Wed Mar 9 10:23:28 2022 |
Jordan | Update | VAC | Leak Testing of New Manual Gate Valve - Attempt #1 |
Jordan, Chub, Paco
Chub and I went into the lab this morning to leak check the new gate valve after pumping over the weekend. This would be done through the RGA and spraying helium around the newly installed flanges. The RGA is set to monitor the partial pressure of helium versus time and we visually watch for any spikes in pressure to indicate an air leak.
So, I mistakenly thought the gate valve was opened and both sides were being pumped on, this was not the case. The vavle was closed so there was a pressure differential whenI turned the handle it tripped the interlocks and closed V4. I closed VM3 to the RGA volume to prevent the filament from being damaged.
Then the medm screen for the vac controls started flashing rapidly, I closed the window and reopened the controls to find all the panels were white, but we could still see the read only vac screen. So Paco restarted c1vac and the controls were restored. V7 closed, but was then reopened.
We now need to restart pumping with the odd pressure differentials between V4 and VM3.
Plan
- TP2 turned off along with the AUX pump
- VM3 opened to vent RGA volume (RGA turned off, filament cooled)
- Open the manual vent screw on TP1 to bring the RGA+TP1 volume back to atmosphere, now no pressure diff. on V4
- Open V4, restart AUX pump to rough out the volume
- Connect RP1/3 line to the pump scroll, turn on RP1/3 RGA+TP1 volume went to mtorr
- Slowly open the manual gate valve
- Connect AUX pump to TP2 and rough out
- Restart TP2, once at speed open V4, close V6 and turn off RP1/3
Now we are pumping on the RGA/TP1 volume with TP2, leak check attempt #2 will happen tomorrow morning |
16716
|
Wed Mar 9 09:35:26 2022 |
Paco | Update | BHD | Re-balance of AS1 |
[Paco]
AS1 is installed, OSEMs balanced, and the optic damped successfully. We should run the free swinging test overnight to validate this re-installation. |
16715
|
Tue Mar 8 19:29:36 2022 |
Paco | Update | BHD | Re-balance of AS1 |
[Paco]
Installed AS1 in vacuum, near the center of the table, and installed the OSEMs. All OSEMS are "balanced" nominally, i.e. their shadow is at the halfway point optimum, but fine tuning is required, which I will attempt tomorrow after restoring the AS1 suspension screen settings. Today, I tried damping the SIDE DOF, but didn't succeed, although there was definitely some oscillating behaviour with high (> 5) gains on the damping, so I believe this is a matter of patience. For now, all OSEMs are looking ok, the SOS is in place, and hopefully it will soon be damped. |
16714
|
Tue Mar 8 12:24:13 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
The gluing seemed to be successful. I assembled the side block with the magnet on the adapter. Paco helped me hang the adapter on the SOS tower.
The height and roll of the adapters were balanced (attachment 1,2).
The QPD was placed at the beam reflection. The beam was centered horizontally on the QPD and then measured vertically. The pitch DOF was balanced using the counterweights. The counterweight was locked. Balance was retained.
I tried to assemble the upper mirror clamp on the tower but for some reason, one of its tap holes was not able to accept screws. I gave it to Jordan for retapping. I measured the motion spectrum using the QPD connected to a scope (attachment 3).
Major peaks are at 668mHz, 942mHz, and 1029mHz.
|
Attachment 1: AS1_Roll_Balance.png
|
|
Attachment 2: AS1_Height_Balance.png
|
|
Attachment 3: FreeSwingingSpectra_new.pdf
|
|
16713
|
Tue Mar 8 12:08:47 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
OMG, it worked! It was indeed a calibration issue and all I had to do was press the "OK" button after selecting the "CAL" tab beside the pressure reading. Wow.
Quote: |
Great trouble shoot!
> I guess, the only remaining issue now is the incorrect atmospheric pressure reading of 1000 Torrs.
This is just a calibration issue. The controller should have the calibration function.
(The other Pirani showing 850Torr was also a calibration issue although I didn't bother to correct it. I think the pirani's typically has large distribution of the calibration values and requires individual calibration)
|
|
Attachment 1: XGS600_calibration.pdf
|
|
16712
|
Mon Mar 7 19:38:47 2022 |
Anchal | Summary | CDS | c1susaux2 slow controls issues |
I tried to perform a simple enabling test of coils using c1susaux2 modbus channels but failed. I'm able to do the enabling of coils using the windows GUI of acromag card but I can not do it when the cards are connected to the computer subnetwork. The issue is two-fold:
- The enable channels such as C1:SUS-LO1_UL_ENABLE are not changing values when their DOL changes a value. In this case, I created a calc channel C1:SUS-LO1_ALL_CALC which takes the AND of all coil's individual CALC channels which are normally used as DOL for the ENABLE channels. But even though the changes are reflected properly to C1:SUS-LO1_ALL_CALC, it does not affect C1:SUS-LO1_UL_ENABLE. See the db files here for more info.
- I tried to directly change the value of C1:SUS-LO1_UL_ENABLE using caput and even though in soft value the channel changes, it does not propagate a change at the output of Acromag card. So my suspicion is that something might be off with the setting of the Acromag card or c1susuaux2.cmd file. I followed this wiki page instructions, but if anyone can find an error, it would be useful.
There's also an issue in reading back the ENABLE_MON channels. Here we suspect that one of the optical isolator box that we have been using might have a short in one of it's output channel. I'll investigate this more tomorrow. Again, the issue is two-fold. The EPICS channel values do not really change. So there is clearly some issue of communicating with the acromag cards. |
16711
|
Mon Mar 7 18:53:16 2022 |
Koji | Update | BHD | Re-susspension of AS1 |
Not sure if that small difference can cause the alignment inability. Particularly, the removed metal was just below the wire. This means that there is no misalignment effect at the first order.
Here is my idea:
You may be able to assist the alignment by adding washers on one side of the four holes to this "H" shaped parts. The holes are away from the center line, adding some weight definitely do some misalignment.
|
16710
|
Mon Mar 7 16:56:08 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
{Paco, Yehonathan}
We tried to roughly balance the adapter with two counterweights at the front, like with the other thin optics using an iris. As before, we couldn't get the beam above the iris hole no matter how much we inserted the counterweights into the adapter. We noticed that one of the side blocks is actually the one where the clearance for the wire was made on the wrong side. So there was clearance on both the up and bottom sides of the side block (see attachment 1).
Could this be the cause of the balancing issue? Running out of ideas on how to fix it we gave it a try and replaced it with a spare side block. We also found that the wire on the other side block was kinked so we replaced the wire on this one as well.
After inserting new wires into the side blocks, we hung the adapter on the winches and the beam was above the iris aperture! How could this tiny amount of missing mass make this much difference?
We were able to roughly balance the adapter.
We then tried to balance the roll of the adapter but accidentally knocked off the side magnet 😫.
We usually glue several side magnets together and they all together support the metallic plate on which the magnets are magnetically attached to. This time we had only one side magnet to glue so instead of trying to glue the magnet vertically we are trying to glue it horizontally using a flat surface and a stage to clamp it (attachments 2,3).
