40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 37 of 344  Not logged in ELOG logo
ID Date Authorup Type Category Subject Text Attachments
  15265   Wed Mar 11 16:46:25 2020 HangUpdateIOOMC2 coil balancingMy old scheme was flawed as I used pitch
as the readback. The pitch signal could not
distinguish the cross-coupling due to coil
  
  15322   Fri May 8 14:27:25 2020 HangUpdateBHDNew SRC gouy phase[Jon, Hang]

After updating the 40 m finesse
file to incorporate the new SRC length (and
 mode_spec_finesse.pdfmode_spec_ideal.pdf 
  15336   Mon May 18 18:00:16 2020 HangUpdateBHDBHD mode-matching study[Jon, Tega, Hang]

We proposed a few BHD mode-matching telescope
designs and then preformed a few monte-carlo
 MM.pdf 
  15339   Wed May 20 18:45:22 2020 HangUpdateBHDBHD mode-matching study--corner plot & adjustment requirementAs Rana suggested, we present the
scattering plot of the AS path
mode matching for various variables.
 AS_MM_scat2.pdfAS_MM_adj2.pdf 
  15357   Tue May 26 19:19:30 2020 HangUpdateBHDBHD MM-- effects of astigmatismPlease see the attached doc. 

I think the conclusion is that
if the AS1 RoC error is not significantly
 MM.pdf 
  15363   Tue Jun 2 14:05:24 2020 HangUpdateBHDMM telescope actuation range requirmentsWe computed the required actuation range
for the telescope design in elog:15357. The result
 TM_OL_spec_1274418500_512.pdfCORNER_OL_spec_1274418500_512.pdf 
  15380   Mon Jun 8 11:50:02 2020 HangUpdateBHDAstigmatism and scattering plotsWe consider the astigmatism effects of
the stock options. The conclusions are:

1. For the AS path, the stock should work
 6x 
  15503   Tue Jul 28 13:55:11 2020 HangUpdateBHDExploring bilinear SRCL->DARM couplingWe explore bilinear SRCL to DARM noise
coupling mechanisms, and show two cases that
by doing BHD readout the noise performance
 SRCL_bilin_DHARD.pdfARM_RIN.pdfSRCL_bilin_SRCL.pdf 
  16373   Mon Oct 4 15:50:31 2021 HangUpdateCalibrationFisher matrix estimation on XARM parameters[Anchal, Hang]

What: Anchal and I measured the
XARM OLTF last Thursday.
 tf_meas.pdffisher_est_vs_data.pdfPxx_evol.pdffisher_evol.pdf 
  16384   Wed Oct 6 15:04:36 2021 HangUpdateSUSPRM L2P TF measurement & Fisher matrix analysis[Paco, Hang]

Yesterday afternoon Paco and I
measured the PRM L2P transfer function.
 prm_l2p_tf_meas.pdfprm_l2p_fisher_vs_data.pdfprm_l2p_Pxx_evol.pdfprm_l2p_fisher_evol.pdf 
  16388   Fri Oct 8 17:33:13 2021 HangUpdateSUSMore PRM L2P measurements[Raj, Hang]

We did some more measurements on
the PRM L2P TF. 
 Screenshot_2021-10-08_17-30-52.png 
  16390   Mon Oct 11 13:59:47 2021 HangUpdateSUSMore PRM L2P measurementsWe report here the analysis results for
the measurements done in elog:16388. 
 prm_l2p_tf_meas_white.pdfprm_l2p_tf_meas_opt.pdfprm_l2p_fisher_vs_data_white_vs_opt.pdfprm_l2p_Pxx_evol_v2.pdf 
  16399   Wed Oct 13 15:36:38 2021 HangUpdateCalibrationXARM OLTFWe did a few quick XARM oltf measurements.
We excited C1:LSC-ETMX_EXC with a broadband
white noise upto 4 kHz. The timestamps for
 Screenshot_2021-10-13_15-32-16.png 
  16467   Tue Nov 16 11:37:26 2021 HangHowToSUSFitting suspension model--large systematic errorsOne goal of our sysID study is to improve
the aLIGO L2A feedforward. Our algorithm
currently improves only the statistical uncertainty
 L2L_L3P_fit.pdfL2L_L3P_residual.pdf 
  16486   Mon Nov 29 15:24:53 2021 HangHowToGeneralFisher matrix vs length of each FFT segmentWe have been discussing how does the parameter
estimation depends on the length per FFT
segment. In other words, after we collected
 setup.pdfFisher_vs_Tperseg.pdffisher_vs_mcmc_offset_removed.pngfisher_vs_mcmc.png 
  17011   Mon Jul 18 15:17:51 2022 HangUpdateCalibrationError propagation to astrophysical parameters from detector calibration uncertainty1. In the error propogation equation, it
should be \Delta \Theta = -H^{-1} M \Delta
\Lambda, instead of the fractional error. 
  
