ID |
Date |
Author |
Type |
Category |
Subject |
Text |
 |
10533
|
Wed Sep 24 16:02:58 2014 |
Jenne | Update | VAC | vent is completed | [Steve, EricQ, Jenne]
ITMY and BS heavy doors are off,
light doors are on. Q is aligning the |
|
10536
|
Thu Sep 25 08:21:15 2014 |
Steve | Update | VAC | vent day 1 |
|
|
10547
|
Mon Sep 29 09:11:25 2014 |
Steve | Update | VAC | vent day 5 RGA scan |
|
|
10548
|
Mon Sep 29 10:29:25 2014 |
Steve | Update | VAC | RGA is not running |
The RGA time stamp was correct
last at 20140527 |
|
10553
|
Tue Sep 30 16:18:57 2014 |
Steve | Update | VAC | pump down #78 has started |
Q checked the earth quake stops of
SRM and we put the ITMY & BS doors on. |
|
10555
|
Tue Sep 30 18:02:53 2014 |
Steve | Update | VAC | pump down #78 stops at 320 Torr |
We stopped pumping just short
of 3 hours at 320 Torr. Pumping speed |
|
10556
|
Wed Oct 1 10:21:19 2014 |
Steve | Update | VAC | pump down #78 stops at 3.5 Torr |
|
|
10557
|
Wed Oct 1 16:26:53 2014 |
Steve | Update | VAC | pump down #78 completed |
Pump down reached "vacuum
normal" state. IFO _P1 pressure 1e-4 |
|
10559
|
Thu Oct 2 09:23:23 2014 |
Steve | Update | VAC | vent 77 |
|
|
10563
|
Fri Oct 3 10:10:37 2014 |
Steve | Update | VAC | cold cathode gauge reading switched |
We have two cold cathode gauges at
the pump spool and one signal cable |
|
10568
|
Mon Oct 6 10:23:43 2014 |
Steve | Update | VAC | Unexpected power shutdown |
|
|
10579
|
Tue Oct 7 16:55:16 2014 |
Steve | Update | VAC | Unexpected sweaty valves | Pump spool valves V5, V4, V3
sweating a lot. VM3 and VC2 not so much.
|
 |
10584
|
Wed Oct 8 08:46:57 2014 |
Steve | Update | VAC | VAT valves actuator lubricant |
|
|
10587
|
Thu Oct 9 11:56:35 2014 |
Steve | Update | VAC | Power outage II & recovery |
|
|
10590
|
Thu Oct 9 17:33:28 2014 |
Steve | Update | VAC | Power outage II & recovery |
|
|
10597
|
Fri Oct 10 14:41:04 2014 |
Steve | Update | VAC | Power outage II & recovery |
|
|
10599
|
Mon Oct 13 14:44:52 2014 |
Steve | Update | VAC | RGA scan pd78 -day 13 | Our first RGA scan since May 27,
2014 elog10585
The Rga is still warming up. |
|
10741
|
Mon Dec 1 17:13:57 2014 |
Steve | Update | VAC | RGA scan pd78 -day 63 |
|
|
10837
|
Tue Dec 23 14:33:24 2014 |
Steve | Update | VAC | Chiara gets UPS |
|
|
10838
|
Tue Dec 23 15:37:32 2014 |
Steve | Update | VAC | TP3 drypump replaced |
|
|
10843
|
Fri Dec 26 17:45:21 2014 |
Steve | Update | VAC | vac pressure rose to 1.3 mTorr | We run out of N2 for the vacuum system.
The pressure peaked at 1.3 mTorr with MC
locked. V1 did not closed because the N2 |
 |
10845
|
Sun Dec 28 07:29:16 2014 |
Steve | Update | VAC | vacuum is normal |
|
|
10968
|
Tue Feb 3 15:20:13 2015 |
Steve, Koji | Update | VAC | Pneumatic pressure is being read again | [Koji, Steve]
Summary
The N2 pressure |
|
11156
|
Sun Mar 22 18:42:40 2015 |
Steve | Update | VAC | vac pressure rose to 1.2mTorr |
|
 |
11159
|
Mon Mar 23 10:36:55 2015 |
ericq | Update | VAC | Pressure watch script | Based on Jenne's chiara disk usage
monitoring script, I made a script that checks
the N2 pressure, which will send an email |
|
11178
|
Fri Mar 27 10:38:40 2015 |
steve | Update | VAC | vac rack UPS battery replaced | Batteries replaced after 3.5 years with
Amstron AP-1250F2, 8x 12V 6Ah
|
|
11184
|
Tue Mar 31 09:03:32 2015 |
Steve | Update | VAC | RGA scan pd78 day 182 |
|
|
11243
|
Fri Apr 24 17:30:32 2015 |
Jenne | Update | VAC | Pressure watch script broken |
|
|
11249
|
Sat Apr 25 18:50:47 2015 |
ericq | Update | VAC | Pressure watch script broken | Ugh, this turns out to be because cron
doesn't source the controls bashrc that
defines where to find caget and all that |
|
11260
|
Mon Apr 27 14:37:55 2015 |
Steve | Update | VAC | N2 pneumatic pressure watch |
|
|
11261
|
Mon Apr 27 21:42:07 2015 |
rana | Update | VAC | Summary pages | We want to have a VAC page in the summaries,
so Steve |
|
11262
|
Tue Apr 28 09:49:26 2015 |
Steve | Update | VAC | Vac Summery Channels |
|
|
11264
|
Thu Apr 30 16:30:25 2015 |
Steve | Update | VAC | N2 pneumatic pressure watch set up | We have 2 transduser PX303-3KG5V
http://www.omega.com/pressure/pdf/PX303.pdf
They will be installed on the out put |
|
11274
|
Tue May 5 16:02:57 2015 |
Steve | Update | VAC | Vac Summery Channels with discription | As it was requested by the Bos.