BTW, the HeNe was not working when we came into the cleanroom. We realized it was the old HeNe that we already determined to be broken but there was no sign on it. I attached a "BAD" sign on it and replaced it with the new HeNe. The OpLeve beam was realigned. All of this happened before all the things described above |
Attachment 1: signal-2022-03-07-171520_001.png
|
|
Attachment 2: signal-2022-03-07-172659_001.jpeg
|
|
Attachment 3: signal-2022-03-07-172659_003.jpeg
|
|
16709
|
Mon Mar 7 16:44:15 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement |
Now that the ETMSUS is back up and running I reran my measurements from the beginning of the process. The results below show a change in gain between the before and after measurements. I have given values of the low-frequency section below.
Average Gain difference from the TFs: 18.867 (excluding thee side change)
|
Before |
After |
Gain difference |
UL |
-31 dB |
-5 dB |
19.952 |
UR |
-36 dB |
-10 dB |
19.952 |
LR |
-27 dB |
-2 dB |
17.782 |
LL |
-37 dB |
-12 dB |
17.782 |
SIDE |
-48 dB |
-45 dB |
1.412 |
I also am noting the new values for the OSEM DC output: average gain increase: 9.004
OSEM |
DC OFFSET Before |
DC OFFSET After |
Gain increase |
UL |
557 |
5120 |
9.19 |
UR |
568 |
5111 |
8.99 |
LR |
780 |
7041 |
9.02 |
LL |
385 |
3430 |
8.91 |
SD |
328 |
2922 |
8.91 |
In addition, the oplev position was:
|
Before |
After |
OPLEV_PERROR |
-16.055 |
-16.715 |
OPLEV_YERROR |
-6.667 |
-16.597 |
All data and settings have been included in the zip file
From the average gain increase of the TFs which indicates the increase of the whole system and the increase in gain from the OSEM we can calculate the gain from the actuators.
18.867/9.004 = 2.09
thus the increase in gain on the actuator is about 2.09
EDIT: I updated the side TF with one with better SNR. I increased the excitation amplitude. |
Attachment 1: UR_TF_Graph.pdf
|
|
Attachment 2: UL_TF_Graph.pdf
|
|
Attachment 3: LR_TF_Graph.pdf
|
|
Attachment 4: LL_TF_Graph.pdf
|
|
Attachment 5: SD_TF_Graph.pdf
|
|
Attachment 6: 20220307_SUSElectronicsAfterTests.zip
|
16708
|
Mon Mar 7 14:55:33 2022 |
Koji | Update | IOO | IMC unlocked again, completely misaligned |
Hmm, the bias values were reset at 2022-03-03-20:01UTC which is 2022-03-03-12:01 PST with no apparent disruption of the data acquisition (= no resetting of the RTS). Not sure how this could happen.
|
16707
|
Mon Mar 7 14:52:34 2022 |
Koji | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
Great trouble shoot!
> I guess, the only remaining issue now is the incorrect atmospheric pressure reading of 1000 Torrs.
This is just a calibration issue. The controller should have the calibration function.
(The other Pirani showing 850Torr was also a calibration issue although I didn't bother to correct it. I think the pirani's typically has large distribution of the calibration values and requires individual calibration) |
16706
|
Mon Mar 7 13:53:40 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
So it appears that my deduction from the pictures of needing a cable swap was correct, however, it turns out that the installed cable was actually the normal RS232 and what we need instead is the RS232 null cable. After the swap was done, the communication between c1vac and the XGS600 controller became active. Although, the data makes it all to the to c1vac without any issues, the scope view of it shows that it is mainly utilizing the upper half of the voltage range which is just over 50% of the available range. I don't know what to make of this.
I guess, the only remaining issue now is the incorrect atmospheric pressure reading of 1000 Torrs.
Quote: |
Following repeated failure to establish communication between c1vac and the XGS600 controller via Perle IOLAN serial device server, I decided to monitor the signal voltage of the communication channels (pin#2, pin#3 and pin#5) using an oscilloscope. The result of this investigation is presented in the attached pdf document. In summary, it seems I have used a crossed wired RS232 serial cable instead of a normal RS232 serial cable, so the c1vac read request command is being relayed on the wrong comm channel (pin#2 instead of pin#3). I will swap out the cable to see if this resolves the problem.
|
|
Attachment 1: iolan_xgs_comm_live.pdf
|
|
16705
|
Mon Mar 7 10:06:32 2022 |
Yehonathan | Update | IOO | IMC unlocked again, completely misaligned |
Came this morning and saw that the IMC is unlocked.
Went into MC Lock screen and see that the watchdog is down and the PSL shutter is closed. I tried to open the shutter but nothing happened - no REFL signal or beam on the MC REFL camera .
Thinking this has something to do with the watchdog I upped the watchdog:
ezcawrite C1:SUS-MC2_LATCH_OFF 1
The watchdog on the MEDM screen became green but the shutter still seemed unresponsive. I went to the PSL table and made sure that the shutter is working. I opened the AS table and saw there no MC REFL beam anywhere.
Thinking that MC1 must be completely misaligned I opened the MC align screen to find that indeed all the alignment values has been zeroed! (attachment).
I burt restore c1iooepics from Mar 4th 00:19. Didn't help.
I try to burt restore c1susepics from Mar 1st 13:19. Still zero.
I try to burt restore c1susaux from Mar 1st 00:19 -> seems like alignment values have been restored.
I open the shutter. Beam is flying! MC Watchdogs tripped! I close the shutter. OK, I need to wait until the MCs are dampped enough. MC2 and MC3 have relaxed so I enable their watchdogs. MC1 is still swinging a bit. I turn on damping for MC1 as well.
MC locked immediately but the REFL is still high like 1.2. Is it normal?
I turn on the WFSs and the REFL went down to 0.3 nice. I run the MC WFS relief script. |
Attachment 1: Screenshot_2022-03-07_10-15-19.png
|
|
16704
|
Sun Mar 6 18:14:45 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
Following repeated failure to establish communication between c1vac and the XGS600 controller via Perle IOLAN serial device server, I decided to monitor the signal voltage of the communication channels (pin#2, pin#3 and pin#5) using an oscilloscope. The result of this investigation is presented in the attached pdf document. In summary, it seems I have used a crossed wired RS232 serial cable instead of a normal RS232 serial cable, so the c1vac read request command is being relayed on the wrong comm channel (pin#2 instead of pin#3). I will swap out the cable to see if this resolves the problem. |
Attachment 1: iolan_xgs_comm_investigation.pdf
|
|
16703
|
Sat Mar 5 02:03:46 2022 |
Koji | Update | SUS | ETMY 1Y4 Electronics Replacement |
Oplev saga
Summary
- The new coil driver had not enough alignment range to bring the oplev beam back to the QPD center
- The coil driver output R was reduced from 1.2k to 1.2k//100 = 92.3 +/- 0.4 Ohm
- Now the oplev spot could be moved to the center of the QPD
- The damping gains (POS/PIT/YAW) and the oplev gains were reduced by a factor of 1/10.
- The damping and the oplev servos work now. Fine gain tuning is necessary.
To Do:
- DC value / TF measurements
- Adjust damping gains
- RFM issue
- Connection check
- Cable labeling
== Alignment Range ==
- Since c1auxey was removed, we no longer have C1:SUS-ETMY_PIT_COMM and C1:SUS-ETMY_YAW_COMM . At this moment, all the alignment is taken with the offset input from the fast real-time system via C1:SUS-ETMY_PIT_OFFSET and C1:SUS-ETMY_YAW_OFFSET.
