03:23:00 2021 AnchalUpdateCDSTime synchornization not runningI tried to read a bit and understand the
NTP synchronization implementation in FE
computers. I'm quite sure that NTP synchronization
  
  16284   Thu Aug 19 14:14:49 2021 KojiUpdateCDSTime synchornization not running131.215.239.14 looks like Caltech's
NTP server (ntp-02.caltech.edu)
https://webmagellan.com/explore/caltech.edu/28415b58-837f-4b46-a134-54f4b81bee53
  
  16285   Fri Aug 20 00:28:55 2021 AnchalUpdateCDSTime synchornization not runningI added ntpserver as a known host name
for address 192.168.113.201 (fb1's address
where ntp server is running) in the martian
  
  16286   Fri Aug 20 06:24:18 2021 AnchalUpdateCDSTime synchornization not runningI read on some stack exchange that 'NTP
synchornized' indicator turns 'yes'
in the output of command timedatectl only
  
  16289   Mon Aug 23 15:25:59 2021 Ian MacMillanUpdateCDSSUS simPlant modelI am adding a State-space block to the
SimPlant cds model using the example Chris
gave. I made a new folder in controls
 SimPlantStateSpace.zip 
  16294   Tue Aug 24 18:44:03 2021 KojiUpdateCDSFB is writing the frames with a year old dateDan Kozak pointed out that the new frame
files of the 40m has not been written in
2021 GPS time but 2020 GPS time.
 Screen_Shot_2021-08-24_at_18.46.24.png 
  16297   Wed Aug 25 11:48:48 2021 YehonathanUpdateCDSc1auxey assemblyAfter confirming that, indeed, leaving
the RTN connection floating can cause reliability
issues we decided to make these connections
  
  16298   Wed Aug 25 17:31:30 2021 PacoUpdateCDSFB is writing the frames with a year old date[paco, tega, koji]

After invaluable assistance from
Jamie in fixing this yearly offset in the
 TRX_noise_2021-08-25_17-40-55.pngTRX_TRY_power_spectra.pdf 
  16299   Wed Aug 25 18:20:21 2021 JamieUpdateCDSGPS time on fb1 fixed, dadq writing correct frames againI have no idea what happened to the GPS
timing on fb1, but it seems like the issue
was coincident with the power
  
  16300   Thu Aug 26 10:10:44 2021 PacoUpdateCDSFB is writing the frames with a year old date[paco, ]

We went over the X end to check
what was going on with the TRX signal. We
 Screenshot_from_2021-08-26_10-09-50.pngTRXTRY_Spectra.pdf 
  16302   Thu Aug 26 10:30:14 2021 JamieConfigurationCDSfront end time synchronization fixed?I've been looking at why the front
end NTP time synchronization did not seem
to be working.  I think it might not
  
  16309   Thu Sep 2 19:47:38 2021 KojiUpdateCDSThis week's FB1 GPS Timing Issue SolvedAfter the reboot daqd_dc was not working,
but manual starting of open-mx / mx services
solved the issue.
  
  16310   Thu Sep 2 20:44:18 2021 KojiUpdateCDSChiara DHCP restartedWe had the issue of the RT machines rebooting.
Once we hooked up the display on c1iscex,
it turned out that the IP was not given at
  
  16311   Thu Sep 2 20:47:19 2021 KojiUpdateCDSChiara DHCP restarted[Paco, Tega, Koji]

Once chiara's DHCP is back,
things got much more straight forward.
  
  16321   Mon Sep 13 14:32:25 2021 YehonathanUpdateCDSc1auxey assemblySo we agreed that the RTNs points on the
c1auxex Acromag chassis should just be grounded
to the local Acromag ground as it just needs
 ETMX_OSEMS_Noise.png 
  16325   Tue Sep 14 15:57:05 2021 jamieFrogsCDSfb1 /var full after reboot, caused all sorts of problems/var on fb1 filled up today, which caused
all sorts of CDS issues.  I found out
about the problem by reading the logs of
  
  16327   Tue Sep 14 16:44:54 2021 jamieFrogsCDSfb1 /var full after reboot, caused all sorts of problemsJonathan Hanks pointed me to this fix to
the gpstime kernel module that was unfortunately
put in after the 3.4 release that we're
  
  16330   Tue Sep 14 17:22:21 2021 AnchalUpdateCDSAdded temp sensor channels to DAQ list[Tega, Paco, Anchal]

We attempted to reboot fb1 daqd
today to get the new temperature sensor channels
  
  16332   Wed Sep 15 11:27:50 2021 YehonathanUpdateCDSc1auxey assembly{Yehonathan, Paco}

We turned off the ETMX watchdogs
and OpLevs. We went to the X end and shut
 c1auxex_Grounding_OSEM_comparison1.pdfc1auxex_Grounding_OSEM_comparison2.pdf 
  16351   Tue Sep 21 11:09:34 2021 AnchalSummaryCDSXARM YARM UGF Servo and Oscillators addedI've updated the c1LSC simulink model
to add the so-called UGF servos in the XARM
and YARM single arm loops as well. These
  