  17029   Sun Jul 24 08:56:01 2022 HangUpdateCalibrationError propagation to astrophysical parameters from detector calibration uncertaintySorry I forgot to put tc & phic in
the example. 

I modified astroFisherLib.py
  
  10083   Fri Jun 20 18:33:53 2014 HarryUpdateGeneralRazorblade Beam Analysis Setup Eric Q and I set up the optical configuration
for razorblade beam analysis on SP table
for future use.
  
  10098   Wed Jun 25 09:16:52 2014 HarryUpdateGeneralRazorblade MeasurementsPurpose
To use a razorblade to measure beam
waist at multiple points along the optical
 40m.zip 
  10100   Wed Jun 25 09:30:44 2014 HarryUpdateGeneralWeekly UpdateSee attached weekly update  Weekly_Update—June_25_thru_July_1.pdf 
  10103   Wed Jun 25 17:49:36 2014 HarryUpdateGeneralRazorblade Analysis Pt. 2Reconfigured razorblade analysis setup
on the PD table as per instructions. Used
it to collect data to calculate beam waist
 RazorbladeSetup.pdf 
  10106   Fri Jun 27 10:09:10 2014 HarryUpdateGeneralBeam Waist MeasurementPurpose
To use a razorblade to measure beam
waist at four points along  the optical axis,
 BeamWaist.zip 
  10139   Mon Jul 7 14:42:33 2014 HarryUpdateGeneralBeam WaistsI was finally able to get a reasonable
measurement for the beam waist(s) of the
spare NPRO.
 16x 
  10158   Tue Jul 8 23:59:49 2014 HarryUpdateGeneralWeekly Plan (7.8.14) Last Week:
-I continued to struggle with the
razorblade beam analysis, though after a
  
  10170   Wed Jul 9 23:02:33 2014 HarryUpdateGeneralBeam Waists via Beamscan Today, I borrowed the beam profiler
from Brian (another SURF) in order to double
check my razor blade measurement figures,
 beamscanSetup.pngbeamProfileX.pngbeamProfileY.pngbeamScanWaist.zipbeamScanWaists.png 
  10175   Thu Jul 10 15:27:09 2014 HarryUpdateGeneralCoupling telescope design I designed this telescope to couple
the 1064 NPRO into the PM980 fiber, using
lenses from the Thorlabs LSB04-C kit.
 telescopeBeamProfile.pngtelescope.zip 
  10185   Fri Jul 11 15:29:26 2014 HarryUpdateGeneralTelescope With Collimator I used a la mode to make a design
for the coupling telescope with a 3.3um target
waist, that included the collimator in the
 telescope_2.ziptelescopeWCollimator.png 
  10204   Tue Jul 15 18:26:40 2014 HarryUpdateGeneralBeam Waist, Telescope, and Fiber Coupling Goal
To design an optical setup (telescope
/ lens) to couple 1064nm NPRO light into
 fiberTestingSetup.pngfiberTestingCouplingDesign.pngfiberTesting.zip 
  10205   Tue Jul 15 18:39:04 2014 HarryUpdateGeneralWeekly Plan (7.16.14) The Past Week
 
Attempted to design coupling telescope,
  
  10218   Wed Jul 16 17:34:11 2014 HarryUpdateGeneralFiber Coupled Purpose
To couple the spare NPRO into our
Panda PM980 fibers, in order to carry out
 fiberTestCouplingSchematic.png 
  10230   Thu Jul 17 17:08:58 2014 HarryUpdateGeneral1X2 Rack Changes Purpose
 