- The oplev spot could not be moved on the center of the QPD without exceeding the DAC output range (~+ or -32000) for the coils. (Attachment 1)
- This is because the old system had a slow but large current range (Rout = 100) and a small current range for the fast control. Until we commission the new HV BIAS Driver, we have to deal with the large DC current with the HAM-A coil driver.
== Modification to the output resistances ==
The following units and the channels were modified. Each channel had a differential current driver and two output resistances of 1.2K. 100Ohm (OHMITE 43F100, 3W) wire wound resistors were added to them in parallel, making the resulting output R of ~92Ohm.
- ETMY HAM A Coil Driver 1: S2100622 (Attachments 2/3) CH1/2/3
- ETMY HAM A Coil Driver 2: S2100621 (Attachments 4/5) CH3
- This modification allowed me to align the oplev spot to the center of the QPD. C1:SUS-ETMY_PIT_OFFSET and C1:SUS-ETMY_YAW_OFFSE are + 2725 (8%FS) and -2341 (7%FS), respectively.
- The previous alignment slider values were -0.9392 and 0.7615 (out of 10). These are the reasonable numbers, considering the change of the Rout from 100 to 92Ohm, and the sign flip.
(By the way, autoBurt files for c1auxex and c1auxey were not properly configured and the history of C1:SUS-ETM*_*_COMM was not recorded.)
== Damping Servos ==
- Now, the POS/PIT/YAW servos experience ~x10 gains. So temporarily these gains were reduced (POS 20->2, PIT 6->0.6, YAW 4->0.4) and the loops are stable when engaged.
- Also the gains of the OPLEV servos were reduced from -4.5 to -0.45. The loops are stable when engaged.
== Snapshot of the working condition ==
Attachent 6 shows the screenshot for the snapshot of the working condition.
To Do
- The damping servos were tested without proper PD whitening compensation.
-> It turned out this is not necessary as our modified PD whitening has the pole and zero at the same freqs as before.
- Compare the DC values of the OSEM outputs and compensate for the gain increase by the "cts2um" filter.
- The end RTS suffers from the RFM issue. There is no data transmitted from the vertex to the end. I suspect we need to restart the c1rfm process. But this will likely suspend all the vertex real-time machines. Careful execution is necessary.
- c1iscey has all the necessary analog connections. But they are not tested. When we lock the green/IR cavity, we'll need them.
- The cable labeling is only half done. |
Attachment 1: oplev_spot.jpg
|
|
Attachment 2: PXL_20220305_090003837.jpg
|
|
Attachment 3: PXL_20220305_090023436.jpg
|
|
Attachment 4: PXL_20220305_091232290.jpg
|
|
Attachment 5: PXL_20220305_091306604.MP.jpg
|
|
Attachment 6: Screenshot_2022-03-05_01-37-26.png
|
|
16702
|
Sat Mar 5 01:58:49 2022 |
Koji | Summary | CDS | paola rescue |
ETMY end ThinkPad "paola" could not reboot due to "Fan Error". It seems that it is the failure of the CPU fan. I really needed a functional laptop at the end for the electronics work, I decided to open the chassis. By removing the marked screws at the bottom lid, the keyboard was lifted. I found that the CPU fan was stuck because of accumulated dust. Once the fan was cleaned, the laptop starts up as before. |
Attachment 1: PXL_20220305_035255834.jpg
|
|
Attachment 2: PXL_20220305_034649120.MP.jpg
|
|
16701
|
Fri Mar 4 18:12:44 2022 |
Koji | Update | VAC | RGA pumping down |
1. Jordan reported that the newly installed Pirani gauge for P2 shows 850Torr while PTP2 show 680 Torr. Because of this, the vacuum interlock fails when we try to open V4.
2. Went to c1vac. Copied the interlock setting file interlock_conditions.yaml to interlock_conditions_220304.yaml
3. Deleted diffpressure line and pump_underspeed line for V4
4. Restarted the interlock service
controls@c1vac:/opt/target/python/interlocks$ sudo systemctl status interlock.service
controls@c1vac:/opt/target/python/interlocks$ sudo systemctl restart interlock.service
controls@c1vac:/opt/target/python/interlocks$ sudo systemctl status interlock.service
5. The above 2~4 was unnecessary. Start over.
Let RP1/3 pump down TP1 section through the pump spool. Then let TP2 pump down TP1 and RGA.
1. Open V7. This made P2 a bit lower (P2 is alive) and P3.
2. Connected the main RP tube to the RP port.
2. Started RP1/3. PRP quickly reaches 0.4Torr.
3. Opened V6 this made P3 and O2 below 1Torr.
4. Close V6. Shutdown RP1/3. Disconnect the RP tube.
5. Turn on auxRP at the wall powe
6. Turn on TP2. Wait for the starting up.
7. Open V4. Once the pressure is below Pirani range, open VM3.
8. Keep it running over the weekend.
9. Once TP2 reached the nominal speed, the "StandBy" button was clicked to lower the rotation speed (for longer life of TP2) |
Attachment 1: Screenshot_2022-03-04_19-38-11.png
|
|
16700
|
Fri Mar 4 11:04:34 2022 |
Anchal | Summary | CDS | c1susaux2 system setup and running |
I took the c1teststand computer from teststand and converted it into c1susaux2. To do so, I installed a fresh copy of debian 10 on it and followed the steps on this wiki page. I did some parts slightly differently though. The directory /cvs/cds/caltecg/c1susaux2 is a repository and contains the service unit file modbusIOC.service as well. A symbolic link is created at /etc/systemd/system to use this service file for creating the modbusIOC service. All db files are generated by parsing the acromag chassis wiring file using this python script.
The service file is running without any errors now and all channels are available. The leftmost bench on EEshop at 40m is now ready to do LO1 slow controls and monitor testing. If someone gets time today, they can hookup an unused coil driver to the chassis and verify ENABLE switching and monitoring through the optical isolators. We can also drive some voltage on the PD monitors and verify the functioning of our ADCs. Once this test passes, it is straight forward to finish the remaining 6 SOS wiring and we would be good to install the chassis.
Attaching wiring diagram of c1susuaux2 acromag chassis. Any comments/modification suggestions should come soon as we'll go ahead and wire it soon.
Note: While accessing channels using caget on c1susuaux2, you might get a warning "Identical process variable names on multiple servers". You can safely ignore it. It just means that channel is accessible on that particular computer via two different network interfaces (martian network eno1 and acromag subnetwork eno2) and it will just pick one of them. |
Attachment 1: 40mBHD_C1SUSAUX2_Acromag_Chassis.pdf
|
|
16699
|
Thu Mar 3 17:21:11 2022 |
Ian MacMillan | Update | SUS | ETMY 1Y4 Electronics Replacement |
[Koji, Ian]
1) We attached the 30 coil driving cables to the vacuum feed through to the sat amp [40m wiki] they run along the cable tray then up and down into the rack.
2) we checked all DB and power cables. We found that the anti-imaging filter had a short and got very hot when plugged in. the back power indicator lights turned on fine but the front panel stayed off. We removed it and replaced it with the one that was on the test stand marked for the BHD. This means we need to fix the broken one and Koji mentioned getting another one.
