  6188   Thu Jan 12 07:44:21 2012 steveUpdateSUSsapphire wire standoff quote


On Tue, Dec 20, 2011 at 11:37 PM, Rana Adhikari wrote:
Steve,

We are thinking of replacing a few metal parts in the suspension with sapphire/ruby. We want to replace the standoff (which is Al) with sapphire and also the top plate where the wire is clamped.

For the top, we should make a cutout for a little plate in the existing crossbar. In the cutout would go a little mounting plate. Then the clamping plate (which is now steel) we would also replace with a Sapphire one.

I don't think it matters whether its sapphire or ruby, just has to be very hard.

Can you please get some quotes on the parts?

rana

Attachment 1: standoff.PDF
standoff.PDF
  6190   Thu Jan 12 10:11:59 2012 steveUpdatePEMoptical table top with 1" wall

Quote:

3/4 " thick colored acrylic material will be used in this air tight design. Surgical tubing for o-ring. We may have to put an o-ring into the bottom to have it really air tight.

Feedtrouhs: www.roxtec.com

The top drawing is not ready. It will have handle and industrial grade L-handle lock pin to hold cover down. There will be 2  one inch od post in the midle of the table to hold the cover and lock  the ball pin.

I'm waiting for your inputs, so I can send this preliminary design out for quote.

 

 The present plan to go with clear cast acrylic plexiglass 1" thick side wall  and  two  clear  1/2 thick top cover.

The inside would be lined with light braun YAG safety window sheets 0.14"  VLT ~25%  OD 4 @ 532nm & OD 5 @ 1064nm

 

Attachment 1: 101122012atc.PDF
101122012atc.PDF
  6210   Wed Jan 18 12:38:44 2012 steveUpdatePEMAcrylic plexiglass transmittance

Acrylite-Gp.pptx

Transparent- clear plexyglass from tree different sources were measured in 1064 and 532 nm light.

Samples: a, clear Acrylic-GP 0F00  from Ridout Plastics in thickness 0.7" ,  made by  Evonic Ind

                   b, clear cast acrylic from Mc Master Carr in thickness 0.94" , likely  made by Reynolds-Cast

                   c, clear cell cast  plexyglass from Delvie's Plastics - Utah in thickness  0.93" , maker not known

PMC reflected beam was used at 92 mW and 6 mm diameter at incident angle 0-25 degrees.

All tree samples agreed on Transmittance of ~90%, Reflectivity ~3-4% and calculated Adsorption ~6-7%

 

Transparent Colored Acrylic orange-amber #2422   from www.eplastics.com in 0.12" thickness gave  T 96%,  R 1% and  Ab-calc ~3% in the beam of 92 mW 1064  nm at 6 mm diameter.

 

Transparent , colored   Light Red #26 thin film filter   policarbonate-polyester   0.002" thick   from Roscolux measured T 81% of 115 mW 1064 nm

 

Now I changed power meter FieldMate to Ophir and the light source to laser pointer 2.2 mW  ~532 nm  with 1-2 mm beam diameter.

Orange - amber #2422  sample, 0.12" thick,  T 1% ,  R 4%  and  Ab-calculated ~95%, estimated visibility  ~50% It does cut out the green at this low power level.

 Light red #26  sample  T 0.5%  at 2.5 mW of 532 nm . The transparent green is not visible.  The softening point of this sandwiched polycarbonate-polyecter filter is 160C. Estimated VLT of this film ~40%

 

SUMMERY:

Clear and colored acrylics'  @ 1064 nm  transmittance 90% or higher  regardless of thickness. Softenig point 115 degrees C

Colored acrylic and colored policarbonate film are adsorbing the low power green and they  transmit the 1064nm beam.

