8401
|
Wed Apr 3 14:46:17 2013 |
Gabriele | Summary | LSC | Error signal simulation in PRMI |
Here is a summary of a simulation of the error signal behavior in the PRMI configuration. The main parameters are:
L_PRC = 6.7538 m
Schnup = 0.0342 m
fmod1 = 11.065399e6 Hz
fmod2 = 5 * fmod1
 
These two plots shows the response of the POP22 and POP110 signals (in almost arbitrary units) to a PRCL sweep around the resonance. The splitting of the 55 sideband peaks is well visible in the second plot. It is due to the fact that the 55MHz sidebands are not perfectly matched to the PRC length
 
The same thing when sweeping MICH. The peaks are wider and it is not possible to see the splitting.
 
These are the error signals (REFL11_I/Q and REFL_55_I/Q) as a function of the PRCL (left) and MICH (right) sweep. Here the demodulation phases are not properly tuned. This is just to show that when the phase is wrong, you can get multiple zero crossings (in this case only in the Q signals, but in general also in I) close to resonance.
 
If the phases are tuned in order to maximize the slope of the I signals with respect to PRCL, one gets these "optimized phase" responses. It is that the phase does not correspond to the one that makes the PRCL peak to peak signal small in Q. The Q signals are indeed flat around resonance for a PRCL motion, but they deviate quite a lot from zero when moving more far from resonance. Moreover, both the REFL_55 error signals (I for PRCL and Q for MICH) are crossing again zero at two additional positions, but those are quite far from the resonance point.
 
These plots just show the PRCL and MICH error signals together with the POP22 and POP110 signals, to give an idea of the level of triggering that might be needed to be inside the linear range. It seems that if we trigger on POP22 when using the REFL55 signal we loose a bit of linear range, but not that much.
 
If you reached this point it means you're really interested in this topic, or maybe you have nothing better to do... However, this plot shows the effect of linearization of the error signal, obtained dividing them by the proper POP22/110 signal. The linear range is increased, but unfortunately for the 55 signals, the additional zero crossing I was mentioning before creates two sharp features. Those are however quite outside the triggering region, so they should not be harmful. |
8402
|
Wed Apr 3 15:00:24 2013 |
Jamie | Summary | Electronics | Sorensen supplies in LSC rack (1Y2) |
I investigated the situation of the two Sorensen supplies in the LSC rack (1Y2). They are there solely to supply power to the LSC LO RF distribution box. One is +18 V and the other is +28 V. All we need to do is make a new longer cable with the appropriate plug on one end (see below), long enough to go from the bottom of the 1Y3 rack to the top of 1Y2, and we could move them over quickly. Some sort of non-standard circular socket connector is used on the distribution box:

It could probably use thicker conduction wire as well.
If someone else makes the cable I'll move everything over. |
8411
|
Thu Apr 4 10:12:55 2013 |
Gabriele | Summary | LSC | POP22 and POP110 |
I had a look at the POP110 signal, with the PRMI flashing.
1) The LSCoffset script does not zero any more POP22_I_ERR offset. I did it by hand
2) The gain of POP22 is changed a lot, as well as the sign: now sidebands are resonant when POP22_I is negative
3) POP110 seems to deliver good signals. The plot attached shows that when we cross the sideband resonance, there is a clear splitting of the peak. If we rely on the simulations I posted in entry 8401, the full width at half height of the POP_22 peak is of the order of 5 nm. Using this as a calibration, we find a splitting of the order of 7 nm, which is not far from the simulated one (5 nm) |
8412
|
Thu Apr 4 10:32:42 2013 |
Gabriele | Summary | LSC | REFL55 error signals |
The attached plot shows that also the behaviour of the REFL11 and 55 signals is qualitatively equal to the simulation outcome. |
8413
|
Thu Apr 4 10:46:54 2013 |
Koji | Summary | LSC | REFL55 error signals |
Beautiful double peaks. I don't see the triple zero-crossings. Is this because you adjusted the phase correctly (as predicted)?
Don't you want to have a positive number for POP22? Should we set the demod phase in the configuration script for the positive POP22, shouldn't we? |
8448
|
Fri Apr 12 10:33:42 2013 |
Charles | Summary | ISS | DC-Coupled ISS Servo Design |
General ISS Design
Signals through the ISS are directed as follows: an error signal is obtained by summing the ~5 V signal from the PD with a -5 V signal from a high precision voltage regulator (which is first filtered with an ~ 30 mHz low-pass Sallen-Key filter). It is this signal that is processed/amplified by the servo. The output from the servo is then used to drive an AOM (it is not known exactly how this is done and whether or not any preamplifier/extra circuitry is necessary). The resulting modulation, hopefully, reduces fluctuations in the laser intensity incident on the PD, lowering the relative intensity noise.
Servo Design
Almost the entirety of my focus has been directed toward designing the servo portion of the ISS. Speaking in general terms, the currently proposed design consists of stages of active op-amp filters, but now the stages will have internal switches that allow them to switch between ‘flat’ gain buffers and more complicated filters with our desired behavior. Consider some Example Filter Stages where I have demonstrated a typical switching filter with the switch open and closed. When the switch is closed, the capacitor is shorted and we simply have a variable gain buffer (variable in the sense that its gain can be tuned by proper choice of the resistances) with no frequency dependence. When the switch is open, the capacitor introduces a pole at ~100 Hz and a zero at ~1 kHz.
CircuitLab has decent analysis capabilities and attached are plots generated by CircuitLab. The first plot corresponds to a frequency analysis of the voltage gain of op-amp U1 and the ‘flat’ ~20 dBV gain filter with the switch closed and the capacitor shorted. The second plot is the same frequency analysis, but now with op-amp U2 and the filter with the switch open and the capacitor introduced into signal processing. This particular combination of resistors and capacitors produce a DC gain of 60 dBV, a pole at ~100 Hz, a zero at ~10 kHz and high frequency behavior of ~constant gain of 20 dBV. In this simulation, the gain-bandwidth product of the simulated op-amp (the standard op-amp CircuitLab uses) was artificially increased in order to see more ideal behavior in the higher frequency domain.
Switches like the above can be used to add boosts to some initial filter state (which could be like the above or possibly a simple integrator to achieve high DC gain) and change it into a more complex and more useful filter state advantageous for desired noise suppression. Cascades of these switching filters could be used to create very complicated transfer function behavior. No general servo has yet been designed as the exact details of the intensity noise requirements are still being determined.
With regards to the implementation of the switches, some ‘smart’ signal will be used to trigger a switch opening and the boost being introduced to the signal processing. The switches will be opened (open corresponds to adding the boost) in a manner that maintains stability of the servo circuit. Essentially, some sort of time delay or power monitor induced signal (power from the PD output) will be used to modify the servo's behavior.
AOM
How exactly the signal will drive the AOM for correct noise suppression is unknown currently.
|
8469
|
Mon Apr 22 11:46:09 2013 |
Koji | Summary | IOO | MC locked/aligned. MC WFS offloading by ezcaservo |
Еру ьс шы тщц дщслув фтв фдшптувю
Фдыщ ш кфт еру ащддщцштп ыскшзе ещ щаадщфв еру ЬС ЦАЫ ыукмщю
I blame Den for russian keyboard installation on the control machines.
ezcaservo -r 'C1:SUS-MC2_ASCPIT_OUT16' -g '0.00001' -t 60 C1:SUS-MC2_PIT_COMM&
ezcaservo -r 'C1:SUS-MC2_ASCYAW_OUT16' -g '0.00001' -t 60 C1:SUS-MC2_YAW_COMM&
ezcaservo -r 'C1:SUS-MC1_ASCPIT_OUT16' -g '0.00001' -t 60 C1:SUS-MC1_PIT_COMM&
ezcaservo -r 'C1:SUS-MC1_ASCYAW_OUT16' -g '0.00001' -t 60 C1:SUS-MC1_YAW_COMM&
ezcaservo -r 'C1:SUS-MC3_ASCPIT_OUT16' -g '0.00001' -t 60 C1:SUS-MC3_PIT_COMM&
ezcaservo -r 'C1:SUS-MC3_ASCYAW_OUT16' -g '0.00001' -t 60 C1:SUS-MC3_YAW_COMM& |
8471
|
Mon Apr 22 17:06:42 2013 |
rana | Summary | IOO | MC locked/aligned. MC WFS offloading by ezcaservo |
Why use the PSL beam as a reference? Don't we want to keep the MC pointing in a good direction through the Faraday instead??? |
8491
|
Thu Apr 25 10:19:10 2013 |
Koji | Summary | LSC | Locking activity on Apr 24th |
Last night I worked on the several locking configurations:
General preparations / AS table inspection
- The AS beam looked clipped. I went to the AP table and confirmed this is a clipping in the chamber.
This may be fixed by the invacuum PZTs.
Modulation frequency tuning
RFPD Mon of the MC demodulator was check with the RF analyzer. Minimized the 25.8MHz (=55.3-29.5MHz) peak by changing the marconi freq.
This changed the modulation freq from 11.066147MHz to 11.066134MHz. This corresponds to the change of the MC round-trip length from
27.090952m to 27.090984m (32um longer).
Michelson tests
- I wonder why I could not see good Michelson signal at REFL ports.
- I roughly aligned the Michelson. On the AP table, the RF analyzer was connected to the REFL11 RF output.
By using "MAX HOLD" function of the analyzer, I determined that the maximum output of the 11.07MHz peak
was -61.5dBm.
- I went to the demodboard rack. I injected -61dBm from DS345 into the RFEL11 demodboard. This produced
clean sinusoidal wave with the amplitude of 4 count. The whitening gain was 0dB.
- The output from the PD cable was -64.0dBm. So there is ~2.5dB loss in the cable. Despite this noise, the demodulation
system should be sufficiently low noise. i.e. the issue is optical
- The Michelson was locked with AS55Q. And the REFL11 error signals were checked.Fringe like feature was there.
This suggested the scattering from the misaligned PRM. The PRM was further misaligned. Then some reasonable
(yet still noisy) Michelson signal appeared. (Usual misaligned PRM is not at the right place)
Q. How much scattering noise (spurious cavity between PRM and the input optics) do we have when the PRM is aligned?
Q. Where should we put the glass beam dumps in the input optics?
Q. Can we prepare "safe" misaligned place for the PRM with the beam dump?
- The Michelson was locked with REFL11Q. From the transfer function measurement, the gain difference between AS55Q (whitening gain 24dB)
and REFL11Q was 32dB. The whitening gain was 0dB. In fact I could not lock the Michelson with the whitening gain 33dB (saturation???)
The element in the Input matrix was 1, The gain of the servo was +100. BS was actuated.
Coupled cavity tests
- At least REFL11 is producing reasonable signals. So what about the other REFL ports? The Michelson signals in the other frequencies
were invisible. So I decided to use three-mirror coupled cavity with the loss PRC.
- Aligned X arm, Misaligned ETMX, ITMY. Aligned PRM.
- Locked the PRM-ITMX cavity with REFL11 and REFL33.
- Aligned ETMX. If I use REFL11I for the PRC locking, I could not lock the coupled cavity. But I could with REFL33I.
This is somewhat familiar to me as this is the usual feature of the 3f signal.
- The coupled cavity could be locked "forever". To realize this I needed to tweak the normalization factor from 1.0 to 1.6.
Q. How does the coupled cavity change the response of the cavity? Can we compensate it by something?
Q. Measure open loop transfer functions to check if there is any issue in the servo shapes.
- Transmission during the lock is 3.2 while the nominal TRX with PRM misaligned was 0.93.
This corresponds to power recycling gain of 0.17.
- X arm:
- Source: POX11I, phase 79.5 deg, whitening gain 36dB
- Input matrix: POX11I->1.0->XARM, Normalization TRX*1.60
- XARM servo gain +0.8, actuation ETMX
- XARM trigger 0.25 up, 0.05 down. XARM Filter trigger untouched.
- PRC: (sideband locking)
- Source: REFL33I, phase -34.05 deg, whitening gain 30dB
- Input matrix: REFL33I->1.0->PRCL, Normalization None
- PRCL servo gain +4.0, actuation PRM
- PRCL trigger None
- Same test for the Y arm. At the moment ETMY did not have the OPLEV.
Same level of transmission (~3.3)
- Y arm:
- Source: POY11I, phase -61.00 deg, whitening gain 36dB
- Input matrix: POY11I->1.0->YARM, Normalization TRX*2.1
- YARM servo gain +0.25, actuation ETMX
- YARM trigger 0.25 up, 0.05 down. YARM Filter trigger untouched.
- PRC: (sideband locking)
- same as above
Sideband PRMI attempt
- Now I got some kind of confidence on the REFL33 signal.
- So I tried to get any stable setup for sb PRMI, then to find any reasonable MICH signals anywhere else than AS55Q.
- With REFL33I(PRCL) & AS55Q(MICH), I got maximum ~10sec lock. It regularly locked. It was enough long to check
the spectrum on DTT. But it was not enough long to find anything about the MICH signals at the REFL ports.
- I tried REFL33Q for MICH. The lock was even shorter but could lock for 1~2 sec.
Q. What is the cause of the lock loss? I did not see too much angluar fluctuation. The actuation was also quiet (below 10000).
- PRCL: (sideband locking)
- Same as above except for
- the PRCL servo gain +0.05, No limitter at the servo output.
- Trigger POP22I (low pass filtered by LP10) 20 up, 3 down
- MICH:
- AS55Q -24.125 24dB -> x1.0 -> MICH -0.7, No limitter -> ITMX/Y differential
or
- REFL33Q -34.05dB -> x2.0 -> MICH same as above
- For both case, trigger POP22I (low pass filtered by LP10) 20 up, 3 down
At this point Jenne came back from dinner. Explained what I did and handed over the IFO. |
8537
|
Tue May 7 16:21:01 2013 |
Jenne | Summary | LSC | Error signal simulation in PRMI |
I asked Gabriele why it looked like for the PRCL sweep REFL 55 I&Q were zero at zero, but for the MICH sweep only REFL55 I was zero. He took a look at his code, and found that he was not at the correct locking point. Here is his email back to me:
I found the reason for the not zero value. Indeed, if you could zoom into the PRCL sweep, you would see that the error signals does not cross zero exactly at PRCL=0, but instead some 50 pm away from zero. This is enough to change a lot the PRCL signal when sweeping MICH. If I put PRCL to the correct zero point, and I sweep MICH, I now get everything at zero. I'm sending again the plots.
The fact that such a small detuning is enough to change PRCL signal when sweeping MICH is due, I believe, to the fact that MICH optical gain is much smaller than PRCL one.
Here are the redone plots:
Phase not tuned:
 
