40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 37 of 344  Not logged in ELOG logo
    Reply  Tue Apr 2 22:39:17 2013, Annalisa, Update, Auxiliary locking, Beat note between "Alberto" NPRO laser and PSL laser Beat_note_3col.jpg
 
 The beat note between the PSL laser and the "Alberto" NPRO laser has been measured. In particular, for each PSL temperature, more
than one Aux laser frequency has been found. 
Entry  Mon Mar 25 23:20:57 2013, Annalisa, Summary, Auxiliary locking, Beat note found! 
[Annalisa, Manasa]
The beat note between the main PSL and the auxiliarly NPRO has been found!
The setup didn't change with respect to the one described on the previous note on the elog.
    Reply  Thu Mar 28 19:32:22 2013, Annalisa, Summary, Auxiliary locking, Beat note found! BeatFreq.jpg
 
 I plot the variation of the beat note frequency as a function of "Alberto" NPRO laser's temperature.
After some discussion, now I'm going to vary the PSL temperature and find the auxiliary NPRO temperature matching to have the beat note between
Entry  Sat Jan 17 15:07:28 2015, manasa, Update, General, Beat note frequency discrepancy 
I was around the PSL table and the X end table today.

X end table:

I was at the X end table making some distance measurements for the telescope to the fiber coupler. I have NOT moved anything
    Reply  Thu Jan 29 11:59:48 2015, manasa, Update, General, Beat note frequency discrepancy Freq_compare1.pngFreq_compare2.png
Below is the set of plots comparing measurements for the green and IR beat notes frequencies. The measurements were made on the spectrum analyzer at
the same time. So I have not taken measurement error into account. 
From the plots, the discrepancy is not very large.
Entry  Wed Mar 27 21:53:21 2013, Annalisa, Update, ABSL, Beat note of ATF auxiliary laser 
After measuring the beat note, the "Alberto" NPRO auxiliary laser has been moved from the PSL table to the POY table. Its beam profile is going
to be measured. It's going to be used as green laser on the END table, in place of the broken one.
The auxiliary laser borrowed form ATF lab (which will be used for the ABSL measurement) has been set on the PSL table to make a measurement of
    Reply  Thu Mar 28 23:00:30 2013, Annalisa, Update, ABSL, Beat note of ATF auxiliary laser found 
 
The beat note for the ATF lab laser has been found. 
The measurement has been carried out in the same way as described in elog 8368.
    Reply  Fri Mar 22 23:23:38 2013, Annalisa, Update, Locking, Beat note still missing Beat_note_-_new_setup.jpg
[Annalisa, Manasa, Koji]
I updated the setup for the beat note. The main reason is that I needed to keep the ADJ to 0 in way to operate at the nominal laser power.
Now the input power of the laser is increased (about 315 mW) and needs to be dumped so as not to exceed the PD threshold of 1mW. 
Entry  Wed Jul 10 02:28:04 2013, Annalisa, Update, Green Locking, Beat notes lost! 
[Manasa, Jenne, Annalisa]
I was going to find the beat note to start the cavity scan, but I couldn't.
These are the steps I followed:
Entry  Mon Jan 22 12:48:48 2018, Koji, Summary, General, Beat setup for aLIGO EOM test IMG_3048.JPG
I'm planning to construct a beat setup between the PSL and AUX beams. I am going to make it in the area shown in a blue square in the attached photo.
This does not disturb Johannes' and PSL setups. The beams are obtained from the PBS reflection of the PSL and the dumped beam of the aux path (0th
or 1st order beam of the AOM).
Entry  Thu Nov 10 05:54:23 2011, kiwamu, Update, Green Locking, Beat-note detected : PSL vs Y arm DSC_3606_small.jpg
[Katrin / Kiwamu]
The beat-note between the PSL green laser and the Y end green laser
was successfully detected.
Entry  Mon Jun 15 22:36:40 2015, rana, koji, Update, Green Locking, BeatBox Assay: not looking good 
Because the ALS beatbox schematic is out-of-date and misleading, we pulled the box to photograph the current implementation and figure out how to proceed.
The box is out on the EE bench right now. Schematic Doc added to 40m Document tree: https://dcc.ligo.org/LIGO-D1102241.
Some notes:
    Reply  Fri Jun 19 01:24:26 2015, rana, koji, Update, Green Locking, BeatBox Assay: not looking good 
We had decided a few days ago, to bypass the IF part of the BeatBox board and put some of the Haixing Maglev generic filter boards in there so that we
could get more whitening and also have it be low noise.

