Entry  Mon Aug 8 15:52:39 2011, steve, Update, VAC, back ground rga scan bg-d5.jpgpresBg-d5.jpg
The RGA region is beeing monitored during the vent. This will tell us how clean the RGA itself is.
 
Entry  Tue Aug 9 16:40:12 2011, steve, Update, General, clean room coat counts  P1080155.JPG
I measured the particles coming off of new- unused clean room coat. Tyvek, Convertors, Allegiance #9393 measured  10 counts of 0.3 micron at 1 minute
and 0 count at 7 minute.
The 0.5 micron size  measured 0 at both times. Jenne's used coat measured 20 counts of 0.3 micron  and 0 counts for 0.5 micron at 1 minute.
Entry  Tue Aug 9 16:54:47 2011, steve, Update, VAC, scratch on vac door P1080152.JPG
Jenne found a big scratch on the north vac door of ITMY. Fortunately it does not reach the inner annulos 0 -ring seal.
This is precisely what we have to avoid to preserve our vacuum
system!
    Reply  Wed Aug 10 07:59:33 2011, steve, Update, SUS, this is too dam nice 


    
        
            Quote:
        
        
            

           Suresh and I tweaked the OSEM angles in ETMX yesterday.  Last night the ETMs were left free swinging, and today I ran Rana's peakFit
Entry  Thu Aug 11 08:29:28 2011, steve, Update, Computers, dataviewer at Rosalba gpsfmb.png
I'm having this problem with DTV every morning at Rosalba only. It wants to start with a negative GPS time and it can not connect to the frame builder.
Normally after a few time of starting it, it will work.
    Reply  Thu Aug 11 15:31:43 2011, steve, Update, General, malfunctioning crane at ETMY 


    
        
            Quote:
        
        
            

           The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take
Entry  Thu Aug 11 16:07:37 2011, steve, Update, SUS, ETMY rack cables strain releived P1080159.JPGP1080160.JPG
 
Entry  Fri Aug 12 16:52:49 2011, steve, Update, General, Chamber illuminator switch 
I'm looking for an ether net based  power switch for turning lights on and off for the vacuum system from MEDM screen.  This
is what I found
Jamie please take a look at it.
    Reply  Thu Aug 18 10:59:04 2011, steve, Update, Treasure, Prepared for drag wiping 


    
        
            Quote:
        
        
            

           While waiting for the IFO team to align things (there were already ~5 people working on a ~1 person job...), I got all of our supplies prepped
Entry  Thu Aug 18 15:54:35 2011, steve, Update, SUS, damped and undamped OSEMs 5susLL.jpg8freeSUSLL.jpg
damped sus at atm1 and freeswingging sus at atm2
 
Entry  Fri Aug 19 09:14:56 2011, steve, Update, IOO, degrading laser power at atm power@atm.jpg
Light into the MC is 20 mW at atm, MC_Transmitted ~10 MW = 400 count
The PMC_T is OK but something else is drifting.
Entry  Fri Aug 19 10:26:53 2011, steve, Update, SUS, OSEM sensor spectra 8freeSUSosemSensors.jpg
Free swingging OSEM sensors LL at atm
Entry  Fri Aug 19 15:31:53 2011, steve, Update, General, power interruption rescheduled to 10-1-2011 
                UTILITY
& SERVICE INTERRUPTION
**PLEASE  POST**
Entry  Mon Aug 22 11:40:08 2011, steve, Update, VAC, cranes checked 
Cranes are  checked and they are ready for lifting. At the east end will use the manual Genei-lift to put door on.
    Reply  Thu Aug 25 08:12:09 2011, steve, Update, SUS, ETMY & ITMY wiped, replaced, vac door on 


    
        
            Quote:
        
        
            
            We've closed up ETMX:
Entry  Thu Aug 25 17:14:49 2011, steve, Update, VAC, clean room fashion changes P1080185.JPGP1080183.JPG
Jamie is modeling our next generation  in-vac
& clean room bonny suit  that Jenne and myself already tested.
It is quite bearable with our traditional cleanroom beret-bouffant cap. Please use these in the future.
    Reply  Fri Aug 26 07:53:59 2011, steve, Update, SUS, Broken UL magnet on ITMX 


    
        
            Quote:
        
        
            

           Dmass just reminded me that the usual procedure is to bake the
    Reply  Fri Aug 26 10:45:17 2011, steve, Update, General, crane at ETMY is fixed P1080190.JPGP1080188.JPG


    
        
            Quote:
        
        
            