3) we reassigned the ADC and DAC channels in the iscey model and the asy model. we committed a version before we made any changes.
4) Finally we tested the setup to make sure the ETM was being damped.
Next step:
1) Measure the change of the PD gains and the actuator gains. See pervious elog |
16698
|
Thu Mar 3 17:09:46 2022 |
Paco | Update | BHD | Re-susspension of AS1 |
[Anchal, Paco]
Wire clamp spare was installed, furthermore AS1 was reinstalled on adapter, attached wire clamps, and cleaned using ionized air gun. Finally, we suspended it on the SOS tower and left it resting on the bottom earthquake stops; ready for balancing.
Quote: |
Yesterday, I rebuilt the OpLev setup in the cleanroom in order to suspend AS1. It took me a while to find all the necessary parts but I found them in the end.
The HeNe laser was placed on the optical table and turned on. The beam was aimed to bounce off a folding mirror to the SOS tower.
The beam's height was controlled by the HeNe laser stage and made to be 5+14/32". The beam from the folding mirror was made parallel to the table, first with an iris and then with the QPD connected to a scope.
Preparing the SOS tower for the suspension I noticed that the wire clamp is scratched on both sides from previous suspensions. I discarded that wire clamp but couldn't find the spares. Time ran out and I had to stop.
|
|
16697
|
Thu Mar 3 15:37:40 2022 |
Anchal | Summary | CDS | c1teststand restructured |
c1teststand has been restructured. There is no port computer called 'c1teststand' anymore. When you ssh into the c1teststand network using ssh c1teststand from inside martian or from outside network using the method mentioned in this wiki page , you would land into chiara (clone) computer and you can navigate into any teststand network computer from there.
I'll be repurposing 1U c1teststand computer into the new c1susaux2 slow machine now. All files from home directory and from /etc directory of former c1teststand have been zipped and stored in /home/controls of chiara (clone). Just a aside, the network configuration of teststand can be done from inside the teststand network, by going to a browser on either fb1 (clone) or chair (clone) and going to address 10.0.1.1. The login and password are same as our usual workstation username and password. |
16696
|
Thu Mar 3 04:24:23 2022 |
Koji | Update | SUS | ETMY 1Y4 Electronics Replacement |
The DC power strip at Y-end was connected to the bottom two Sorensen power supplies. They are configured to provide +/-18V.
|
Attachment 1: PXL_20220303_094421604.jpg
|
|
Attachment 2: PXL_20220303_094435127.jpg
|
|
16695
|
Thu Mar 3 04:11:36 2022 |
Koji | Update | SUS | ETMY 1Y4 Electronics Replacement |
For the Y-end electronics replacement, we want to remove unused power supplies. In fact, we already removed the +/-5V supplies from the stack. I was checking what supply voltages are used by the Eurocard modules. I found that D990399 QPD whitening board had the possible use of +/-5V.
The 40m Y-end version can be found here D1400415. The +/-5V supply voltages are used at the input stage AD620 and the QPD bias voltage of -5V.
AD620 can work with +/-15V. Also the bias voltage can easily be -15V. So I decided to cut the connector legs and connected +5V line to +15V, and -5V line to -15V.
With this modification, I can say that the eurocards only use the +/-15V voltages and nothing else.
The updated schematics can be found as D1400415-v6 |
Attachment 1: PXL_20220303_082726693.jpg
|
|
Attachment 2: PXL_20220303_082752494.jpg
|
|
Attachment 3: PXL_20220303_082744464.jpg
|
|
16694
|
Wed Mar 2 14:02:43 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
Yesterday, I rebuilt the OpLev setup in the cleanroom in order to suspend AS1. It took me a while to find all the necessary parts but I found them in the end.
The HeNe laser was placed on the optical table and turned on. The beam was aimed to bounce off a folding mirror to the SOS tower.
The beam's height was controlled by the HeNe laser stage and made to be 5+14/32". The beam from the folding mirror was made parallel to the table, first with an iris and then with the QPD connected to a scope.
Preparing the SOS tower for the suspension I noticed that the wire clamp is scratched on both sides from previous suspensions. I discarded that wire clamp but couldn't find the spares. Time ran out and I had to stop. |
16693
|
Wed Mar 2 12:40:08 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
Connector Test:
A quick test to rule out any issue with the Ethernet to Serial adapter was done using the setup shown in Attachment 1. The results rule out any connector problem.
IOLAN COMM test (as per Koji's suggestion):
The next step is to swap the controller with a laptop set up to receive serial commands using the same settings as the XGS600 controller. Basically, run a slightly modified version of python script where we go into listening mode. Then send commands to the TCP socket on the IOLAN SDS unit using c1vac and check what data makes its way to the laptop USBserial terminal. After working on this for a bit, I realized that we do not need to do anything on the c1vac machine. We only need to start the service as it would work normally. So I wrote a small python code for a basic XGS-600 controller emulator, see Attachment 4. The outputs from the laptop and c1vac terminals are Attachments 5 and 6 respectively.
These results show that we can communicate via the assigned IP address "192.168.114.22" and the commands that are sent from c1vac reaches the laptop in the correct format. Furthermore, the serial_XSG service, a part modbusIOC_XGS service, which usually exits with an error seems fine now after successfully communicating with the laptop. I don't know why it did not die after the tests. I also found a bug in my code as a result of the test, where the status field for the fourth gauge didn't get written to.
Pressure reading issue:
I noticed that the pressure reading was not giving the atmospheric value of ~760 Torrs as expected. Looking through my previous readouts, it seems the unit showed this atm value of ~761 Torrs when the first gauge was attached. However, a closer look at the issue revealed a transient behavior, i.e. when the unit is turned on the reading dips to atm value but eventually rises up to 1000 Torrs. I don't think this is a calibration problem bcos the value of 1000 Torrs is the maximum value for the gauge range. I also found out that when the XGS-controller has been running for a while, a power cycle does not have this transient behavior. So maybe a faulty capacitor somewhere? I have attached a short video clip that shows what happens when the XGS-controller unit is turned on. |
Attachment 1: IMG_20220302_123529382.jpg
|
|
Attachment 2: XGS600_Serial2Ethernet2Serial2USB_comm_test_result.txt
|
$ python3 XGS600_comm_test.py
----- Multiple Sensor Read Commands -----
Sent to XGS-600 -> #0001\r : Read XGS contents
response : >FE4CFE4CFE4C
Sent to XGS-600 -> #0003\r : Read Setpoint States
response : >0000
... 73 more lines ...
|
Attachment 3: VID-20220302-WA0001.mp4
|
Attachment 4: comm_test_c1vac_to_laptop_via_iolansds.py
|
#!/usr/bin/env python
#Created 3/2/22 by Tega Edo
'''Script to emulate XGS-600 controller using laptop USBserial port'''
import serial
import sys,os,math,time
ser = serial.Serial('/dev/cu.usbserial-1410') # open serial port
... 19 more lines ...