Tonight we wondered if we can ditch the whole BeatBox and just use the quad aLIGO demod box (D0902745)
    Reply  Fri Jun 19 01:55:35 2015, ericq, Update, Green Locking, BeatBox Assay: not looking good 



Quote:



But, it has no whitening.
Can we do the whitening part externally? Perhaps we can run the RF signals from the output of the beat RF Amps over to the LSC rack and then put the outputs
    Reply  Wed Jun 13 22:07:31 2018, gautam, Update, ALS, BeatMouth PDFR measurement IMG_7056.JPGBeatMouthPDFR.pdfBeatMouth_PDFRdata.tgz
Summary:

Neither of the Menlo
FPD310 fiber coupled PDs in the beat mouth have an optoelectronic response (V/W) as advertised. This possibly indicates a damaged RF amplification
    Reply  Fri Jun 15 14:22:05 2018, gautam, Update, ALS, BeatMouth PDFR measurement BeatMouthPDFR.pdfBeatMouth_PDFRdata.tgz
I did the measurement with the BeatMouth open today. Main changes:


Directly pipe the RF output of the Menlo PDs to the Agilent, bypassing the 20dB coupler inside the BeatMouth.
Directly pipe
the unused port of the Fiber Beamsplitter used to send light to the Menlo PD to an in-air collimator, which then sends the beam to the NF1611 reference
    Reply  Thu Mar 28 19:40:02 2019, gautam, Update, ALS, BeatMouth with NF1611s assembled IMG_7381.JPGIMG_7382.JPGrelTF_schem.pdfdarkNoise.pdf
Summary:

The parts I was waiting for arrived. I finished the beat mouth assembly, and did some characterization. Everything looks to be working as expected.

Details:
    Reply  Fri Mar 29 21:00:06 2019, gautam, Update, ALS, BeatMouth with NF1611s installed IMG_7384.JPGIMG_7385.JPG

Newfocus 15V current limited supply was taken from bottom NE corner of the ITMY Oplev table to power the BeatMouth on the PSL table
BeatMouth
was installed on top shelf on PSL table [Attachment #1].
Light
levels in fibers were checked:

PSL: initially, only ~200uW / 4mW was coupled in. This was improved to 2.6mW/4mW (~65% MM) which was deemed
Entry  Mon Jul 8 21:46:31 2013, manasa, Update, Green Locking, Beatbox 
[Koji, Manasa]
I wanted to investigate on the ALS electronics(in particular the beatbox and the phase tracker) and find out if the beatbox is showing a linear
behavior 
Entry  Tue Jul 16 10:16:23 2013, Koji, Update, , Beatbox XARM whitening modified Beatbox_mod.pdfALS_whitening.pdfBeatbox_input_dependence.pdf
The X arm whitening filters of the beatbox were modified.
Now we have about 10 times better floor level above 100Hz and ~3 better at 1Hz.


- The previous whitening was zero@1Hz, pole@10Hz, and the DC gain of the unity.
    Reply  Wed Jul 10 02:10:41 2013, manasa, Update, Green Locking, Beatbox gets a makeover IMG_0842.JPGIMG_0844.JPGIMG_0845.JPGIMG_0846.JPG


    
        
            Quote:
        
        
            

           [Koji, Manasa]
    Reply  Fri Jun 12 17:09:58 2015, ericq, Update, LSC, Beatbox needs whitening gain beatGain.png
Short entry just to preview a new development; more detail about this investigation will soon follow. 

The beatbox I and Q signals are too small at the ADC! I was able to
reduce the RMS out of loop ALS sensitivity (arm locked on POX) by 300Hz with G=10 SR560s between the beatbox output and the ADC whitening
    Reply  Sat Jun 13 18:37:46 2015, ericq, Update, LSC, Beatbox needs whitening gain ADCdiagnosis.png
Here are the promised details!

I was worried about the lack of whitening gain, as I saw that the DC phase tracker Q output (which is the magnitude of the signal in the beatbox's
I-Q plane) was no more than 200 ADC counts for X (~120mV), and 800 for Y (~500mV). I.e. this is the
    Reply  Sat Jun 13 23:52:14 2015, ericq, Update, LSC, Beatbox needs whitening gain 
Nice find!

We ought to use our noise model of the ALS signal chain to determine what the right gains are, rather than hunt and peck. More likely we'll
start from the right gains.​
Entry  Wed Mar 6 20:21:34 2013, Jamie, Update, General, Beatbox pulled from rack 2013-03-06-195232_1060x927_scrot.png
I pulled the beatbox from the 1X2 rack so that I could try to hack in some output whitening filters. 
These are shamefully absent because of my mis-manufacturing of the power on the board.
Right now we're just using the MON output.  The MON output buffer (U10) is the only chip in the output section that's stuffed:
    Reply  Thu Mar 14 11:51:14 2013, Jamie, Update, General, Beatbox upgraded with output whitening, reinstalled 


    
        
            Quote:
        
        
            

           I pulled the beatbox from the 1X2 rack so that I could try to hack in some output whitening filters. 
    Reply  Fri May 4 21:17:57 2018, gautam, Configuration, ALS, BeathMouth pulled out of PSL table 
I have been puzzled about the beat note level we get out of the BeatMouth for some time.