Options to consider: a, acrylic laser safety shield liner of  0.125" thick inside of 1" thick clear acrylic  box, OD +5 @1064 and OD +4 @ 532nm,  amber color VLT 27%,  150$/sqft

                                      b, thin metal liner for 1" wall acrylic box, VLT 0%

                                     

 

Attachment 2: roscogel_red#26_film.pdf
roscogel_red#26_film.pdf
  6223   Wed Jan 25 17:32:03 2012 steveUpdateGreen Lockinggeen pointing into y arm is misaligned

I  placed an other Y2-LW-1-2050-UV-45P/AR steering mirror into the beam path of the green beam launching in order to avoid the ~30 degrees use of the 45 degrees mirror. The job is not finished.

  6226   Thu Jan 26 08:36:38 2012 steveUpdateSAFETYevacuation drill

It started with fire alarm test yesterday at 14:50 All alarms are  functioning VERY loud and their flashers are bright. Evacuation drill followed. We assembled at north west corner of the 40m building and counted 6 heads.

Nobody was left sleeping inside. Bob carried the success report of the drill to PMA office immediately.

  6227   Thu Jan 26 10:17:01 2012 steveUpdateGreen Lockinggeen pointing into y arm is realigned

Quote:

I  placed an other Y2-LW-1-2050-UV-45P/AR steering mirror into the beam path of the green beam launching in order to avoid the ~30 degrees use of the 45 degrees mirror. The job is not finished.

 The alignment is finished after the realization that the 3rd steering mirror had to be adjusted too.

The input power increased from 1.2 to 1.4 mW

Attachment 1: P1080513.JPG
P1080513.JPG
  6232   Fri Jan 27 09:05:41 2012 steveUpdateSUSOSEM locking plunger

Quote:

Our existing 300 series SS plungers from McMastercar #8476A43 are silver plated as Atm2 shows.

Problems:  1, they become magnetized after years being close to the magnets

                     2, they oxidize by time so it is hard to turn them

                    

I looked around to replace them.

Titanium body, nose and beryllium copper spring. None magnetic for UHV enviorment.

Can be made in 7 weeks at an UNREASONABLE $169.00 ea at quantity of 50

 In order to get a better price from Vlier's Tom Chen I changed Ti body back to SS304L-siver plated and music wire spring. The price is still ~$120 ea. at quantity 50

I will talk to Mike G about modifying the  McMaster plunger with a hex nut.

  6236   Mon Jan 30 08:17:06 2012 steveUpdateIOOPMC

Quote:

I have realigned the beam pointing to PMC. The transmitted light increased from 0.74 to 0.83.

The misalignment was mainly in pitch.

 The PMC pointing has changed, so MC is resonating in high order modes.

Attachment 1: sickPMC.png
sickPMC.png
  6237   Mon Jan 30 16:18:51 2012 steveUpdatePEMRoscolux colored film transmittance at 1064 nm

 

 Roscolux filter films  #74 night blue,  0.003" thick  and  #26 light red, 0.002" thick  were measured in the beam path of  ~6 mm diameter,  1W 1064 nm .

T 90%  + - 5% at 0-30 degrees of  incident angles and R ~10 % 

These sandwitched thin films of policarbonate-polyester filters are not available in thicker forms. Rosco is recommending them to be cooled by air if used in high power beam.

These filters did not get warm at all in 1W, so absorption must be very small.

  6239   Tue Jan 31 08:44:10 2012 steveUpdateIOOlaser shuts down

 

 The 2W Innilight shutdown shut when I opened side door for safety scan. This was not a repeatable by opening -closing side doors later on. Turned laser on, locked PMC and MC locked instantly. The MC was not locked this moring and it seemed that the MC2 spot was still some high order mode

like yesterday. MC lock was lost when the janitor bumped something around the MC.

Attachment 1: laseroff.png
laseroff.png
  6241   Tue Jan 31 17:13:49 2012 steveUpdateIOOonly the PSL laser is off

 

The 2W PSL laser is turned off.  The danger laser lights are not illuminated at the entry doors because of malfunctioning electronic circuit!!!

Laser safety glasses are still required!  Other lasers are in operation!

  6242   Wed Feb 1 17:00:57 2012 steveUpdateIOOlaser is back ON

Quote:

 

The 2W PSL laser is turned off.  The danger laser lights are not illuminated at the entry doors because of malfunctioning electronic circuit!!!