  16354   Wed Sep 22 12:40:04 2021 AnchalSummaryCDSXARM YARM UGF Servo and Oscillators shifted to OAFTo reduce burden on c1lsc, I've shifted
the added UGF block to to c1oaf model. c1lsc
had to be modified to allow addition of an
  
  16365   Wed Sep 29 17:10:09 2021 AnchalSummaryCDSc1teststand problems summary[anchal, ian]

We went and collected some information
for the overlords to fix the c1teststand
 c1teststand_issues_summary.pdf 
  16367   Thu Sep 30 14:09:37 2021 AnchalSummaryCDSNew way to ssh into c1teststandLate elog, original time
Wed Sep 29 14:09:59 2021

We opened a new port (22220) in
  
  16372   Mon Oct 4 11:05:44 2021 AnchalSummaryCDSc1teststand problems summary[Anchal, Paco]

We tried to fix the ntp synchronization
in c1teststand today by repeating the steps
  
  16376   Mon Oct 4 18:00:16 2021 KojiSummaryCDSc1teststand problems summaryI don't know anything about
mx/open-mx, but you also need open-mx,don't
you?
  
  16381   Tue Oct 5 17:58:52 2021 AnchalSummaryCDSc1teststand problems summaryopen-mx service is running successfully
on the fb1(clone), c1bhd and c1sus.

  
  16382   Tue Oct 5 18:00:53 2021 AnchalSummaryCDSc1teststand time synchronization working nowToday I got a new router that I used to
connect the c1teststand, fb1 and chiara.
I was able to see internet access in c1teststand
  
  16391   Mon Oct 11 17:31:25 2021 AnchalSummaryCDSFixed mounting of mx devices in fb. daqd_dc is running now.

  
  16392   Mon Oct 11 18:29:35 2021 AnchalSummaryCDSMoving forward?The teststand has some non-trivial issue
with Myrinet card (either software or hardware)
which even teh experts are saying they don't
  
  16395   Tue Oct 12 17:10:56 2021 AnchalSummaryCDSSome more informationChris pointed out some information displaying
scripts, that show if the DAQ network is
working or not. I thought it would be nice
  
  16396   Tue Oct 12 17:20:12 2021 AnchalSummaryCDSConnected c1sus2 to martian networkI connected c1sus2 to the martian network
by splitting the c1sim connection with a
5-way switch. I also ran another ethernet
  
  16397   Tue Oct 12 23:42:56 2021 KojiSummaryCDSConnected c1sus2 to martian networkDon't you need to add the new
hosts to /diskless/root/etc/rtsystab at fb1?
--> There looks many elogs talking about
  
  16398   Wed Oct 13 11:25:14 2021 AnchalSummaryCDSRan c1sus2 models in martian CDS. All good!Three extra steps (when adding
new models, new FE):


Chris
pointed out that the sudo command in c1sus2
 CDS_screens_running.png 
  16414   Tue Oct 19 18:20:33 2021 Ian MacMillanSummaryCDSc1sus2 DAC to ADC testI ran a DAC to ADC test on c1sus2 channels
where I hooked up the outputs on the DAC
to the input channels on the ADC. We used
 data3_Plots.pdfdata2_Plots.pdf 
  16415   Tue Oct 19 23:43:09 2021 KojiSummaryCDSc1sus2 DAC to ADC test(Because of a totally unrelated reason)
I was checking the electronics units for
the upgrade. And I realized that
 P_20211019_224433.jpgP_20211019_224122.jpgP_20211019_224400.jpgP_20211019_224411.jpg 
  16417   Wed Oct 20 11:48:27 2021 AnchalSummaryCDSPower supple configured correctly.This was horrible! That's my bad, I
should have checked the configuration before
assuming that it is right.
  
  16430   Tue Oct 26 18:24:00 2021 Ian MacMillanSummaryCDSc1sus2 DAC to ADC test[Ian, Anchal, Paco]

After the Koji found that there
was a
 data2_Plots.pdfdata3_Plots.pdf 
  16495   Thu Dec 9 00:32:56 2021 TegaUpdateCDSNew SUS medm screen updateThe new SUS screen can be reached via sitemap
-> IFO SUS button -> NEW
ETMX dropdown menu link. Please use and provide
 Screen_Shot_2021-12-09_at_12.29.30_AM.pngScreen_Shot_2021-12-09_at_12.42.35_AM.png 