Steve and I moved some things around
 1X2Changes.png 
  10240   Sat Jul 19 01:59:34 2014 HarryUpdateGeneralFiber Mode MeasurementPurpose 
We wanted to measure the mode coming
out of the fibers, so we can later couple
 fiberModeMeasurement.pngfiberModeMeasurement1.pngfiberModeMeasurement1.zip 
  10244   Mon Jul 21 10:30:38 2014 HarryUpdateGeneralFiber Mode Measurement Purpose
The idea was to measure the profile
of the light coming out of the fiber, so
 fiberModeMeasurement.pngfiberModeFit2.pngfiberModeMeasurement2.zip 
  10249   Mon Jul 21 18:08:19 2014 HarryUpdateGeneralFiber Mode Measurement

    

       
            
 fiberModeMeasurement3.png 
  10255   Tue Jul 22 16:26:04 2014 HarryUpdateGeneralFiber Mode MeasurementI repeated this process once more, this
time using the computer controlled stage
that the beam profiler is designed to be
 fiberModeProfile3.pngfiberModeMeasurement4.zip 
  10256   Tue Jul 22 17:45:11 2014 HarryUpdateGeneralWeekly Update The Past Week
 
I spent the past week coupling NPRO
  
  10264   Wed Jul 23 17:54:51 2014 HarryUpdateGeneralCoupling Improvements plus PER Measurement Setup Purpose
We wanted to improve the coupling
into the fibers, because it's very rarely
 PERSetup.png 
  10271   Thu Jul 24 17:37:19 2014 HarryUpdateGeneralCoupling Improvements plus PER Measurement Setup

    

       
            
  
  10273   Fri Jul 25 17:28:31 2014 HarryUpdateGeneralFOL Box and PER Update Purpose
We're putting together a box to go
into the 1X2 rack, to facilitate the frequency
 FOLBox.png 
  10282   Mon Jul 28 17:25:32 2014 HarryUpdateGeneralFiber Mode With Collimators Purpose
We want a measurement of the fiber
modes at either end, with the collimators,
 end1Profiles.pngend2Profiles.pngFiberModeWCollimators.zip 
  10287   Tue Jul 29 18:52:42 2014 HarryUpdateGeneralPSL and AUX Coupling Waist Measurement//edit Manasa//  Harry will update
this elog with before/after pictures of the
table and power of the 1064nm rejected beam
 9x 
  10289   Tue Jul 29 19:00:40 2014 HarryUpdateGeneralWeekly Plan (7.29.14) The Past Week
In the past week, I have improved
the coupling in the fiber testing setup on
  
  10299   Wed Jul 30 18:09:08 2014 HarryUpdateGeneralAUX and PSL Coupling Telescopes Purpose
These telescopes will be used to
mode match//couple the dumped SHG light from
 PSLTelescope.pngAUXTelescope.pngtelescopes.zipAUXTelescopePlot.pngPSLTelescopePlot.png 
  10308   Thu Jul 31 15:25:51 2014 HarryUpdateGeneralPolarizationExtinction Ratio of Fibers Purpose
We wanted to measure the PER of the
polarization maintaining fibers, so we could
 PERSetup.pngPERBeamPower.png 
  10336   Wed Aug 6 10:10:45 2014 HarryUpdateGeneralWeekly Plan 8.6.14Last Week
 
Took first round of PER measurements
  
  10347   Thu Aug 7 14:50:43 2014 HarryUpdateGeneralPER Measurement Purpose
I wanted to do a more robust measurement
of PER of PM fibers for FOL, so I thought
 PERFinalSetup.pngPDCalibration.pngPolarization_Intensity_Variation.pngExtrema.pngPEReport.zip 
  10349   Thu Aug 7 17:09:53 2014 HarryUpdateGeneralAUX Coupling In Progress I'm currently in the process of coupling
dumped SHG light from the Y arm end table
into fibers for FOL.
  
  10360   Sun Aug 10 00:54:54 2014 HarryUpdateGeneralAUX CoupingThe Y End laser dumped SHG light has been
coupled into the yellow fiber that terminates
at the PSL table.
  
  10371   Tue Aug 12 23:07:24 2014 HarryUpdateGeneralPSL TelescopeI put the PSL telescope in place, and started
coupling to it.
Unfortunately, I was only able to
 PSLTelescopePic.png 
  10373   Wed Aug 13 10:49:39 2014 HarryUpdateGeneralWeekly Update In the past week, I designed and
assembled coupling telescopes for the PSL
and Y Arm Lasers
  
ELOG V3.1.3-