Laser safety glasses are still required!  Other lasers are in operation!

 BEN fixed the interlock.  The laser is turned ON. Thanks for all, Rich and Sam who came over to help. Atm1

All emergency shut- off switches, lights and door indicators are working at this moment. More about this tomorrow.

Atm2, PSL enclosure interlock jungle without REAL schematic drawing.....at this point.... We all agreed it is easier to redo the hole thing than find the problem

Atm3, Emergency shut off switches and illuminated signs from  entry doors to AC on-off box  ( Use this switches in emergency ONLY,  otherwise leave alone , even it is labeled obsolete !)

Summery: I still do not really know what was wrong.

 

Attachment 1: P1080525.JPG
P1080525.JPG
Attachment 2: P1080514.JPG
P1080514.JPG
Attachment 3: P1080518.JPG
P1080518.JPG
  6245   Fri Feb 3 14:47:51 2012 steveUpdatePEMlaser interlock drawing

 Rough draft of      updated interlock drawing by Ben is here.

 

  6247   Fri Feb 3 16:13:49 2012 steveUpdatePEMLED lights for chamber illumination

Cold LED lights replaced hot halogen ones. Flat LED MYAL 6S,  model #112560002  24VAC

This is a LATE ENTRY.  They were purchased  in Jan 2010 and installed 6 of them around May 2010

Attachment 1: P1080526.JPG
P1080526.JPG
  6250   Fri Feb 3 17:31:09 2012 steveUpdateGeneralgreen welding glass

Schott, green welding glass, shade 14, 3 mm thick  was measured in the beam path of 1.2W, S polarization of 1064nm at ~1 mm diameter size as MC reflected path.

Absorption 95%, R 5% at incident angle 25-50 degrees. It looks like the perfect material for beam trap.

 

Attachment 1: 02031202.JPG
02031202.JPG
  6256   Mon Feb 6 11:07:21 2012 steveUpdateSUSOplev QPD Sum Trends are suspicious

Quote:

The attached trend shows a problem with the QPD sums.

Why are ETMX and ITMX so much lower than ETMY and ITMY? Are the laser's dying? Or is it the gain inside the QPD? Or the reflectivity of the coatings?

Steve - please check on Monday the laser powers and the ETM/ITM reflectivity for HeNe lasers. Maybe we have to increase the transimpedance gain in the heads.

ETMX and ETMY have 0.2 mW returning to their QPDs........so the gain must lower at  ETMX

ITMX laser 1103P has only 0.67 mW output and 0.025 mW returning to the QPD.

ITMX and ETMX oplev lanching paths  have lenses without AR coating. This is my fault. I will buy them.

          ETMX   0.2 mW           900 counts

         ITMX     0.025           1300

         SRM      0.04            2600

          BS      0.05            3500

          PRM     0.06            4000

         ETMY     0.2             9000

        ITMY      0.3            14500

Attachment 1: oplevsums.png
oplevsums.png
  6257   Mon Feb 6 17:06:11 2012 steveUpdateSUSoplev transferfunction of PIT: ETMY & ETMX

Kiwamu showed me how to do transferfunction of oplev pitch

Attachment 1: ETMX_PIT.png
ETMX_PIT.png
Attachment 2: ETMY_PIT.png
ETMY_PIT.png
  6259   Tue Feb 7 16:30:46 2012 steveUpdateSUS ALL oplev PIT transfer functions

SUS- BS, ITMX, ITMY, PRM, SRM, ETMX & ETMY_OLPIT transfer funtion with sine wave excitation 0.1 amplitude:

 

Attachment 1: BS_PIT.png
BS_PIT.png
Attachment 2: ITMX_PIT.png
ITMX_PIT.png
Attachment 3: itmy_pit.png
itmy_pit.png
Attachment 4: PRM_PIT.png
PRM_PIT.png
Attachment 5: SRM_PIT.png
SRM_PIT.png
Attachment 6: ETMX_PIT.png
ETMX_PIT.png
Attachment 7: ETMY_PIT2.png
ETMY_PIT2.png
  6260   Wed Feb 8 16:37:02 2012 steveUpdateSUSITMX OSEM LL is sick

I'm driving C1:SUS-ITMX_OLYAW and PIT_EXC with amplitude 0,1-0.3 while taking transfer funtions of oplev.