|
Attachment 5: laptop_terminal.txt
|
(base) tega.edo@Tegas-MBP serial % python3 comm_test_c1vac_to_laptop_via_iolansds.py
----- Listen for USBserial command and asynchronously send data in XGS600 format -----
Command received from c1vac [1] :
Data sent to c1vac [1] : >1.000E+00,NOCBL ,NOCBL ,NOCBL ,2.00E+00,NOCBL\r
Command received from c1vac [2] :
Data sent to c1vac [2] : >2.000E+00,NOCBL ,NOCBL ,NOCBL ,3.00E+00,NOCBL\r
... 54 more lines ...
|
Attachment 6: c1vac_terminal.txt
|
controls@c1vac:/opt/target/python/serial$ caget C1:Vac-FRG1_status && caget C1:Vac-FRG2_status && caget C1:Vac-FRG3_status && caget C1:Vac-FRG4_status && caget C1:Vac-FRG5_status
C1:Vac-FRG1_status 1.530E+02
C1:Vac-FRG2_status OFF
C1:Vac-FRG3_status OFF
C1:Vac-FRG4_status NO COMM
C1:Vac-FRG5_status 1.55E+02
controls@c1vac:/opt/target/python/serial$ caget C1:Vac-FRG1_status && caget C1:Vac-FRG2_status && caget C1:Vac-FRG3_status && caget C1:Vac-FRG4_status && caget C1:Vac-FRG5_status
C1:Vac-FRG1_status 1.630E+02
C1:Vac-FRG2_status OFF
C1:Vac-FRG3_status OFF
... 70 more lines ...
|
16692
|
Wed Mar 2 11:50:39 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
Here is the IOLAN SDS TCP socket setting and the USBserial setting for comparison.
I have also included the python script and output from the USBserial test from earlier. |
Attachment 1: XGS600_IOLAN_settings_1.png
|
|
Attachment 2: XGS600_IOLAN_settings_2.png
|
|
Attachment 3: XGS600_USBserial_settings.png
|
|
Attachment 4: XGS600_comm_test.py
|
#!/usr/bin/env python
#Created 2/24/22 by Tega Edo
'''Script to read/write to the XGS-600 Gauge Controller'''
import serial
import sys,os,math,time
ser = serial.Serial('/dev/cu.usbserial-1410') # open serial port
... 74 more lines ...
|
Attachment 5: XGS600_comm_test_result.txt
|
----- Multiple Sensor Read Commands -----
Sent to XGS-600 -> #0001\r : Read XGS contents
response : >FE4CFE4CFE4C
Sent to XGS-600 -> #0003\r : Read Setpoint States
response : >0000
Sent to XGS-600 -> #0005\r : Read software revision
response : >0206,0200,0200,0200
... 69 more lines ...
|
16691
|
Tue Mar 1 20:38:49 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
During my investigation, I inadvertently overwrote the serial port configuration for the connected devices. So I am now working to get it all back. I have attached screenshots of the config settings that brought back communication that is not garbled. There is no physical connection to port 6, which I guess was initially used for the UPS serial communication but not anymore. Also, ports 9 and 10 are connected to Hornet and SuperBee, both of which have not been communicating for a while and are to be replaced, so there is no way to confirm communication with them. Otherwise, the remaining devices seem to be communicating as before.
I still could not establish communication with the XGS-600 controller using the serial port settings given in the manual, which also happen to work via Serial to USB adapter, so I will revisit the problem later. My immediate plan is to do a Serial Ethernet, then Ethernet to Serial, and then Serial to USB connection to see if the USB code still works. If it does then at least I know the problem is not coming from the Serial to Ethernet adapters. Then I guess I will replace the controller with my laptop and see what signal comes through when I send a message to the controller via the IOLAN serial device server. Hopefully, I can discover what's wrong by this point.
Note to self: Before doing anything, do a sanity check by comparing the settings on the IOLAN SDS and the config settings that worked for the Serial to USB communication and post an elog for this for reference. |
Attachment 1: Working_Serial_Port_List_1.png
|
|
Attachment 2: Working_Serial_Port_List_2.png
|
|
Attachment 3: Working_Config_Ports#1-5.png
|
|
Attachment 4: Working_Config_Ports#7-8.png
|
|
16690
|
Tue Mar 1 19:26:24 2022 |
Koji | Update | SUS | ETMY SUS Electronics Replacement |
The replacement key switches and Ne Indicators came in. They were replaced and work fine now.
The power supply units were tested with the X end HeNe display. It turned out that one unit has the supply module for 1350V 4.9mA while the other two do 1700V 4.9mA.
In any case, these two ignited the HeNe Laser (1103P spec 1700V 4.9mA).
The 1350V one is left at the HeNe display and the others were stored in the cabinet together with spare key SWs and Ne lamps. |
Attachment 1: PXL_20220302_025723651.jpg
|
|
Attachment 2: PXL_20220302_030102033.MP.jpg
|
|
16689
|
Tue Mar 1 16:01:14 2022 |
Paco | Update | Electronics | RFSoC 2x2 board -- setup for remote work & BALUN saga |
[Tommy, Paco]
Since last week I've worked with tommy on getting the RFSoC 2x2 board to get some TFs from simple minicircuits type filters. The first thing I did was set up the board (which is in the office area) for remote access. I hooked up the TCP/IP port to a wall ethernet socket (LIGO-04) and the caltech network assiggned some IP address to our box. I guess eventually we can put this behind the lab network for internal use only.
After fiddling around with the tone-generators and spectrum analyzer tools in loopback configuration (DAC --> ADC direct connection), we noticed that lower frequency (~ 1 MHz) signals were hardly making it out/back into the board... so we looked at some of the schematics found here and saw that both RF data converters (ADC & DAC) interfaces are AC coupled through a BALUN network in the 10 - 8000 MHz band (see Attachment #1). This is in principle not great news if we want to get this board ready for audio-band DSP.
We decided that while Tommy works on measuring TFs for SHP-200 all the way up to ~ 2 GHz (which is possible with the board as is) I will design and put together an analog modulation/demodulation frontend so we can upconvert all our "slow" signals < 1MHz for fast, wideband DSP. and demodulate them back into the audio band. The BALUN network is pictured in Attachment #2 on the board, I'm afraid it's not very simple to bypass without damaging the PCB or causing some other unwanted effect on the high-speed DSP. |
Attachment 1: balun_adc1.png
|
|
Attachment 2: PXL_20220302_001734121.jpg
|
|
16688
|
Mon Feb 28 19:15:10 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
I decided to create an independent service for the XGS data readout so we can get this to work first before trying to integrate into current system. After starting the service, I noticed that the EPICS channel were not updating as expected. So I started to debug the problem and managed to track it down to an ip socket connect() error, i.e. we get a connection error for the ip address assigned to the LAN port to which the XGS box was connected. After trying a few things and searching the internet, I think the error indicates that this particular LAN port is not yet configured. I reached this conclusion after noting that only a select number of LAN ports connected without issues and these are the ports that already had devices connected. So it must be the case that the LAN ports were somehow configured. The next step is to look at the IOLAN manual to figure out how to configure the ip port for the XGS controller. Fingers crossed. |
16687
|
Mon Feb 28 15:51:07 2022 |
Ian MacMillan | Update | SUS | ETMY 1Y4 Electronics Replacement |
[Paco, Ian]
paco helped me wire the ETMY 1Y4 rack. We wired the following (copied from Koji's email):
- Use DB9-DB9 to complete the wiring between
- 16bit DAC AI Chassis - End DAC Adapter (4 cables)
- End DAC Adapter - HAM-A Coil Driver (2 cables)
- AA Chassis - End ADC Adapter (2 cables)
- Koji already brought two special DB9-DB15 cables (in plastic bags) to the end. They connect the HAM-A coil drivers to the satellite amp. At this time, we skip Low Noise HV Bias Driver.