           The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take
Entry  Fri Aug 26 15:59:33 2011, steve, Update, General, drawer cabinet moves in P1080192.JPG
manufacturer info
Jamie and Shuresh moved in Jenne's 11 drawers cabinet and relocated old note book boxes on the inside of the vac tube.
Entry  Tue Aug 30 10:49:29 2011, steve, Update, SUS, BS & PRM damping restored 
I have restored the damping of BS and PRM. Today is janitor day. He is shaking things around the lab.
Entry  Wed Aug 31 11:37:56 2011, steve, Update, Green Locking, shutter in place at east end 
Uniblitz mechanical shutter installed in the green beam path at ETMY-ISCT  The remote control cable has not been connected.
Entry  Thu Sep 1 11:03:22 2011, steve, Update, VAC, back ground rga scan v70-bg-d29scan.jpgvent70tempres.jpg
The RGA back ground at day 29 of this vent.
Entry  Thu Sep 1 15:07:45 2011, steve, Update, SUS, why ITMY is moving more ? itmy_srm_etmx.jpgP8220152.JPG
Atm1, ITMY and the SRM are on the same isolation stack.  So why does  the SRM move twice as much?
Atm2, We should check the ITMY  SIDE_OSEM before pump down. Anatomically correct, beautiful picture taken by Kiwamu on August 22
Entry  Thu Sep 1 15:59:46 2011, steve, Update, SUS, light doors on at the ITMs 
Suresh, Kiwamu and Steve
Heavy chamber doors replaced by light ones at  ITMX-west and ITMY-north locations.
    Reply  Tue Sep 6 08:05:53 2011, steve, Update, SUS, ITMX must be touching osemITMX.jpg


    
        
            Quote:
        
        
            
            The new ITMX was aligned by changing the DC
Entry  Wed Sep 7 08:14:01 2011, steve, Update, SUS, final OSEM check finalcheck.jpgITMX10min.jpgfinalsum.png
All fine, except ITMX_sensor_UL's  60  counts deep hoop for an hour.
Entry  Wed Sep 7 12:03:49 2011, steve, Update, VAC, Pump down has started 
Jamie and Steve
We closed ITMX and ITMY chambers and started pumping around 11am
What we did before:
    Reply  Wed Sep 7 13:28:25 2011, steve, Summary, Cameras, All Camera setups a need upgrading 


    
        
            Quote:
        
        
            
            I just tried to adjust the ETMY camera and its not very user friendly = NEEDS FIXING.
    Reply  Wed Sep 7 18:19:37 2011, steve, HowTo, VAC, stop pump down for overnight 


    
        
            Quote:
        
        
            
            Jamie and Steve
    Reply  Thu Sep 8 10:50:44 2011, steve, Update, VAC, pumpdown #71 comleted pd#71.jpgmaglevpd.jpg
Vacuum status:  " vacuum normal"   CC1 pressure 5e-5 Torr
Slow roughing through butterfly valve took 11 hours at 1-2 Torr/min speed.  We had a leak at the metal hose connecting o-ring  that became
visible  at 200 Torr.
Entry  Thu Sep 8 13:20:07 2011, steve, Update, PEM, vac envelope at atm for 36 days vent70.jpg
We tried not to open chambers above 10,000 particles of 0.5 micron cf/min
New items going in:                       2 rasor beam traps, 5 badly oxidized
old silver plated  setscrew with spring loaded tips......to be replaced in the future, viton tips for eq screws....some are lose, gold plated allen
    Reply  Tue Sep 13 16:57:44 2011, steve, Update, LSC, ITMX oplev ** 


    
        
            Quote:
        
        
            

           I lined up the Y Arm for locking and then centered the oplevs for ETMY and ITMY.
Entry  Wed Sep 14 07:51:20 2011, steve, Update, SUS, PRM damping restored 
The PRM damping was restored at side sensor var 1050
Entry  Wed Sep 14 15:05:52 2011, steve, Update, PEM, air cond maintenance  ac1y.jpg
Jeff has changed our AC filters inside the lab this morning. Now he is checking on the main filters at CES.  He will finish the roof units tomorrow.
Met One #1 counter is on the top of IOO chamber.  It is measuring 1 and 0.5 micron size particles.  One year of lab condition plot below.
 