The transfer functions are normal. However I noticed that the LL osem is not responding to this excitations

Healthy sensor respons should be like Atm3

Attachment 1: ITMXsensorLL.png
ITMXsensorLL.png
Attachment 2: ITMXsensorLLyaw.1.png
ITMXsensorLLyaw.1.png
Attachment 3: ETMX.2amplPIT2.png
ETMX.2amplPIT2.png
  6263   Thu Feb 9 16:46:02 2012 steveUpdateSUS ALL oplev YAW transfer functions

Quote:

SUS- BS, ITMX, ITMY, PRM, SRM, ETMX & ETMY_OLPIT transfer funtion with sine wave excitation 0.1 amplitude:

 

OL_YAW transfer functions are here.

 I had two PHDs helping me to overlap the EXML files in DTT. We failed. This job requires professorial help.

Attachment 1: BS-OLYAW.1ampl.png
BS-OLYAW.1ampl.png
Attachment 2: ITMX-OLYAW.1ampl.png
ITMX-OLYAW.1ampl.png
Attachment 3: ITMY-OLYAW.1ampl.png
ITMY-OLYAW.1ampl.png
Attachment 4: PRM_OLYAW.1.png
PRM_OLYAW.1.png
Attachment 5: SRM_OLYAW.1ampl.png
SRM_OLYAW.1ampl.png
Attachment 6: ETMX-OLYAW.1ampl.png
ETMX-OLYAW.1ampl.png
Attachment 7: ETMY-OLYAW.1ampl.png
ETMY-OLYAW.1ampl.png
  6264   Thu Feb 9 17:11:05 2012 steveUpdateSUSmore OSEM problems

These observations of the OSEMs  were taken while taking transfer functions of oplev YAW at excitation amplitude 0.1

 

 Atm1,  C1:SUS-ETMX_SENSOR_SIDE cross coupling

Atm2,   C1:SUS-ITMX_SENSOR_LL   not excitable

Atm3-4,   BS and PRM  insensitive

 

Good OSEM list: ITMY, ETMY and SRM

 

Attachment 1: ETMXosemEXC0.1ampl.png
ETMXosemEXC0.1ampl.png
Attachment 2: IMTX-YAW0.1.png
IMTX-YAW0.1.png
Attachment 3: BSosemEXC.1ampl.png
BSosemEXC.1ampl.png
Attachment 4: PRMosemEXC.1ampl.png
PRMosemEXC.1ampl.png
  6268   Fri Feb 10 11:01:31 2012 steveUpdateIOOcrazy ground motion

Quote:

I gave up tonight's locking activity because the MC can't stay locked.

It seems that somehow the seismic noise became louder from about 1:00 AM.

I walked around the outside of the 40-m building to see what's going on, but no one was jumping or partying.

I am leaving the MC autolocker disabled so that the laser won't be driven crazy and the WFS won't kick the MC suspensions.

 

The attachment is a 3-hour trend of the seismometer outputs and the MC trans.

MCunlock.png

 Something has started shaking last night.  Everybody is claiming to be innocent next door.

I turned off the 40m AC at 11:06

Attachment 1: seism1davg.png
seism1davg.png
  6269   Fri Feb 10 11:46:44 2012 steveUpdateIOOseismic noise back to normal

The shaking has stopped at 9:32am  The AC was turned back on at 11:30am  We still do not have any explanation

 

Attachment 1: seism4h.png
seism4h.png
Attachment 2: seis60s.png
seis60s.png
Attachment 3: oneday.png
oneday.png
  6270   Fri Feb 10 15:46:59 2012 steveUpdateSUSruby wire standoff

Finally I found a company who can do Koji's improved  -hard to make-  specification on ruby or sapphire wire standoff.