The beat PD used is the Menlo FPD310.
But the version we have is an obsolete version of the product, for which a manual is
hard to find.
Hence, I don't know the transimpedance/electrical characteristics of this PD.
The optical
    Reply  Tue May 8 00:40:51 2018, gautam, Configuration, ALS, BeathMouth pulled out of PSL table IMG_6986.JPGIMG_6987.JPG
Summary:

I did some more BeatMouth characterization. My primary objective was to do a power budget. Specifically, to measure the insertion loss of the
mating sleeves and of the fiber beam splitters. All power numbers quoted in this elog are measured with the fiber power
    Reply  Thu May 24 19:41:28 2018, gautam, Configuration, ALS, BeathMouth reinstalled on PSL table 
Summary:


DC light power incident on beat PD is ~400uW
from the PSL and ~300uW from EX.
These
Entry  Thu Jul 3 18:48:09 2008, Alberto, Update, General, Beats of the two lasers in the absolute length measurement observed  
I adjusted the alignment of the flipper mirror as suggested by Koji making the two beam spots match. I also aligned all the IFO mirrors (ITMs, PRM, SRM,
ETMs) to have more power for the IFO signal at the AS port. When I did that I could see the beats at the AS OSA. Then I explored the range of temperature
of the NPRO from 35deg (C) to 51.2807deg and at that point I could observe a peak corresponding to the beat at about 10MHz on the network analyzer. The
    Reply  Thu Jul 3 22:54:45 2008, Koji, Update, General, Beats of the two lasers in the absolute length measurement observed  
Great! Conguraturation! I wish if I could see it! It's nice if you can put the photo or anything of the RF spectrum analyzer.

Next step:
    Reply  Mon Jul 7 11:22:02 2008, Alberto, Update, General, Beats of the two lasers in the absolute length measurement observed  
I didn't post a screenshot from the RF SA because I had troubles with the interface with the computer (unfortunately the network SA cannot export the data
either).
    Reply  Mon Jul 7 13:06:38 2008, Koji, Update, General, Beats of the two lasers in the absolute length measurement observed  
One may need an RF filter after the mixer. I expect the SR560 does work for this purpose.
If it does not, a passive LPF can be used.
Entry  Mon Oct 25 13:23:45 2021, Anchal, Summary, BHD, Before photos of BSC 
[Yehonathan, Anchal]

On thursday Oct 21 2021, Yehonathan and I opened the door to BSC and took some photos. We setup the HEPA stand next to the door with anti-static
curtains covering all sides. We spend about 15 minutes trying to understand the current layout and taking photos and a video. Any suggestions on improvement
Entry  Thu Jun 16 18:08:57 2016, Johannes, Update, PSL, Before the AOM installation IMG_20160615_154556392e.jpg
There was only one razor blade beam dump labeled for atmospheric use left, but that's all we need. Steve is working on restocking. I placed the modified
AOM mount on the PSL table near its intended location (near the AOM where it doesn't block any beams).

Things to keep in mind:
Entry  Mon Feb 14 18:27:39 2011, josephb, Update, CDS, Began updating to latest CDS svn, reverted to previous state 
[Joe, Alex]
This morning I began the process of bringing our copy of the CDS code up to date to the version installed at Livingston. The motivation was to
get fixes to various parts, among others such as the oscillator part.   This would mean cleaning up front end model .mdl files without having
Entry  Tue May 24 11:04:53 2022, JC, Update, VAC, Beginning Pumpdown PUMPDOWN-2022-05-24_16-57-59.pngC1VAC_Screenshot_2022-05-24_16-59-27.png
[JC, Jordan, Paco, Chub]

We began with the pumpdown this morning. We started with the annulus volume and proceeded by using the following:

1. Isolate the RGA Volume by closing of valves VM3 and V7.
    Reply  Wed Jun 15 07:26:35 2022, JC, Update, VAC, Beginning Pumpdown 
[Jordan, JC]

Jordan and I went in to retore the Vacuum System back to it's original state before the power loss on June 8, 2022. The process went smoothly
as we first closed V7 and opened VM3 (in that order).
Entry  Thu Feb 5 16:21:07 2009, Jenne, Frogs, Computers, Belladonna connects to the wireless Martian network again 
Symptoms:  Belladonna could not (for a while) connect to the wireless network, since there was a driver problem for the wireless card. 
This (I believe) started when Yoichi was doing updates on it a while back.
The system: Belladonna is a Dell Inspirion E1505 laptop, with a Broadcom Corporation Dell Wireless 1390 WLAN Mini-PCI Card (rev 01)
    Reply  Sun Aug 11 11:47:42 2019, gautam, Update, CDS, Bench test of c1iscaux 
I bench tested the functionality of all the c1iscaux Acromag crate channels. Summary: we are not ready for a Monday install, much debugging
remains.