- Bring two 30ft DB9 (called #1, aka D2100675-01) cables from the office area to the end. I collected one end and left them there.
- All the new units have +/-18V DC supply in the back. Find the orange cables behind the 40m vacuum duct around Y-end and connect the units and the DC power strip. Use short cables if possible to save the longer ones.
the cables we used:
Number Used |
Type of Cable |
Length |
8 |
DB9 to DB9 |
2.5 ft |
2 |
DB9 to DB9 |
5 ft |
2 |
DB9-DB15 |
|
2 |
DB9 (called #1, aka D2100675-01) |
30ft |
9 |
Orange Power Cables |
~ 3 ft |
I attached pictures below. |
Attachment 1: IMG_0865.jpg
|
|
Attachment 2: IMG_0867.jpg
|
|
Attachment 3: IMG_0868.jpg
|
|
Attachment 4: IMG_0866.jpg
|
|
16686
|
Sun Feb 27 01:12:46 2022 |
Koji | Update | General | IMC manual alignment procedure |
We expect that the MC sus are susceptible to the temperature change and the alignment drifts away with time.
Here is the proper alignment procedure.
0) Assume there is no TEM00 flash or locking, but the IMC is still flashing with higher-order modes.
1) Use the CCD camera and WFS DC spots to bring the beam to the nominal position.
2) Use only MC2 and MC3 to align the cavity to have low-order modes (TEM00,01,02 etc)
3) You should be able to lock the cavity on one of these modes. Minimize the reflection (maximize the transmission) for that mode.
4) This should allow you to jump to a better lower-order mode. Continue alignment optimization only with MC2/3 until you get TEM00.
5) Optimize the TEM00 alignment only with MC2/3
6) Look at the MC end QPD. use one of the scripts in scripts/MC/moveMC2 . Note that the spot moves opposite to the name of the scripts. i.e. MC2_spot_down moves the spot up, MC2_spot_right moved the spot left, etc...
These scripts move MC1/2/3 and try to keep the good MC transmission.
7) moveMC2 scripts are not perfect. As you use them, it makes the MC alignment gradually degraded. Use MC2 and MC3 to recover good transmission.
8) If MC2 spot is satisfactory, you are done.
-------------
Step 6-8 can be done with the WFS on. This way, you can skip step 7 as the WFS servo takes care of it. But if the spot move is too fast, the servo can't keep up with the change. If so, you have to wait for the settling of the servo. Once the spot position is satisfactory, MC servo relief should be run so that the servo offset (in actuation) can be offloaded to the bias slider.
|
Attachment 1: PXL_20220226_100859871.jpg
|
|
16685
|
Sun Feb 27 00:37:00 2022 |
Koji | Update | General | IMC Locking Recovery |
Summary:
- IMC was locked.
- Some alignment change in the output optics.
- The WFS servos working fine now.
- You need to follow the proper alignment procedure to recover the good alignment condition.
Locking:
- Basically followed the previous procedure 40m/16673.
- The autolocker was turned off. Used MC2 and MC3 for the alignment.
- Once I hit the low order modes, increased the IN1 gain to acquire the lock. This helped me to bring the alignment to TEM00
- Found the MC2 spot was way too off in pitch and yaw.
- Moved MC1/2/3 to bring the MC2 spot around the center of the mirror.
- Found a reasonably good visibility (<90%) at a MC2 spot. Decided this to be the reference (at least for now)
SP Table Alignment Work
- Went to the SP table and aligned the WFS1/2 spots.
- I saw no spot on the camera. Found that the beam for the camera was way too weak and a PO mirror was useless to bring the spot on the CCD.
- So, instead, I decided to catch an AR reflection of the 90% mirror. (See Attachment 1)
- This made the CCD vulnerable to the stronger incident beam to the IMC. Work on the CCD path before increasing the incident power.
MC2 end table alignment work
- I knew that the focusing lens there and the end QPD had inconsistent alignment.
- The true MC2 spot needs to be optimized with A2L (and noise analysis / transmitted beam power analysis / etc)
- So, just aligned the QPD spot using today's beam as the temporary target of the MC alignment. (See Attachment 2)
Resulting CCD image on the quad display (Attachment 3)
WFS Servo
- To activate the WFS with the low transmitted power, the trigger threshold was reduced from 5000 to 500. (See Attachment 4)
- WFS offset was reset with /opt/rtcds/caltech/c1/scripts/MC/WFS/WFS_RF_offsets
- Resulting working state looks like Attachment 5 |
Attachment 1: PXL_20220226_093809056.jpg
|
|
Attachment 2: PXL_20220226_093854857.jpg
|
|
Attachment 3: PXL_20220226_100859871.jpg
|
|
Attachment 4: Screenshot_2022-02-26_01-56-31.png
|
|
Attachment 5: Screenshot_2022-02-26_01-56-47.png
|
|
16684
|
Sat Feb 26 23:48:14 2022 |
Koji | Update | SUS | ETMY SUS Electronics Replacement |
[Ian, Koji] - Activity on 25th (Fri)
We continued working on the ETMY electronics replacement.
- The units were fixed on the rack along with the rack plan.
- Unnecessary Eurocard modules were removed from the crate.
- Unnecessary IDC cables and the sat amp were removed from the wiring chain. The side cross-connects became obsolete and they also were removed.
- A 18V DC power strip was attached to one of the side DIN rails.
Warning:
- Right now the ETMY suspension is free and not damped. We are relying on the EQ stops.
Next things to do:
- Layout the coil driving cables from the vacuum feedthru to the sat amp (2x D2100675-01 30ft ) [40m wiki]
- Layout DB cables between the units
- Layout the DC power cables from the power strip to the units
- Reassign ADC/DAC channels in the iscey model.
- Recover the optic damping
- Measure the change of the PD gains and the actuator gains. |
Attachment 1: PXL_20220226_023111179_2.jpg
|
|
16683
|
Sat Feb 26 15:45:14 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
I have attached a flow diagram of my understanding of how the gauges are connected to the network.
Earlier today, I connected the XGS-600 gauge controller to the IOLAN Serial Device Server on port 192.168.114.22 .
The plan is a follows:
1. Update the serial device yaml file to include this new ip entry for the XGS-600 gauge controller
2. Create a serial gauge class "serial_gauge_xgs.py" for the XGS-600 gauge controller that inherits from the serial gauge p arent class for EPICS communication with a serial device via TCP sockets.
- Might be better to use the current channels of the devices that are being replaced initially, i.e.