NOT POLISHED excimer laser cut, wire groove radius R 0.0005" + - 0.0002"

$250 ea at 50 pieces of order

  6273   Fri Feb 10 15:54:27 2012 steveUpdatePEMAC turned back ON

The air cond was off for 2 hrs.  I just switched it back on at 15:51

  6278   Tue Feb 14 08:22:27 2012 steveUpdateSUSsus damping restored

ITMX, PRM and BS watchdogs are tripped. They were restored.

Stable MC was disabled so I can use MC_REFL 1 W beam to measure green glass .

  6280   Tue Feb 14 17:09:05 2012 steveUpdateGeneralreflectivity of green welding glass

 Schott, Athermal green welding glass, shade #14 reflectivity was measured in 1.2W,  ~1 mm diameter  beam of MC reflected.

The P polarization measurement  was done with the help of half wave plate and PBC

 

 

Attachment 1: gg#14ref.pdf
gg#14ref.pdf
Attachment 2: gg14refl1Wps.png
gg14refl1Wps.png
  6282   Wed Feb 15 11:34:01 2012 steveUpdate under the shouth end optical table

I added an U channel based bottom shelf at the south end today.

Attachment 1: P1080535.JPG
P1080535.JPG
Attachment 2: P1080537.JPG
P1080537.JPG
  6288   Thu Feb 16 09:59:16 2012 steveUpdateASCIP- ANG

Initial pointing or IP-ANG is a pointing monitor of the MC. This beam is launched after the second  pzt  steering mirror.

IP-ANG  is missing the pick up mirror by a few inches at ETMYchamber

1000 days plot show last appearance in Feb 2010

Attachment 1: lastIPang.png
lastIPang.png
  6299   Tue Feb 21 08:33:16 2012 steveUpdateGreen Lockingperiscope adapter plate

Two extender plates ready for cleaning. The existing optical table tops have 38" OD. Using two of these the OD will be 44"

Attachment 1: percplate.PDF
percplate.PDF
  6308   Thu Feb 23 09:09:33 2012 steveUpdateSAFETYsafety checks

Emergency exit lights were inspected:  2 out of 13 batteries have to be replaced

One of the Halon fire extinguishers needs to be recharged out of 8

Please do participate in preparation for the upcoming safety audit on Feb 28

  6309   Thu Feb 23 10:53:54 2012 steveUpdateSUS oplev PIT transfer functions

 Oplev transfer functions PIT UGF were optimized to be at 2-3 Hz with 60 degree minimum phase margin by adjuting oplev gains.

Additional Notes by KI:

  • The PRM oplev has a tailored 3.3 Hz resonant gain in order to calm down a wobble during the lock acquisitions.
  • Also in the PRM oplev a 35 Hz elliptic cut-off filter wasn't  activated at the time when Steve measured it.
  • In both ITMs, elliptic cut-off filters seem to have higher corner frequencies compered with the others.
    • I guess these settings are from the old days.
  • ETMs and ITMs have whitening filters while the rest of the suspensions don't.
    • Without the whitening filters, normally the signals above 30 Hz are covered by some electrical noises or perhaps He-Ne laser intensity noise (#5630).
      • This is why we usually use the 35 Hz elliptic filters to roll off the control noises.
    • Since the ETMs and ITMs have whitening filters they potentially can have slightly higher corner frequencies in the elliptic filters.
      • Of course the corner frequencies need to be re-designed in terms of the amount of noise injection to the longitudinal motion.

 

Attachment 1: oplPITtrans.pdf
oplPITtrans.pdf
  6312   Fri Feb 24 08:06:52 2012 steveUpdateSUSsus restored

Quote:
The following optics were kicked:
MC1 MC2 MC3 ETMX ETMY ITMX ITMY PRM SRM BS
Fri Feb 24 04:11:15 PST 2012
1014120690
 
Steve (or anyone), can you restore the watchdogs when you come to the lab in the morning ?