    Reply  Thu Sep 15 07:28:08 2011, steve, Update, SUS, PRM damping restored 


    
        
            Quote:
        
        
            
            The PRM damping was restored at side sensor var 1050
Entry  Fri Sep 16 16:07:28 2011, steve, Update, SAFETY, visitors safety training P1080241.JPG
Paul, Mirko and Katrin visiting grad students received the 40m basic safety training.
Entry  Wed Sep 21 11:35:07 2011, steve, Update, VAC, RGA scan rgascanpd71m14d.jpg
RGA scan with maglev pumping speed at day 14 of the pump down.
The larger inserted box contains the tuning parameters of the SRS  200 amu RGA
Entry  Fri Sep 23 16:25:01 2011, steve, Update, VAC, vac rack UPS battery replaced P1080252.JPGP1080254.JPG
APC Smart -UPS 2200   model: SUA2200RM2U   batteries were replaced by compatible RBC43, 8x  12V5A
Entry  Mon Sep 26 11:35:44 2011, steve, Update, Cameras, arms cameras upgraded P1080251.JPG
The arm's CCD cameras were reset as picture shows last week.
The height adjustment only works at ITMX. Short post holders are ordered to make this feature  available on the rest.
The 75 ohms video and power supply cables are stress relieved. Solid cover can be attached now without  miss aligning cameras.
Entry  Wed Sep 28 07:36:41 2011, steve, Update, SUS, ITMY damping restored 
ITMY suspention damping restored
    Reply  Wed Sep 28 07:46:42 2011, steve, Update, SUS, Dsub connections at the back of 1X5 are fixed 
 
 I'm turning  the sus damping  off for Dsub connection fix at  the back of 1X5 rack
The plan is to install 4-40 jack screws to lock connector positions.
Entry  Fri Sep 30 17:01:29 2011, steve, Update, SUS, ETMX oplev intensity effect on noise ETMXoplintens.jpgP1080267.JPG
Kiwamu and Steve,
This is one of the better oplev set up we have. Single bounce from TM, no mirror on stack. One lens and one  mirror on ISCT. Old Uniphase
1103P laser on heafty 3" od mounts.
    Reply  Mon Oct 3 08:38:21 2011, steve, Summary, VAC, Recovery from the power shutdown is completed poweroutage.jpg
 
I failed to start Rana's favorit anciant desktop computer at the vac rack. He has to baby this old beast just a little bit more.
Vacuum status: Vac Normal was reached through Farfalla: Rga was switched back to IFO and and Annuloses are beeing pumped now.
    Reply  Tue Oct 4 07:51:36 2011, steve, Update, CDS, c1lsc and c1sus are still down 
 
 c1lsc and c1sus are still down. Only ETMX and ETMY are damped
    Reply  Wed Oct 5 11:33:25 2011, steve, Update, SUS, PRM and BS oplev laser replaced 
 
JDSU 1103P died after 4 years of service. It was replaced with new identical head of 2.9 mW output. The power supply was also changed.
The return spots of 0.04 mW  2.5 mm diameter on qpds are BS  3,700 counts and PRM 4,250 counts.
Entry  Wed Oct 5 17:08:49 2011, steve, Update, SUS, BS oplev spectra BSoplservON2.png
Kiwamu and Steve,
The He/Ne oplev shows no coherece so relative intensity noise is not limiting factor for the oplev servo
Entry  Fri Oct 7 14:04:48 2011, steve, Update, SUS, He/Ne intensity noise effect on oplevs 7x
The SRM bounce peak 18.33 Hz. Suresh helped me to retune filter through Foton, but we failed to remove it.
ITMX_OPLEV_PERROR shows high coherence with oplev laser. This is our lousiest set up. I will work on it next week.
Generally speaking we can say that JDSU-Uniphase 1103P and 1125P laser intensity noise do not limit oplev servo performance.
    Reply  Mon Oct 10 16:32:18 2011, steve, Update, SUS, UL sensor of ETMY is recovered ETMY_UL.pngETMY_OSEM_UL.pngnoisyETMY_UL.png
 I lost UL osem voltage this morning when I was checking the actual connection at rack ETMY
This after noon I disconnected  the 64 pins IDE connector from satelite amp at the rack, and the two 25 pins Dsubs at this juction board.
UL OSEM recovered after reconnecting these three connectors.
    Reply  Wed Oct 12 08:43:30 2011, steve, Update, SUS, ITMX oplev improved a bit ITMXoplev.pngITMXoplevservo_ON.png
 Atm2 is before optical path adjustment. The idea was to remove possible clipping in vacuum.
Coherense significantly reduced below 4 Hz
Today I will replace the He/Ne laser 1125P with 1103P
    Reply  Thu Oct 13 14:23:09 2011, steve, Update, SUS, ITMX oplev with 3 mm beam on qpd ITMXoplev1103p#2.pngITMXoplev.png
 I replaced the JDSU-Uniphase 1125P by 1103P He/Ne laser. This new laser had 2.8 mW output yesterday. It degraded to 0.5 mW by this morning.
The beam size on the QPD is ~3 mm  This should give us  better sensitivity. These are not the perfect lenses at all, but we have them
here.