-
C1:Vac-FRG1_pressure |
C1:Vac-CC1_pressure |
C1:Vac-FRG2_pressure |
C1:Vac-CCMC_pressure |
C1:Vac-FRG3_pressure |
C1:Vac-PTP1_pressure |
C1:Vac-FRG4_pressure |
C1:Vac-CC4_pressure |
C1:Vac-FRG5_pressure |
C1:Vac-IG1_pressure |
3. Modify the launcher file to include the XGS gauge controller. Following the same pattern used to start the service for the other serial gauges, we can start the communication between the XGS-600 gauge controller and the IOLAN serial server and write data to EPICS channels using
controls@c1vac> python launcher.py XGS600
If we are able to establish communication between the XGS-600 gauge controller and write it gause data to EPICS channels, go on to steps 4.
4. Create a serial service file "serial_XGS600.service" and place it in the service folder
5. Add the new EPICS channels to the database file
6. Add the "serial_XGS600.service" to line 10 and 11 of modbusIOC.service
7. Later on, when we are ready, we can restart the updated modbusIOC service
For vacuum signal flow and Acromag channel assignments see [1] and [2] respectively. For the 16 port IOLAN SDS (Serial Device Server) ethernet connections, see [3].
[1] https://wiki-40m.ligo.caltech.edu/Vacuum-Upgrade-2018?action=AttachFile&do=view&target=40m_Vacuum_System_Signal_Flow.pdf
[2] https://wiki-40m.ligo.caltech.edu/Vacuum-Upgrade-2018?action=AttachFile&do=view&target=AcromagChannelAssignment.pdf
[3] https://git.ligo.org/40m/vac/-/blob/master/python/serial/serial_devices.yaml |
Attachment 1: Vac-gauges-flow-diagram.png
|
|
16682
|
Sat Feb 26 01:01:40 2022 |
Tega | Update | VAC | Ongoing work to get the FRG gauges readouts to EPICs channels |
I will make a detailed elog later today giving a detailed outline of the connection from the Agilent gauge controller to the vacuum subnet and the work I have been doing over the past two days to get data from the unit to EPICs channels. I just want to mention that I have plugged the XGS-600 gauge controller into the serial server on the vacuum subnet. I check the vacuum medm screen and I can confirm that the other sensors did not experience and issues are a result of this. I also currently have two of the FRG-700 connected to the controller but I have powered the unit down after the checks. |
16681
|
Fri Feb 25 14:48:53 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement |
I moved the network-enabled power strip from above the power supplies on rack 1y4 to below. Nothing was powered through the strip when I unplugged everything and I connected everything to the same port after. |
Attachment 1: Before.jpg
|
|
Attachment 2: After.jpg
|
|
16680
|
Fri Feb 25 14:00:08 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement |
[Koji, Ian]
We looked at a few power supplies and we found one that was marked "CHECK IF THIS WORKS" in yellow. We found that the power supply worked but the indicator light didn't work. I tried a two other lights from other power supplies that did not work but they did not work. Koji ordered a new one. |
Attachment 1: IMG_0853.jpg
|
|
Attachment 2: IMG_0852.jpg
|
|
16679
|
Thu Feb 24 19:26:32 2022 |
Anchal | Update | General | IMC Locking |
I think I have aligned the cavity, including MC1 such that we are seeing flashing of fundamental mode and significant transmission sum value as well.However, I'm unable to catch lock following Koji's method in 40m/16673. Autolocker could not catch lock either. Maybe I am doing something wrong, I'll pickup again tomorrow, hopefully the cavity won't drift too much in this time. |
16678
|
Thu Feb 24 18:05:58 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
{Yehonathan, Anchal, Paco}
Yesterday, Anchal and Paco removed AS1 from the vacuum chamber and moved it into the cleanroom. The suspension wires were cut and the AS1 optic was put on the table.
Two things were noticed:
1. One of the wires was not sitting inside the side block groove (attachment 1)
2. One of the face magnets was grossly tilted (attachment 2). Probably due to uneven polishing of the dumbbell.
We put new wires into the side blocks making sure they sit in their grooves and we removed the tilted magnet. A different, more straight magnet was picked from the remaining spare magnets. The dumbbell and adapter were cleaned from glue residues and a batch of glue was prepared.
In the process of gluing a different magnet was knocked off. We cleaned that magnet too. The 2 magnets were glued on the adapter.
Today I came and saw that the gluing failed completely. One of the magnets was completely away from its socket and the other one wasn't glued at all.
I prepared a new batch of glue and glued the two magnets. |
Attachment 1: signal-2022-02-24-173933_003.jpeg
|
|
Attachment 2: signal-2022-02-24-173933_002.jpeg
|
|
16677
|
Thu Feb 24 14:32:57 2022 |
Anchal | Update | General | MC RFPD DCMON channel got stuck to 0 |
I found a peculiar issue today. The C1:IOO-MC_RFPD_DCMON remains constantly 0. I wonder if the RFPF output is being read properly. I opened the table and used an oscilloscope to confirm that the DC output from the MC REFL photodiode is coming consistently but our EPICs channel is not reading it. I tried restarting the modbusIOC service but that did not affect anything. I power cycled the acromag chassis while keeping the modbusIOC service off, and then restarted teh modbusIOC service. After this, I saw more channels got stuck and became unresponsive, including the PMC channels. So then I rebooted c1psl without doing anything to the acromaf chasis, and finally things came back online. Everything looks normal to me now but I'm not sure if one of the many channels is not in the right state. Anyways, problem is solved now.
|
16676
|
Wed Feb 23 15:08:57 2022 |
Anchal | Update | General | Removed extra beamsplitter in MC WFS path |
As discussed in the meeting, I removed the extra beam splitter that dumps most of the beam going towards WFS photodiodes. This beam splitter needs to be placed back in position before increasing the input power to IMC at nominal level. This is to get sufficient light on the WFS photodiodes so that we can keep IMC locked for more than 3 days. Currently IMC is unlocked and misaligned. I have marked the position of this beam splitter on the table, so putting it back in should be easy. Right now, I'm trying to align the mode cleaner back and start the WFS loops once we get it locked. |
16675
|
Tue Feb 22 18:47:51 2022 |
Ian MacMillan | Update | SUS | ETMY SUS Electronics Replacement |
[Ian, Koji]
In preparation for the replacement of the suspension electronics that control the ETMY, I took measurments of the system excluding the CDS System. I took transfer functions from the input to the coil drivers to the output of the OSEMs for each sensor: UL, UR, LL, LR, and SIDE. These graphs are shown below as well as all data in the compressed file.
We also had to replace the oplev laser power supply down the y-arm. The previous one was not turning on. the leading theory is that it's failure was caused by the power outage. We replaced it with one Koji brought from the fiber display setup.
I also am noting the values for the OSEM DC output
OSEM |
Value |
UL |
557 |
UR |
568 |
LR |
780 |
LL |
385 |
SIDE |
328 |
In addition the oplev position was:
OPLEV_POUT |
4.871 |
OPLEV_YOUT |
-0.659 |
OPLEV_PERROR |
-16.055 |
OPLEV_YERROR |
-6.667 |
(KA ed) We only care about PERROR and YERROR (because P/YOUT are servo output)
Edit: corrected DC Output values |
Attachment 1: ALL_TF_Graph.pdf
|
|
Attachment 2: 20220222_SUSElectronicsReplacement.7z
|
16674
|
Wed Feb 16 15:19:41 2022 |
Anchal | Update | General | Reconfigured MC reflection path for low power |
I reconfigured the MC reflection path for low power. This meant the following changes:
- Replaced the 10% reflection BS by 98% reflection beam splitter
- Realigned the BS angle to get maximum on C1:IOO-MC_RFPD_DCMON when cavity is unlocked.