Phase tuned:
 
POP22 resonance for MICH and PRCL:
 
POP110 resonance for MICH and PRCL:
 
|
8555
|
Thu May 9 00:05:12 2013 |
rana, Koji, Jenne | Summary | LSC | AA and AI change |
We would like to increase the UGF of the PRC loop so as to allow more suppression of the PRC signal and less pollution of the MICH signal (remember that the PRC/MICH optical gain ratio is huge).
We were already losing phase because of delay in the LSC - SUS digital link. In addition to that, a major source of delay is the analog anti-aliasing (on the LSC error signals before they enter the ADC) and the analog anti-imaging (between the SUS DAC and the coil driver).
IN addition to these, the other major sources of phase lag in the system are the FM5 filter in the LSC-PRC filter bank, the digital upsampling and downsampling filters, and the DAC sample and hold.
In the near term, we want to modify these analog filters to be more appropriate for our 64 kHz ADC/DAC sample rate. Otherwise, we are getting the double phase lag whammy.
Staring at the schematics for the AA (D000076-01) and the AI (D000186-A), we determined a plan of action.
For the AA, we want to remove the multi-pin AA chip filter from Frequency Devices, Inc. and replace it with a passive LC low pass. Hopefully, these chips are socketed. Rana will design an appropriate LC combo and elog; we should make the change on a Wednesday afternoon so that we have enough soldering help.
For the AI, the filter is a dual bi-quad using discrete components and LT1125 opamps. Not so clear what to do with these. The resistors are all the noisy thick film kind and maybe should be replaced. Koji will find some online design tool for these or do it in LISO. Changing the TF is easy; we can just scale the capacitors. But we also want to make sure that the noise of the AI does not destroy the noise reduction action of the dewhitening board which precedes it.
Jenne should figure out how low the noise needs to be at the input to the coil driver.
P.S. the matlab code which defines these filters
>> [z,p,k] = ellip(4,4,60,2*pi*7570,'s');
>> misc.ai = zpk(z,p,k*10^(4/20)) * zpk([],-2*pi*13e3,2*pi*13e3);
>>
>> % Fudged Anti-Imaging Filter
>> [z,p,k] = ellip(8,0.001,80,2*pi*7570,'s');
>> misc.aa = zpk(z,p,k*10^(0.001/20)) * zpk([],-2*pi*32768,2*pi*32768); |
8575
|
Tue May 14 20:30:29 2013 |
Jamie | Summary | IOO | MC error spectrum at various FSS gain settings. |
I used the Agilent 4395A and the GPIB network bridge to measure the MC error spectrum at the MC servo board.
I looked at various settings of the FSS Common and FAST gains.
Here is the spectrum of various Common gain settings, with a fixed FAST setting of 23.5:

The peak at 34k is smallest at the largest Common gain setting of 13.0 (probably expected). The other higher frequency peaks are higher, though, such as the ones at 24.7k, 29.6k, 34.5k, etc.:

Here's a blow up of the peak at 1.06M, which peaks at about 9dB of common gain:

Here's the spectrum with a fixed Common gain of 10.5, and various FAST gains:

and here's a zoom around that 1.06 MHz peak, which is smallest at a FAST gain of 23.5 dB:

I'm not sure yet what this points to as the best gain settings. We can of course explore more of the space. I'm going to leave it at 13/23.5, which leaves the PC RMS at ~1.5 and the FAST Monitor at ~6.0.
If this does turn out to be a good setting we'll need to adjust some of the alarm levels.
Various settings:
MCS
in1 gain: 15
offset: 1.174
boost enabled
super boost: 2
VCO gain: 25
FSS:
input offset: -0.8537
slow actuator: 0.6304
I include the python scripts I used to remotely control the AG4395 to take the measurements, and make the plots.
PS: I made some changes/improvements to the netgpib stuff that I'll cleanup and commit tomorrow.
|
8580
|
Wed May 15 17:17:05 2013 |
Jamie | Summary | CDS | Accounting of ADC/DAC channel availability |
We need ADC and DAC channels for a couple of things:
- POP QPD: 3x ADC
- ALS PZTs: 3x 2x 2x DAC (three pairs of PZTs, at ends and vertex, each with two channels for pitch and yaw)
- Fibox: 1x DAC
What's being used:
- c1iscex/c1iscey:
- DAC_0: 7/16 = 9 free
- ADC_0: 17/32 = 15 free
- c1sus:
- c1ioo
- DAC_0: 0/16 = 16 free ?? This one is weird. DAC in IO chassis, half it's channels connected to cross connect (going ???), but no model links to it
- ADC_0: 23/32 = 9 free
- ADC_1: 8/32 = 24 free
- c1lsc
- DAC_0: 16/26 = 0 free
- ADC_0: 32/32 = 0 free
What this means:
- We definitely have enough DACs for the ALS PZTs. The free channels are also in the right places: at the end stations and in the c1ioo FE, which is close to the PSL and hosts the c1als controller.
- We appear to have enough ADCs for the QPD in c1ioo.
- We don't have any available DAC outputs in c1lsc for the Fibox. If we can move the Fibox to the IOO racks (1X1, 1X2) then we could send LSC channels to c1ioo and use c1ioo's extra DAC channels.
Of course we'll have to investigate the AA/AI situation as well. I'll try to asses that in a follow up post.
PS: this helps to identify used ADC channels in models:
grep adc_ sus/c1/models/c1scx.mdl | grep Name | awk '{print $2}' | sort | uniq
|
8581
|
Wed May 15 17:38:49 2013 |
Jamie | Summary | CDS | AA/AI requirements |
Quote: |
What this means:
- We definitely have enough DACs for the ALS PZTs. The free channels are also in the right places: at the end stations and in the c1ioo FE, which is close to the PSL and hosts the c1als controller.
- We appear to have enough ADCs for the QPD in c1ioo.
- We don't have any available DAC outputs in c1lsc for the Fibox. If we can move the Fibox to the IOO racks (1X1, 1X2) then we could send LSC channels to c1ioo and use c1ioo's extra DAC channels.
Of course we'll have to investigate the AA/AI situation as well. I'll try to asses that in a follow up post.
|
It looks like we have spare channels in the AA chassis for the existing c1ioo ADC inputs to accommodate the POP QPD. 
We need AI interfaces for the ALS PZTs. What we ideally need is 3x D000186, which are the eurocard AI boards that have the flat IDC input connects that can come straight from the DAC break-out interfaces. I'm not finding any in the spares in the spare electronics shelves, though. If we can't find any we'll have to make our own AI interfaces. |
8583
|
Wed May 15 19:32:04 2013 |
rana | Summary | CDS | Accounting of ADC/DAC channel availability |
- What are we using 16 DAC channels for in the LSC?
- What are the functions of those IOO DAC channels which go to cross-connects? If they're not properly sending, then we may have malfunctioning MC or MCWFS.
- Can we just use the SLOW DAC (4116) for the ALS PZTs? We used this for a long time for the input steering and it was OK (but not perfect).
|
8585
|
Wed May 15 22:47:11 2013 |
Jamie | Summary | CDS | Accounting of ADC/DAC channel availability |
Quote: |
- What are we using 16 DAC channels for in the LSC?
|
For the new input and output tip-tilts. Two input, two output, each requires four channels.
Quote: |
- What are the functions of those IOO DAC channels which go to cross-connects? If they're not properly sending, then we may have malfunctioning MC or MCWFS.
|
I have no idea. I don't know what the hardware is, or is supposed to be, connected to. DAC for WFS?? Was there at some point supposed to be fast output channels in the PSL?
Quote: |
- Can we just use the SLOW DAC (4116) for the ALS PZTs? We used this for a long time for the input steering and it was OK (but not perfect).
|
Probably. I'm not as familiar with that system. I don't know what the availability of hardware channels is there. I'll investigate tomorrow. |
8587
|
Thu May 16 01:41:31 2013 |
Jenne | Summary | IOO | FSS gain settings set |
Quote: |
I'm not sure yet what this points to as the best gain settings. We can of course explore more of the space. I'm going to leave it at 13/23.5, which leaves the PC RMS at ~1.5 and the FAST Monitor at ~6.0.
|
I changed the value of the nominal FSS common gain in the PSL Settings screen (It's called the 'FSS global gain' there). To get to this screen: sitemap -> PSL_main -> PSL_settings. The MC autolocker reads these settings from the screen and uses those values. Now the FSS returns to this value of 13 that Jamie has chosen. For the past few days, it's been going back to the old value of 10.1 . The FAST gain was already set to this 23.5 value. |
8598
|
Fri May 17 18:58:58 2013 |
Jamie, Koji | Summary | CDS | Weird DAC bit flipping at half integer output values |
While looking at the DAC anti-imaging filters, Koji noticed an odd feature of the DAC output:

What you see here is 16kHz double data from a model right before the DAC part ('C1:SUS-PRM_ULCOIL_OUT', blue), and the full 64kHz int data sent to the physical DAC as reported by the IOP ('C1:X02-MDAC0_TP_CH0', green). The balls are the actual sample values (as expected there are four green balls for every blue). The output data is being ramped continuously between 0 and 1.
As the output data crosses the half-count level, the integer DAC output oscillates continuously at every 64kHz sample between the bounding integer values (in this case 0 and 1).
Here's the data as we hold the output continuously at the half-count level; the integer DAC output just oscillates continuously:

After some probing I found that the oscillation happens between [-0.003 +0.004] of the half-count level.
The result of this is a fairly strong 32 kHz line in the DAC analog output.
We looked in the controller.c and couldn't identify anything that would be doing this. This is the output procedure as I can see it (controller.c lines):
- The double from the model is passed to the IOP
- The IOP applies a sample-and-hold or zero-pad if the model is running at a slower speed than the IOP (1799)
- The data is then anti-image filtered (1801)
- A half-integer is added/subtracted before casting such that the cast is a round instead of a floor (1817)
- The data double is cast to an int (1819)
- The data is written to the DAC (1873)
There's nothing there that would indicate this sort of bit flipping. |
8599
|
Fri May 17 19:56:52 2013 |
Koji | Summary | CDS | Weird DAC bit flipping at half integer output values |
Let me make a complimentary comment on this effect.
Because of this oscillation feature, we have a 32kHz peak in the DAC spectrum rather than a 64kHz peak.
For advanced LIGO, the universal AI (D070081) was made to have 3rd-order 10kHz LPF with 64kHz notch.
If we have a higher peak at 32kHz (where the rejection is not enough) than at 64kHz, the filter does not provide
enough filtering of the DAC artifacts.
For the 40m, the original filter had the cut off of 7kHz as the sampling rate was 16kHz.
If we want to extend the frequency range by 4times, the correspoding cut off should be 28kHz.
The rejection is again not enough at 32kHz.
If this peak is an avoidable feature by using simple sample&hold the peak freq is pushed up to
64kHz and we can use the AI filters as planned. |
8612
|
Wed May 22 00:42:13 2013 |
Koji | Summary | SUS | violin Q |
While looking at the decay of the violin mode of the PRM, I made a simple measurement of the decay rate.
Error signal: REFL33I
The peak @628Hz became 0.372 to 0.303 in 60 sec.
-> Half life of the amplitude T_{1/2} is 203sec.
Q = 4.53 f0 T_{1/2} = 5.8 x10^5 |
8613
|
Wed May 22 11:09:33 2013 |
Jamie | Summary | CDS | Weird DAC bit flipping at half integer output values |
After querying CDS folks about this issue, I got some responses that indicated the problems was likely limit-cycle oscillations due to zero-padding of the data when upsampling. Tobin ran some Matlab tests to confirm this issue.
Starting in RCG 2.5 there is a new "no_zero_pad=1" cdsParameters option turns zero padding OFF. I tried enabling this option c1scy to see how the behavior changed. Sure enough, the 32 kHz oscillations mostly went away. There are no oscillations for outputs held at the half-count value, and the oscillations around the half-count transitions went away as well.
The only thing I could see is a bit of oscillation when converging on a constant half-count value that went away after a couple of milliseconds:

So we might consider adding the no_zero_pad=1 option to all of our coil driver outputs, which might eliminate the need to add notches at the Nyquist in the analog anti-image filters |
8614
|
Wed May 22 11:21:28 2013 |
Koji | Summary | CDS | Weird DAC bit flipping at half integer output values |
Is this limit cycle caused by the residual of the digital AI filtering at the half sampling freq and that his the threshold?
Or is this some nonlinear effect? If this is a linear effect associated with the zero-padding, the absolute
value of the DC may affect the amplitude of the oscillation. (Or equivalently the range of the DC where we get this oscillation.)
Anyway, it seems that we should use no-zero-padding.
You pointed out the ringdown of the digital AI filter in the sample-hold case (i.e. no-zero-padding case).
How does it look like in the conventional zero-padding case? |
8615
|
Wed May 22 11:35:06 2013 |
Jamie | Summary | CDS | Weird DAC bit flipping at half integer output values |
Quote: |
Is this limit cycle caused by the residual of the digital AI filtering at the half sampling freq and that his the threshold?
Or is this some nonlinear effect? If this is a linear effect associated with the zero-padding, the absolute
value of the DC may affect the amplitude of the oscillation. (Or equivalently the range of the DC where we get this oscillation.)
|
This is a good question. We may be able to test if it's a linear effect if we have enough DAC range to get the oscillation to be more than a single sample.
Quote: |
You pointed out the ringdown of the digital AI filter in the sample-hold case (i.e. no-zero-padding case).
How does it look like in the conventional zero-padding case?
|
In the zero-pad case the oscillation just continues indefinitely at the half-count value, so it never dies out (at least as far as I can tell). |
8616
|
Wed May 22 15:08:37 2013 |
Koji | Summary | CDS | Weird DAC bit flipping at half integer output values |
It seems that the effect is from the (linear) residual fluctuation of the digital AI filter for the zero-padded signal.
Namely, if we give the larger constant number, we get more oscillation. |
8624
|
Thu May 23 01:27:11 2013 |
Manasa | Summary | LSC | Xarm beat note search continues |
Towards finding the x-arm beat note:
The green would not lock to a maximum GTRX this morning. In the course of aligning the green stably to the X arm, somewhere down the line, the input pointing got messed up (reasons unknown). To set this right, Koji tried to lock the Yarm with POY DC but it wouldn't work. The transmon for Y had to be set up temporarily and the Y arm was locked with TRY. This restored the input pointing and the arms locked with transmission TRX/TRY > 0.9 counts. The transmon path along the Y arm was then re-configured as mentioned in Annalisa's elog.
I still had trouble getting the X-green locked in TEM00 (similar situation mentioned by Jenne in elog). The arm cavity mirrors were tweaked to get the green to resonate in TEM00 but it wouldn't stay locked when the temperature of the x-end NPRO was changed. Koji helped recover missing links to filters for the ALS_X_SLOW servo from the archives. Enabling the filters helped keep the green locking stable for laser temperature changes (which corresponds to 'offset' change in ALS_X_SLOW servo screen).
PSL green alignment was checked once again and the X-end laser temperature was scanned trying to find the beatnote. RFMON from the beatbox was connected to the spectrum analyzer. I have scanned through the whole range of offset but have not been able to find the beat note yet.
The search will continue tomorrow 
|
8626
|
Thu May 23 10:24:23 2013 |
Jamie | Summary | CDS | c1scy model continues to run at the hairy edge |
c1scy, the controller model at the Y END, is still running very long, typically at 55/60 microseconds, or ~92% of it's cycle. It's currently showing a recorded max cycle time (since last restart or reset) of 60, which means that it has actually hit it's limit sometime in the very recent past. This is obviously not good, since it's going to inject big glitches into ETMY.
c1scy is actually running a lot less code than c1scx, but c1scx caps out it's load at about 46 us. This indicates to me that it must be some hardware configuration setting in the c1iscey computer.
I'll try to look into this more as soon as I can. |
8629
|
Thu May 23 13:14:34 2013 |
Koji | Summary | LSC | Xarm beat note search continues |
We should consider to hook up the temperature monitors of the NPROs to the ADCs. |
8631
|
Thu May 23 17:51:39 2013 |
Koji | Summary | LSC | My usual locking procedure |
For purpose of the automation and my record, I summarized my locking procedure as a chart. |
8686
|
Thu Jun 6 15:46:10 2013 |
Max Horton | Summary | General | Smart UPS 2200 Batteries Replaced |
Replaced the batteries successfully in the control room. We just had to switch the clips from the old batteries to the new one, which we didn't know was possible until now. |
8694
|
Tue Jun 11 22:16:56 2013 |
Manasa | Summary | Green Locking | ALS for X arm |
I discussed with Yuta about the ALS servo and phase tracker and found that there was a lot of information lying around from last year but there aren't any clear elogs on how to enable ALS and obtain IR resonance.
Guide to enabling the ALS servo and find IR resonance:
The steps will explain in detail how to ressurrect the ALS servo for green X-arm and find IR resonance using ALS. The medm screens are very confusing right now.
(i) Finding the beat note
1. Get the IR to flash in TEM00 for the arm and lock it by enabling LSC (Locking the arm to IR keeps the arm cavity mirrors stable so that you can scan the temperature of the X-end laser to find the beat note).
2. Steer the X-green into the arm cavity such that the arm cavity locks in TEM00 for green as well. At this point you should also have the X-green reaching the PSL table.
3. Align the PSL doubled green (PSL-green) and the X-green in near-field (at the camera) and far-field (letting the beams to propagate beyond the Green-TRX PD).
4. Check cabling of the RF beat PD.
5. Change the X-laser temperature by sweeping the offset (C1: ALS-SLOW_SERVO2_OFFSET) in steps of 10.
6. Find the beat note and tune the alignment at the beat PD to maximize the beatnote amplitude. Disable LSC for X arm.
(ii) The GREEN HORNET explained
'Input signal conditioning' block takes I and Q signals after the delay frequency discriminator (DFD) in the beat box and these signals pass through C1ALS_BEATX_FINE filter banks. The output signal then enters the phase rotation matrix of the phase tracker. The phase tracker gives 'PHASE_OUT' which is the error signal that is fed to the ETM servo filter module (DOF filters) through the 'Input matrix' in the medm.
An offset can also be fed to the phase tracker which will scan the beat frequency (used to find IR resonance).
(iii) Scripts
1. easyALS.py - This runs from 'ON plus' or 'ON minus' buttons in the C1ALS_COMPACT.
The script clears history of 'fine_phase' filter module and increases gain of the servo in steps ('ON plus' for positive gain and 'ON minus' for negative gain).
2. findIRresonance.py - This runs from 'IRres' button in the C1ALS_COMPACT.
It adds offset to the phase tracker in steps which scans the beat frequency to find IR resonance.
P.S. Check the scripts before enabling the servo so that the right filter modules are being turned ON. Using the wrong set of filter modules can kick the ETM.
____________________________________________________________________________________________________________________________________________________
X arm ALS progress:
I found the beat note and got ALS to work reasonably for the Xarm without kicking the ETM. I did this by manually toggling buttons and changing gains. The scripts need editing.
To do:
Modify the scripts to work as we want them to.
The ALS medm is SSSOOOO confusing. It definitely needs to be fixed (remove all unwanted parts of the screen that existed 'pre-phase tracker').
Find IR resonance.
|
8703
|