Quote from #6305

Kiwamu (or whoever is here last tonight): please run the free-swing/kick script (/opt/rtcds/caltech/c1/scripts/SUS/freeswing) before you leave, and I'll check the matrices and update the suspensions tomorrow morning.

 All suspentions were restored and MC locked. PRM side osem  RMS motion was high.

Atm2, Why the PRM is 2x as noisy as the SRM ?

Attachment 1: freePRM.png
freePRM.png
Attachment 2: noisyPRM.png
noisyPRM.png
  6322   Mon Feb 27 10:21:37 2012 steveUpdateSAFETYsafety audit tomorrow morning

Quote:

Emergency exit lights were inspected:  2 out of 13 batteries have to be replaced

One of the Halon fire extinguishers needs to be recharged out of 8

Please do participate in preparation for the upcoming safety audit on Feb 28

 Batteries replaced and cylinder recharged. Please clean up your experimental set up if it is blocking breakers or entry way etc.

I will start the final clean up 2pm today.

 

  6323   Mon Feb 27 14:35:22 2012 steveUpdateSUS oplev YAW transfer functions

 The BS and the PRM have 3.3 Hz resonant gain filters that kill the phase margins.

 

Attachment 1: oplYAWtransf.pdf
oplYAWtransf.pdf
  6329   Tue Feb 28 11:20:51 2012 steveUpdateSAFETYsafety audit 2012

Correction list by visiting safety committee, Haick Issaian is not shown:

1,  update laser, crane operator list and post it

2,  check fire extinguishers monthly, date and initials must be on the tags

3,  move drinking water tower so it does not block fire extinguisher

4,  post updated crane doc at cranes

5,  post present phone lists at IFO room phones

6,  emergency laser shutoff at the south end must be mounted without C-clamp

7,  use heavy cable tie to insure position of  mag-fan on cabinet top

 

Additional to do list:

a,  safety glasses to be cleaned

b, let the electrical shop fix Rack-AC power to optical tables at the ends

c, measure transmission of  laser safety glasses

d, update  IFO outside door info signs

e, update laser inventory and post it

f,  schedule annual crane inspection and renew maintenance contract

g, PSL enclosure  inner shelf needs a good clean up so it is earthquake safe

 

Attachment 1: safety12.JPG
safety12.JPG
  6344   Thu Mar 1 09:26:50 2012 steveUpdateSUSSOS baffle plates are ready

 Green welding glass 7" x 9"   shade #14 with 40 mm hole and mounting fixtures are ready to reduce scatter light on SOS

PEEK 450CA shims and U-shaped clips  will keep these plates damped.

 

Attachment 1: 03011201.PDF
03011201.PDF 03011201.PDF
  6370   Wed Mar 7 11:20:23 2012 steveUpdateSUSoplev qpd offsets zeroed

All oplev qpd quadrons were zeroed by offset  in blocked dark condition.

  6389   Fri Mar 9 10:17:06 2012 steveHowToPEMdo not leave tables open

Type: How not to

Please do not leave optical tables open! You will be held responsible for creating dirty optics.

Attachment 1: P1080547.JPG
P1080547.JPG
  6390   Fri Mar 9 10:44:57 2012 steveUpdateRF SystemOSA

Optical spectrum analyzers like the Attachment made by Coherent , Meles Griot- CVI and Spectral Product are all discontinued.

The 40m have Coherent models C240 analyzer with controller C251 Their Finesse measured in 2004: sn205408  F302,  sn205409  F396,

Jenne borrowed Jan's Meles Griot model 13SAE006, Peter King has the same model. FSR 300 MHZ, finnees 200 minimum

Attachment 1: OSA.pdf
OSA.pdf
  6395   Fri Mar 9 16:00:46 2012 steveUpdateGreen Lockinglaser emergency shut down switch replaced at the south end

Over-sized local laser emergency switch was held by large C clamp at the south end. This was replaced by a smaller one and it is mounted with magnets.

The Innolight laser was turned off, while the interlock was wired.

  6413   Wed Mar 14 10:06:26 2012 steveUpdateGreen Lockingdichroic mirror quotes

Dichroic mirror quotes are in the wiki.