- Then realigned the steering mirrors for WFS1 and WFS2.
- I tried to align the light for MC reflection CCD but then I realized that the pickoff for the camera is too low for it to be able to see anything.
Note, even the pick-off for WFS1 and WFS2 is too low I think. The IOO WFS alignment does not work properly for such low levels of light. I tried running the WFS loop for IMC and it just took the cavity out of the lock. So for low power scenario, we would keep the WFS loops OFF.
|
16673
|
Tue Feb 15 19:40:02 2022 |
Koji | Update | General | IMC locking |
IMC is locking now. There was nothing wrong: just a careful alignment + proper gain adj
=== Primary Alignment ===
- I used WFS error signals as the indicator of the PDH error signals. Checked C1:IOO-WFS1_(I/Q)n_ERR and ended up C1:IOO-WFS1_I4_ERR as it showed the largest PDH error PP.
- Then used MC2 and MC3 to align the IMC by maximizing the PDH error and the MC trans (C1:IOO-MC_TRANS_SUM_ERR)
=== Locking procedure ===
Note that the MC REFL path is still configured for the full power input
- (Only at the beginning) Run scripts/MC/mcdown for initialization / Run scripts/MC/MC2tickleOFF just in case
- Enable IOO-MC-SW1 (MC SERVO switch right after "IN1 Gain (dB)").
- Disable 40:4000 boost
- Increase VCO Gain from -15 to 0
- Jiggle IN1 Gain from low to +31 until the lock is achieved
- As soon as the lock is acquired, enable 40:4000
- Increase VCO Gain to +10
- Turn up "SUPER BOOST" from 0 to 3
=== Lock loss procedure ===
Note that the MC REFL path is still configured for the full power input
- Disable IOO-MC-SW1
- Disable 40:4000 boost
- Reduce VCO Gain 0
- Turn down "SUPER BOOST" to 0
- Then jiggle IN1 Gain again to lock the IMC
=== MC2 spot ===
- It was obvious that the MC2F spot was not on the center of the optic.
- I tried to move the spot on the camera as much as possible, but this did not make the trans beam to the center of the MC end QPD
- I had the impression that the trans beam started to be clipped when the beam was moved towards the end QPD,
We need to reestablish the reasonable/consistent MC2 spot on the mirror, the MC end optics, and the QPD.
We will need to use MC2 dithering and A2L coupling to determine the center of the mirror
But as long as the transmission is maximized, the transmitted beam thru MC1 and MC3 follows the input beam. So we can continue the vent work
The current maximized transmission was ~1300. MC1 refl CCD view was largely off -> The camera path was adjusted.
=== MC2 alignment note ===
During the alignment, I noticed a sudden change of the MC2 alignment. There might be some hysteresis in the MC2 suspension. If you are locking the IMC and noticed significant misalignment, the first thing to try is to touch MC2 alignment. |
16672
|
Tue Feb 15 19:32:50 2022 |
Koji | Update | General | Scheduled power outage recovery - IMC recovery progress |
Reduced the IMC power to 100mW
Setup: The power meter was placed right before the final aperture (Attachment 1)
Before the adjustment: Initial position of the HWP was 37.29deg and the input power was 987mW (Attachments 2/3)
After the adjustment: Initial position of the HWP was 74.00deg and the input power was 100mW (Attachments 4/5)
This made the MCREFL reading 0.549.
The MC refl path optics has not been modified. |
Attachment 1: PXL_20220216_001731377.jpg
|
|
Attachment 2: Screen_Shot_2022-02-15_at_16.18.16.png
|
|
Attachment 3: PXL_20220216_001727465.jpg
|
|
Attachment 4: Screen_Shot_2022-02-15_at_16.22.16.png
|
|
Attachment 5: PXL_20220216_002229572.jpg
|
|
16671
|
Mon Feb 14 21:03:25 2022 |
Koji | Update | General | Scheduled power outage recovery |
I opened the boxes. Allegra has obvious vent of at least 4 caps. And the power supply did not respond even a paper clip test was performed. https://www.silverstonetek.com/downloads/QA/PSU/PSU-Paper%20Clip-EN.pdf (Paper Clip Test)
=> The mother board and the PSU are dead.
Then Ottavia was also checked. The mother board looked OK, but the PSU did not respond. I quickly opened the PSU and it had a bunch of bulged capacitors in it. => PSU dead
Conclusion: Save the cards/memory etc as much as possible. Migrate the allegra HDD to any other healthy PC or obtain a new used PC from Larry. Otherwise, we just want to buy another WS and copy the disk in it.
|
Attachment 1: PXL_20220215_025325118.jpg
|
|
16670
|
Mon Feb 14 18:43:49 2022 |
Paco | Summary | General | SOS materials clean room cleared |
[Yehonathan, Paco]
We put away most items used / involved in SOS assembly and characterization. Many were stored in the left-most cabinet in the clean area. The OpLev test setup and optics were stored in the upper cabinets above the microscope area, and several screws and other general components were collected in clean bags or wrapped in foil, labeled and put away. |
16669
|
Mon Feb 14 18:31:50 2022 |
Paco | Update | General | Scheduled power outage recovery - IMC recovery progress |
[Paco, Anchal, Tega]
We have been realigning the IMC as of last Friday (02/11). Today we made some significant progress (still at high input power), but the IMC autolocker is unable to engage a stable mode lock. We have made some changes to reach this point, including re-centering of the MC1 REFL beam on the ccd, centering of MC2 QPD trans (using flashes), and centering of the MC REFL RFPD beam. The IMC is flashing to peak transmission of > 50% its max (near 14,000 counts average on 2021), and all PDs seem to be working ok... We will keep the PSL shutter closed (especially with high input power) for now. |
16668
|
Fri Feb 11 17:07:19 2022 |
Anchal | Update | CDS | [Solved] Huge random numbers flowing into ETMX/ETMY ASC PIT/YAW |
The autoBurt file for FE already has the C1:ASC-ETMX_PIT_SW2 (and other channels for ETMY, ITMX, ITMY, BS and for YAW) present, and I checked the last snapshot file from Feb 7th, 2022, which has 0 for these channels. So I'm not sure why when FE boots up, it does not follow the switch configuration. Fr safety, I changed all the gains of these filter modules, named like C1:ASC-XXXX_YYY_GAIN (where XXXX is ETMX, ETMY, ITMX, ITMY, or BS , and YYY is PIT or YAW) to 0.0. Now, even if the FE loads with switches in ON configuration, nothing should happen. In future, if we use this model for anything, we can change the gain values which won't be hard to track as the reason why no signal moves forward. Note, the BS connections from this model to BS suspension model do not work.
Quote: |
you can hand edit the autoBurt file which the FE uses to set the values after boot up. Just make a python script that amends all of the OFF or ZERO that are needed to make things safe. This would be the autoBurt snap used on boot up only, and not the hourly snaps.
|
Yeah, this is a known issue actually. We go to ASC screen and manually swich off all the outputs after every reboot. We haven't been able to find a way to set default so that when the model comes online, these outputs remain switched off. We should find a way for this.
|
|
|