Thu Jun 13 22:31:12 2013 |
Manasa | Summary | Green Locking | X arm ALS |
Stabilized ALS and beat frequency sweep realized.
Procedure:
1. Enable appropriate filter modules and set appropriate servo gains.
Servo module
|
Filters |
Gain |
C1:ALS-BEATX_FINE_I |
FM2 FM3 |
1.0 |
C1:ALS-BEATX_FINE_Q |
FM2 FM3 |
1.0 |
C1:ALS-BEATX_FINE_PHASE |
FM1 |
300.0 |
C1:ALS-BEATX_FINE_OFFSET |
- |
1.0 |
C1:ALS-XARM |
FM4 FM5 |
-0.25 |
C1:ALS-ETMX |
- |
1.0 |
2. Clear history of C1:ALS-BEATX_FINE_PHASE
3. Enable the servo loop. I had set limits on the servo loop and ramp time for gain switching so that I don't kick the ETMY hard.
Gains were decided such that the error signal C1:ALS-BEATX_FINE_PHASE_OUT was minimized.
4. Beat frequency sweep is realized by stepping up on C1:ALS-BEATX_FINE_OFFSET_OFFSET (from 0 to 2100 in this case).
Video1 shows the difference that can be seen at the RF spectrum analyzer when ALS is enabled.
Video2 shows the beat frequency sweep as seen on the spectrum analyzer.
I could not get 'getdata' to work as I wanted. So I have attached the error signal trend before and after the ALS servo loop is enabled.
Thank you Jenne for helping retrieve more sensible data!
More info:
The beat note is very strong and we can clearly see its harmonics as well. Attached is the picture showing the several harmonics.
________________________________
Peak frequency(MHz) Power(dBm)
________________________________
47 -2.77
93.5 -27.56
139 -32.75
185.4 -45.64
231.9 -57.10
278.4 -59.42
________________________________
To do:
1. Obtain IR resonance.
2. Check the digital anti-whitening filter after the beatbox.
3. The effect of the harmonics should be figured out.
4. Write scripts to enable ALS and findIRresonance. |
8708
|
Fri Jun 14 07:06:19 2013 |
Koji | Summary | Green Locking | X arm ALS |
It's nice that we are now able to scan the cavity again. We got close to PRMI+one arm one step further.
The calibration of the scan frequency and the evaluation of the in-loop/out-of-loop error signal in terms of (Hz/rtHz) would be necessary.
The beat amplitude looks actually huge aIthough I don't know where you are monitoring.
Talk to Jamie to figure out how much the signal should be at the monitoring point.
If it is more than we are supposed to have, put an attenuator somewhere. |
8717
|
Tue Jun 18 10:37:00 2013 |
Steve | Summary | General | new laser pointers |
Red-green laser pointers added to the depleted stock of 2011
The two pointers output measured 4.4 mW green and 2 mW red
|
8738
|
Mon Jun 24 16:06:17 2013 |
Manasa | Summary | Green Locking | ALS model |
I am working on the basic ALS servo model. The simulink model for the same is attached. The loop is not yet complete (I'm still debugging it) ; but this is just an update of where I am right now.
Attached is the simulink and matlab file.
|
8752
|
Wed Jun 26 01:30:31 2013 |
rana | Summary | PEM | Variation in 10-30 Hz seismic RMS |
For quite a while (no one knows how long), we've seen fluctuations in the 10-30 Hz seismic motion. This shows up as the purple trace on the seismic BLRMS on the wall projector.
The second plot shows that this is not only a periodic increase in the usual 29.5 Hz HVAC peak, but also an anomolous 32.2 Hz peak. Probably some malfunctioning machinery - maybe in the 40m or maybe on the roof. |
8776
|
Thu Jun 27 22:52:38 2013 |
Rana, Gabriele, Francesco | Summary | Computer Scripts / Programs | LIGO-DV installed |
I installed ligoDV in the /ligo/apps/ligoDV/
Now, by pointing the tool at the local NDS2 server (megatron:31200) you can access the recent local data (raw, trends, etc.)
by running /ligo/apps/ligoDV/ligodv from the command line. |
8795
|
Wed Jul 3 11:07:17 2013 |
Alex | Summary | | Photodetector Characterization |
[Alex, Koji]
We characterized Koji's BBPD MOD for REFL165 (see attachment).
First, we calibrated the Agilent 4395 Network Analyzer (NA) to account for differences in cable features between the Ref PD and Test PD connections. This was done using the 'Cal' softkey on the NA.
Then we performed transimpedance measurements for the test PD and reference PD relative to the RF output of the NA and relative to each other (see 2nd attachment. Note that the NA's RF output is split and sent to both the IR Laser and the NA's Ref input).
Next, we made DC measurements of the outputs of the photodetectors to estimate the photocurrent distribution of the transimpedance setup (like the 2nd attachment, but with the outputs of the PDs going to a multimeter). By photocurrent distribution, we mean how the beamsplitter and respective quantum efficiencies/generalized impedance/etc. of the PDs influence how much current flows through each PD at with a DC input.
Finally, we measured the output noise as a function of photocurrent (like the 2nd attachment, but with a lightbulb instead of the IR Laser). Input voltages for the lightbulb ranged from 0mV to 6V. Data was downloaded from the NA using netgpibdata from the scripts directory. Analysis is currently in progress; graphs to come soon.
|
8797
|
Wed Jul 3 14:33:46 2013 |
Koji | Summary | LSC | Test result for the REFL165 photodetector |
P.1 Circuit diagram
Added components are indicated by red symbols.
- The diode on the board is HAMAMATSU S3399. It is a Si PIN diode with φ3.0 mm.
- Based on prototype version of aLIGO BBPD D1002969-v8 (although the board says v7, It is v8.)
- The input impedance of the MAR-6SM amplifier (50Ohm) provides the transimpedance.
- The first notch (Lres and Cresa/b) is actually not notch but a LF rejection with DC block.
- The second and third notches are tuned to 11MHz and 55MHz.
- Another notch is implemented between the RF amps. The 33MHz signal is weak so I expected
to have no saturation at the first amplifier.
- As you see from the DC path, the transimpedance of the DC path is 2k V/A. If this is too high,
we need to replace R9 and R11 at the same time. TP1 is providing +10V such that the total
reverse bias becomes 25V without bringing a special power supply.
P.2 Transimpedance
The transimpedance is measured with an amplitude modulated diode laser.
The transimpedance is 1k V/A ish. It is already at the edge of the bandwidth.
If we need more transimpedance at 165MHz, we should replace
the PD with FFD-100 (I have one) and apply 100V of reverse bias.
P.3 Current noise spectrum
The measured dark noise voltage spectrum was converted to the equivalent current noise at the diode.
The measured transimpedance is ~1.2kV/A.
The reduction of the transimpedance above 100MHz has been seen as 165MHz is already at the edge of the bandwidth.
If we need more transimpedance at 165MHz, we should replace the diode with FFD-100 (I have one) and apply 100V of reverse bias.
P.4 Shot-noise intercept current
Shot-noise intercept current was measured with a white light from a light bulb.
This measurement suggests the shot-noise intercept current of 1mA, and transimpedance of 1.5kV/A. |
8816
|
Tue Jul 9 23:27:17 2013 |
Koji | Summary | LSC | MICH: ITMX/Y <=> PRM/BS |
The MICH actuation with PRM/BS was investigated again.
(ITMX -1 / ITMY +1) is equivalent to (PRM -0.267 and BS +0.50).
- PRMIsb was locked with REFL33I&AS55Q.
- Using the locking module in the LSC model, actuate ITMX (-1) and ITMY (+1) at 580.1Hz. Note that the notch filters in the MICH/PRCL servos were on.
- Look at the peak in the AS55Q spectrum. Tune the BS element in the output matrix of the lock-in to minimize the peak height.
=> The peak was minimized at BS = -0.50.
- Look at the peak in the REFL33I spectrum. Tune the PRM element in the output matrix of the lock-in to minimize the peak height.
=> The peak was minimized at PRM = +0.267
- These measurement leads to the conclusion mentioned above. |
8882
|
Fri Jul 19 22:35:06 2013 |
Koji | Summary | LSC | Various Arm signal (Yarm) |
The StripTool plot attached below shows various arm signals measured with the Y arm cavity swept using ALS.
Yellow: TRY
Blue: ALS additive OFFSET to the error signal
Red: Raw PDH error signal (POY11I)
Purple: Linearized PDH error (POY11/TRY)
Green: 1/Sqrt(TRY)-5 (No normalization)