DAC channels were tested using 4 ch oscilloscope and stepping the whitening gain sliders through their 15 gain settings

Response
Entry  Tue Feb 4 18:13:01 2020, Yehonathan, Update, PSL, Bench testing of PSL ai channels c1psl_wire_testing_-_By_Connector.pdf
{Yehonathan, Jon, Jordan}

I tested the ai channels of the new PSL Acromag by looping an already-tested ao channel (C2:PSL-FSS-INOFFSET) back to the different ai channels.

I use Jon's IFOTest with /users/jon/ifotest/PSL.yaml.
    Reply  Wed Feb 5 08:57:11 2020, Yehonathan, Update, PSL, Bench testing of PSL ai channels 
I checked the failed channels against the EPICS database definitions and the yaml file inputted to IFOTest. The channels where the readings are
something other than +10/0 V, but the high/low values do change, I think can be attributed to one of two things:


An incorrect gain and/or offset conversion parameter in the yaml file
The EPICS SMOO parameter (smoothing) is set to some
    Reply  Wed Feb 5 21:04:10 2020, Yehonathan, Update, PSL, Bench testing of PSL ai channels 
{Yehonathan, Jon}

We retested the failed ai channels. Most of them got fixed by applying the inverse calibration in the yaml file.

We still find some anomalous channels, mostly in the DB25 connector. Turns out, their limits were ill-defined in the EPICS database.
Entry  Mon Apr 22 21:33:17 2019, Jon, Update, SUS, Bench testing of c1susaux replacement 
Today I bench-tested most of the Acromag channels in the replacement c1susaux. I connected a DB37 breakout board to each chassis feedthrough connector
in turn and tested channels using a multimeter and calibrated voltage source. Today I got through all the digital output channels and analog input
channels. Still remaining are the analog output channels, which I will finish tomorrow.
    Reply  Fri Aug 21 15:27:00 2020, gautam, Update, BHD, Better BHD mode-matching IMG_7548.JPGBHD_MM.pngPRMI_DCPDs.pdf
Summary:

The mode-matching between the LO and AS beams is now ~50%. This isn't probably my most
average mode-matching in the lab, but I think it's sufficient to start doing some other characterization and we can try squeezing out hopefully another
Entry  Wed Apr 4 21:43:16 2012, Mike J., Update, Computers, Better Hysteresis Model SHO_hyst.png
A better hysteresis model based on the simple harmonic oscillator equation. Useless variables have been removed and output can now be saved to workspace
for plotting. The model is at "/users/mjenson/matlab/SHO_hyst.mdl".
Entry  Tue Jan 29 19:50:49 2013, Jenne, Update, Locking, Better POP layout, no extra PRM motion with locked cavity HalfPRCL_PRM-flatMirror_RefsAreLocked_OthersUnlocked.pdfHalfPRCL_PRM-flatMirror_RefsAreLocked_OthersUnlocked.png
[Jenne, Jamie, Manasa]
Today's activities focused on getting the POP layout improved, so that we could get clean data for the mode scan measurement. 
As Jamie and Koji pointed out yesterday, the beam was still a little too big on the POP DC PD, and was falling off the diode when the beam moved
Entry  Thu Dec 12 21:23:15 2013, ericq, Update, Green Locking, Better Xarm OLTF Dec12_Xarm_OLTF.pdf
With the newly repaired PDH board, I spent some time with the x arm green PDH loop. I found it SO MUCH EASIER to measure the OLTF by injecting before
the servo, instead of after it. (i.e. I added a swept sine from the SR785 to the mixer output (error signal) before the servo input). This is likely because
the error signal is much flatter. I used a 10mV excitation across the whole frequency range (30-100kHz). 
    Reply  Thu Dec 12 22:12:17 2013, Koji, Update, Green Locking, Better Xarm OLTF 
OK, the next question will be "why the loop bandwidth is so miserable?" 
In other words, what is preventing us to have the bandwidth of 20~30kHz?
Your breaking down will give us the answer. 
    Reply  Fri Dec 13 11:47:11 2013, Gabriele, Update, Green Locking, Better Xarm OLTF 
I'm not as good as a fit, but it seems that there is a loop delay of about 30 microseconds, looking at the high frequency phase. This
might explain the limitation on the BW. Eric should be able to get the delay out of the fit with some care.
ELOG V3.1.3-