It would be nice to read from the
epic screen C1:Vac-state_mon.......Current |
|
11276
|
Fri May 8 14:30:09 2015 |
Steve | Update | VAC | CC1 cold cathode gauges are baked now | CC1s are not reading any longer.
It is an attempt to clean them over the weekend
at 85C |
|
11287
|
Tue May 12 14:57:52 2015 |
Steve | Update | VAC | CC1 cold cathode gauges are baked now | Baking both CC1 at 85 C for 60 hrs did
not help.
The temperature is increased to |
|
11292
|
Fri May 15 16:18:28 2015 |
Steve | Update | VAC | Vac Operation Guide | Vacuum
Operation Guide is up loaded into the
40m-wiki. This is an old master copy. Not |
|
11352
|
Wed Jun 10 15:54:14 2015 |
Steve | Update | VAC | Vacuum comp. rebooted | Koji and Steve succeded rebooting C1vac1,
C1vac2 and pressure reading is working now
More tomorrow ......... |
|
11353
|
Thu Jun 11 19:40:59 2015 |
Koji | Update | VAC | Vacuum comp. rebooted | The serial connections to the vacuum gauges
were recovered by rebooting c1vac1 and c1vac2.
Steve claimed that the vacuum screen |
|
11354
|
Fri Jun 12 08:40:17 2015 |
Steve | Update | VAC | Vacuum comp. rebooted | Koji and Steve,
One computer expert and one vacuum
expert required. |
 |
11387
|
Wed Jul 1 10:01:25 2015 |
Steve | Update | VAC | RGA scan pd78 day 275 |
|
|
11483
|
Thu Aug 6 09:16:57 2015 |
Steve | Update | VAC | Chiara gets new Ethernet card |
Yesterday we did put the vacuum
system into
|
 |
11593
|
Mon Sep 14 10:41:03 2015 |
Steve | Update | VAC | TP2 dry pump replaced | TP2 dry fore pump sn PLE10082 was replaced
at pressure 717 mTorr, TP2 50K rpm
0.33A @ 112,677 hrs |
|
11625
|
Mon Sep 21 11:12:14 2015 |
Steve | Update | VAC | cold cathode is flaky | CC4 cold cathode gauge jump triggered interlock
to close VM1 valve to protect the RGA.
The IFO pressure is 1e-5 Torr |
|
11633
|
Tue Sep 22 08:58:38 2015 |
Steve | Update | VAC | cold cathode is flaky | The cold cathode gauge is back to normal.
cc4 is the last gauge is "functioning"
MKS is not responding. The spare |
|
11644
|
Fri Sep 25 17:00:38 2015 |
Steve | Update | VAC | cold cathode is flaky | The IFO pressure is estimated ~1E-6 Torr
with modified Vac normal valve configuration.
CC4 is in a jumping mode between |
|
11666
|
Mon Oct 5 10:11:35 2015 |
Steve | Update | VAC | RGA scan pd78 day 386 | RGA background scan.
The IFO is closed off from RGA
with VM1 |
|
11678
|
Fri Oct 9 11:41:09 2015 |
ericq | Update | VAC | N2 Pressure fell | At 10:02AM, the N2 Pressure fell below
60 PSI. The watch script saw this happen,
but I did not recieve the email it is supposed |
|
11695
|
Fri Oct 16 18:27:52 2015 |
Steve | Update | VAC | vacuum normal condition | Vacuum normal configuration with VM1-closed,
VM2-open valve positions. Power load normal
24V 0.2A |
|
11697
|
Mon Oct 19 11:20:34 2015 |
gautam | Update | VAC | RGA scan reset | Steve pointed out that in the aftermath
of the Nitrogen running out a couple of times
last week, the RGA had shut itself off thinking |
|