Inverse Sqrt of the TRY had been implemented when this LSC controller was first coded.
It is confirmed that the calculation is working correctly. |
8885
|
Fri Jul 19 23:35:52 2013 |
Koji | Summary | LSC | TRY DC locking |
The Y arm was locked with the TRY DC signal.
The handing off process is too complicated because there is no path from ALS to the LSC error.
The TRY DC error signal & the gain determination
- The error signal was produced by the operation 1/SQRT(TRY) - OFFSET. The initial offset was -5.
- The sign of the TRY DC error signal depends on which side of the resonance the arm is.
By looking at the strip chart, I determined that the sign is opposite of the ALS.
The ALS had the gain of -25, so the TRY control gain was to be positive.
- From the strip chart on the previous entry , the slope difference between the PDH error and the TRY DC error was x500.
The arm control with POY11 PDH had the gain of 0.2. So the target gain for the TRY DC was determined to be +100.
Handing off
- The arm was stabilized by ALS. The ALS gain was -25 with FM2/3/5/6/7/10
- YARM configuration: no trigger / no FM trigger / gain =+0 / FM5 ON / OFFSET -5
- Start handing off:
YARM: Turned up the gain to +50
- ALS: Turned off FM6/7
- YARM: Turned on FM6/7
- ALS: Turned off FM2
- YARM: Turned on FM4
- ALS: Turned off FM3/10
- YARM: Turned on FM2/3/8/9 ON
- ALS: Reduced the gain to -15
- YARM: Increased the gain to +70
- ALS: Reduced the gain to 0
- YARM: Increased the gain to +100
HANDING OFF - DONE
Changing the offset
The offset of -5 gave the TRY of <0.1.
The detuning was reduced by giving the offset of -4. TRY went up to ~.1
The offset of -3 made TRY 0.13
The offset of -2 made TRY 0.25
The offset of -1.5 made TRY 0.4. And the arm could not be held by this error signal anymore.
|
8887
|
Mon Jul 22 03:10:41 2013 |
rana | Summary | lore | Angel of the Y End Table? |
Trying to take an image or movie of the ETMY Transmon cam, we got instead this attached image.
I think it is just some scattered green light, but others in the control room think that it is a message from somewhere or someone... |
8888
|
Mon Jul 22 06:58:17 2013 |
Lisa | Summary | lore | Angel of the Y End Table? |
Quote: |
Trying to take an image or movie of the ETMY Transmon cam, we got instead this attached image.
I think it is just some scattered green light, but others in the control room think that it is a message from somewhere or someone...
|
It is not an angel, it is clearly a four leaf clover (also known as "quadrifoglio"). It is very rare, it brings good luck! |
8905
|
Tue Jul 23 13:29:58 2013 |
Koji | Summary | General | |
record of the initial state |
8987
|
Thu Aug 8 18:43:12 2013 |
Sujan | Summary | PEM | Optimally subtracting signals from two seismometers. |
An exercise of optimally subtracting one seismometer signal by another using weiner filters was done. Results have been summarized document attached. |
8988
|
Thu Aug 8 18:47:41 2013 |
Sujan | Summary | PEM | Using weiner filters for subtracting signals MC_L and GUR2_X |
I used MC_L signal from the Mode Cleaner as the desired signal with GUR2_X as witness signals. I observed good subtraction where coherence is high. But there was noise added in other frequency bands. I am not sure how to avoid that.
Please find attached documents that contains relevant plots. |
9004
|
Tue Aug 13 11:40:19 2013 |
Alex Cole | Summary | Electronics | RFPD Demod Filter Frequency Response Measurement |
For the RF PD Frequency Response Measurement project, we get each PD signal from the "PD RF Mon" output of each demodulator board corresponding to our PD under test. Therefore we can't neglect the frequency response of various filters inside the demodulator board. I used our Agilent 4395 Network Analyzer to gather frequency response data for each demodulator board being considered for the RFPD frequency response project (AS55, REFL11, REFL33, REFL55, REFL165, POX11, POP22, POP110).
The NA swept over a frequency range of 1-500 MHz. Data was collected using NWAG4395A (from the netgpibdata directory). It should be noted that the command line options -a 16 -x 15 (averaging=16 and excitation amplitude=15 dBm[the max]), in addition to the usual command line options described in the help file, were used to minimize noise.
The data is located in /users/alex.cole. The file names are in the format [PDNAME]DemodFilt_1000000.dat (e.g. REFL11DemodFilt_1000000.dat). Results for POP110 are shown below. |
9009
|
Tue Aug 13 21:49:32 2013 |
Koji | Summary | General | Testing new AG4395A network analyzer |
New AG4395, sn MY41101114 for West Bridge Labs was delivered. For the test purpose it is at the 40m now.
I made a series of tests in order to find anything broken.
Network analyzer test
- RF out / Rch test
RF out directly connected to R input channel.
The received power at the R-ch was measured while the output was swept from 10Hz to 500MHz.
The RF power was changed from -50dBm to +15dBm with +10dBm increment (but the last one).
The attenuator setting was changed from 50dB to 0dB.
=> The configured output power was properly detected by the R channel.
=> RF output is producing the signal properly. R-ch is detecting the produced signal properly.
- Ach/Bch test
Same test as above for Ach and Bch
=> Same result as above
=> A-ch and B-ch are detecting the produced signal properly.
- Transfer function test
Connect a power splitter to the RF out. Detect the split signals by R-ch and A-ch
=> Measurement is at around 0dB +/- 1dB up to 500MHz.
Same measurement for B-ch
=> Same result
=> A/R and B/R indicates proper transfer function measurements.
- Calibration
RF out was split in to two. One was connected to R-ch. The other was connected to A-ch.
The thru response calibration was run.
=> The thru calibration was performed properly.
- Practical tranfer function measurements.
In the above calibration setup, various RF filters were inserted in the Ach path.
The measured data was extracted via GPIB connection.
=> Practical transfer function measurements were performed.
=> GPIB connectivity was confirmed
External reference test
- External 10MHz reference from an SRS frequency counter was connected to Ext Ref In
=> Ext Ref indicator on the screen appeard
=> The internal oscillator seemed to be locked to the external reference in
Spectrum analyzer test
- Measured the signals from DS345 by R/A/B ch
Sinusoidal signal (1V) swept from 10MHz to 30Mhz
=> Corresponding moving peak was detected in each case
- Noise level measurement
R/A/B channels were terminated. The attenuation at each port was set to 0dB.
Frequency span was changed between 500MHz, 10MHz, 100kHz, 1kHz.
=> Noise level of ~10nV/rtHz between 0.1-500MHz was confirmed. All R/A/B channels have the same performance.
|
9010
|
Tue Aug 13 22:21:12 2013 |
Koji | Summary | General | Minicircuit Filter TFs (AG4395A test) |
As a part of the network analyzer test in the previous entry, the transfer functions of Mini-Circuits filters we have at the 40m were measured.
<<List of the filters>>
- LPF (SMA): SLP1.9, SLP5, SLP21.4, SLP30, SLP50, SLP100, SLP150, SLP750
- LPF (BNC): BLP1.9, BLP2_5, BLP5, BLP30
- BPF (SMA): SBP10.7, SBP21.4, SBP70
- HPF (SMA): SHP25, SHP100, SHP150, SHP200, SHP500
|
9012
|
Thu Aug 15 01:51:50 2013 |
Koji | Summary | General | RFM<->Dolphin bridge distributed to c1rfm and c1mcs |
Since the RFM-Dolphin bridges for the ASX model was added to the c1rfm model, c1rfm kept timing-out from the single sample time of 60us.
The model had 19 dolphin accesses, 21 RFM accesses, and 9 shared memory (SHM) accesses.
At the beginning 2 RFM and 2 SHM accesses were moved to c1sus (i.e. they were mistakenly placed on c1rfm).
But this actually made the c1sus model timed out. So the model was reverted.
The current configuration is that the WFS related bridges were accommdated in the c1mcs model.
This made the timing of c1rfm ~40us. So it is safe now.
On the other hand, the c1mcs model has the time consumption of ~59us. This is marginal now.
We need to understand why any RFM access takes such huge delay. |