ATF is pricy.

We got a good price from Laseroptik, but the wedges are 5 arcminutes. The fused silica grade is 0F, meaning the  homogeneity is 5 ppm instead of 1ppm.  I requested an other large wedge quote on the substrates.We may have to get substrates from somebody else and ship it to Germany

MLT quote is outrageously high

REO is not interested in this low volume job.

 

  6429   Tue Mar 20 09:59:01 2012 steveUpdateIOOLaser tripped off

Today is janitor day. It still does not explain why the 2W Innolight tripped off about an hour ago. All back to normal.

.......................................................I asked Keven later, he admitted hitting the emergency shut off next to the chemical storage cabinet.

  6430   Tue Mar 20 16:53:48 2012 steveUpdatePEMcranes maintenance & certified inspection of 2012

Fred Goodbar of Konecrane has completed the annual certified crane inspection and maintenance of our cranes as required in safety document.

They are in good working condition and safe to use.

  6432   Wed Mar 21 10:31:28 2012 steveUpdateSAFETYAC power to ETMY turned off

ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4

The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.

There will be no more daisy chaining this way. 

  6433   Wed Mar 21 11:32:47 2012 steveUpdateSAFETYAC power to ETMX turned off

Quote:

ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4

The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.

There will be no more daisy chaining this way. 

 The power is back on at ETMY . c1iscey has not been restarted.

Now I'm turning ac power off at ETMX for the same job to be done.

  6434   Wed Mar 21 19:12:27 2012 steveUpdateComputersAC power back on both ends

Quote:

Quote:

ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4

The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.

There will be no more daisy chaining this way. 

 The power is back on at ETMY . c1iscey has not been restarted.

Now I'm turning ac power off at ETMX for the same job to be done.

 The power was turned back on at 4pm It took some time for Suresh to restart the computers. We have damping but things are not perfect yet. Auto BURTH did not work well.

  6435   Thu Mar 22 08:14:21 2012 steveUpdateGreen Lockingdichroic mirror quotes with large wedge

Quote:

Dichroic mirror quotes are in the wiki.

ATF is pricy.

We got a good price from Laseroptik, but the wedges are 5 arcminutes. The fused silica grade is 0F, meaning the  homogeneity is 5 ppm instead of 1ppm.  I requested an other large wedge quote on the substrates.We may have to get substrates from somebody else and ship it to Germany

MLT quote is outrageously high

REO is not interested in this low volume job.

 

 The Laseroptik quote is here.The 2 degrees wedge cost is $40 on each optics!  See wiki

  6469   Fri Mar 30 08:58:37 2012 steveUpdateSAFETYSafety glasses checked

Bob cleaned all safety glasses in 10 % Liquinox soap in water solution first.  The  transmittance of glasses were checked at 100 mW 1064 nm S polarization, beam diameter  0.5 mm at 0-20 deg incident.

Coherent FieldMate power meter measured T= < 1 mW of all glasses.

Attachment 1: safety_glasses_checked.jpg
safety_glasses_checked.jpg
  6470   Fri Mar 30 09:37:13 2012 steveUpdateSAFETYsafety audit 2012 CORRECTIONS

Quote:

Correction list by visiting safety committee, Haick Issaian is not shown:

1,  update laser, crane operator list and post it

2,  check fire extinguishers monthly, date and initials must be on the tags

3,  move drinking water tower so it does not block fire extinguisher

4,  post updated crane doc at cranes

5,  post present phone lists at IFO room phones

6,  emergency laser shutoff at the south end must be mounted without C-clamp

7,  use heavy cable tie to insure position of  mag-fan on cabinet top

 

Additional to do list:

a,  safety glasses to be cleaned

b, let the electrical shop fix Rack-AC power to optical tables at the ends

c, measure transmission of  laser safety glasses

d, update  IFO outside door info signs

e, update laser inventory and post it

f,  schedule annual crane inspection and renew maintenance contract

g, PSL enclosure  inner shelf needs a good clean up so it is earthquake safe

 

 Completed with the exception of d and g